The MicroScan Dried Gram-Negative MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly and facultative anaerobic gram-negative bacilli. After inoculation, panels are incubated for 16 - 20 hours at 35°C +/ - 1°C in a non-CO2 incubator, and read either visually or with MicroScan instrumentation, according to the Package Insert. This particular submission is for the antimicrobial ceftolozane/tazobactam at concentrations of 0.25/4 to 16/4 ug/mL to the test panel.
Device Story
MicroScan Dried Gram-Negative MIC/Combo Panels are miniaturized, dehydrated broth dilution susceptibility tests. The device uses Mueller Hinton broth supplemented with calcium and magnesium to test antimicrobial concentrations. Inoculation is performed via Prompt or turbidity methods. Panels are incubated for 16–20 hours at 35°C in a non-CO2 incubator. Results are determined by observing the lowest antimicrobial concentration inhibiting growth, either visually or via MicroScan WalkAway or autoSCAN-4 instrumentation. The device provides MIC values and categorical interpretations (S, I, R) to assist clinicians in selecting appropriate antibiotic therapy for Gram-negative bacterial infections. Limitations exist for specific organisms (e.g., E. cloacae, M. morganii) requiring manual interpretation or retesting with turbidity methods when using specific instrument/inoculation combinations.
Clinical Evidence
Performance evaluated using 575 clinical isolates and 118 challenge isolates (total 693). Study compared MicroScan results to CLSI frozen broth microdilution reference. Essential Agreement (EA) and Category Agreement (CA) were generally >90%. Specific performance issues identified for E. cloacae, M. morganii, P. rettgeri, and S. liquefaciens when using Prompt inoculation with WalkAway/autoSCAN-4, resulting in labeling limitations. MIC trending observed for Enterobacteriaceae (one doubling dilution higher) with WalkAway/Prompt. Reproducibility >95% across all methods.
Technological Characteristics
Miniaturized, dehydrated broth microdilution panels. Energy source: non-CO2 incubator. Inoculation: Prompt or turbidity (0.5 McFarland). Read methods: Visual, MicroScan WalkAway, or autoSCAN-4. Connectivity: Instrument-based automated reading. Materials: Mueller Hinton broth supplemented with calcium and magnesium. Software: Automated growth detection and MIC calculation.
Indications for Use
Indicated for determining quantitative and/or qualitative antimicrobial susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-negative bacilli, including Enterobacter cloacae, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Proteus mirabilis, Pseudomonas aeruginosa, Citrobacter koseri, Morganella morganii, Proteus vulgaris, Providencia rettgeri, Providencia stuartii, Serratia liquefaciens, and Serratia marcescens.
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
Predicate Devices
MicroScan Dried Gram-Negative MIC/Combo Panels with Imipenem (K162740)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K172255
B. Purpose for Submission:
To obtain a substantial equivalence for the addition of ceftolozane/tazobactam at concentrations 0.25/4-16/4 µg/mL to the MicroScan Dried Gram negative MIC/Combo Panel for susceptibility testing of non-fastidious Gram negative organisms.
C. Measurand:
Ceftolozane/tazobactam 0.25/4-16/4 µg/mL
D. Type of Test:
Quantitative Antimicrobial Susceptibility Test (AST) growth based detection
E. Applicant:
Beckman Coulter
F. Proprietary and Established Names:
MicroScan Dried Gram-Negative MIC/Combo Panels with Ceftolozane/tazobactam (0.25/4-16/4 µg/mL)
G. Regulatory Information:
1. Regulation section:
21 CFR 866.1640 Antimicrobial Susceptibility Test Powder
2. Classification:
Class II
3. Product codes:
LTT – Panels, Test, Susceptibility, Antimicrobial
{1}
LRG – Instrument for Auto Reader and Instrumentation of Overnight Susceptibility Systems
JWY – Manual Antimicrobial Susceptibility Test Systems
4. Panel:
83 - Microbiology
H. Intended Use:
1. Intended use(s):
For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo panels. MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.
2. Indication(s) for use:
The MicroScan Dried Gram-Negative MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-negative bacilli. After inoculation, panels are incubated for 16 – 20 hours at 35°C +/- 1°C in a non-CO2 incubator, and read either visually or with MicroScan instrumentation, according to the Package Insert.
This particular submission is for the addition of the antimicrobial ceftolozane/tazobactam at concentrations of 0.25/4 to 16/4 µg/mL to the test panel.
Ceftolozane/tazobactam has been shown to be active both clinically and in vitro against the following organisms:
Enterobacter cloacae, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Proteus mirabilis, and Pseudomonas aeruginosa
In vitro data is available for the following organisms, but their clinical significance is unknown:
Citrobacter koseri, Morganella morganii, Proteus vulgaris, Providencia rettgeri, Providencia stuartii, Serratia liquefaciens, and Serratia marcescens
2
{2}
3. Special conditions for use statement(s):
For prescription use only
The following information is included in labeling
- Results obtained with *E. cloacae* and ceftolozane/tazobactam have shown discrepant MICs when compared with the reference method when read by the MicroScan instruments. Therefore, prior to reporting, these isolate results should be interpreted manually.
- Results obtained *P. rettgeri* and *S. liquefaciens* and ceftolozane/tazobactam have shown discrepant MICs when compared with the reference method when read by the WalkAway instrument and inoculated using the Prompt. Therefore, prior to reporting, these isolate results should be interpreted manually.
- Ceftolozane/tazobactam results obtained with *M. morganii* and *P. rettgeri* isolates using the Prompt Inoculation system have shown potential major errors. Therefore, prior to reporting, isolates with resistant results should be retested using the turbidity inoculation method.
- Ceftolozane/Tazobactam is not active against *Enterobacteriaceae* bacteria that produce metallo-beta lactamases and serine carbapenemases.
4. Special instrument requirements:
MicroScan panels can be read either manually or automatically on the WalkAway or autoScan-4 instrument systems. Inoculum suspensions can be determined manually or using the Prompt.
I. Device Description:
The MicroScan Dried Gram Negative MIC/Combo Panel with ceftolozane/tazobactam is used to determine the quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative Gram-negative bacilli. After inoculation, panels are incubated for 16-20 hours at $35^{\circ}\mathrm{C} \pm 1^{\circ}\mathrm{C}$ in a non- $\mathrm{CO}_{2}$ incubator, and read either visually or with MicroScan instrumentation, according to the package insert.
Primary inoculation method: Prompt; Alternate inoculation method: Turbidity
Primary read method: MicroScan WalkAway system; Alternate read methods: MicroScan autoSCAN-4 (AS4) instrument and manual reads
{3}
J. Substantial Equivalence Information:
1. Predicate device name(s):
MicroScan Dried Gram-Negative MIC/Combo Panels with Imipenem (0.25 - 8 µg/mL)
2. Predicate 510(k) number(s):
K162740
3. Comparison with predicate:
Table 1. Comparison with Predicate Device
| Similarities | | |
| --- | --- | --- |
| Item | Device: MicroScan Dried Gram-Negative MIC/Combo Panels – Ceftolozane/tazobactam (K172255) | Predicate: MicroScan Dried Gram-Negative MIC/Combo Panels - Imipenem (K162740) |
| Technology | Overnight microdilution MIC susceptibility | Same |
| Read Method | Manual and automated | Same |
| Inoculation Methods | Turbidity and Prompt | Same |
| Result Reported | Report results as minimum inhibitory concentration (MIC) and categorical interpretation (S,I,R) | Same |
| Instrumentation | WalkAway or autoSCAN-4 Instrument systems | Same |
| Incubation | 16-20 hours | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device: MicroScan Dried Gram-Negative MIC/Combo Panels – Ceftolozane/tazobactam (K172255) | Predicate: MicroScan Dried Gram-Negative MIC/Combo Panels - Imipenem (K162740) |
| Antimicrobial agent | Ceftolozane/tazobactam | Imipenem |
| Concentration Range | 0.25/4-16/4 µg/mL | 0.25 – 8 µg/mL |
K. Standard/Guidance Document Referenced (if applicable):
1. Guidance for Industry and FDA - Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems – August 28, 2009.
2. CLSI M100-S027: Performance Standards for Antimicrobial Susceptibility Testing;
{4}
Twenty-Seventh Informational Supplement
3. CLSI M7-A10: Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically; Approved Standard - Tenth Edition
# L. Test Principle:
The MicroScan Dried Gram-Negative MIC/Combo Panels with Ceftolozane/tazobactam are antimicrobial susceptibility tests that are miniaturizations of the broth dilution susceptibility test which have been dehydrated. Various antimicrobial agents are diluted in Mueller Hinton broth supplemented with calcium and magnesium to concentrations bridging the range of clinical interest. Breakpoint Combo panels use concentrations equivalent to the categorical breakpoints of FDA and/or CLSI. After inoculation and rehydration with a standardized suspension of organism and incubation of $35^{\circ}\mathrm{C} + / - 1^{\circ}\mathrm{C}$ for a minimum of 16 hours, the minimum inhibitory concentration (MIC) for the test organism is determined by observing the lowest antimicrobial concentration showing inhibition of growth.
# M. Performance Characteristics:
# 1. Analytical performance:
# a. Precision/Reproducibility:
A reproducibility study was performed at three sites using 17 isolates of non-fastidious Gram negative organisms. The isolates were tested in triplicate over three different days using both inoculation methods (Prompt, turbidity) for each reading method (WalkAway, autoSCAN-4, manual). The isolates tested in the reproducibility study included C. koseri (1), E. cloacae (1), E. coli (5), K. oxytoca (2), K. pneumoniae (3), P. aeruginosa (3), P. mirabilis (1) and S. marcescens (1). The mode MIC value was pre-determined and the reproducibility was calculated based on MIC values falling within $\pm 1$ dilution of the mode MIC value. There were no "off-scale" MIC results for any reading method versus inoculation method; therefore, only best case scenario is reported in Table 2 below. Reproducibility was calculated and was reported as greater than $95\%$ for all combinations of reading methods and inoculation methods. The results (Table 2) were acceptable.
Table 2. Reproducibility Results with all inoculation and read methods.
| Read Method | Prompt Inoculation | Turbidity Inoculation |
| --- | --- | --- |
| WalkAway | 453/459 (98.7%) | 455/459 (99.1%) |
| autoSCAN-4 | 447/459 (97.4%) | 451/459 (98.3%) |
| Manual | 449/459 (97.8%) | 447/459 (97.4%) |
# b. Linearity/assay reportable range:
Not Applicable
{5}
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
# Quality Control Testing:
The QC isolates recommended by both FDA and CLSI, $E.$ coli ATCC 25922, $E.$ coli ATCC 35218, $K.$ pneumoniae ATCC 700603, and $P.$ aeruginosa ATCC 27853 were tested a sufficient number of times (i.e., at least 20/site) at each of three testing sites. They were tested using both turbidity and Prompt inoculation methods for the manual read, WalkAway instrument and autoSCAN 4 instruments. The results are summarized in Table 3 below. Greater than $95\%$ of the results obtained for each inoculation method with each read method were acceptable.
Table 3. Quality Control Results - Ceftolozane/tazobactam
| Organism | Conc (μg/mL) | Ref | Prompt Inoculation Method | | | Turbidity Inoculation Method | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | Manual | Walk Away | AS4 | Manual | Walk Away | AS4 |
| E. coli ATCC 25922 | 0.06 | | - | - | - | - | - | - |
| Expected | 0.12 | 44 | 11 | 32 | 20 | 19 | 46 | 37 |
| Result: 0.12/4 - 0.5/4 μg/mL | 0.25 | 119 | 147 | 124 | 137 | 142 | 116 | 125 |
| | 0.5 | 1 | 6 | 6 | 6 | 2 | 1 | 1 |
| | 1 | - | - | - | - | 1 | 1 | 1 |
| | | | | | | | | |
| E. coli ATCC 35218Expected Result: 0.06/4 - 0.25/4 μg/mL | ≤0.03 | - | - | - | - | - | - | - |
| | 0.06 | 42 | 1 | 2 | 1 | - | - | 1 |
| | 0.12 | 112 | 142 | 151 | 144 | 153 | 160 | 157 |
| | 0.25 | 10 | 16 | 6 | 14 | 11 | 4 | 6 |
| | 0.5 | - | 3 | 2 | 2 | - | - | - |
| | 1 | - | 2 | 1 | 1 | - | - | - |
| | 2 | - | - | - | - | - | - | - |
| | | | | | | | | |
| K. pneumoniae ATCC 700603Expected Result: 0.5/4 - 2/4 μg/mL | 0.25 | - | - | - | - | - | - | - |
| | 0.5 | 10 | - | - | - | 1 | 3 | - |
| | 1 | 144 | 136 | 155 | 153 | 145 | 160 | 155 |
| | 2 | 10 | 25 | 5 | 8 | 16 | 1 | 7 |
| | 4 | - | 3 | 1 | 3 | 2 | - | 2 |
| | 8 | - | - | - | - | - | - | - |
| | | | | | | | | |
| P. aeruginosa ATCC 27853Expected Result: 0.25/4 - 1/4 μg/mL | 0.12 | - | - | - | - | - | - | - |
| | 0.25 | 9 | - | - | - | 1 | 2 | - |
| | 0.5 | 151 | 158 | 161 | 164 | 162 | 162 | 163 |
| | 1 | 4 | 6 | - | - | 1 | - | - |
| | 2 | - | - | - | - | - | - | - |
| | | | | | | | | |
# Inoculum Density Check:
A spectrophotometric device, the Beckman Coulter Turbidity Meter, was used to ensure quality control of the turbidity. The turbidity method inocula were prepared using the MicroScan Turbidity Meter with a reading of $0.08 \pm 0.02$ (equivalent to a $0.5\mathrm{McFarland}$ barium sulfate turbidity standard). The digital reading was recorded
{6}
for each organism tested. Colony counts were performed using the QC strain *E. coli* ATCC 25922 for both Prompt and Turbidity inoculation methods during QC testing.
In addition, data was also collected during the Reproducibility phase using the Prompt inoculum preparation method. Colony counts were performed weekly on the quality control strain *E. coli* ATCC 25922. Colony counts were also done once at each site for each Reproducibility isolate. All colony counts were acceptable.
## Growth Failure Rate
The growth rate for all combinations of inoculation and read methods was 100%.
## Purity Check:
Purity check plates were performed on all isolates from each inoculum preparation.
## Prompt Inoculation Validation:
An additional study was conducted to demonstrate that additional time after inoculating the Prompt would not have an impact on susceptibility testing prior to panel inoculation. Prompt bottles were inoculated with *C. koseri* and *S. liquefaciens* and tested at 0 (baseline) and 4 hours after inoculation. Panels were read by the WalkAway, AutoScan-4 and manually and compared to the reference method. When compared to baseline and reference results, the 4 hour incubation did not appear to affect *C. koseri* susceptibility results when read by any method. In addition, the 4 hour incubation did not appear to affect *S. liquefaciens* susceptibility results when compared to the reference and baseline when read manually only.
d. Detection limit:
Not Applicable
e. Analytical specificity:
Not Applicable
f. Assay cut-off:
Not Applicable
2. Comparison studies:
a. Method comparison with predicate device:
Results obtained with the MicroScan Dried Gram Negative MIC/Combo Panel for ceftolozane/tazobactam were compared to results obtained with the CLSI frozen broth
7
{7}
microdilution reference panel. The range of dilutions evaluated with the MicroScan panels was 0.25/4 to 16/4 µg/mL.
Two inoculation methods were used: the Prompt Inoculation System and the turbidity method. For the Prompt Inoculation System, inocula were prepared using the instructions of the manufacturer. For the turbidity method, inocula were standardized using the MicroScan Turbidity Meter to approximate a 0.5 McFarland turbidity standard. Reference panels were inoculated using the turbidity method. Each isolate was tested using both inoculation methods and read by the WalkAway instrument, autoSCAN-4 instrument, and read manually; however, the primary read that was identified as being used by most users was the WalkAway instrument with the Prompt inoculation.
## Clinical:
Clinical testing was conducted at three sites using 394 fresh, 162 recent, and 19 stock Gram negative organisms for a cumulative of 575 isolates. These consisted of 496 Enterobacteriaceae isolates to include E. cloacae (48 isolates), E. coli (77), S. liquefaciens (1), S. marcescens (32), C. koseri (49), K. oxytoca (47), K. pneumoniae (88), M. morganii (41), P. mirabilis (56), P. stuartii (21), P. rettgeri (19) and P. vulgaris (17) and 79 P. aeruginosa isolates. All isolates tested are listed in the Indications for Use. Of these, 23 isolates were resistant to cefotolozane/tazobactam by the reference method.
## Challenge:
Additional stock challenge isolates were tested at an external study site. These isolates consisted of organisms with known resistance mechanisms (e.g., isolates obtained from FDA/CDC AR bank) to challenge the ability of AST system to correctly identify the susceptibility category. Challenge testing was conducted using 118 Gram negative organisms, consisting of 20 P. aeruginosa isolates, and 98 Enterobacteriaceae including E. cloacae (13 isolates), E. coli (20), S. liquefaciens (10), S. marcescens (4), C. koseri (13), K. oxytoca (8), K. pneumoniae (18), M. morganii (3), P. mirabilis (7), P. stuartii (1), and P. vulgaris (1). Of these 118 Gram negative organisms, 40 were resistant to ceftolozane/tazobactam by the reference method.
594 Enterobacteriaceae and 99 P. aeruginosa isolates (Clinical and Challenge) isolates were evaluated. Tables 3, 4, 5, and 6 below illustrate the performance of testing Ceftolozane/tazobactam on the WalkAway, autoSCAN-4, and manual reads using the Prompt and turbidity inoculation methods, respectively.
8
{8}
Table 3. Combined (Clinical and Challenge) Performance Summary of Gram Negative Panels with Enterobacteriaceae – Prompt Inoculation Method, All Read Methods
| Panel Reading Method | Tot | EA N | %EA Total | Total Eval | EA Eval | %EA Eval | CA N | % CA | #R | min (%) | maj (%) | vmj (%) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| WalkAway | 594 | 545 | 91.8 | 93 | 68 | 73.1 | 563 | 94.8 | 55 | 10 (1.7%) | 19 (3.6%) | 2 (3.6%) |
| autoSCAN-4 | 594 | 561 | 94.4 | 97 | 78 | 80.4 | 576 | 97 | 55 | 9 (1.5%) | 7 (1.3%) | 2 (3.6%) |
| Manual | 594 | 565 | 95.1 | 105 | 92 | 87.6 | 577 | 97.1 | 55 | 10 (1.7%) | 6 (1.1%) | 1 (1.8%) |
Table 4. Combined (Clinical and Challenge) Performance Summary of Gram Negative Panels with P. aeruginosa – Prompt Inoculation Method, All Read Methods
| Panel Reading Method | Tot | EA N | %EA Total | Total Eval | EA Eval | %EA Eval | CA N | % CA | #R | min (%) | maj (%) | vmj (%) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| WalkAway | 99 | 98 | 99 | 73 | 72 | 98.6 | 97 | 98 | 8 | 2 (2%) | 0 | 0 |
| autoSCAN-4 | 99 | 97 | 98 | 74 | 72 | 97.3 | 97 | 98 | 8 | 2 (2%) | 0 | 0 |
| Manual | 99 | 95 | 96 | 74 | 70 | 94.6 | 95 | 96 | 8 | 4 (4%) | 0 | 0 |
Table 5. Combined (Clinical and Challenge) Performance Summary of Gram Negative Panels with Enterobacteriaceae – Turbidity Inoculation Method, All Read Methods
| Panel Reading Method | Tot | EA N | %EA Total | Total Eval | EA Eval | %EA Eval | CA N | % CA | #R | min (%) | maj (%) | vmj (%) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| WalkAway | 594 | 569 | 95.8 | 96 | 76 | 79.2 | 581 | 97.8 | 55 | 10 (1.7%) | 2 (0.4%) | 1 (1.8%) |
| autoSCAN-4 | 594 | 568 | 95.6 | 102 | 83 | 81.4 | 583 | 98.1 | 55 | 7 (1.2%) | 3 (0.6%) | 1 (1.8%) |
| Manual | 594 | 573 | 96.5 | 116 | 102 | 87.9 | 583 | 98.1 | 55 | 5 (0.8%) | 6 (1.1%) | 0 |
Table 6. Combined (Clinical and Challenge) Performance Summary of Gram Negative Panels with P. aeruginosa – Turbidity Inoculation Method, All Read Methods
| Panel Reading Method | Tot | EA N | %EA Total | Total Eval | EA Eval | %EA Eval | CA N | % CA | #R | min (%) | maj (%) | vmj (%) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| WalkAway | 99 | 99 | 100 | 70 | 70 | 100 | 97 | 98 | 8 | 2 (2%) | 0 | 0 |
| autoSCAN-4 | 99 | 94 | 94.9 | 73 | 68 | 93.2 | 97 | 98 | 8 | 2 (2%) | 0 | 0 |
| Manual | 99 | 96 | 97 | 74 | 71 | 95.9 | 96 | 97 | 8 | 3 (3%) | 0 | 0 |
EA - Essential Agreement
CA - Category Agreement
R - resistant isolates
maj - major discrepancies
vmj - very major discrepancies
min - minor discrepancies
Evaluable results are those that are on-scale for both the reference panel and the MicroScan Gram negative panel. Essential agreement (EA) occurs when there is agreement between the MIC result of the reference method and that of MicroScan panel within plus or minus one serial two-fold dilution of the antibiotic. Category agreement occurs when the interpretation of the result of the reference method agrees exactly with the interpretation of the MicroScan panel result.
As seen in Tables 3, 4, 5, and 6, the percent EA overall for combined clinical and challenge isolates was above 90% for all read methods and inoculation methods with the exception of some organisms within the Enterobacteriaceae family when inoculated with the Prompt. Specifically, E. cloacae demonstrated a performance of 83.6% (51/61) and 86.9% (53/61) when read by the WalkAway and AS-4 instruments but when read manually, the EA increased to 90.2% (55/61). Therefore, for this organism, the following limitation is included in the labeling:
{9}
Results obtained with *E. cloacae* and ceftolozane/tazobactam have shown discrepant MICs when compared with the reference method when read by the MicroScan instruments. Therefore, prior to reporting, these isolate results should be interpreted manually.
*M. morganii*, *P. rettgeri*, and *S. liquefaciens* isolates also demonstrated less than 90% EA [82.0% (50/61), 77.3% (34/44), 78.9% (15/19), and 36.4% (4/11) respectively] when read by the WalkAway and inoculated using the Prompt. When the same plates were reviewed manually, the EA when testing *P. rettgeri* and *S. liquefaciens* isolates surpassed 90% [100% (19/19) and 100% (11/11) respectively]. Therefore, for these two isolates, the following limitation is included in the labeling:
Results obtained *P. rettgeri* and *S. liquefaciens* and ceftolozane/tazobactam have shown discrepant MICs when compared with the reference method when read by the WalkAway instrument and inoculated using the Prompt. Therefore, prior to reporting, these isolate results should be interpreted manually.
In addition, isolates of *M. morganii* demonstrated high major error (MAJ) rates when inoculated with the Prompt and read by the WalkAway [15.9% (7/44)]. When these isolates were retested using the turbidity method, they were able to meet 90% EA with the WalkAway [95.5% (42/44)] and demonstrated no MAJ errors (0%). Therefore, the following limitation is included in the labeling:
Ceftolozane/tazobactam results obtained with *M. morganii* isolates using the Prompt Inoculation system have shown potential major errors. Therefore, prior to reporting, isolates with resistant results should be retested using the turbidity inoculation method.
The MAJ error rate for *Enterobacteriaceae* with the WalkAway and Prompt is 3.6%; however, when *M. morganii* isolates were retested using the turbidity method, the MAJ error rate decreased to 2.3% which is acceptable. In addition, the VMJ error rate decreased from 3.6% to 1.8% when *P. rettgeri* isolates were read manually and therefore, acceptable.
As seen in Tables 3, 4, 5, and 6, the percent CA overall for combined clinical and challenge isolates was above 90%, and is therefore acceptable as described in the Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA, August 2009. However, when interpreted by the WalkAway and inoculated with the Prompt, *M. morganii*, *P. rettgeri*, and *S. liquefaciens* isolates demonstrated lower than adequate CA performance when analyzed separately [i.e., 81.8% (36/44), 84.2% (16/19), and 45.5% (5/11) respectively] and yielded high major error rates as mentioned above. The CA rates significantly improved when *M. morganii* isolates were repeated with the turbidity inoculation (100%), and when *P. rettgeri* (100%) and *S. liquefaciens* (100%) isolates were interpreted manually.
10
{10}
MIC Trends:
Table 7. Summary of Evaluation of MIC Trends (combined clinical and challenge data)
| Ceftolozane/ tazobactam | Total | ≥1 dil lower | Exact | ≥1 dil higher |
| --- | --- | --- | --- | --- |
| WalkAway/Prompt | | | | |
| Enterobacteriaceaea | 576 | 100 (17.36%) | 299 (51.96%) | 177 (30.73%) |
| P. aeruginosa | 98 | 14 (14.29%) | 75 (76.53%) | 9 (9.18%) |
aDifference between the higher and lower dilutions is 13.37%; 95% C.I. (8.46% to 18.20%)
As illustrated in Table 7, there was a difference in in results trending above the reference for Enterobacteriaceae using the WalkAway read method and Prompt inoculation method. The data demonstrates a trend for one doubling dilution higher for these results compared to the reference method and therefore, the following is included as a footnote in the labeling:
Ceftolozane/tazobactam MIC values for Enterobacteriaceae isolates tended to be one doubling dilution higher when read by the WalkAway and using the Prompt inoculation method.
b. Matrix comparison:
N/A
3. Clinical studies:
a. Clinical Sensitivity:
N/A
b. Clinical specificity:
N/A
c. Other clinical supportive data (when a. and b. are not applicable):
N/A
4. Clinical cut-off:
N/A
5. Expected values/Reference range:
The FDA susceptibility interpretive criteria for Ceftolozane/tazobactam are as listed in Table 8.
{11}
Table 8. Interpretive Criteria for Ceftolozane/tazobactam (μg/mL)
| | Susceptible (S) | Intermediate (I) | Resistant (R) |
| --- | --- | --- | --- |
| Enterobacteriaceae | ≤2/4 | 4/4 | ≥8/4 |
| P. aeruginosa | ≤4/4 | 8/4 | ≥16/4 |
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.