The PROPEL Device is indicated for manually drilling holes in tissue and bone for orthodontic and dental operative procedures including: (1) initiation holes for drill bits, implants, screws, plates, and other orthodontic appliances; and (2) for use as a tool to create micro-osteoperforations.
Device Story
Manual, single-use, hand-held bone drill; stainless steel (17-4) drill tip (1.6 mm OD); plastic (ABS/polycarbonate) handle with retractable tube for depth control (0, 3, 5, 7 mm). Device lacks motor; battery-powered red LED signals when target depth is reached. Used by oral surgeons/orthodontists in operating room/clinical settings for alveolar osteotomies without prior tissue retraction. Provides precise depth control to prevent over-penetration; benefits patient via minimally invasive access for orthodontic/dental procedures.
Clinical Evidence
Clinical studies demonstrate the device can be used to manually drill holes in tissue and bone for orthodontic and dental procedures and as a tool to create micro-osteoperforations (1.6 mm OD; 0, 3, 5, 7 mm depths).
Technological Characteristics
Manual drill; 17-4 stainless steel bit; ABS/polycarbonate body. Depth control via retractable tube. Battery-powered LED depth indicator. Single-use; provided sterile. Complies with ISO 10993-1 (biocompatibility), ISO 11137-1/2/3 (sterilization), ASTM F1980-07 (shelf-life), IEC 60601-1 (electrical safety), and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for patients requiring manual drilling of tissue and bone for orthodontic and dental procedures, including initiation holes for appliances/implants and creation of micro-osteoperforations.
Regulatory Classification
Identification
A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Propel Orthodontics LLC c/o Jeffrey K. Shapiro Director Hyman, Phelps & McNamara, P.C. 700 13th Street NW, Suite 1200 Washington, DC 20005
January 17, 2018
Re: K172164
Trade/Device Name: PROPEL Device Regulation Number: 21 CFR 872.4120 Regulation Name: Bone Cutting Instrument and Accessories Regulatory Class: Class II Product Code: DZJ Dated: December 6, 2017 Received: December 7, 2017
Dear Jeffrey K. Shapiro:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Mary S. Runner -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K172164
Device Name
PROPEL Device
Indications for Use (Describe)
The PROPEL Device is indicated for manually drilling holes in tissue and bone for orthodoutic and dental operative procedures including: (1) initiation holes for drill bits, implants, screws, plates, and other orthodontic appliances; and (2) for use as a tool to create micro-osteoperforations.
| Type of Use (Select one or both, as applicable) | <span> [X] Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
| FORM FDA 3881 (7/17) | Page 1 of 1 | PSC Publishing Services (301) 443-6740 | EP |
|----------------------|-------------|----------------------------------------|----|
|----------------------|-------------|----------------------------------------|----|
{3}------------------------------------------------
## 510(k) Summary PROPEL Device
| Submitter Name: | Propel Orthodontics LLC |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Address: | 394 South Abbott Ave<br>Milpitas, CA 95035 |
| Contact Person: | Bryce Way |
| Phone Number: | 408-394-5851 |
| Fax Number: | 914-560-2056 |
| Date Prepared: | January 9, 2018 |
| Device Trade Name: | PROPEL Device |
| Classification Name: | Bone Cutting Instrument and Accessories |
| Device Common Name: | Manual Bone Drill |
| Product Code: | DZJ |
| Classification: | Class II pursuant to 21 C.F.R. § 872.4120 |
| Predicate Device: | PROPEL Device (K150392) |
| Indications for Use: | The PROPEL Device is indicated for manually drilling holes in<br>tissue and bone for orthodontic and dental operative procedures<br>including: (1) initiation holes for drill bits, implants, screws,<br>plates, and other orthodontic appliances; and (2) for use as a tool to<br>create micro-osteoperforations. |
| Device Description: | The PROPEL Device has a stainless steel drill tip with an outer<br>diameter (drill bit OD) of 1.6 mm that can be manually adjusted to<br>the length needed for treatment or cutting depth. The plastic<br>handle incorporates a retractable plastic tube over the cutting drill<br>instrument. This enables the drill tip to be locked into a specific<br>depth (0, 3, 5, or 7 mm), thus preventing the drill tip from<br>penetrating the bone beyond the targeted depth when used for<br>osteotomies during dental operative procedures. The handle<br>contains the red LED operated by batteries. The LED is<br>illuminated when the tip of the drill tip reaches the set depth.<br>There is no motor in the device. The batteries are only for<br>signaling when the prescribed depth has been achieved. |
{4}------------------------------------------------
Performance data: The nonclinical testing cleared as part of the predicate device's 510(k) submission supports substantial equivalence of the proposed device because the intended use, technological features, design, and materials of the proposed device are similar to the predicate device. Because no changes have been made to the PROPEL Device, aside from the addition of a specific indication for use as a tool to create micro-osteoperforations (diameter and range of depth dimensions noted above), no additional nonclinical testing was conducted for purposes of this submission.
> Biocompatibility data for the predicate PROPEL Device demonstrates that the PROPEL Device is in compliance with the FDA Guidance Document, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process."
Sterilization validation according to ISO 11137-1, -2, and -3 to ensure a SAL of 106 and shelf-life validation according to ASTM F1980-07 were conducted to validate the 2-year shelf-life of the predicate PROPEL Device.
Electrical safety testing for the predicate PROPEL Device was conducted in compliance with IEC 60601-1: Medical electrical equipment – Part 1: General requirements for basic safety and essential performance and IEC 60601-1-2: Medical electrical equipment – Part 1-2: General requirements for safety – Collateral standard: Electromagnetic compatibility -Requirements and tests.
Bench testing for the predicate PROPEL Device has demonstrated that the PROPEL Device is in compliance with its product specifications, the expectations of the medical community, and the product labeling. The following bench tests were conducted to demonstrate the PROPEL Device's performance:
- Bit reliability, including deflection resistance and hole depth repeatability
- Various tensile and compression tests of components
- Dial and bit depth reliability
- LED light reliability
Clinical studies demonstrate that the PROPEL Device (1) can be used to manually drill holes in tissue and bone for orthodontic and dental operative procedures and (2) can be used as a tool to create micro-osteoperforations (OD 1.6 mm and depth of 0, 3, 5, and 7 mm).
{5}------------------------------------------------
## Substantial Equivalence: The intended use and technological characteristics of the proposed PROPEL Device are similar to the predicate PROPEL Device. There is a change in the indications for use statement between the predicate and proposed devices for use as a tool to create microosteoperforations. A comparison of the proposed and predicate devices is shown in Table VI.1, below.
{6}------------------------------------------------
Table VI.1: Comparison to Predicate
| Specification | PROPOSED<br>PROPEL Device | PREDICATE<br>PROPEL Device<br>(K150392) | Similarities and<br>Differences |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|
| Intended Use | Intended for<br>manually drilling<br>holes in tissue and<br>bone for orthodontic<br>and dental operative<br>procedures. | Intended for<br>manually drilling<br>holes in tissue and<br>bone for orthodontic<br>and dental operative<br>procedures. | Identical |
| Indications for<br>Use | The PROPEL Device<br>is indicated for<br>manually drilling<br>holes in tissue and<br>bone for orthodontic<br>and dental operative<br>procedures including:<br>(1) initiation holes for<br>drill bits, implants,<br>screws, plates, and<br>other orthodontic<br>appliances; and (2)<br>for use as a tool to<br>create micro-<br>osteoperforations. | The PROPEL Device<br>is indicated for<br>manually drilling<br>holes in tissue and<br>bone for orthodontic<br>and dental operative<br>procedures including<br>initiation holes for<br>drill bits, implants,<br>screws, plates and<br>other orthodontic<br>appliances. | Similar. Addition in<br><b>bold italics</b> . |
| FDA Product<br>Code | DZJ | DZJ | Identical |
| Operating<br>Principle<br>(Technology) | Hand-held manual<br>drill | Hand-held manual<br>drill | Identical |
| Drill Angle | Drills holes at a 0°<br>angle<br>(i.e., inline) with the<br>handle | <b>Design</b><br>Drills holes at a 0°<br>angle<br>(i.e., inline) with the<br>handle | Identical |
| Powered or<br>Manual | Manually operated<br>drill | Manually operated<br>drill | Identical |
| Direct Drive or<br>Geared | Direct drive, device is<br>turned by rotating<br>proximal end | Direct drive, device is<br>turned by rotating<br>proximal end | Identical |
| Method for<br>Creating Holes | Stainless Steel Drill<br>Bit | Stainless Steel Drill<br>Bit | Identical |
| Interchangeable<br>Drill Bits | No. Drill bit built-in. | No. Drill bit built-in. | Identical |
| Drill Bit Outer<br>Diameter (OD) | 1.6 mm | 1.6 mm | Identical |
| Specification | PROPOSED<br>PROPEL Device | PREDICATE<br>PROPEL Device<br>(K150392) | Similarities and<br>Differences |
| Drill Bit<br>Working<br>Length (depth) | 0, 3, 5, and 7 mm | 0, 3, 5, and 7 mm | Identical |
| Depth Stops | Yes, by adjusting the<br>depth stop on the<br>drill. | Yes, by adjusting the<br>depth stop on the<br>drill. | Identical |
| Types of<br>Screwdriver<br>Bits | None. Device cannot<br>be converted into a<br>screwdriver | None. Device cannot<br>be converted into a<br>screwdriver | Identical |
| Indicator of<br>Achievement of<br>Proper Depth | LED light comes on<br>to prevent<br>unwarranted pressure<br>on tissue | LED light comes on<br>to prevent<br>unwarranted pressure<br>on tissue | Identical |
| Materials: Drill<br>Bits | Stainless Steel (17-4) | Stainless Steel (17-4) | Identical |
| Materials:<br>Device Body | ABS and<br>polycarbonate | ABS and<br>polycarbonate | Identical |
| Principal<br>Operator | Oral<br>surgeon/orthodontist | Oral<br>surgeon/orthodontist | Identical |
| Can Be Used As<br>Minimally<br>Invasive? | Yes. Can be used for<br>alveolar osteotomies<br>without prior tissue<br>retraction. | Yes. Can be used for<br>alveolar osteotomies<br>without prior tissue<br>retraction. | Identical |
| Use Location | Operating room | Operating room | Identical |
| Single Use? | Single-use only | Single-use only | Identical |
| Sterilization | Provided sterile | Provided sterile | Identical |
{7}------------------------------------------------
## Substantial Equivalence Conclusion:
The intended use and technological characteristics for the subject PROPEL Device are identical to those of the predicate PROPEL Device. The addition of an indication as a tool to create micro-osteoperforations is a specific bone cutting indication and performance testing supports the subject device for the proposed indications for use. Therefore, the information in this submission demonstrates that the subject PROPEL Device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.