K172238 · Fresenius Medical Care Renal Therapies Group, LLC · KOC · Dec 14, 2017 · Gastroenterology, Urology
Device Facts
Record ID
K172238
Device Name
CAREline Airless Hemodialysis Blood Tubing Sets
Applicant
Fresenius Medical Care Renal Therapies Group, LLC
Product Code
KOC · Gastroenterology, Urology
Decision Date
Dec 14, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5820
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
CAREline Twister: The blood tubing set is indicated for use with a prescribed hemodialyzer. The suitability of a particular bloodline/ hemodialyzer configuration is the responsibility of the physician. • The Blood Tubing Set is intended for acute and chronic hemodialysis therapy. • The Fresenius Access Flow Reversing Connector (AFRC) is for use during hemodialysis to reverse the blood flow to and from the arterial and venous vascular access devices during hemodialysis in order to obtain an access flow measurement. The AFRC facilitates the test procedure by eliminating the need to disconnect bloodlines during the test procedure. • Blood Tubing Set is intended to be used with the Fresenus Medical Care 2008® Series, K, K2 and T Hemodialysis Machines. CAREline Standard: The blood tubing set is indicated for use with a prescribed hemodialyzer. The suitability of a particular bloodline/ hemodialyzer configuration is the resonsiblity of the physician. • The Blood Tubing Set is intended for acute and chronic hemodialysis therapy. • Blood Tubing Set is intended to be used with the Fresenus Medical Care 2008® Series, K, K.2 and T Hemodialysis Machines.
Device Story
Extracorporeal blood tubing set for hemodialysis; transports arterial blood from patient access through hemodialyzer and back to venous access. Includes Pressure Output Device (POD) for pressure transmission to machine transducers; needleless access ports for medication/sampling; and optional Access Flow Reversing Connector (AFRC) for flow measurement without disconnecting lines. Used in clinical hemodialysis settings by trained personnel. Device facilitates therapy delivery and monitoring; AFRC simplifies access flow testing. Single-use, ethylene oxide sterilized.
Clinical Evidence
No clinical studies were performed. Substantial equivalence supported by bench testing, including structural integrity (pressure/pull testing), connector performance, clamp testing, biocompatibility (ISO 10993-1), and usability evaluation.
Technological Characteristics
Single-use extracorporeal blood tubing set. Materials: PVC, PC, PP, PE, MABS, TPE, acrylic-based polymer, polyisoprene, siloxanes, silicones. Features POD for pressure monitoring and needleless access ports. Sterilized via ethylene oxide. Complies with ISO 8638:2010. No electrical or software components.
Indications for Use
Indicated for patients requiring acute or chronic hemodialysis therapy. Used with prescribed hemodialyzer and Fresenius Medical Care 2008 Series K, K2, and T machines.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
Predicate Devices
CombiSet Blood Tubing with Access Flow Reversing Connector (Twister) (K022536)
Streamline Airless System Set with Locksite Needleless Access Site (K080807)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 14, 2017
Fresenius Medical Care Renal Therapies Group, LLC Denise Oppermann Senior Director, Regulatory Affairs 920 Winter Street Waltham, MA 02451
Re: K172238
Trade/Device Name: CAREline Airless Hemodialysis Blood Tubing Sets Regulation Number: 21 CFR& 876.5820 Regulation Name: Hemodialysis System and Accessories Regulatory Class: II Product Code: KOC, FJK Dated: November 10. 2017 Received: November 13, 2017
Dear Denise Oppermann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Charles Viviano -S
Benjamin R. Fisher, Ph.D. For Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172238
### Device Name
CAREline Airless Hemodialysis Blood Tubing Sets
Indications for Use (Describe)
CAREline Twister:
The blood tubing set is indicated for use with a prescribed hemodialyzer. The suitability of a particular bloodline/ hemodialyzer configuration is the responsibility of the physician.
· The Blood Tubing Set is intended for acute and chronic hemodialysis therapy.
• The Fresenius Access Flow Reversing Connector (AFRC) is for use during hemodialysis to reverse the blood flow to and from the arterial and venous vascular access devices during hemodialysis in order to obtain an access flow measurement. The AFRC facilitates the test procedure by eliminating the need to disconnect bloodlines during the test procedure.
· Blood Tubing Set is intended to be used with the Fresenus Medical Care 2008® Series, K, K2 and T Hemodialysis Machines.
CAREline Standard:
The blood tubing set is indicated for use with a prescribed hemodialyzer. The suitability of a particular bloodline/ hemodialyzer configuration is the resonsiblity of the physician.
• The Blood Tubing Set is intended for acute and chronic hemodialysis therapy.
· Blood Tubing Set is intended to be used with the Fresenus Medical Care 2008® Series, K, K.2 and T Hemodialysis Machines.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Fresenius Medical Care. On the left is a blue symbol that looks like three downward-pointing chevrons stacked on top of each other. To the right of the symbol is the text "FRESENIUS MEDICAL CARE" in blue, with "FRESENIUS" on top and "MEDICAL CARE" on the bottom.
K172238 Page 1 of 6
#### 5. 510(K) SUMMARY
This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR 807.92.
#### 5.1. Submitter's Information
| Name: | Fresenius Medical Care Renal Therapies Group, LLC |
|-------------------|-------------------------------------------------------------------|
| Address: | 920 Winter Street<br>Waltham, MA<br>02451-1457 |
| Phone: | (781) 699-4479 |
| Fax: | (781) 699-9635 |
| Contact Person: | Denise Oppermann, Senior Director<br>Regulatory Affairs – Devices |
| Preparation Date: | 24 July 2017 |
#### 5.2. Device Name
| Trade Name: | CAREline Airless Hemodialysis Blood Tubing Set |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Common Name: | KOC – Accessories, Blood Circuit,<br>Hemodialysis<br>FJK - Set, Tubing, Blood, With and Without<br>Anti-regurgitation Valve |
| Classification Name : | Hemodialysis system and accessories |
| Regulatory Class: | Class II per 21 CFR 876.5820 |
| Product Code: | KOC / FJK |
| Classification Panel: | Gastroenterology-Urology |
#### 5.3. Legally Marketed Predicate Device
### Primary Predicate - CombiSet® Blood Tubing with Access Flow 5.3.1. Reversing Connector (Twister®)
CombiSet Blood Tubing with Access Flow Reversing Connector (Twister), K022536, is the primary predicate for the CAREline Airless Hemodialysis Blood Tubing Sets (hereafter referred to as "CAREline bloodlines").
### Secondary Predicate - Streamline Airless System Set with Locksite® 5.3.2. Needleless Access Site
Streamline Airless System Set with Locksite Needleless Access Site, K080807, is the secondary predicate for the CAREline bloodlines.
{4}------------------------------------------------
K172238 Page 2 of 6
#### 5.4. Device Description
The CAREline bloodlines are intended for use in acute and chronic hemodialysis therapy. The bloodline is part of the hemodialysis extracorporeal circuit, which transports arterial blood from the patient's arterial access (e.g., fistula or catheter) through a hemodialyzer and back to the patient's venous access.
#### 5.4.1. Device Identification
The proposed CAREline bloodlines include the following devices (Table 1):
#### Table 1: Proposed CAREline Bloodlines
| Product | Alternate name |
|------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| CAREline Airless Hemodialysis Blood Tubing Set with<br>Attached Priming Set | CAREline Standard bloodline |
| CAREline Airless Hemodialysis Blood Tubing Set with<br>Access Flow Reversing Connector (Twister) and Attached<br>Priming Set | CAREline Twister® bloodline |
#### 5.4.2. Environment of Use
The CAREline bloodlines are used in environments where acute and chronic hemodialysis are performed.
#### 5.4.3. Brief Written Description of the Device
The CAREline bloodlines are intended for use in acute and chronic hemodialysis therapy. The bloodline is part of the hemodialysis extracorporeal circuit, which transports arterial blood from the patient's arterial access (e.g., fistula or catheter) through a hemodialyzer and back to the patient's venous access.
#### 5.4.4. Materials of Use
The CAREline bloodlines are classified as an external communicating medical device with prolonged exposure (>24 hours < 30 days) to circulating blood in accordance with ISO 10993- 1:2009, FDA G95-1, and FDA Guidance on Hemodialysis blood tubing sets (2008).
Materials used in the manufacture of the CAREline bloodlines include:
- Tubing and Components:
- -Polyvinylchloride (PVC)
- -Polycarbonate (PC)
- Polypropylene (PP) -
- -Polyethylene (PE)
- Methyl Isobutyl Ketone (TetraMEK) -
{5}------------------------------------------------
- । Cyclohexanone
- Pressure Output Device (POD): ●
- -Methyl Methacrylate Acrylonitrile Butadiene Styrene (MABS)
- Thermoplastic Elastomer (TPE) made of Medalist® MD |
- Twister (Twister model only): ●
- Acrylic-based polymer (Cryo XT) —
- Polyisoprene Rubber -
- -Siloxanes
- Silicones -
- Dimethyl alcohol -
- Isopropyl alcohol |
#### Key Performance Specifications/Characteristic 5.4.5.
The CAREline bloodlines are single-use devices sterilized by ethylene oxide (EO).
The CAREline bloodlines have the following performance characteristics:
- Transmission of arterial and venous pressures to the hemodialysis (HD) machine's pressure transducer via a Pressure Output Device (POD)
- . Collection of blood samples and administration of prescribed medications via a needleless access port (T-port)
- Prevention of stretched or kinked tubing via a dialyzer holder lock sleeve (a ● class I exempt accessory)
- Measurement of a patient's access flow via the Twister component, which is used to reverse blood flow to and from the arterial and venous vascular access sites.
### 5.5. Intended Use
The CAREline Airless Hemodialysis Blood Tubing Sets are indicated for use in acute and chronic hemodialysis therapy.
#### Indications for Use 5.6.
# CAREline Twister
The blood tubing set is indicated for use with a prescribed hemodialyzer. The suitability of a particular bloodline/hemodialyzer configuration is the responsibility of the physician.
- The Blood Tubing Set is intended for acute and chronic hemodialysis therapy. ●
- The Fresenius Access Flow Reversing Connector (AFRC) is for use during
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for Fresenius Medical Care. The logo consists of a blue symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right, also in blue. The symbol is made up of three downward-pointing chevrons stacked on top of each other.
hemodialysis to reverse the blood flow to and from the arterial and venous vascular access devices during hemodialysis in order to obtain an access flow measurement. The AFRC facilitates the test procedure by eliminating the need to disconnect bloodlines during the test procedure.
- Blood Tubing Set is intended to be used with Fresenius Medical Care ● 2008® Series K, K2 and T Hemodialysis Machines.
# CAREline Standard
The blood tubing set is indicated for use with a prescribed hemodialyzer. The suitability of a particular bloodline/hemodialyzer configuration is the responsibility of the physician.
- The Blood Tubing Set is intended for acute and chronic hemodialysis therapy. .
- Blood Tubing Set is intended to be used with Fresenius Medical Care 2008® Series K, K2 and T Hemodialysis Machines.
### 5.7. Comparison of Technological Characteristics with the Predicate Device
The CAREline bloodlines and the predicate devices share the following characteristics:
- Similar intended use including similar indications for use
- Similar design and configuration .
- . Same scientific technology and principles of operation
- . Same sterilization method, packaging, and sterility label claims
- Same materials - Polyvinylchloride (PVC), Polypropylene (PP), Polyethylene (PE)
### 5.8. Performance Data
Performance testing was conducted in accordance with ISO 8638:2010. Results of the testing listed below support the determination of substantial equivalence.
- Performance testing of connectors (hemodialyzer, vascular access device, and ancillary components)
- . Structural integrity testing
- Biological safety testing (biocompatibility) .
- . Sterility and pyrogenicity testing
- . Structural integrity - Pressure and pull testing
- Pressure testing Access ports and pressure output devices (PODs) ●
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the Fresenius Medical Care logo. The logo consists of a blue symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right. The symbol is made up of three downward-pointing chevrons stacked on top of each other. The text is in a bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.
K172238
- Endurance testing Conducted with Fresenius Medical Care 2008® Series K, K2, . and T Hemodialysis Machines
- Clamp testing Ability to occlude and effects of repeated use .
- . Usability evaluation
#### 5.8.1. Biocompatibility Testing
The following testing was performed to support the biological safety of the CAREline bloodlines:
- Chemical analysis Extractables and leachables ●
- Cytotoxicity
- Sensitization ●
- Irritation ●
- Systemic Toxicity ●
- Material Mediated Pyrogenicity
- Genotoxicity
- Hemocompatibility .
A toxicological risk assessment was also performed.
#### 5.8.2. Electrical Safety and Electromagnetic Compatibility (EMC)
No electrical safety and electromagnetic compatibility (EMC) tests were performed.
#### 5.8.3. Software Verification and Validation Testing
No software verification and validation tests were performed.
#### 5.8.4. Mechanical and Acoustic Testing
No mechanical or acoustic tests were performed.
#### 5.8.5. Animal Studies
No animal studies were performed.
#### 5.8.6. Clinical Studies
No clinical studies were performed.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for Fresenius Medical Care. The logo consists of a blue symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right. The symbol is made up of three downward-pointing chevrons stacked on top of each other. The text is in a bold, sans-serif font and is also blue.
#### 5.9. Conclusions
Based on the information and data provided in this Traditional 510(k) submission, the CAREline bloodlines are substantially equivalent in intended use, indications for use, design, principle of operation, technology, materials, and performance to the predicate devices (K022536, CombiSet Blood Tubing with Access Flow Reversing connector (Twister) and K080807, Streamline Airless System Set with Locksite Needleless Access Site). The devices are safe and effective for their intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.