K172312 · Eemagine Medical Imaging Solutions GmbH · GWL · Mar 30, 2018 · Neurology
Device Facts
Record ID
K172312
Device Name
eego amplifiers
Applicant
Eemagine Medical Imaging Solutions GmbH
Product Code
GWL · Neurology
Decision Date
Mar 30, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1835
Device Class
Class 2
Indications for Use
The eego amplifier is intended to be used by or under the direction of a physician for acquisition of EEG signals and to transmit them digitally to a computer. The device is intended for use on humans. The device is intended for use in a clinical environment (e.g., hospital, physician's office, etc). The device is not intended for use in life support systems.
Device Story
Multi-channel, plastic-enclosed EEG amplifier; models EE-2xx (battery-powered) and EE-4xx (USB-powered). Inputs: EEG signals from external electrodes (e.g., Waveguard EEG Cap); referential DC input channels; parallel trigger input channel. Operation: Amplifies physiological signals; performs low-pass filtering; measures electrode impedance; transmits digital data to a computer via USB. Used in clinical environments by physicians or under their direction. Output: Digital EEG signal stream for computer storage/analysis. Benefits: Enables clinical EEG monitoring and recording. Healthcare providers use output for neurophysiological assessment and diagnosis.
Clinical Evidence
Bench testing only. No clinical data provided. Performance confirmed via compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-26, and IEC 62133 standards.
Technological Characteristics
Multi-channel EEG amplifier; plastic enclosure. Models: EE-2xx (rechargeable battery), EE-4xx (USB-powered). Features: Active shielding, referential DC input channels, parallel trigger input. Sampling rates: 0.5 kHz to 16 kHz. Connectivity: USB to computer. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-26, IEC 62133.
Indications for Use
Indicated for acquisition of EEG signals in humans in clinical environments (e.g., hospitals, physician offices) under physician direction. Not for use in life support systems.
Regulatory Classification
Identification
A physiological signal amplifier is a general purpose device used to electrically amplify signals derived from various physiological sources (e.g., the electroencephalogram).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 30, 2018 eemagine Medical Imaging Solutions GmbH % Gary J. Syring Principal Consultant Ouality & Regulatory Associates, LLC 800 Levanger Lane Stoughton, Wisconsin 53589
Re: K172312
Trade/Device Name: eego amplifiers Regulation Number: 21 CFR 882.1835 Regulation Name: Physiological Signal Amplifier Regulatory Class: Class II Product Code: GWL, GWQ, OMC Dated: July 25, 2017 Received: March 1, 2018
Dear Gary J. Syring:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
William J. Heetderks -S 2018.03.30 16:43:21 -04'00"
for Carlos L. Peña. PhD. MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172312
Device Name eego™ amplifiers
Indications for Use (Describe)
The eego amplifier is intended to be used by or under the direction for acquisition of EEG signals and to transmit them digitally to a computer. The device is intended for use on humans. The device is intended for use in a clinical environment (e.g., hospital, physician's office, etc). The device is not intended for use in life support systems.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------|---------------------------------------------|
|----------------------------------------------|---------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K173312 510(k) Summary
## Submitter Information
| Submitter's Name: | eemagine Medical Imaging Solutions GmbH |
|----------------------|---------------------------------------------------------|
| Submitter's Address: | Gubenerstr. 47, Fabrik |
| | D-10243, Berlin, Germany |
| Submitter's Phone: | Phone: +49 30 29 04 84 04 |
| Contact Person: | Frank Zanow, Chief Executive Officer |
| | or Ralf Hauffe, PhD, Chief Operations Executive Officer |
| Date Summary | |
March 29, 2018
# Subject Device
Prepared:
| Trade Name: | eego™ amplifiers |
|----------------------|----------------------------------------------------------------------------------------------------------|
| Common Name: | Physiological signal amplifier |
| Classification Name: | 21 CFR 882.1835 Physiological signal amplifier<br>Class II |
| Product Code: | GWL, amplifier, physiological signal |
| Subsequent Code: | GWQ, full-montage standard electroencephalograph<br>OMC, reduced- montage standard electroencephalograph |
# Predicate Device
| 510(k) Number: | K060374 |
|----------------------|-------------------------------|
| Trade Name: | REFA, Physiological Amplifier |
| Submitter Name: | TMS International BV |
| Classification Name: | 21 CFR 882.1835 |
| Product Code: | GWL |
### Device Description
The eego amplifier is a multi-channel, plastic enclosed EEG amplifier. There are two ranges for the variations of amplifiers (models):
- 1. EE-2xx, and
- 2. EE-4xx.
In total, there are 8 models, as follows:
| Model | EEG<br>Channels<br>(Referential) | Sampling rate configuration<br>(down sampling in API software,<br>internal 32 kHz for all) | Power Source |
|--------|----------------------------------|--------------------------------------------------------------------------------------------|-------------------------------|
| EE-211 | 64 | 2 kHz | Rechargeable battery |
| EE-212 | 32 | 2 kHz | |
| EE-213 | 16 | 2 kHz | |
| EE-221 | 32 | 16 kHz | |
| EE-222 | 64 | 16 kHz | |
| EE-224 | 32 | 16 kHz | |
| EE-411 | 8 | 2 kHz | USB Connection to<br>Computer |
| EE-430 | 8 | 0.5 kHz | |
{4}------------------------------------------------
The EE-411 and EE-430 amplifiers are built using identical components and share identical internal structure, functionality and performance.
The eego amplifier electroencephalography (EEG) input is from commercially available EEG electrodes from the Waveguard™ EEG Cap cleared in K110223.
The eego amplifiers support EEG acquisition and storage by:
- 1. referential DC input channels,
- 2. parallel trigger input channel.
Electrode Impedance values can be measured for all referential electrodes, reference and patient ground electrode. Signal sampling rate and gain are set through the user interface of the Software Development Kit (SDK).
The EE-2XX series of eego amplifier is powered by a built-in, rechargeable battery. The battery provides a minimum of 4 hours of operation time. The amplifier can be applied for data acquisition with the 12 V battery charger source applied. The USB port supports interface to the computer running the SDK.
The EE-4XX series of eego amplifier is powered by the USB port connection to a computer.
### Statement of Intended Use
The eego amplifier is intended to be used by or under the direction of a physician for acquisition of EEG signals and to transmit them digitally to a computer. The device is intended for use on humans. The device is intended for use in a clinical environment (e.g., hospital, physician's office, etc). The device is not intended for use in life support systems.
### Summary of Technological Characteristics & Comparison to Predicate
The following table provides a comparison the eego amplifiers to the predicate device identified to demonstrate substantial equivalence.
| Table 1: Substantial Equivalence Technical Characteristics | | | |
|------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| Feature | eego™ amplifiers<br>Subject Device K172312 | REFA Physiological Amplifier<br>Predicate K060374 | Comment |
| Intended Use | Used to electrically amplify signals<br>derived from electroencephalogram. | Used to electrically amplify signals<br>derived from various physiological<br>sources, including the<br>electroencephalogram. | Both devices are<br>amplifiers of signals<br>derived from physiologic<br>sources. |
| Indications for Use | The eego amplifier is intended to be<br>used by or under the direction of a<br>physician for acquisition of EEG signals<br>and to transmit them digitally to a<br>computer. The device is intended for<br>use on humans. The device is intended<br>for use in a clinical environment (e.g.,<br>hospital, physician's office, etc). The<br>device is not intended for use in life<br>support systems. | The REFA Physiological Amplifier<br>family is intended to be used by or<br>under the direction of a physician<br>for acquisition of EEG, polygraphy<br>and polysomnography signals and<br>transmission of the signals to a PC<br>during recording of neurophysical /<br>physiological research and exams.<br>Polygraphy and Polysomnography<br>may besides EEG, include<br>physiological information such as<br>EMG, ECG, EOG, EGG, PH,<br>Respiration, Temperature and<br>Oxygen Saturation. | Amplifiers are intended<br>for acquisition of EEG<br>signals to a computer. |
| Type of use | Prescription Use | Prescription Use | Same |
| Modes of Operation | Electroencephalography (EEG) | EEG, EMG, ECG, EOG<br>EGG, PH<br>Respiration, Temperature, | Both acquire EEG signals. |
| Table 1: Substantial Equivalence Technical Characteristics | | | |
| Feature | eego™ amplifiers<br>Subject Device K172312 | REFA Physiological Amplifier<br>Predicate K060374 | Comment |
| | | Oxygen Saturation | |
| Environment of Use | Clinical environment<br>(Example: hospital, physician's office,<br>etc) | Clinical environment<br>(Example: hospital, physician's<br>office, etc) | Same |
| Data Storage | Yes | Yes | Same |
| Electrode<br>impedance<br>measurement | Yes | Yes | Same |
| Filter | Low pass filtering in amplifier | Low pass filtering in amplifier | Same |
| Patient Contact | No<br>Patient contact is made by the<br>Waveguard™ EEG Cap (K110223). | No<br>Patient contact is made by<br>commercially available electrodes<br>and transducers. | Equivalent |
| Sampling rate | EE-2XX Series:<br>Up to 16384 Hz Set through the user<br>interface of the Software Development<br>Kit (SDK) software<br>EE-4XX Series: Up to 2048 Hz | Up to 20000 Hz | Equivalent |
| Video | Optional | No | Option to allow video |
| Amplifier active<br>shielding technique | Yes<br>(Protects the referential EEG inputs from<br>environmental noise) | Yes | Same |
| Amplifier CMRR<br>(referential) | >100 dB | 90 dB | Equivalent |
| Amplifier Input<br>Channels | Up to 64 | Up to 64<br>Special mode supports 128 EEG<br>inputs | Equivalent |
| Amplifier input<br>impedance<br>(referential) | >109 Ohm | >1012 Ohm | Equivalent for low<br>impedance of<br>physiological signal<br>inputs. |
| Amplifier power | EE-2XX Series: Rechargeable<br>integrated battery<br>EE-4XX Series: USB from Computer | AC Mains | DC power source for eego<br>amplifiers. |
| Amplifier operating<br>time on battery<br>power | EE-2XX Series: Up to 4 hours | Not Applicable | DC power source for eego<br>amplifiers. |
| Amplifier referential<br>DC input channels | Yes<br>8, 16, 32, or 64 channels | Yes<br>16, 32, 64 or 128 channels | Equivalent |
| Amplifier trigger<br>input | Yes,<br>EE-2XX Series: Optional 8 bit TTL<br>EE-4XX Series: Optional 2 bit TTL | Yes,<br>Optional 8 bit TTL | Equivalent |
| Amplifier trigger<br>input channel | Yes<br>(Parallel) | Yes | Equivalent |
| Amplifier USB<br>interface | Yes | Yes | Same |
{5}------------------------------------------------
{6}------------------------------------------------
| Table 1: Substantial Equivalence Technical Characteristics | | |
|------------------------------------------------------------|--|--|
|------------------------------------------------------------|--|--|
| Feature | eego™ amplifiers<br>Subject Device K172312 | REFA Physiological Amplifier<br>Predicate K060374 | Comment |
|--------------------------|-------------------------------------------------------------|---------------------------------------------------|-------------------------------|
| Safety standards applied | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-26<br>IEC 62133 | IEC 60601-1<br>IEC 60601-1-2 | Appropriate standards applied |
## Performance Testing
To establish the equivalency of the eego amplifiers, evaluations were conducted to confirm compliance with performance requirements and standards including:
- IEC 60601-1:2005 (DIN EN 60601-1:2007), Medical electrical equipment Part 1: General 0 requirements for basic safety and essential performance.
- . IEC 60601-1-2:2007, Medical electrical equipment - part 1-2: general requirements for basic safety and essential performance - collateral standard: electromagnetic compatibility requirements and tests.
- IEC 60601-1-2:2014 Medical electrical equipment Part 1-2: General requirements for basic . safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests.
- . IEC 60601-2-26: 2012 Medical electrical equipment, Particular requirements for the basic safety and essential performance of electroencephalographs.
- IEC 62133 Edition 2.0 2012-12. Secondary cells and batteries containing alkaline or other . non-acid electrolytes - safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications.
### Conclusion
The eego amplifiers meet performance requivalent to the predicate device. The intended use and technology of the eego amplifiers are equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.