The EUROIMMUN IFA: Crithidia luciliae (anti-dsDNA) EUROPattern test kit is intended for the quantitative determination of human antibodies of immunoglobulin class IgG against anti-double stranded DNA (dsDNA) in human serum with the EUROPattern Microscope and Software automated instrument. It is used as an aid in the diagnosis of systemic lupus erythematosus (SLE), in conjunction with other laboratory and clinical findings. All suggested results obtained with the EUROPattern system must be confirmed by trained personnel.
Device Story
The device is an indirect immunofluorescence (IFA) assay using Crithidia luciliae substrate to detect anti-dsDNA antibodies in human serum. The test utilizes BIOCHIP slides containing Crithidia luciliae smears; patient samples are incubated on these slides, followed by incubation with fluorescein-labelled anti-human IgG conjugate. The EUROPattern Microscope and Software system captures fluorescence images of the kinetoplast. The software provides automated evaluation of fluorescence intensity and titer, which the operator must verify. The device is used in clinical laboratories to assist in SLE diagnosis. By identifying specific kinetoplast fluorescence, the system provides qualitative or semi-quantitative results that support clinical decision-making when combined with other laboratory and clinical findings.
Clinical Evidence
Clinical study evaluated 364 samples from patients with SLE and various differential diagnoses (AAV, PM/DM, SSc, SS, MCTD, RA, APS, DIL, AIH, AITD, HCV, CP, EBV). Sensitivity and specificity were assessed across three modes (A: automated, B: automated imaging/manual reading, C: manual). Overall sensitivity for SLE was 33.0% (Mode C) to 38.8% (Mode A), and specificity for non-SLE samples was 91.6% (Mode A) to 97.0% (Mode C). Analytical studies confirmed reproducibility and precision across sites and operators.
Technological Characteristics
The system comprises an automated fluorescence microscope and image analysis software. It uses the Crithidia luciliae substrate for IFA. The device is designed for the detection of IgG class antibodies. It operates as a networked or standalone automated workstation. The software performs image acquisition and pattern recognition to assist in the identification of anti-dsDNA antibodies.
Indications for Use
Indicated for the quantitative determination of IgG anti-dsDNA antibodies in human serum to aid in the diagnosis of systemic lupus erythematosus (SLE). For prescription use only. Results require confirmation by trained personnel.
Regulatory Classification
Identification
An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION MEMORANDUM
A. 510(k) Number:
K172244
B. Purpose for Submission:
New Device on a previously cleared instrument
C. Measurand:
Anti-double stranded DNA (ds DNA) antibodies
D. Type of Test:
Qualitative or semi-quantitative
E. Applicant:
EUROIMMUN US Inc
F. Proprietary and Established Names:
EUROIMMUN IFA: Crithidia luciliae (anti-dsDNA) EUROPattern
G. Regulatory Information:
1. Regulation section:
21 CFR 866.5100, Antinuclear antibody immunological test system
21 CFR 866.4750, Automated indirect immunofluorescence microscope and software-assisted system for clinical use
2. Classification:
Class II
3. Product code:
KTL, anti-DNA indirect immunofluorescent solid phase
PIV, automated indirect immunofluorescence microscope and software-assisted system for clinical use
4. Panel:
82, Immunology
H. Intended Use:
1. Intended use(s):
The EUROIMMUN IFA: Crithidia luciliae (anti-dsDNA) EUROPattern test kit is intended for the qualitative and semi-quantitative determination of human antibodies of immuno-globulin class IgG against anti-double stranded DNA (dsDNA) in human serum with the EUROPattern Microscope and Software automated instrument. It is used as an aid in the diagnosis of systemic lupus erythematosus
{1}
(SLE), in conjunction with other laboratory and clinical findings. All suggested results obtained with the EUROPattern system must be confirmed by trained personnel.
2. Indication(s) for use:
Same as intended use
3. Special conditions for use statement(s):
1. For prescription use only
2. This device is only for use with reagents that are indicated for use with the device.
3. The device is for use by a trained operator in a clinical laboratory setting.
4. All software-aided results must be confirmed by the trained operator.
5. Special instrument requirements: for use only with EUROPattern Microscope and Software (cleared in K141827)
I. Device Description:
Materials provided. The kit contains enough for 50 determinations:
- Slides, each containing 5 BIOCHIPs coated with a smear of Crithidia luciliae
- Conjugate, Fluorescein-labelled anti-human IgG (goat), ready for use with Evans Blue
- Positive control: autoantibodies against dsDNA, human, ready for use
- Negative control: autoantibody-negative, human, ready for use
- Salt for PBS pH 7.2
- Tween 20
- Mounting medium, glycerol, ready for use
- Cover glasses (62 mm x 23 mm)
- Instruction booklet
Needed and not provided:
- Reagent trays for slides containing up to 10 fields (order #ZZ 9999-0110)
- Fluorescence microscope (EUROPattern microscope & software): Equipped with a 450-490 nm excitation filter; 510 nm color separator & 515-565 nm bandpass filter with LED Bluelight
- Distilled or de-ionized water for wash buffer production
- Pipettes with a range of 10 µl to 200 µl
- Cuvettes or wash/staining dishes for PBS wash step
- Lint free towelling
J. Substantial Equivalence Information:
1. Predicate device name(s):
Quanta Lite® dsDNA ELISA
2. Predicate 510(k) number(s):
K903898
{2}
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device EUROIMMUN IFA: Crithidia luciliae (anti-dsDNA) EUROPattern | Predicate QUANTA Lite dsDNA ELISA (K903898) |
| Intended Use | For the qualitative or semi-quantitative determination of human antibodies of immuno-globulin class IgG against anti-double stranded DNA (dsDNA) in human serum with the EUROPattern Microscope and Software automated instrument. It is used as an aid in the diagnosis of systemic lupus erythematosus (SLE), in conjunction with other laboratory and clinical findings. | Enzyme-linked immunosorbent assay (ELISA) for the quantitative detection of ds DNA autoantibodies in human serum. The presence of antibodies to ds DNA can be used in conjunction with clinical findings and other laboratory tests to aid in the diagnosis of Systemic Lupus Erithematosys (SLE) |
| Assay format | Qualitative or semi-quantitative | Semi-quantitative |
| Controls | 1 Positive control 1 Negative control | Same |
| Sample matrix | Serum | Same |
| Procedure | Standard immunoassay: Serum incubation with antigen, followed by a wash step, incubation with labelled anti-human globulin, wash step, measurement/evaluation | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Technology | IFA | ELISA |
| Assay platform | BIOCHIP TITERPLANE technology | 96-well microtiter plates |
| Antigen | Crithidia luciliae cells | Purified dsDNA antigen from calf thymus |
| Conjugate | Fluorescein labeled goat anti-human IgG | Horseraddish peroxidase, TMB chromogen reaction with 0.5 M sulphuric acid stopping |
{3}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Calibrators | None | 1 calibrator |
| Sample dilution | 1:10 | 1:101 |
| Measurement/ evaluation | Visual evaluation of fluorescence under microscope, IFA EUROPattern automatic evaluation with user verification | Photometric reading |
| Cutoff | Negative at <1:10 dilution
Positive at 1:10 dilution | Sample OD / Calibrator OD
0-200 IU/ml: negative
201-300 IU/ml: equivocal
301-800 IU/ml: moderate
positive ≥800 IU/ml: strong
positive |
| Reported results | Qualitative, Titer | Qualitative, IU/ml |
## K. Standard/Guidance Document Referenced:
N/A
## L. Test Principle:
For the detection of autoantibodies against native, double-stranded DNA (dsDNA, nDNA) by indirect immunofluorescence, EUROIMMUN uses as the assay substrate the Crithidia luciliae species. This protozoon possesses a giant mitochondrium containing double-stranded DNA (kinetoplast) which shows none of the remaining antigens contained in human cell nuclei. Antinuclear antibodies reacting with the kinetoplast are directed against dsDNA antigens.
Patient samples, controls and in separate steps conjugate and mounting medium are applied to the reaction fields of a reagent tray. The BIOCHIP slides are then placed into the recesses of the reagent tray, where all BIOCHIPs of the slide come into contact with the fluids, and the individual reactions commence simultaneously. The fluids are confined to the recessed wells eliminating the need to use a conventional "humidity chamber".
Patient samples are diluted 1:10 in PBS-Tween, 30 μl of each diluted patient sample are added to each reaction field of the reagent tray. Reactions are started by fitting the BIOCHIP slides containing the sections from the substrate (Crithidia luciliae smears) into the corresponding recesses of the reagent tray and incubated for 30 minutes at room temperature. Specific antibodies attach to the Crithidia luciliae antigens. After incubation the BIOCHIP slides are washed with PBS-Tween to remove unbound antibodies. 25 μl of fluorescein-labelled anti-human globulin are added to each reaction field of a clean reagent tray and the BIOCHIP slides placed into the recesses of the tray. After a 30 minutes incubation at room temperature, the BIOCHIPs are again washed with PBS-Tween to remove any unbound fluorescein-labelled reagent. 10 μl of mounting medium are placed for each reaction field on a cover glass and the BIOCHIP slides, with the
{4}
BIOCHIPs facing downwards, placed onto the prepared cover glass. Fluorescence is read by EUROPattern fluorescence microscopy.
# M. Performance Characteristics:
# 1. Analytical performance:
Modes of comparison:
| Mode | Imaging | Reading |
| --- | --- | --- |
| A | Automated | Automated |
| B | Automated | Manual, images on the computer monitor |
| C | Manual | Manual, on the microscope |
Results of the analytical and clinical studies are given separately for the three modes (A, B and C) to demonstrate automated performance using the EUROPattern microscope and software compared to manual performance. For Mode B and C, manufacturer's predetermined acceptance criteria for performance characteristics were met. Trained operators must confirm all assessments made by the EUROPatter microscope and software.
# Nomenclature:
CLIFT - Crithidia lucilae indirect immunofluorescence test
IIFT - Indirect immunofluorescence test
IF - Immunofluorescence
FI - Fluorescence Intensity
Titer:
| Titer | Consideration |
| --- | --- |
| Negative | Less than 1:10 |
| Low positive | 1:10, 1:20, 1:40, 1:80 |
| Medium positive | 1:160, 1:320, 1:640 |
| High positive | 1:1280, 1:2560, 1:5120 |
# Interpretation of results:
For a positive result, a distinct, homogeneous, in parts circular fluorescence of the kinetoplast can be identified, the same pattern is found as for the positive control. For negative samples, the kinetoplast shows no staining. If the positive control shows no specific fluorescence pattern or the negative control shows a clear specific fluorescence, the results are not to be used and the test is to be repeated.
Qualitative evaluation
| Anti-dsDNA reactivity (IgG) | Evaluation |
| --- | --- |
| No reaction at 1:10 | Negative. No antibodies against dsDNA detected in the patient sample. |
| Positive reaction at 1:10 | Positive. Indication of antibodies to dsDNA in the patient sample. |
{5}
Semi-quantitative evaluation: The endpoint titer is defined as the highest sample dilution factor for which specific fluorescence is identifiable. For a semi-quantitative evaluation all dsDNA positive samples should be reported with endpoint titers, depending on the user established titering protocol. It is stated in the labeling: "For diagnosis, the clinical symptoms of the patient should always be taken into account."
# a. Precision/Reproducibility:
Repeatability/Reproducibility was investigated using a panel of serum samples representing the measurement range (range of patterns and titers) in a laboratory setting at multiple sites. Titer agreement and percent positive agreement and percent negative agreements were evaluated and summarized. The following samples were tested:
| Sample | Titer |
| --- | --- |
| Negative | <10 |
| Low positive | 1:40 |
| Medium positive | 1:160 |
| High positive | 1:2560 |
# Repeatability:
Repeatability was determined by repeated measurements using four saples on five different days with two runs per day and two replicates per run (80 reads per mode). The IFA: Crithidia luciliae (anti-dsDNA) EUROPattern assays were processed according to the package insert. Each result was reported in three modes (A, B, C). Positive samples were not found negative and vice versa.
Within mode agreement
| Expected result | Mode A | Mode B | Mode C |
| --- | --- | --- | --- |
| | % titer agreement (within ±1) | | |
| Negative | 100 | 100 | 100 |
| 1:40 | 100 | 100 | 95 |
| 1:160 | 85 | 90 | 90 |
| 1:2500 | 100 | 100 | 100 |
Mode to mode agreement
| N=80 | Mode A/B | Mode A/C | Mode B/C |
| --- | --- | --- | --- |
| % positive agreement (95% CI) | 100 (94.0–100.0) | 100 (94.0–100.0) | 100 (94.0–100.0) |
| % negative agreement (95% CI) | 100 (83.2–100.0) | 100 (83.2–100.0) | 100 (83.2–100.0) |
| % overall agreement (95% CI) | 100 (95.5–100.0) | 100 (95.5–100.0) | 100 (95.5–100.0) |
| % titer agreement (95% CI) | 96.3 (89.4–99.2) | 87.5 (78.2–93.8) | 97.5 (91.3–99.7) |
{6}
Reproducibility:
Reproducibility was evaluated with the same four samples as repeatability and was determined by using four samples on five days with two runs per day and two replicates per run, performed at three different sites (60 replicates per sample and 240 reads per mode), at EUROIMMUN's laboratory and two external US clinical laboratories. The IFA: Crithidia luciliae (anti-dsDNA) EUROPattern assays were processed according to the package insert and evaluated by three different EUROPattern microscope and software systems. Each result was reported in three modes (A, B, C). Positive samples were not found negative and vice versa for mode B and C. The results are summarized in the tables below:
Agreement with expected results within mode
| Sample N=60 | Site | Mode A | | Mode B | | Mode C | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | % positive agreement | % titer agreement within±1 | % positive agreement | % titer agreement within±1 | % positive agreement | % titer agreement within±1 |
| Negati ve | 1 | 0 | 100 | 0 | 100 | 0 | 100 |
| | 2 | 5 | 95 | 0 | 100 | 0 | 100 |
| | 3 | 25 | 75 | 0 | 100 | 0 | 100 |
| 1:40 | 1 | 100 | 100 | 100 | 100 | 100 | 95 |
| | 2 | 100 | 100 | 100 | 100 | 100 | 100 |
| | 3 | 100 | 100 | 100 | 100 | 100 | 100 |
| 1:160 | 1 | 100 | 85 | 100 | 90 | 100 | 90 |
| | 2 | 95 | 85 | 100 | 100 | 100 | 100 |
| | 3 | 100 | 80 | 100 | 100 | 100 | 100 |
| 1:2560 | 1 | 100 | 100 | 100 | 100 | 100 | 100 |
| | 2 | 100 | 95 | 100 | 100 | 100 | 100 |
| | 3 | 100 | 100 | 100 | 100 | 100 | 100 |
Mode to mode agreement at each site
Mode to mode agreement is summarized in the tables below
Site 1 mode to mode agreement
| Site 1, N=240 | A/B | A/C | B/C |
| --- | --- | --- | --- |
| % positive agreement (95% CI) | 100 (94.0–100) | 100 (94.0–100) | 100 (94.0–100) |
| % negative agreement (95% CI) | 100 (83.2–100) | 100 (83.2–100) | 100 (83.2–100) |
| % overall agreement (95% CI) | 100 (95.5–100) | 100 (95.5–100) | 100 (95.5–100) |
| % titer agreement (95% CI) | 96.3 (89.4–99.2) | 87.5 (78.2–93.8) | 93.8 (86.0–97.9) |
{7}
Site 2 mode-to-mode agreement
| Site 2, N=240 | A/B | A/C | B/C |
| --- | --- | --- | --- |
| % positive agreement (95% CI) | 98.3 (91.1–100) | 98.3 (91.1–100) | 100 (94.0–100) |
| % negative agreement (95% CI) | 95.0 (75.1–99.9) | 95.0 (75.1–99.9) | 100 (83.2–100) |
| % overall agreement (95% CI) | 97.5 (91.3–99.7) | 97.5 (91.3–99.7) | 100 (95.5–100) |
| % titer agreement (95% CI) | 93.8.0 (86.0–97.9) | 92.5 (84.4–97.2) | 100 (95.5–100) |
Site 3 mode-to-mode agreement
| Site 3, N=240 | A/B | A/C | B/C |
| --- | --- | --- | --- |
| % positive agreement (95% CI) | 100 (94.0–100) | 100 (94.0–100) | 100 (94.0–100) |
| % negative agreement (95% CI) | 75.0 (50.9–91.3) | 75.0 (50.9–91.3) | 100 (83.2–100) |
| % overall agreement (95% CI) | 93.8 (96.0–97.9) | 93.8 (96.0–97.9) | 100 (95.5–100) |
| % titer agreement (95% CI) | 90.0 (81.2–95.6) | 91.3 (82.8–96.4) | 98.8 (93.2–100) |
Summary, mode to mode overall agreement at each site
| N=240 | Overall agreement | A/B | A/C | B/C |
| --- | --- | --- | --- | --- |
| Site 1 | % overall agreement (95% CI) | 100 (95.5–100) | 100 (95.5–100) | 100 (95.5–100) |
| | % titer agreement (95% CI) | 96.3 (89.4–99.2) | 87.5 (78.2–93.8) | 93.8 (86.0–97.9) |
| Site 2 | % overall agreement (95% CI) | 97.5 (91.3–99.7) | 97.5 (91.3–99.7) | 100 (95.5–100) |
| | % titer agreement (95% CI) | 93.8.0 (86.0–97.9) | 92.5 (84.4–97.2) | 100 (95.5–100) |
| Site 3 | % overall agreement (95% CI) | 93.8 (96.0–97.9) | 93.8 (96.0–97.9) | 100 (95.5–100) |
| | % titer agreement (95% CI) | 90.0 (81.2–95.6) | 91.3 (82.8–96.4) | 98.8 (93.2–100) |
| Total | % overall agreement (95% CI) | 97.1 (94.1–98.8) | 97.1 (94.1–98.8) | 100 (98.5–100) |
| | % titer agreement (95% CI) | 93.3 (89.4–96.1) | 90.4 (86.0–93.8) | 98.8 (96.4–99.7) |
{8}
Site to site agreement for each mode
Site to site agreement is summarized in the tables below for each mode. Study was evaluated using four samples on five days with two runs per day and two replicates per run, (80 reads per site).
Mode A, site to site comparison
| Mode A, N=80 | Site 1 vs. 2 | Site 1 vs. 3 | Site 2 vs. 3 |
| --- | --- | --- | --- |
| % positive agreement (95% CI) | 98.3 (91.1–100) | 92.3 (83.0–97.5) | 92.3 (83.0–97.5) |
| % negative agreement (95% CI) | 95.0 (75.1–99.1) | 100 (78.2–100) | 100 (78.2–100) |
| % overall agreement (95% CI) | 97.5 (91.3–99.7) | 93.8 (86.0–97.9) | 93.8 (86.0–97.9) |
| % titer agreement (95% CI) | 90.0 (81.2–95.6) | 85.0 (75.3–92.0) | 90.0 (81.2–95.6) |
Mode B, site to site comparison
| Mode B, N=80 | Site 1 vs. 2 | Site 1 vs. 3 | Site 2 vs. 3 |
| --- | --- | --- | --- |
| % positive agreement (95% CI) | 100
94.0–100 | 100
94.0–100 | 100
94.0–100 |
| % negative agreement (95% CI) | 100.0
83.2–100 | 100
83.2–100 | 100
83.2–100 |
| % overall agreement (95% CI) | 100
95.5–100 | 100
92.5–100 | 93.8
95.5–100 |
| % titer agreement (95% CI) | 95.0
87.7–98.6 | 92.5
84.4–97.2 | 97.5
91.3–99.7 |
Mode C, site to site comparison
| Mode C, N=80 | Site 1 vs. 2 | Site 1 vs. 3 | Site 2 vs. 3 |
| --- | --- | --- | --- |
| % positive agreement (95% CI) | 100 (94.0–100) | 100 (94.0–100) | 100 (94.0–100.0) |
| % negative agreement (95% CI) | 100 (83.2–100) | 100 (83.2–100) | 100 (83.2–100.0) |
| % overall agreement (95% CI) | 100 (95.5–100.0) | 100 (95.5–100) | 100 (95.5–100) |
| % titer agreement (95% CI) | 95.0 (87.7–98.6) | 91.3 (82.8–96.4) | 93.8 (86.0–97.9) |
{9}
Summary: site to site agreement for each mode
| Mode N=80 | % Agreement | Site 1 vs. 2 | Site 1 vs. 3 | Site 2 vs. 3 |
| --- | --- | --- | --- | --- |
| Mode A | % overall agreement (95% CI) | 97.5 (91.3–99.7) | 93.8 (86.0–97.9) | 93.8 (86.0–97.9) |
| | % titer agreement (95% CI) | 90.0 (81.2–95.6) | 85.0 (75.3–92.0) | 90.0 (81.2–95.6) |
| Mode B | % overall agreement (95% CI) | 100 (95.5–100) | 100 (92.5–100) | 93.8 (95.5–100) |
| | % titer agreement (95% CI) | 95.0 (87.7–98.6) | 92.5 (84.4–97.2) | 97.5 (91.3–99.7) |
| Mode C | % overall agreement (95% CI) | 100 (95.5–100) | 100 (95.5–100) | 100 (95.5–100) |
| | % titer agreement (95% CI) | 95.0 (87.7–98.6) | 91.3 (82.8–96.4) | 93.8 (86.0–97.9) |
# Operator to Operator Comparison
An additional study was performed to evaluate between operator precision by repeated measurements using three samples on five days with two runs per day and two replicates per run (60 replicates per operator). The assays were processed according to the package insert. Each result was reported in two modes (B and C). Each mode B and each mode C reading was performed by two different technicians independently. Acceptance criteria were met; positive samples were not found negative and vice versa. The samples used in the study are summarized in the table below:
| Sample | Titer |
| --- | --- |
| Negative | <10 |
| Low positive | 1:40 |
| Medium positive | 1:320 |
Operator to operator agreement with expected results within mode
| N=60 | Mode B Two operators | Mode C Two operators |
| --- | --- | --- |
| % positive agreement (95% CI) | 100 (91.2–100) | 100 (91.2–100) |
| % negative agreement (95% CI) | 100 (83.2–100) | 100 (83.2–100) |
| % overall agreement (95% CI) | 100 (94.0–100) | 100 (94.0–100) |
| % titer agreement (95% CI) | 98.3 (91.1–100) | 87.5 (94.0–100) |
{10}
# b. Linearity/assay reportable range:
To investigate linearity, three positive samples were serially diluted and assayed with The IFA: Crithidia luciliae (anti-dsDNA) EUROPattern system. Assays were processed according to the package insert and evaluated by the EUROPattern microscope and software. Individual results per dilution were read visually and reported using a fluorescence intensity scale from "0" (negative) to "5" (high positive). The EUROPattern software creates one result based on all dilutions, it does not allow individual reading of single dilutions. The fluorescence intensities (FI) decreased with dilutions until negative. The results are shown in the table below.
Three sample were used in the study:
| Sample No. | Sample | End titer results erstimate per sample | | |
| --- | --- | --- | --- | --- |
| | | Mode A | Mode B | Mode C |
| 1 | High positive | 1:2560 | 1:5120 | 1:5120 |
| 2 | Medium positive | 1:80 | 1:160 | 1:160 |
| 3 | Low positive | 1:20 | 1:20 | 1:20 |
The dilution fluorescence intensity results (FI) are summarized in the table below:
| Sample No. | Result per dilution (FI level) Mode C | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | 1:10 | 1:20 | 1:40 | 1:80 | 1:160 | 1:320 | 1:640 | 1:1280 | 1:2560 | 1:5120 |
| 1 | 5 | 5 | 5 | 5 | 5 | 5 | 5 | 3 | 2 | 1 |
| 2 | 5 | 5 | 4 | 2 | 1 | 0 | 0 | 0 | 0 | 0 |
| 3 | 2 | 1 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
# c. Traceability, Stability, Expected values (controls, calibrators, or methods):
# Traceability:
A recognized standard or reference material for anti-dsDNA antibodies is not available.
# Stability:
Stability studies were conducted in accordance with the international standard DIN EN 13640 / DIN EN ISO 23640: Stability testing of in vitro diagnostic reagents. Six positive and four negative samples were tested in an accelerated stability study. The data supports a stability claim of 18 months after the date of manufacturer if stored properly. The slides and reagents should be stored at $+2^{\circ}\mathrm{C}$ to $+8^{\circ}\mathrm{C}$ . The tests included three different kit lots and were performed according to the package insert with imaging and reading both being done manually (i.e. mode C). Kits must not be used beyond the expiration date noted on the kit label. After initial opeing, the reagents are stable until the expiry date when stored at $+2^{\circ}\mathrm{C}$ to $+8^{\circ}\mathrm{C}$ and protected from contamination. Real time stability was tested on three lots at 0, 12 months, 24 months and 36 months. The stability claims are summarized in the table below:
| | Temperature | Stability claim |
| --- | --- | --- |
| Shelf life | +2°C to +8°C | 18 months |
| Open | +2°C to +8°C | 18 months |
| Shipping | 37°C | ≤14 days |
{11}
12
Control:
Negative and positive controls are included in the kit, ready for use. The positive controls are ready to use human serum that contain anti-dsDNA antibodies that exhibit a distinct, homogeneous, in parts circular fluorescence of the kinetoplast. There is no target titer assigned to the positive controls. The negative control contains ready-to-use autoantibody-negative human sera and should exhibit a negative result. The controls are derived from native samples that are diluted with stabilizing buffer to reach the appropriate ready-to-use concentration. The reactivity of the positive and negative controls is assured by functional and stability testing. The manufacturer recommends using the positive and negative controls as stated within the labeling (Instructions for Use).
d. Detection limit:
Not applicable
e. Analytical specificity:
Cross reactivity:
The specificity of the IFA: Crithidia luciliae (anti-dsDNA) EUROPattern was verified using the ANA reference panel and ANCA (PR3, MPO) positive reference samples obtained from the CDC Centers for Disease Control and Prevention, Atlanta, USA (16 samples). The samples were assayed according to the package insert and evaluated by the EUROPattern microscope and software. Each result was reported in three modes (A, B, C). The CDC sample specific for nDNA were positive by the assay; all other samples were negative (for mode B or C); RNP, SP and Ribosomal P-protein were positive for Mode A. The other samples represent other ANA-type antigens such as Scl-70, SS-A, SS-B, Jo-1, etc.
In addition, the World Health Organization (WHO) Reference Reagent Wo/80 for anti-dsDNA was evaluated by The IFA: Crithidia luciliae (anti-dsDNA) EUROPattern. The WHO sample specific for dsDNA was positive by the assay.
Additional information on cross-reactivity can be seen in the clinical study differential diagnosis samples that were evaluated for specificity, please see below table in Sensitivity/specificity section, Summary: Sensitivity/Specificity, Correlation with Clinical Diagnosis
Interference:
The effect of interfering substances on assay results were tested by spiking normal and positive serum samples with endogenous serum components and drugs. The samples consisted of negative (< 1:10), weak positive, medium positive, and strong positives samples. The spiked samples were processed according to the package insert and evaluated by the EUROPattern microscope and software. Each result was reported in three modes (A, B, C). For Modes B and C the deviation in titer level did not exceed +/- 1 titer within modes and positive samples were not found negative and vice versa. In the triglycerides study the results for one sample were in a range of four titer levels between mode A (at 1:160) and modes B and C (at 1:20). The summary results are shown in the
{12}
tables below. Rheumatoid Factor (RF) was tested and interfering effect was obtained with RF IgG (at 61 U/ml) and with RF IgM (at 1200 U/mL), so interference occurs with RF and with RF IgM. No significant interference was observed up to the levels indicated in the table below:
| | Substance | No interference up to |
| --- | --- | --- |
| Endogenous | Hemoglobin | 500 mg/dL |
| | Bilirubin | 40 mg/dL |
| | Triglyceride | 2000 mg/dL |
| | Cyclic Citrullinated Peptide (CCP) | Ratio 8.7 (sample over calibrator) |
| Drugs | Cyclophosphamide monohydrate | 14.4 nmol/L |
| | Methotrexate hydrate | 9.1 mg/mL |
| | Azathioprine | 29.0 mg/L |
| | Mycophenolate mofetil | 350 mg/L |
| | Prednisone | 3.0 mg/L |
| | Hydroxychloroquine sulfate | 33.0 μg/mL |
| | Ibuprofen | 5.0 mg/mL |
| | Naproxen sodium | 21.7 mmol/L |
| | Rituximab | 1.1 mg/mL |
| | Belimumab | 1.1 mg/mL |
f. Assay cut-off:
The recommended starting dilution, above which the result is reported as positive and below which the result is reported as negative, is 1:10. The manufacturer suggests performing two-fold dilutions and also recommends that each laboratory establish its own titering protocol. Assay cut-off of 1:10 was determined from Renato Tozzoli, MD, Nicola Bizzaro, MD, Elio Tonutti, MD, Danilo Villalta, MD, Danila Bassetti, MD, Fabio Manoni, MD, Anna Piazza, MD, Marco Pradella, MD, and Paolo Rizzotti, MD. Guidelines for the Laboratory Use of Autoantibody Tests in the Diagnosis and Monitoring of Autoimmune Rheumatic Diseases. Am J Clin Pathol 2002;117:316 - 324.
{13}
14
2. Comparison studies:
Method comparison and clinical evaluation of sensitivity and specificity was performed with 364 samples, the samples are detailed in the table below:
| Panel/Disease | Acronym | N (men, women, unknown) | Mean age (range) |
| --- | --- | --- | --- |
| Systemic lupus erythematosus | SLE | 98 (20, 78) | 49 (18-85) |
| ANCA Associated Vasculitis | AAV | 15 (9, 6) | 67 (31-92) |
| Poly-dermatomyositis | PM/DM | 10 (4, 6) | 56 (42-86) |
| Systemic sclerosis | SSc | 12 (6, 6) | 52 (39-79) |
| Sjögren’s syndrome | SS | 17 (0, 17) | 53 (34-88) |
| Mixed connective tissue disease | MCTD | 18 (8, 10) | 48 (19-75) |
| Rheumatoid arthritis | RA | 49 (11, 38) | 56 (22-87) |
| Anti-Phospholipid Syndrome | APS | 30 (20, 10) | 53 (30, 84) |
| Drug-Induced Lupus | DIL | 30 (2, 28) | 52 (25, 84) |
| Autoimmune hepatitis | AIH | 20 (6, 6, 8) | 56 (33-77) |
| Autoimmune thyroid disease | AITD | 16 (6, 10) | 45 (23-63) |
| Hepatitis C virus | HCV | 29 (14, 15) | 48 (19-70) |
| Clamydia pneumoniae | CP | 10 (9, 1) | 54 (32-73) |
| Epstein-Barr virus | EBV | 10 (8, 2) | 44 (27-72) |
| Total | | 364 | |
a. Method comparison with predicate device:
Samples were tested with the IFA: Crithidia luciliae (anti-dsDNA) EUROPattern assay and the predicate QUANTA Lite ds DNA ELISA assay. The QUANTA Lite ds DNA assay has positive, negative and equivocal assay range. The QuantaLite equivocal results were evaluated as positive or as negative. The results are summarized in the tables below:
Correlation between new device and predicate
| N=364 | Quanta Lite dsDNA ELISA | | | | |
| --- | --- | --- | --- | --- | --- |
| | | Positive | Equivocal | Negative | Total |
| IFA: Criedia lucilae (anti-dsDNA) EUROPattern Mode C | Positive | 31 | 3 | 6 | 40 |
| | Negative | 20 | 12 | 292 | 324 |
| | Total | 51 | 15 | 298 | 364 |
Equivocal as positive:
% positive agreement 34/66 = 51.5% 95% CI: 38.9–64.0
% negative agreement 292/298 = 98.0% 95% CI: 95.7–99.3
% overall agreement 326/364 = 89.6% 95% CI: 86.0–92.5
Equivocal as negative:
% positive agreement 31/51 = 60.8% 95% CI: 46.1–74.2
% negative agreement 304/313 = 97.1% 95% CI: 94.6–98.7
% overall agreement 335/364 = 92.0% 95% CI: 88.8–94.6
{14}
Comparison between modes:
| Mode | A/B | A/C | B/C |
| --- | --- | --- | --- |
| % positive agreement (95% CI) | 44/45=97.8% (88.2–99.9) | 40/40=100% (91.2–100) | 40/40=100% (91.2–100) |
| % negative agreement (95% CI) | 286/319=89.7% (85.8–92.8) | 287/324=88.6% (84.6–91.8) | 319/324=98.5% (96.4–99.5) |
| % overall agreement (95% CI) | 330/364=90.7% (87.2–93.4) | 327/364=89.8% (86.3–92.7) | 359/364=98.6% (96.8–99.6) |
| %titer agreement ±1 titer (95% CI) | 312/364=85.7% (81.7–89.1) | 315/364=86.5% (82.6–89.0) | 356/364=97.8% (95.7–99.0) |
b. Matrix comparison:
Not applicable, any serum is used in the assay.
3. Clinical studies:
Clinical studies were performed at three sites (two external US clinical laboratories and EUROIMMUN's laboratory). In total 364 clinically characterized samples, diagnosed according to the American College of Rheumatology (ACR) or other appropriate classification criteria, were investigated for anti-dsDNA antibodies. Samples were assayed with the IFA: Crithidia luciliae (anti-dsDNA) EUROPattern according to the package insert and evaluated by the EUROPattern microscope and software. Each result was reported in three modes (A, B, C). The overall sensitivity and specificity of the system was evaluated, $95\%$ CI are calculated by the exact method.
a. Clinical Sensitivity/Specificity:
Sensitivity and specificity was evaluated for each mode at each site and at the three sites combined using 364 clinical samples (364 x 3 sites=1092 total samples; 98 positive x 3 sites=294; 266 negative X 3 Sites=798)
Correlation to clinical diagnosis for each mode at the three sites
| N=1092 | Clinical diagnosis | | |
| --- | --- | --- | --- |
| | | Positive/per total SLE | Negative/per total non SLE |
| IFA: Critedia lucilae(anti- dsDNA) EUROPattern | Mode A | 114/294 | 731/798 |
| | Mode B | 105/294 | 765/798 |
| | Mode C | 97/294 | 774/798 |
Summary: clinical overall agreement for all three sites for each mode
| Overall sensitivity/specificity (95% CI) | Mode A | Mode B | Mode C |
| --- | --- | --- | --- |
| Sensitivity N=294 | 114/294=38.8% (33.2–44.6) | 105/294=35.7% (30.2–41.5) | 97/294=33.0% (27.6–38.7) |
| Specificity N=798 | 731/798=91.6% (89.5–93.4) | 765/798=95.9% (94.2–97.1) | 774/798=97.0% (95.6–98.1) |
{15}
Correlation with SLE clinical diagnosis in each mode at each site:
| Sensitivity (95% CI) N=98 | Mode A | Mode B | Mode C |
| --- | --- | --- | --- |
| Site 1 | 46/98=46.9% (36.8–57.3) | 35/98=35.7% (26.3–46.0) | 32/98=32.7% (23.5–42.9) |
| Site 2 | 32/98=32.7% (23.5–42.9) | 33/98=33.7% (24.4–43.9) | 29/98=29.6% (20.8–39.7) |
| Site 3 | 36/98=36.7% (27.2–47.10) | 37/98=37.8% (28.2–48.1) | 36/98=36.7% (27.2–47.1) |
Correlation with differential diagnosis in each mode at each site:
| Specificity (95% CI) N=266 | Mode A | Mode B | Mode C |
| --- | --- | --- | --- |
| Site 1 | 235/266=88.3% (83.3–91.90) | 256/266=96.2% (93.2–98.2) | 258/256=97.0% (94.2–98.7) |
| Site 2 | 243/266=91.4% (87.3–94.4) | 252/266=94.7% (91.3–97.1) | 258/266=97.0% (94.2–98.7) |
| Site 3 | 253/266=95.1% (91.8–97.4) | 257/266=96.6% (93.7–98.4) | 258/266=97.0% (94.2–98.7) |
Summary: sensitivity/specificity, correlation with clinical diagnosis by disease
| Disease | N | Site 1 | | | Site 2 | | | Site 3 | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | Mode A | Mode B | Mode C | Mode A | Mode B | Mode C | Mode A | Mode B | Mode C |
| | | Sensitivity SLE: | | | | | | | | |
| SLE | 98 | 32.7 | 35.7 | 32.7 | 32.7 | 33.7 | 29.6 | 36.7 | 37.8 | 36.7 |
| | | Specificity Non-SLE: | | | | | | | | |
| AAV | 15 | 80.0 | 93.3 | 100 | 66.7 | 86.7 | 100 | 100 | 100 | 100 |
| PM/DM | 10 | 100 | 100 | 100 | 70.0 | 90.0 | 100 | 100 | 100 | 100 |
| SSc | 12 | 91.7 | 100 | 100 | 100 | 100 | 100 | 83.3 | 100 | 100 |
| SS | 17 | 94.1 | 100 | 100 | 76.5 | 94.1 | 100 | 94.1 | 94.1 | 100 |
| MCTD | 18 | 94.4 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 |
| RA | 49 | 87.8 | 98.0 | 100 | 98.0 | 100 | 100 | 95.9 | 100 | 100 |
| APS | 30 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 |
| DIL | 30 | 73.3 | 86.7 | 86.7 | 86.7 | 86.7 | 86.7 | 93.3 | 86.7 | 86.7 |
| AIH | 20 | 90.0 | 90.0 | 90.0 | 90.0 | 90.0 | 90.0 | 90.0 | 90.0 | 90.0 |
| AITD | 16 | 87.5 | 93.8 | 93.8 | 93.8 | 93.8 | 93.8 | 93.8 | 93.8 | 93.8 |
| HCV | 29 | 89.7 | 96.6 | 96.6 | 96.6 | 96.6 | 96.6 | 96.6 | 96.6 | 96.6 |
| CP | 10 | 90.0 | 100 | 100 | 100 | 100 | 100 | 100 | 100 | 100 |
| EBV | 10 | 70.0 | 100 | 100 | 80 | 80 | 100 | 80.0 | 90.0 | 100 |
| Total | 266 | 88.3 | 96.2 | 97.0 | 91.4 | 94.7 | 97.0 | 95.1 | 96.6 | 97.0 |
{16}
b. Clinical specificity:
See Clinical Sensitivity
c. Other clinical supportive data (when a. and b. are not applicable):
N/A
4. Clinical cut-off:
See assay cut-off
5. Expected values/Reference range:
Positive dsDNA results detected by IFA are not commonly seen in normal populations and the expected value in the normal population is “negative” at a 1:10 starting dilution. Anti-dsDNA antibodies were analyzed with the IFA: Crithidia luciliae (anti-dsDNA) EUROPattern system using a panel of 124 sera from normal U.S. healthy adult blood donors of mixed age and gender (47 men, 77 women, mean age 37 years, range 20-50 years). The samples were assayed according to the package insert and evaluated by the EUROPattern microscope and software. Each result was reported in three modes (A, B, C). At the cut-off dilution of 1:10, the prevalence was found to be 0% for mode B or C and 9.7% for mode A. For mode B and C samples were 100% negative.
The manufacturer recommends that each laboratory establish its own normal range based on the population and equipment used.
N. Instrument Name:
EUROPattern: The EUROPATTERN Microscope and Software were cleared in K141827
O. System Descriptions:
1. Modes of Operation:
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ☐ X or No ☐
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ☐ or No ☐ X
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No ☐
{17}
3. Specimen Identification:
Human serum
4. Specimen Sampling and Handling:
Starting dilution is 1:10
5. Calibration:
Samples are evaluated on the microscop, there is no calibrator in the assay.
6. Quality Control:
Positive and negative control are part of the assay kit
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
N/A
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
1. The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
18
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.