The ECOM Cardiac Output Monitoring System is intended for the monitoring of cardiac output by impedance cardiography. The ECOM System displays the R-Wave Detection and the Impedance Waveforms as well as the patient's Cardiac Output (CO), Stroke Volume (SV), Heart Rate (HR), Systolic and Diastolic Pressure. The ECOM System is indicated for use in patients who are expected to be intubated for 24 hours or less and in whom an arterial pressure line is used. The ECOM with the endotracheal cardiac output monitor system is intended for the monitoring of cardiac output by impedance cardiography while providing airway management by oral intubation with an ECOM endotracheal tube. The ECOM Double Lumen Endobronchial Tube is used to isolate the left lung of a patient for surgery, one lung ventilation or one lung anesthesia. The ECOM Endobronchial Tube is indicated for use in thoracic surgery, bronchspirometry, for the administration of endobronchial anesthesia and other uses commonly associated with endobronchial tubes. The tube allows isolation and selective insufflation or deflation of either lung.
Device Story
System monitors cardiac output via impedance cardiography; utilizes high-frequency, low-amplitude electrical current applied to electrodes on endotracheal/endobronchial tubes. Inputs include impedance signals from tube electrodes and arterial pressure signals. System calculates and displays CO, CI, SV, SVV, HR, SVR, SVRI, and blood pressures. Used in clinical settings (e.g., surgery) by clinicians; provides real-time hemodynamic data to assist in patient management and airway control. Double-lumen tube design allows selective lung isolation/ventilation. Benefits include integrated airway management and hemodynamic monitoring.
Clinical Evidence
Bench testing only. Design verification confirmed system monitors and displays R-Wave detection, impedance waveforms, and hemodynamic parameters (CO, SV, HR, BP). Material analysis confirmed compliance with ISO 10993-1:2009.
Technological Characteristics
Impedance plethysmograph; high-frequency, low-amplitude electrical current sensing. Patient-contacting materials compliant with ISO 10993-1:2009. System includes monitor and double-lumen endobronchial tubes. Connectivity via arterial pressure line integration.
Indications for Use
Indicated for patients expected to be intubated for 24 hours or less requiring arterial pressure monitoring; also indicated for thoracic surgery, one-lung ventilation/anesthesia, and bronchspirometry requiring lung isolation.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
ECOM Endotracheal Cardial Output Monitor System (K131765)
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of two main elements: a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around it, and an abstract symbol in the center. The symbol resembles a stylized human figure or bird in flight, composed of three overlapping profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 19, 2017
ECOM Medical, Inc., % Cheryl Blake Regulatory Affairs 27392 Capricho Mission Viejo, California 92692
Re: K172196
Trade/Device Name: ECOM™ Cardiac Output Monitoring System Regulation Number: 21 CFR 870.2770 Regulation Name: Impedance Plethysmograph Regulatory Class: Class II Product Code: DSB Additional Product Codes: CBI Dated: July 22, 2017 Received: July 25, 2017
Dear Cheryl Blake:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
M.A. Hillebrenner
for _ Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
#### 510(k) Number (if known) K172196
Device Name ECOM Cardiac Output Monitoring System
#### Indications for Use (Describe)
The ECOM Cardiac Output Monitoring System is intended for the monitoring of cardiac output by impedance cardiography.
The ECOM System displays the R-Wave Detection and the Impedance Waveforms as well as the patient's Cardiac Output (CO), Stroke Volume (SV), Heart Rate (HR), Systolic and Diastolic Pressure.
The ECOM System is indicated for use in patients who are expected to be intubated for 24 hours or less and in whom an arterial pressure line is used.
The ECOM with the endotracheal cardiac output monitor system is intended for the monitoring of cardiac output by impedance cardiography while providing airway management by oral intubation with an ECOM endotracheal tube.
The ECOM Double Lumen Endobronchial Tube is used to isolate the left lung of a patient for surgery, one lung ventilation or one lung anesthesia. The ECOM Endobronchial Tube is indicated for use in thoracic surgery, bronchspirometry, for the administration of endobronchial anesthesia and other uses commonly associated with endobronchial tubes. The tube allows isolation and selective insufflation or deflation of either lung.
Type of Use (Select one or both, as applicable)
ال Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary of Safety and Effectiveness
# ECOM™ Cardiac Output Monitoring System
In accordance with the requirements of the Safe Medical Device Act of 1990 and 21CFR 807.92, ECOM Medical, Inc. is hereby submitting the 510(k) Summary of Safety and Effectiveness for 510(k) number K172196 as of September 14, 2017.
#### A. Submitter
ECOM Medical, Inc 27127 Calle Arroyo, Suite 1905 San Juan Capistrano, CA 92576 USA
Establishment Registration: 13010770019
#### B. Applicant Correspondent
Cheryl Blake Regulatory Consultant 27392 Capricho Mission Viejo, CA 92692 cherylblake@cox.net 949-285-3517
#### C. Device Name
| Proprietary Name: | ECOM™ Cardiac Output Monitoring System |
|----------------------|----------------------------------------|
| Common Name: | Plethysmograph, impedance |
| Classification Name: | Impedance plethysmograph |
| Regulation Number: | 870.2770 |
| Product Code: | DSB |
| Regulatory Class: | II |
| Panel: | Cardiovascular |
#### D. Predicate Device
| Primary Device Name: | ECOM Endotracheal Cardial Output Monitor System |
|------------------------|---------------------------------------------------------|
| Company Name: | ECOM Medical via acquisition of ECOM division of Conmed |
| 510(k): | K131765 |
| Secondary Device Name: | Well Lead Endobronchial Tubes |
| Company Name: | Well Lead Medical Device Instruments Ltd. |
| 510(k): | K092886 |
{5}------------------------------------------------
## E. Device Description
The ECOM Cardiac Output Monitor System consists of a monitor, and various accessories. The line extension of the ECOM Double Lumen Endobronchial Tube to the systems line of Endotrachel tubes continues to apply a high frequency. low amplitude electrical current to a series of electrodes applied to the endobronchial tube. The resulting signals, when used in conjunction with an arterial pressure signal, allow for the calculation and display of Cardiac Output (CO), Cardiac Input (CI), Stroke Volume (SV), Stroke Volume Variation (SVV), Heart Rate (HR), Systemic Vascular Resistance (SVR), Systemic Vascular Resistance Index (SVRI), and systolic, diastolic and mean blood pressures. No changes have been made to the monitor system in K131765 or the endobronchial tube design in K092886.
## F. Intended Use / Indications for Use
The ECOM Cardiac Output Monitoring System is intended for the monitoring of cardiac output by impedance cardiography.
The ECOM System displays the R-Wave Detection and the Impedance Waveforms as well as the patient's Cardiac Output (CO), Stroke Volume (SV), Heart Rate (HR), Systolic and Diastolic Pressure.
The ECOM System is indicated for use in patients who are expected to be intubated for 24 hours or less and in whom an arterial pressure line is used.
The ECOM with the endotracheal cardiac output monitor system is intended for the monitoring of cardiac output by impedance cardiography while providing airway management by oral intubation with an ECOM endotracheal tube.
The ECOM Double Lumen Endobronchial Tube is used to isolate the left lung of a patient for surgery, one lung ventilation or one lung anesthesia. The ECOM Endobronchial Tube is indicated for use in thoracic surgery, bronchspirometry, for the administration of endobronchial anesthesia and other uses commonly associated with endobronchial tubes. The tube allows isolation and selective insufflation or deflation of either lung.
### G. Non-clinical Performance Testing
Non-clinical bench testing demonstrated the ECOM Endobronchial Cardiac Output Monitor System is substantially equivalent to the predicate with regard to intended use, materials, technology, and performance of the monitoring. Design verification testing, demonstrates the system, using the previously cleared monitor and software, monitors and displays the R-Wave Detection and the Impedance Waveforms as well as the patient's Cardiac Output (CO), Stroke Volume (SV), Heart Rate (HR), Systolic and Diastolic Pressure. Material analysis demonstrates the patient contacting materials comply with the requirements of ISO 10993-1:2009.
{6}------------------------------------------------
## H. Substantial Equivalence
The differences between the predicate and the new tube design do not raise any new risks of safety or efficacy. No changes have been made to the devices represented in K131765 and K092886. No changes have been made to the endobronchial tube design and only IFU has been updated to include the endobronchial tube supporting information. Additional supporting information per this premarket submission confirms that the ECOM Endobronchial Cardiac Output Monitor System is safe and effective for the intended use and is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.