← Product Code [EBC](/productcode/EBC) · K172169

# Pulpdent (Activa) Pit and Fissure Sealant with MCP (K172169)

_Pulpdent Corporation · EBC · Jan 5, 2018 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K172169

## Device Facts

- **Applicant:** Pulpdent Corporation
- **Product Code:** [EBC](/productcode/EBC.md)
- **Decision Date:** Jan 5, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3765
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

Pulpdent's Activa Pit and Fissure Sealant with MCP is intended to seal pit and fissure depressions/faults in the enamel, in the biting surfaces of teeth. For dental professional use only.

## Device Story

Activa Pit and Fissure Sealant is a light-cured, flowable dental material supplied in a single-barrel syringe. It consists of methacrylate resins, barium glass, silica, sodium fluoride, and modified calcium phosphate (MCP). The device is applied by dental professionals to tooth enamel to fill and seal pits and fissures. Upon light curing, the material hardens to provide a physical barrier; it also releases calcium, phosphate, and fluoride ions into the oral environment. The device is used in a clinical dental setting to prevent or treat enamel faults, potentially benefiting the patient by protecting biting surfaces from decay.

## Clinical Evidence

No clinical data. Evidence consists of bench testing, including physical property measurements (compressive strength, flexural strength, diametral tensile strength, water absorption/solubility, depth of cure) and biocompatibility testing (ISO 10993-5:2009 cytotoxicity).

## Technological Characteristics

Light-cured, glass-filled methacrylate resin sealant. Composition includes diurethane dimethacrylate, Bis(2-(Methacryloyloxy) Ethyl) Phosphate, barium glass, silica, sodium fluoride, and MCP. Form factor: single-barrel syringe. Standards: ISO 4049:2009 (polymer-based restorative materials), ISO 10993-5:2009 (cytotoxicity), ISO 7405:1999/2009. Energy source: visible light curing unit.

## Regulatory Identification

A pit and fissure sealant and conditioner is a device composed of resin, such as polymethylmethacrylate, intended for use primarily in young children to seal pit and fissure depressions (faults in the enamel) in the biting surfaces of teeth to prevent cavities.

## Predicate Devices

- Embrace Pit and Fissure Sealant/Pulpdent Dentition Integrating Material ([K020287](/device/K020287.md))

## Reference Devices

- Pulpdent Activa Solo Flowable Composite with MCP ([K153249](/device/K153249.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 5, 2018

Pulpdent Corporation % Roger Mastrony President MedTek LLC 2516 Kettle Creek Court Lincolnton, North Carolina 28092

Re: K172169

Trade/Device Name: Pulpdent (Activa) Pit and Fissure Sealant with MCP Regulation Number: 21 CFR 872.3765 Regulation Name: Pit And Fissure Sealant And Conditioner Regulatory Class: Class II Product Code: EBC, EBF Dated: November 27, 2017 Received: December 4, 2017

Dear Roger Mastrony:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mary S. Runner -S

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K172169

Device Name

Pulpdent Activa Pit and Fissure Sealant with MCP

Indications for Use (Describe)

Pulpdent's Activa Pit and Fissure Sealant with MCP is intended to seal pit and fissure depressions/faults in the enamel, in the biting surfaces of teeth. For dental professional use only.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# PULPDENT CORPORATION

80 Oakland Street Watertown, MA 02472 USA

Tel: 617 926 6666 Fax: 617 926 6262 Email: Pulpdent@pulpdent.com

### 510(k) SUMMARY K172169

### DATE OF SUBMISSION: 07/11/2017, resubmitted 11-27-17.

OWNER: Pulpdent Corporation 80 Oakland Street Watertown, MA 02472 USA

- CONTACT: Roger S. Mastrony 2516 Kettle Creek Court Lincolnton NC 28092 Tel: 1 203 640 5047 / Email: rogermastrony@gmail.com
TYPE OF SUBMISSION: Traditional 510(k).

#### DEVICE:

Trade Name: Pulpdent Activa Pit and Fissure Sealant with MCP Device Class: II Review Panel: Dental Classification Name: Pit and Fissure Sealant FDA Product Code: EBC, 21 CFR Part 872.3765

### PREDICATE DEVICE: Pulpdent's Embrace Pit and Fissure Sealant/Dentition Integrating Material K020287

Please Note: Pulpdent Dentition Integrating Material was the working name for Pulpdent Embrace Pit and

Fissure Sealant, before a marketing name was decided upon. It is the same formula.

### REFERENCE DEVICE: Pulpdent's Activa Solo Flowable Composite with MCP K153249

DESCRIPTION: Activa Pit and Fissure Sealant is a light cured, flowable dental sealant material, available in a single barrel syringe. This sealant is designed with MCP (Modified Calcium Phosphate); it releases calcium, phosphate, and fluoride ions into the oral environment. It contains no Bisphenol A, no BisGMA, and no BPA derivatives.

The applicable standards and device-specific guidance documents, special controls documents, and/or requirements in a device-specific requlation, regarding device description that are applicable to the subject device, and are either referenced within this submission or used in consultation for developing this submission, include:

-ISO 4049:2009, Polymer Based Filling Restorative and Luting Materials;

-ANSI/ADA Specification No. 41, Recommended Standard Practices for Biological Evaluation of Dental Materials (Referenced for predicate device);

-ISO 10993-5:2009, Biological Evaluation of Medical Devices-Part 5: tests for in vitro Cytotoxicity; -ISO 7405:1999 and 2009; and

-Guidance for Industry and FDA staff; Dental Composite Resin Devices Premarket Notification [510(k)] Submissions.

INDICATIONS FOR USE: Pulpdent's Activa Pit and Fissure Sealant with MCP is intended to seal pit and fissure depressionsffaults in the enamel, in the biting surfaces of teeth. For dental professional use only.

BIOCOMPATIBILITY EVALUATION: The subject device, Pulpdent Activa Pit and Fissure Sealant with MCP, was tested with a Cytotoxicity test (ISO 10993-5: 2009). Additional chemical characterization is provided in file 012_Biocompatibility Evaluation, and a histopathology test for the predicate device, Embrace Pit and Fissure Sealant, K020287, is also referenced. The formula for Embrace is the same as for Activa except that Activa has MCP added. Biocompatibility of MCP in the formula is supported by presence of MCP in the reference device, Pulpdent Solo, K153249, which has a similar formula.

{4}------------------------------------------------

# PULPDENT CORPORATION

80 Oakland Street Watertown, MA 02472 USA

### DEVICE COMPARISON WITH PREDICATE AND REFERENCE PRODUCTS

The predicate product has been found to be substantially equivalent under the 510(k) Premarket Notification process as Class II Dental Device under CFR 872.3765 Code EBC

|                                     | Pulpdent Activa Pit and<br>Fissure Sealant with MCP                                                                                                                                                          | Embrace Pit and Fissure<br>Sealant/Pulpdent<br>Dentition Integrating<br>Material<br>K020287                                                                                                                    | Pulpdent Solo Flowable<br>Composite with MCP<br>K153249                                                                                                                                     |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| DESCRIPTION                         | Fluoride releasing, light-cured,<br>glass filled, methacrylate resins<br>with modified calcium<br>phosphate                                                                                                  | Fluoride releasing, light-<br>cured, glass filled,<br>methacrylate resins                                                                                                                                      | Fluoride releasing, Light-cured,<br>glass filled, methacrylate resin<br>with modified calcium<br>phosphate                                                                                  |
| TECHNOLOGY /<br>MECHANISM OF ACTION | Light cured, fluoride releasing,<br>resin sealant for filling/sealing<br>faults in the tooth enamel                                                                                                          | Light cured, fluoride<br>releasing, resin sealant for<br>filling/sealing faults in the<br>tooth enamel                                                                                                         | Light cured, fluoride releasing,<br>glass filled resin for tooth<br>restoration                                                                                                             |
| INDICATIONS FOR USE                 | Pulpdent (Activa) Pit and Fissure<br>Sealant with MCP is intended to<br>seal pit and fissure depressions/<br>faults in the enamel, in the biting<br>surfaces of teeth.  For dental<br>professional use only. | Pulpdent dentition integrating<br>material is a fluoride-releasing,<br>light-cured, resin-based material<br>that bonds tightly to dentition and<br>is used to fill and seal the pits<br>and fissures in teeth. | Pulpdent Solo Flowable Composite<br>with MCP is a fluoride, calcium, and<br>phosphate-releasing flowable<br>composite used by dental<br>professionals as a restorative, base,<br>and liner. |
| COMPOSITION                         | Diurethane dimethacrylate, Bis<br>(2-(Methacryloyloxy) Ethyl)<br>Phosphate, Barium glass,<br>Silica, Sodium Fluoride, MCP<br>(modified calcium phosphate)                                                    | Diurethane<br>dimethacrylate, Bis (2-<br>(Methacryloyloxy) Ethyl)<br>Phosphate, Barium glass,<br>Silica, Sodium Fluoride                                                                                       | Diurethane dimethacrylate, Bis (2-<br>(Methacryloyloxy) Ethyl)<br>Phosphate, Barium glass, Silica,<br>Camphorquinone, Sodium Fluoride<br>(optional), MCP (Modified calcium<br>phosphate)    |
| PHYSICAL PROPERTIES                 |                                                                                                                                                                                                              |                                                                                                                                                                                                                |                                                                                                                                                                                             |
| Appearance                          | Tooth shade paste/resin                                                                                                                                                                                      | Tooth shade paste/resin                                                                                                                                                                                        | tooth shade paste/resin                                                                                                                                                                     |
| Specific gravity                    | 1.39-1.42 g/mL                                                                                                                                                                                               | 1.575 g/mL                                                                                                                                                                                                     | 1.81 ± 0.03 g/mL                                                                                                                                                                            |
| pH                                  | 3.5 - 4                                                                                                                                                                                                      | --                                                                                                                                                                                                             | --                                                                                                                                                                                          |
| Compressive Strength                | 304 MPa (19.5)                                                                                                                                                                                               | 240.0 MPa                                                                                                                                                                                                      | 330±47.9 MPa                                                                                                                                                                                |
| Flexural Strength                   | 98 (12) MPa                                                                                                                                                                                                  | 76 MPa                                                                                                                                                                                                         | 129±18.7 MPa                                                                                                                                                                                |
| Diametral Tensile strength          | 49 (1.4) MPa                                                                                                                                                                                                 | 43.4 MPa                                                                                                                                                                                                       | 46.5 MPa                                                                                                                                                                                    |
| Water Absorption (7days)            | 39.2 μg/mm³ or 2.99%                                                                                                                                                                                         | --                                                                                                                                                                                                             | 46.5 MPa                                                                                                                                                                                    |
| Water Solubility (7 days)           | 0.14%                                                                                                                                                                                                        | 0.06%                                                                                                                                                                                                          | 0.09%                                                                                                                                                                                       |
| Depth of cure                       | 3.2 mm                                                                                                                                                                                                       | 3.1 mm                                                                                                                                                                                                         | 2.8 mm (20 sec VLC)                                                                                                                                                                         |

{5}------------------------------------------------

## PULPDENT CORPORATION

80 Oakland Street Watertown, MA 02472 USA

### Similarities

Pulpdent ACTIVA PIT AND FISSURE SEALANT and the predicate product, PULPDENT EMBRACE PIT AND FISSURE SEALANT/Dentition Integrating Material, both share similar technology/mechanism of action as a light curing and fluoride releasing glass filled resin. Indications for use are also the same, as a pit and fissure sealant, for depressions/faults on the tooth enamel. The chemical composition and physical properties are very similar, as described in the Device Description of this 510(k) submission.

Pulpdent ACTIVA PIT AND FISSURE SEALANT with MCP and the reference product, Pulpdent Solo Flowable Composite with MCP K153249, both have the same technology/mechanism of action as light curing and fluoride releasing glass filled resin with the addition of calcium and phosphate ions.

The predicate has been found substantially equivalent under the 510(k) Premarket Notification process as Class II Dental Devices under CFR 872.3765.

#### Differences

Pulpdent ACTIVA PIT AND FISSURE SEALANT differs from the predicate product with the addition of MCP (Modified Calcium Phosphate). The subject device, and reference device have similar chemical compositions and physical properties.

#### Discussion of the Differences

With the addition of MCP to Pulpdent's ACTIVA PIT AND FISSURE SEALANT this sealant now has a release of calcium and phosphate ions into the oral environment, in addition to fluoride ion release seen in the predicate product. The difference is not significant because calcium and phospahte ions have a history of use in dental devices, no history of adverse reactions, as well as literature/studies that show the biocomapatability of these ions in sealing the tooth enamel.

#### Conclusion

Taking into account the above similarities and differences. Pulpdent ACTIVA PIT AND FISSURE SEALANT with MCP is substantially equivalent, in design, performance and indications for use, to the predicate product. Pulpdent Embrace Pit and Fissure Sealant/Dentition Integrating Material and to the reference product. Pulpdent's Activa Solo Flowable Composite with MCP.

---

**Source:** [https://fda-staging.innolitics.com/device/K172169](https://fda-staging.innolitics.com/device/K172169)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).