K172325 · Cook Incorporated · GEX · Oct 26, 2017 · General, Plastic Surgery
Device Facts
Record ID
K172325
Device Name
Rhapsody H-30 Holmium Laser System
Applicant
Cook Incorporated
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 26, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Rhapsody H-30 Holmium Laser System is intended for incision, excision, resection, ablation, vaporization, coagulation, hemostasis of soft tissue, or stone fragmentation during open surgical, laparoscopic, percutaneous or endoscopic procedures in urologic, gynecologic, gastroenterological, or pulmonary applications.
Device Story
Rhapsody H-30 Holmium Laser System is a transportable, solid-state Ho:YAG laser unit; produces 2100 nm coherent near-infrared light. Operated by physician via footswitch; laser energy delivered to surgical site through compatible optical fibers. System includes rotatable control panel; detects fiber-embedded resistors to identify fiber size and limit power output; communicates with fiber-embedded PCB to display usage data/error codes. Used in OR/clinic settings for soft tissue management and stone fragmentation. Benefits include precise tissue interaction and stone destruction. Device is reusable; non-sterile.
Clinical Evidence
Bench testing only. Testing included software verification and validation (firmware v1.6), electrical safety per IEC 60601-1, and performance testing (output energy accuracy within ±10% and 30-minute continuous firing at 2.5J/12Hz). All pre-determined acceptance criteria were met.
Technological Characteristics
Solid-state Ho:YAG laser; 2100 nm wavelength; 30W max output; 5-20 Hz repetition rate; 0.5-3.5 J energy range. Fiber communication via embedded resistor/PCB. Components: rotatable control panel, water reservoir, de-ionized filter, metal enclosure. Electrical: 115/230V. Firmware v1.6. Safety: IEC 60601-1 compliant.
Indications for Use
Indicated for patients requiring soft tissue incision, excision, resection, ablation, vaporization, coagulation, hemostasis, or stone fragmentation during urologic, gynecologic, gastroenterological, or pulmonary procedures.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 26, 2017
Cook Incorporated Colin Jacob Capital Equipment Specialist, Regulatory Affairs 750 Daniels Way P.O. Box 489 Bloomington, Indiana 47402
Re: K172325
Trade/Device Name: Rhapsody H-30 Holmium Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: July 31, 2017 Received: August 1, 2017
Dear Colin Jacob:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820):
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
> Sincerely, Jennifer R. Stevenson -53 For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172325
#### Device Name Rhapsody H-30 Holmium Laser System
#### Indications for Use (Describe)
The Rhapsody H-30 Holmium Laser System is intended for incision, resection, ablation, vaporization, coagulation, hemostasis of soft tissue, or stone fragmentation during open surgical, laparoscopic, percutaneous or endoscopic procedures in urologic, gynecologic, gastroenterological, or pulmonary applications.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size:20px">✖</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|-------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:20px">❍</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Cook Medical logo, which is white text on a red background. The word "COOK" is in large, bold letters at the top. Below that, the word "MEDICAL" is in smaller letters. Below the logo is the text "510(k) Summary" in a large, bold font.
COOK
INCORPORATED
750 DANIELS WAY, P.O. BOX 489
BLOOMINGTON, IN 47402-0489 U.S.A.
PHONE: 812.339.2235 TOLL FREE: 800.457.4500
WWW.COOKMEDICAL.COM
## Rhapsody H-30 Holmium Laser System (21 CFR § 878.4810) Date Prepared: October 26, 2017
#### Submitted By:
| Applicant: | Cook Incorporated |
|-----------------------|-------------------------------------------------------------------------------|
| Contact: | Colin S. Jacob |
| Applicant Address: | Cook Incorporated<br>P.O. Box 489<br>750 Daniels Way<br>Bloomington, IN 47402 |
| Contact Phone Number: | (812) 335-3575 x 104965 |
| Contact Fax Number: | (812) 332-0281 |
#### Device Information:
| Trade Name: | Rhapsody H-30 Holmium Laser System |
|----------------------------|------------------------------------|
| Common Name: | Powered Laser Surgical Instrument |
| Classification Name/Panel: | General & Plastic Surgery |
| Regulation: | 21 CFR § 878.4810 |
| Product Code: | GEX |
#### Predicate Devices:
- . H-30 Holmium Laser System (K130444)
#### Device Description:
The Rhapsody H-30 Holmium Laser System is an enclosed, transportable solid-state (i.e., Ho:YAG) laser source unit. The laser system unit produces a beam of coherent nearinfrared light (2100 nm wavelength) upon activation by depressing a footswitch pedal. The beam is directed to the treatment zone by means of a compatible optical delivery system, such as a holmium laser fiber. Components of the unit include a footswitch pedal and rotatable control panel. The system is a reusable, non-sterile device and is supplied in a wooden shipping crate.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Cook Medical. The logo is a red rectangle with the word "COOK" in white, sans-serif font. Below the word "COOK" is the word "MEDICAL" in a smaller, white, sans-serif font. The bottom left corner of the rectangle is cut off at an angle.
COOK INCORPORATED
750 DANIELS WAY, P.O. BOX 489
BLOOMINGTON, IN 47402-0489 U.S.A.
PHONE: 812.339.2235 TOLL FREE: 800.457.4500
WWW.COOKMEDICAL.COM
#### Indications for Use:
The Rhapsody H-30 Holmium Laser System is intended for incision, excision, resection, ablation, vaporization, coagulation, hemostasis of soft tissue, or stone fragmentation during open surgical, laparoscopic, percutaneous or endoscopic procedures in urologic, gynecologic, gastroenterological, or pulmonary applications.
## Comparison to Predicate Devices:
The Rhapsody H-30 Holmium Laser System and the predicate device, H-30 Holmium Laser System (K130444), are substantially equivalent in that these devices have the same intended use and principles of operation. The differences in technological characteristics (i.e., software and minor internal components) do not raise different questions of safety and/or effectiveness when comparing to the predicate device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Cook Medical logo. The word "COOK" is in white, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in white, sans-serif font on a red background.
COOK INCORPORATED 750 DANIELS WAY, P.O. BOX 489 BLOOMINGTON, IN 47402-0489 U.S.A. PHONE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM
# Comparison to Predicate Device
| | PRIMARY PREDICATE | SUBJECT DEVICE |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | H-30 Holmium Laser System | Rhapsody H-30 Holmium Laser System |
| 510(k) | K130444 | Subject of submission |
| Manufacturer | Cook Incorporated | Cook Incorporated |
| Regulation | 21 CFR § 878.4810<br>General & Plastic Surgery | IDENTICAL TO PREDICATE |
| Product Code | GEX | IDENTICAL TO PREDICATE |
| Classification | Powered Laser Surgical Instrument | IDENTICAL TO PREDICATE |
| Indications for Use | The H-30 Holmium Laser System is indicated for use in fragmentation of urinary calculi and soft tissue applications. The H-30 Holmium Laser System is indicated for use in urinary procedures where fragmentation of stones and soft tissue incision, hemostasis, vaporization, and ablation are indicated. | The Rhapsody H-30 Holmium Laser System is intended for incision, excision, resection, ablation, vaporization, coagulation, hemostasis of soft tissue, or stone fragmentation during open surgical, laparoscopic, percutaneous or endoscopic procedures in urologic, gynecologic, gastroenterological, or pulmonary applications. |
| Principles of Operation | Generates pulsed laser energy for delivery to the surgical site via silica core laser fibers | IDENTICAL TO PREDICATE |
| Fundamental Scientific Technology | Solid-state, holmium yttrium aluminum garnet (Ho:YAG), pulsed laser operating at 2100 nanometers | IDENTICAL TO PREDICATE |
| | PRIMARY PREDICATE | SUBJECT DEVICE |
| | H-30 Holmium Laser System | Rhapsody H-30 Holmium Laser System |
| Technological Characteristics | | |
| Laser Wavelength | 2100 nm | IDENTICAL TO PREDICATE |
| Maximum Output | 30 W | IDENTICAL TO PREDICATE |
| Repetition Rate | 5 to 20 Hz | IDENTICAL TO PREDICATE |
| Visible Aiming Beam | Variable Intensity <3 mW,<br>532 nm | IDENTICAL TO PREDICATE |
| Laser Activation | Foot switch | IDENTICAL TO PREDICATE |
| Minimum Energy | 0.5 J at 5 Hz | IDENTICAL TO PREDICATE |
| Maximum Energy | 3.5 J at 5 Hz | IDENTICAL TO PREDICATE |
| Electrical Pulse Width | Short: 350 to 800 μs<br>Long: 800 to 1200 μs | IDENTICAL TO PREDICATE |
| Input Voltage | 115 & 230 V (full power) | IDENTICAL TO PREDICATE |
| Fiber Communication | 1) Detects a resistor embedded within the fiber to identify<br>the size of the attached fiber and limit the laser energy to<br>the fiber's maximum allowable power output.<br>2) Communicates with a printed circuit board (two-way<br>communication) embedded within the laser fiber to<br>display extra information onscreen, or recording use<br>information for the laser system and the fiber such as<br>power setting, number of pulses, and error codes. | IDENTICAL TO PREDICATE |
| Max Flash-lamp Voltage | 700 V | IDENTICAL TO PREDICATE |
| Blast Shield | User accessible | IDENTICAL TO PREDICATE |
| Water Reservoir | Plastic | IDENTICAL TO PREDICATE |
| Filter | De-ionized cartridge & strainer | IDENTICAL TO PREDICATE |
| Enclosure | Metal with cabinet & flat top | IDENTICAL TO PREDICATE |
| Control Panel | Rotatable | IDENTICAL TO PREDICATE |
| Water Pump | Fixed | Fixed with slightly different orientation |
| Firmware | 1.5 version | 1.6 version |
| Fan Shrouds | None | Present |
| Ferrite | None | Present |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is in white, uppercase letters, and above the word "MEDICAL" which is also in white, uppercase letters.
COOK INCORPORATED 750 DANIELS WAY, P.O. BOX 489 BLOOMINGTON, IN 47402-0489 U.S.A. PHONE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM
# Comparison to Predicate Devices (continued)
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image features the logo for Cook Medical. The logo consists of the word "COOK" in large, white, sans-serif font, set against a red square. Below the word "COOK", there is a smaller red rectangle containing the word "MEDICAL" in white, sans-serif font. The overall design is simple and clean, with a focus on the company name and its association with the medical field.
COOK INCORPORATED WW.COOKMEDICA
### Performance Data:
The subject device, Rhapsody H-30 Holmium Laser System, was subjected to applicable testing to assure reliable design and performance under the testing parameters. The tests are listed below:
| | | Table 2.0-2: Rhapsody H-30 Holmium Laser System Testing Summary | |
|--|--|-----------------------------------------------------------------|--|
| | | | |
| Test Report | Result |
|------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Test # RT20110616 | Testing must demonstrate that LCK 1.6 firmware performs as<br>intended. Test results met predetermined criteria. |
| H-30 Litho family: Software Verification and<br>Validation | |
| Test # 28110456-001 | Test article must demonstrate to comply with IEC 60601-1 |
| IEC 60601-1 Medical electrical equipment<br>Part 1: General requirements for basic safety<br>and essential performance | Medical electrical equipment Part 1: General requirements for<br>basic safety and essential performance. Test results met<br>predetermined criteria. |
| Test # 031714_502_PR #95252 | Test articles must demonstrate acceptable output energy within ±<br>10% of the factory setting and each system must fire for a<br>minimum of 30 minutes at maximum output energy of 2.5J@<br>12Hz on short pulse width. Test results met predetermined<br>criteria. |
| Evaluation of Rhapsody H-30 - Addition of<br>Fan Shroud, Optimization of Cooling Lines<br>and Alignment Procedure | |
### Conclusion:
For these tests, all pre-determined acceptance criteria were met. The results of these tests show that the Rhapsody H-30 Holmium Laser System meets the design input requirements based on the intended use. Furthermore, these results support the conclusion that the differences between the subject device, Rhapsody H-30 Holmium Laser System, and the predicate device, H-30 Holmium Laser System, do not raise new questions of safety or effectiveness and support a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.