K172182 · St. Jude Medical (Now Part of the Abbott Medical) · DQK · Aug 18, 2017 · Cardiovascular
Device Facts
Record ID
K172182
Device Name
QUANTIEN Measurement System
Applicant
St. Jude Medical (Now Part of the Abbott Medical)
Product Code
DQK · Cardiovascular
Decision Date
Aug 18, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1425
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K172182 · Aug 18, 2017
QUANTIEN Measurement System
St. Jude Medical (Now Part of the Abbott Medical)
Published clinical literature; Clinical publications
The sponsor utilized existing clinical literature and published data comparing the device's performance (FFR and iFR measurements) to support the substantial equivalence of the subject device to the predicate.
Clinical literature and publication review; Literature review of clinical use and data analysis
Patients undergoing FFR/iFR measurement
Not applicable for this study
Clinical equivalence of FFR and iFR measurements
Indications for Use
The QUANTIENT™ system is intended for use in catheterization and related cardiovascular specialty laboratories to compute and display various physiological parameters based on the output from one or more electrodes, transducers or measuring devices. The QUANTIENT™ system is indicated to provide hemodynamic information for use in the diagnosis and treatment of patients that undergo measurement of physiological parameters with PressureWire™ guidewire.
Device Story
Diagnostic computer system for catheterization labs; records, computes, displays, and stores hemodynamic data. Inputs: signals from PressureWire™ guidewire and external pressure transducers; ECG data. Processing: calculates physiological parameters including systolic/diastolic/mean blood pressure, heart rate, Fractional Flow Reserve (FFR), and Pd/Pa ratio. Output: graphical and numerical data displayed on backlit LCD; exportable to external hemodynamic recording systems, video monitors, or DICOM servers. Used by physicians to assess stenosis severity and guide therapeutic intervention. Benefits: provides objective hemodynamic information to support clinical decision-making regarding coronary or peripheral artery disease treatment.
Clinical Evidence
No new clinical testing performed. Equivalence to FFR and iFR established via existing clinical literature and public data.
Technological Characteristics
Diagnostic computer; backlit LCD display; inputs from PressureWire™ guidewire and external transducers; connectivity via DICOM, USB, and external hemodynamic recording system (HRS) interfaces; software-based data processing and display.
Indications for Use
Indicated for patients undergoing hemodynamic measurement of physiological parameters using PressureWire™ guidewire in catheterization and cardiovascular specialty laboratories for the diagnosis and treatment of coronary or peripheral artery disease.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 18, 2017
St. Jude Medical Julie Dalquist Senior Regulatory Affairs Specialist One St. Jude Medical Drive St. Paul, Minnesota 55117
Re: K172182
Trade/Device Name: OUANTIEN™ Measurement System Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DOK, DSK Dated: July 18, 2017 Received: July 19, 2017
Dear Julie Dalquist:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
M.G. Wilhelm
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K172182
Device Name Quantien Measurement System
Indications for Use (Describe)
OUANTIEN Measurement System is indicated to provide hemodynamic information for use in the diagnosis and treatment of coronary or peripheral artery disease.
QUANTIEN Measurement System is intended for use in catheterization and related cardiovascular specialty laboratories to compute, and display various physiological parameters based on the output from one or more electrodes, transducers, or measuring devices.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
ST. Jude Medical
St. Jude Medical One St. Jude Medical Drive St. Paul, MN 55117 USA Main 651 756 2000
## 6. 510(k) Summary
The 510(k) Summary is provided below, as required by 21 CFR 807.92(c).
| 510(k) Number | K172182 |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | August 18, 2017 |
| Submitter Name and<br>Address: | St. Jude Medical (now part of Abbott Medical)<br>One St. Jude Medical Dr.<br>St. Paul, MN 55117 |
| Contact Person: | Julie Dalquist, RAC<br>Senior Regulatory Affairs Specialist<br>Phone (651) 756-3501<br>Fax (651) 756-3298<br>jdalquist@sjm.com |
| Proprietary Name: | QUANTIENTM Measurement System |
| Common/Usual Name: | Quantien |
| Product Classification<br>Code: | DQK |
| Product Regulation<br>Number and Name: | §870.1425, Programmable Diagnostic computer |
| Subsequent Code: | DSK |
| Product Regulation<br>Number and Name: | §870.1110, Blood Pressure computer |
| Device Class: | II |
| Predicate Device: | QUANTIENTM Measurement System (K123984) |
| Device Description: | The QUANTIENT™ Measurement System is a diagnostic computer designed to<br>record, compute, display and store data from PressureWire™ guidewire and<br>other external transducers. The information is displayed as graphs as well as<br>numerical values on the screen. Data includes, but is not limited to: systolic,<br>diastolic and mean blood pressure, heart rate, Fractional Flow Reserve (FFR),<br>Pd/Pa, and data from ECG. |
| | Fractional Flow Reserve (FFR) is the ratio of distal coronary arterial pressure<br>to aortic pressure, measured during hyperemia. It provides the maximal blood<br>flow in the presence of a stenosis as a fraction of the achievable blood flow<br>that would exist in the hypothetical situation that the stenosis was not present.<br>The physician may use the FFR parameter, along with knowledge of patient<br>history, medical expertise and clinical judgment to determine if therapeutic<br>intervention is indicated. This functionality is achieved when the<br>QUANTIENT™ Measurement System is used in conjunction with the<br>manufacturer's distal intracoronary pressure transducer and a proximal aortic<br>pressure transducer. |
| | Pd/Pa is the ratio of distal coronary arterial pressure to aortic pressure<br>measured at resting conditions. The physician may use Pd/Pa at rest, along<br>with knowledge of patient history, medical expertise and clinical judgment to<br>determine if additional measurement of FFR during hyperemia or therapeutic<br>intervention is indicated. |
| | Information on screen can also be transferred to an external hemodynamic<br>recording system (HRS) or to an external video monitor. Recorded procedures<br>can be viewed on a PC for post-procedural review and analysis with<br>application specific viewing software installed, such as RadiView™ software. |
| | Additional functions let you import a patient work list from the hospital DICOM<br>system, export recorded measurement data to DICOM or to an external server<br>location or save it to a USB memory stick. |
| Intended Use: | The QUANTIENT™ system is intended for use in catheterization and related<br>cardiovascular specialty laboratories to compute and display various<br>physiological parameters based on the output from one or more electrodes,<br>transducers or measuring devices.<br>The QUANTIENT™ system is indicated to provide hemodynamic information<br>for use in the diagnosis and treatment of patients that undergo measurement of<br>physiological parameters with PressureWire™ guidewire. |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the logo for St. Jude Medical. The logo consists of a square grid of smaller squares, with one square slightly offset, followed by the text "ST. JUDE MEDICAL" in a serif font. The color of the logo and text is a dark teal.
St. Jude Medical One St. Jude Medical Drive St. Paul, MN 55117 USA Main 651 756 2000
{5}------------------------------------------------
# K172182
Page 3 of 3
Image /page/5/Picture/1 description: The image shows the logo for St. Jude Medical. The logo consists of a grid of small squares on the left and the text "ST. JUDE MEDICAL" on the right. The text is in a simple, sans-serif font and is horizontally aligned with the grid.
St. Jude Medical One St. Jude Medical Drive St. Paul, MN 55117 USA Main 651 756 2000
| Comparison of Subject<br>to Predicate Device: | The intended use and technological characteristics of the QUANTIENT™<br>Measurement System that is subject to this Special 510(k) submission are the<br>same as the predicate, as summarized in the following table. | | |
|-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Characteristic | QUANTIENT™ Measurement<br>System (subject device) | QUANTIENT™ Measurement<br>System (predicate device) | |
| Intended Use | The QUANTIENT™ system is<br>intended for use in<br>catheterization and related<br>cardiovascular specialty<br>laboratories to compute and<br>display various physiological<br>parameters based on the output<br>from one or more electrodes,<br>transducers or measuring<br>devices.<br>The QUANTIENT™ system is<br>indicated to provide<br>hemodynamic information for<br>use in the diagnosis and<br>treatment of patients that<br>undergo measurement of<br>physiological parameters with<br>Pressure Wire™ guidewire. | The QUANTIENT™ system is<br>intended for use in<br>catheterization and related<br>cardiovascular specialty<br>laboratories to compute and<br>display various physiological<br>parameters based on the output<br>from one or more electrodes,<br>transducers or measuring<br>devices.<br>The QUANTIENT™ system is<br>indicated to provide<br>hemodynamic information for<br>use in the diagnosis and<br>treatment of patients that<br>undergo measurement of<br>physiological parameters with<br>PressureWire™ guidewire. | |
| Display | Backlit LCD | Backlit LCD | |
| Display<br>Features | PressureWire™ distal (Pd) and<br>proximal/aortic (Pa) pressure<br>Mean Pd and Pa | PressureWire™ distal (Pd) and<br>proximal/aortic (Pa) pressure<br>Mean Pd and Pa | |
| | Pd/Pa<br>ECG | Pd/Pa<br>ECG | |
| Summary on Non-<br>Clinical Testing: | No new non-clinical testing was completed, nor relied upon, in support of this<br>Special 510(k) submission. | | |
| Summary of Clinical<br>Testing: | No new clinical testing was completed in support of this Special 510(k)<br>submission. Physician use and data analysis was completed and published<br>publicly. Comparison to FFR was equivalent clinically, and comparison to<br>iFR (as per the guidelines) was shown equivalent in clinical literature and<br>publication. | | |
| Statement of<br>Equivalence: | The QUANTIENT™ Measurement System described in this Special 510(k)<br>submission is substantially equivalent to the currently-marketed predicate<br>device, based on comparisons of the device classifications, technological<br>characteristics, and intended use. | | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.