← Product Code [FGB](/productcode/FGB) · K172298

# OLYMPUS URF-P6/P6R (K172298)

_Olympus Medical Systems Corp. · FGB · Mar 29, 2018 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K172298

## Device Facts

- **Applicant:** Olympus Medical Systems Corp.
- **Product Code:** [FGB](/productcode/FGB.md)
- **Decision Date:** Mar 29, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

This instrument has been designed to be used with an Olympus light source, documentation equipment, monitor, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment within the ureter, kidney and biliary tract (common bile duct and hepatic duct).

## Device Story

Flexible uretero-reno fiberscope; used for endoscopic diagnosis and treatment in ureter, kidney, and biliary tract. Device comprises eyepiece section, control section, and flexible insertion tube; utilizes light fiber bundle for image transfer. Operates with external light source, documentation equipment, monitor, and EndoTherapy accessories. Used in healthcare facilities by physicians. URF-P6R variant features opposite bending direction compared to URF-P6. Provides visual access to internal anatomy; enables diagnostic observation and therapeutic interventions (e.g., stone extraction/fragmentation). Benefits include minimally invasive access to urinary and biliary tracts.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance verified through risk analysis (ISO 14971), biocompatibility (ISO 10993), electrical safety (IEC 60601-1/2-18), and reprocessing validation.

## Technological Characteristics

Flexible fiberscope; distal end (polysulfone), insertion tube (fluoro resin), bending rubber (fluoro rubber), lens (glass). Sterilization: Ethylene oxide and hydrogen peroxide. Connectivity: Eyepiece/video camera connection. Energy: Electricity (for ancillary equipment).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- URETERORENOSCOPE OLYMPUS URF TYPE P2 ([K912120](/device/K912120.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

March 29, 2018

Olympus Medical Systems Corp. % Sheri L. Musgnung Regulatory Affairs Manager Olympus Corporation of the Americas 3500 Corporate Parkway, P.O. Box 610 Center Valley, PA 18034-0610

Re: K172298 Trade/Device Name: Olympus URF-P6/P6R Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FGB. FBN Dated: February 23, 2018

Received: February 26, 2018

Dear Sheri L. Musgnung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Charles Viviano -S

For Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K172298

Device Name Olympus URF-P6/P6R

Indications for Use (Describe)

This instrument has been designed to be used with an Olympus light source, documentation equipment, monitor, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment within the ureter, kidney and biliary tract (common bile duct and hepatic duct).

Type of Use (Select one or both, as applicable)

| <div> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D)    </div> |
|----------------------------------------------------------------------------------------------------------------|
| <div>     Over-The-Counter Use (21 CFR 801 Subpart C)    </div>                                                |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Olympus. The word "OLYMPUS" is written in large, bold, blue letters. Below the word is a thin, horizontal, yellow line. Underneath the line, the text "Your Vision, Our Future" is written in a smaller, dark font.

# 510(k) Summary

February 23, 2018

K172298 Page 1 of 5

#### 1. General Information

- Applicant: OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho, Hachioji-shi, Tokyo, 192-8507, Japan Establishment Registration No.: 8010047
- Official Correspondent: Sheri L. Musgnung Regulatory Affairs Olympus America Inc. 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-3147 FAX: 484-896-7128 Email: sheri.musgnung@olympus.com
- Aizu Olympus Co., Ltd. Manufacturer: 500 Aza-Muranishi, Ooaza-Iidera, Monden-cho, Aizuwakamatsu-shi, Fukushima, Japan 965-8520 Establishment Registration No.: 9610595

### 2. Device Identification

| ■ | Device Trade Name:         | OLYMPUS URF-P6/P6R,                                   |
|---|----------------------------|-------------------------------------------------------|
| ■ | Common Name:               | URETERO-RENO FIBERSCOPE                               |
| ■ | Regulation Number:         | 876.1500                                              |
| ■ | Regulation Name:           | Endoscope and Accessories                             |
| ■ | Regulatory Class:          | II                                                    |
| ■ | Classification Panel:      | Gastroenterology and urology                          |
| ■ | Product Code:              | FGB (Ureteroscope and accessories, flexible/rigid)    |
| ■ | Secondary Product<br>Code: | FBN (Choledochoscope and accessories, flexible/rigid) |

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Olympus. The word "OLYMPUS" is written in large, bold, blue letters. Below the word is a thin, horizontal, yellow line. Underneath the line, the words "Your Vision, Our Future" are written in a smaller font.

## 3. Predicate Device

K912120 URETERORENOSCOPE OLYMPUS URF TYPE P2

### 4. Device Description

This endoscope is composed from eyepiece section, control section and flexible insertion tube, and equipped light fiber bundle for image transfer system. Besides, it has eyepiece frame on eyepiece section for eyepiece and video camera connection for endoscope.

URF-P6R has Identification marking for opposite bending direction on the control section and the bending section of the URF-P6R moves in the direction opposite to the conventional endoscopes, such as the URF-P6. This is the only different point between URF-P6 and URF-P6R.

### 5. Indications for Use

This instrument has been designed to be used with an Olympus light source, documentation equipment, monitor, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment within the ureter, kidney and biliary tract (common bile duct and hepatic duct).

The Indications for Use statement for the OLYMPUS URF-P6/P6R is not literally identical to the predicate device: however, the meanings of their descriptions are the same. Both of the subject and predicate device have the same intended use for the endoscopic diagnosis and treatment of urinary tract and biliary tract.

### 6. Comparison of Technological Characteristics

Compared to the predicate device, the subject device differs as it relates to labeling, technology/ engineering/ performance, and materials. The main modifications are as follows:

- 1. The physical properties of the insertion section and optical parameters are slightly changed.
- 2. The patient-contact material is changed.
- 3. Sterilization method by hydrogen peroxide is added.
- 4. Internal structure is improved to enhance the mechanical durability.

Validation from non-clinical testing demonstrated that these technological features do not raise any new issues of safety or effectiveness of the subject device.

A side by side comparison of the subject device and the predicate device is provided below.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Olympus. The word "OLYMPUS" is written in large, bold, blue letters. Below the word is a thin, yellow line. Underneath the line is the text "Your Vision, Our Future".

#### OLYMPUS MEDICAL SYSTEMS CORP. Traditional 510(k) URETERO-RENO FIBERSCOPE OLYMPUS URF-P6/P6R

| Item                                                          | <Subject Device><br>URF-P6/P6R                                                                                                                                                                                                                                                                                            | <Predicate Device><br>URF-P2 (K912120)                                                                                                                                                                                                                                                                                                                                                                                    |
|---------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use                                           | This instrument has been<br>designed to be used with an<br>Olympus light source,<br>documentation equipment,<br>monitor, EndoTherapy<br>accessories, and other ancillary<br>equipment for endoscopic<br>diagnosis and treatment within<br>the ureter, kidney and biliary<br>tract (common bile duct and<br>hepatic duct). | The Olympus URF Type P2<br>has been designed for<br>endoscopic observation and<br>diagnosis within both the<br>urinary tract (ureter, renal<br>pelvis, and renal calyx) and<br>biliary tract (common bile duct<br>and hepatic duct.) Therapeutic<br>endoscopic procedures using<br>various kinds of accessories<br>are also possible, and include<br>the ability to extract or<br>fragment stones within these<br>organs. |
| Common name                                                   | URETERO-RENO<br>FIBERSCOPE                                                                                                                                                                                                                                                                                                | URETERO-RENO<br>FIBERSCOPE                                                                                                                                                                                                                                                                                                                                                                                                |
| Regulation number                                             | 876.1500                                                                                                                                                                                                                                                                                                                  | 876.1500                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Regulation name                                               | Endoscope and Accessories                                                                                                                                                                                                                                                                                                 | Endoscope and Accessories                                                                                                                                                                                                                                                                                                                                                                                                 |
| Regulatory class                                              | II                                                                                                                                                                                                                                                                                                                        | II                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Classification panel                                          | Gastroenterology and urology                                                                                                                                                                                                                                                                                              | Gastroenterology and urology                                                                                                                                                                                                                                                                                                                                                                                              |
| Product code                                                  | FGB, FBN                                                                                                                                                                                                                                                                                                                  | FBN, FGB                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Environment of use                                            | Healthcare facility/hospital                                                                                                                                                                                                                                                                                              | Healthcare facility/hospital                                                                                                                                                                                                                                                                                                                                                                                              |
| Single/repeat use                                             | Repeat use                                                                                                                                                                                                                                                                                                                | Repeat use                                                                                                                                                                                                                                                                                                                                                                                                                |
| Sterile/non-sterile                                           | Marketed as non-sterile                                                                                                                                                                                                                                                                                                   | Marketed as non-sterile                                                                                                                                                                                                                                                                                                                                                                                                   |
| Sterilization method                                          | Ethylene oxide; Hydrogen<br>peroxide                                                                                                                                                                                                                                                                                      | Ethylene oxide;                                                                                                                                                                                                                                                                                                                                                                                                           |
| Energy source                                                 | Electricity                                                                                                                                                                                                                                                                                                               | Electricity                                                                                                                                                                                                                                                                                                                                                                                                               |
| Material composition<br>of main patient-<br>contact parts and | Material composition of main<br>patient-contact parts                                                                                                                                                                                                                                                                     | Material composition of main<br>patient-contact parts                                                                                                                                                                                                                                                                                                                                                                     |
| duration and type of<br>contact                               | Distal end: polysulfone<br>Insertion tube: fluoro resin<br>Bending rubber: fluoro rubber<br>Lens: glass<br>Adhesive: ES-19MC<br><br>Surface-contacting device in<br>contact with mucosal<br>membranes. The contact<br>duration is limited exposure<br>(i.e. contact is up to 24 hours).                                   | Distal end: stainless<br>Insertion tube: polyurethane<br>resin<br>Bending rubber: fluoro rubber<br>Lens: glass<br>Adhesive: ES-8SC<br><br>Surface-contacting device in<br>contact with mucosal<br>membranes. The contact<br>duration is limited exposure<br>(i.e. contact is up to 24 hours).                                                                                                                             |
|                                                               |                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                           |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Olympus. The word "OLYMPUS" is written in large, bold, dark blue letters. Below the word is a thin, horizontal, yellow line. Below the line, the words "Your Vision, Our Future" are written in a smaller, dark font.

K172298 Page 4 of 5

### 7. Summary of Non-clinical Testing

The technological characteristic differences between the predicate device and the subject device have been confirmed that they are substantially equivalent through the following tests and standards.

·Risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verification tests and their acceptance criteria were identified and performed as a result of this risk analysis assessment.

· Biocompatibility testing is performed in accordance with the FDA Guidance, Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a risk management process" issued on June 16 2016.

•Performance testing was carried out to verify the safety and the effectiveness of the subject device.

·The reprocessing validation test was performed in accordance with the FDA guidance "Guidance for Industry and FDA Staff - Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling issued on March 17, 2015"

·Electric safety has been validated in accordance with FDA`s currently recognized standards.

The results of the above performance testing demonstrated that the subject device has no concerns on safety and effectiveness.

| Standard No.                              | Standard Title                                                                                                                                              |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 14971 Second<br>Edition: 2007-03-01   | Medical Devices - Application Of Risk Management To<br>Medical Devices                                                                                      |
| ISO 10993-1 Fourth<br>Edition:2009-10-15  | Biological Evaluation Of Medical Devices - Part1:<br>Evaluation And Testing Within A Risk Management<br>Process [Including: Technical Corrigendum 1 (2010)] |
| AAMI ANSI ISO<br>10993-5:2009/(R)2014     | Biological Evaluation Of Medical Devices – Part5: Tests<br>For In Vitro Cytotoxicity                                                                        |
| ISO 10993-10 Third<br>Edition: 2010-08-01 | Biological Evaluation Of Medical Devices - Part 10: Tests<br>For Irritation And Skin Sensitization                                                          |

Also, the following standards have been applied to the subject device:

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Olympus. The word "OLYMPUS" is written in large, bold, blue letters. Below the word is a thin, horizontal, yellow line. Underneath the line is the text "Your Vision, Our Future".

#### OLYMPUS MEDICAL SYSTEMS CORP. Traditional 510(k) URETERO-RENO FIBERSCOPE OLYMPUS URF-P6/P6R

| Standard No.                                        | Standard Title                                                                                                                                  |
|-----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| AAMI ANSI ES60601-<br>1:2005/(R)2012 and<br>A1:2012 | Medical Electrical Equipment - Part 1: General<br>Requirements for Basic Safety And Essential Performance<br>+ Amendment 1: 2012                |
| IEC 60601-2-18 Edition<br>3.0: 2009-08              | Medical Electrical Equipment - Part 2-18: Particular<br>Requirements For The Basic Safety And Essential<br>Performance Of Endoscopic Equipment. |

### 8. Conclusion

In comparison to the predicate device, Olympus URF-P6/P6R has the same intended use and the technical differences do not cause any significant changes that could affect the safety or effectiveness of the device. Therefore, the subject device is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K172298](https://fda-staging.innolitics.com/device/K172298)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
