External evaluations of Etest Ceftazidime/Avibactam; Comparative performance evaluation against reference method
Gram-negative aerobic bacteria (Enterobacteriaceae and Pseudomonas aeruginosa)
CLSI broth microdilution reference method
Essential Agreement (EA) and Category Agreement (CA)
Indications for Use
Etest is a quantitative technique for determination of antimicrobial susceptibility of both non-fastidious Gram-negative and Gram-positive aerobic bacteria such as Enterobacteriaceae, Pseudomonas, Staphylococcus, and Enterococcus species and fastidious bacteria, such as anaerobes, N. gonorrhoeae, S. pneumoniae, Streptococcus and Haemophilus species. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC), in µg/mL, of different antimicrobial agents against microorganisms as tested on agar media using overnight incubation. Ceftazidime/avibactam has been shown to be active against the Gram negative aerobic microorganisms listed below, according to the FDA label for this antimicrobial agent. Etest CZA can be used to determine the MIC of ceftazidime/avibactam against the microorganisms listed below: Active both in vitro and in clinical infections: Citrobacter freundii, Enterobacter cloacae, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Proteus mirabilis, Pseudomonas aeruginosa. The following in vitro data are available, but clinical significance is unknown: Citrobacter koseri, Enterobacter aerogenes, Morganella morganii, Providencia rettgeri, Providencia stuartii, Serratia marcescens.
Device Story
Etest is a thin, inert, non-porous plastic strip with a predefined antibiotic gradient on one side and an MIC reading scale on the other. Used in clinical microbiology laboratories; applied to inoculated agar surfaces. Antibiotic gradient transfers into agar, forming a stable, continuous concentration gradient. Following overnight incubation, bacterial growth forms a symmetrical inhibition ellipse. The MIC value is determined by the intersection of the ellipse with the strip scale. Provides quantitative susceptibility data to guide antimicrobial therapy decisions for Gram-negative infections.
Clinical Evidence
Bench testing only. Performance evaluated using 1,153 clinical isolates and 86 challenge isolates across three sites. Comparison against CLSI broth microdilution reference method showed 99.1% Essential Agreement (EA) and 99.6% Category Agreement (CA) for Enterobacteriaceae, and 99.3% EA and 99.3% CA for P. aeruginosa. A trend toward higher MIC readings was observed for Enterobacteriaceae. Very major error rate for Enterobacteriaceae was 4.4% (adjusted to 0% due to EA). Limitations noted for M. morganii and C. koseri due to lack of resistant strains or specific interpretive discrepancies.
Technological Characteristics
Thin, inert, non-porous plastic strip; predefined antibiotic gradient; manual reading of MIC scale; agar diffusion principle; overnight incubation.
Indications for Use
Indicated for determination of antimicrobial susceptibility of Gram-negative aerobic bacteria (Enterobacteriaceae, Pseudomonas aeruginosa) to Ceftazidime/Avibactam. Used for clinical isolates and challenge strains.
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
K172150
B. Purpose for Submission:
To obtain a substantial equivalence determination for ceftazidime/avibactam at concentrations of 0.016 – 256 µg/mL for susceptibility testing of gram negative aerobic microorganisms with Etest.
C. Measurand:
Ceftazidime/avibactam 0.016 – 256 µg/mL. The avibactam concentration is fixed at 4 µg/mL in this combination.
D. Type of Test:
Quantitative AST growth-based detection
E. Applicant:
bioMérieux, Inc.
F. Proprietary and Established Names:
Etest Ceftazidime/Avibactam (CZA) (0.016 – 256 µg/mL)
G. Regulatory Information:
1. Regulation section:
866.1640 Antimicrobial Susceptibility Test Powder
2. Classification:
II
3. Product code:
JWY – Manual Antimicrobial Test Systems
{1}
4. Panel:
83 – Microbiology
H. Intended Use:
1. Intended use(s):
Etest is a quantitative technique for determination of antimicrobial susceptibility of both non-fastidious Gram-negative and Gram-positive aerobic bacteria such as Enterobacteriaceae, Pseudomonas, Staphylococcus, and Enterococcus species and fastidious bacteria, such as anaerobes, N. gonorrhoeae, S. pneumoniae, Streptococcus and Haemophilus species. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC), in µg/mL, of different antimicrobial agents against microorganisms as tested on agar media using overnight incubation.
Ceftazidime/avibactam has been shown to be active against the Gram negative aerobic microorganisms listed below, according to the FDA label for this antimicrobial agent. Etest CZA can be used to determine the MIC of ceftazidime/avibactam against the microorganisms listed below:
Active both in vitro and in clinical infections:
- Citrobacter freundii
- Enterobacter cloacae
- Escherichia coli
- Klebsiella oxytoca
- Klebsiella pneumoniae
- Proteus mirabilis
- Pseudomonas aeruginosa
The following in vitro data are available, but clinical significance is unknown:
- Citrobacter koseri
- Enterobacter aerogenes
- Morganella morganii
- Providencia rettgeri
- Providencia stuartii
- Serratia marcescens
2. Indication(s) for use:
Same as the Intended Use
3. Special conditions for use statement(s):
For prescription use only
2
{2}
3
Limitations:
Due to the lack of an intermediate interpretive category for ceftazidime/avibactam, testing of Morganella morganii with this drug has resulted in very major discrepancies for isolates that are otherwise within essential agreement of the reference method. Testing should be repeated using an alternative testing/reference method prior to reporting results for M. morganii when the Etest MIC is 8 µg/mL.
The ability of Etest to detect resistance with the following microorganisms is unknown because resistant strains were not available at the time of comparative testing. If such a strain is observed, it should be submitted to a reference laboratory for further testing Ceftazidime/Avibactam: Citrobacter koseri
4. Special instrument requirements:
N/A
I. Device Description:
Etest consists of a thin, inert and non-porous plastic strip 5mm wide and 60 mm long. One side of the strip carries a two letter code designating the identity of the antibiotic and is calibrated with MIC values in terms of µg/mL. On the reverse, a predefined exponential gradient of the dried and stabilized antibiotic covers a continuous concentration range across 15 two-fold dilutions of a conventional MIC method.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Etest Ceftolozane/Tazobactam
2. Predicate 510(k) number(s):
K170670
3. Comparison with predicate:
{3}
Table 1. Comparison with the Predicate Device
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | K172150 Etest Ceftazidime/Avibactam | K170670 Etest Ceftolozane/Tazobactam |
| Intended Use | Etest is a quantitative technique for determination of antimicrobial susceptibility of both non-fastidious Gram-negative and Gram-positive aerobic bacteria such as Enterobacteriaceae, Pseudomonas, Staphylococcus, and Enterococcus species and fastidious bacteria, such as anaerobes, N. gonorrhoeae, S. pneumoniae, Streptococcus and Haemophilus species. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC), in μg/mL, of different antimicrobial agents against microorganism as tested on agar media using overnight incubation. | Same |
| Antimicrobial Concentration Range | 0.016 – 256 μg/mL (Avibactam: fixed at 4 μg/mL) | Same (Tazobactam: fixed at 4 μg/mL) |
| Test Design | A predefined exponential gradient of the dried and stabilized antibiotic covers a continuous concentration range across 15 two-fold dilutions of a conventional MIC method. | Same |
| Inoculation | Isolated colonies from culture | Same |
| Incubation | 35°±2°C for 16–20 hours | Same |
| Result | MIC | Same |
{4}
| Differences |
| --- |
| Item | Device | Predicate |
| Claimed organisms | Citrobacter freundii | Citrobacter koseri |
| Citrobacter koseri | Enterobacter cloacae |
| Enterobacter aerogenes | Escherichia coli |
| Enterobacter cloacae | Klebsiella oxytoca |
| Escherichia coli | Klebsiella pneumoniae |
| Klebsiella oxytoca | Morganella morganii |
| Klebsiella pneumoniae | Proteus mirabilis |
| Morganella morganii | Proteus vulgaris |
| Proteus mirabilis | Providencia rettgeri |
| Providencia rettgeri | Providencia stuartii |
| Providencia stuartii | Pseudomonas aeruginosa |
| Pseudomonas aeruginosa | Serratia liquifaciens |
| Serratia marcescens | Serratia marcescens |
| Antimicrobial Agent | Ceftazidime/Avibactam | Ceftolozane/Tazobactam |
## K. Standard/Guidance Document Referenced (if applicable):
Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA
CLSI M07-A10, Method for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard, January 2015.
CLSI M100-S26, Performance Standards for Antimicrobial Susceptibility Testing; Volume 36, No. 1, January 2016.
## L. Test Principle:
The Etest consists of a thin, inert, nonporous plastic strip that is used to determine the antimicrobial susceptibility of bacteria. One side of the strip carries the minimum inhibitory concentration (MIC) reading scale expressed in $\mu \mathrm{g} / \mathrm{mL}$. The other side of the strip contains a predefined continuous gradient of antibiotic concentrations.
When the strip is applied to an inoculated agar surface, the preformed antibiotic gradient immediately transfers into the agar matrix, then forming a stable, continuous and exponential gradient of antibiotic concentrations directly underneath the strip. Bacteria growth becomes visible during incubation, and a symmetrical inhibition ellipse centered along the strip appears. After incubation, the MIC value is read from the scale in terms of $\mu \mathrm{g} / \mathrm{mL}$ at complete inhibition of bacterial growth, where the pointed end of the ellipse intersects the strip. Since Etest generates MIC values which fall between two-fold dilutions for interpretation, the MIC value read must be recorded to the next two-fold dilution.
The MIC gradient on Etest Ceftazidime/Avibactam ranges from 0.016 to $256~\mu \mathrm{g / mL}$. The
{5}
avibactam concentration is fixed at $4\ \mu\mathrm{g/mL}$ in this combination.
## M. Performance Characteristics:
### 1. Analytical performance:
#### a. Precision/Reproducibility:
A reproducibility study was conducted at three external sites using 25 isolates of gram negative bacilli that were consistent with the intended use. The isolates tested included *Citrobacter freundii* (two isolates), *Citrobacter koseri* (two isolates), *Enterobacter aerogenes* (two isolates), *Enterobacter cloacae* (two isolates), *Escherichia coli* (five isolates), *Klebsiella oxytoca* (three isolates), *Klebsiella pneumoniae* (three isolates) and *Pseudomonas aeruginosa* (six isolates).
All results were within $\pm 1$ doubling dilution of the mode MIC value for ceftazidime/avibactam. Eleven results were on scale. The reproducibility was acceptable at $100.0\%$.
#### b. Linearity/assay reportable range:
Not applicable
#### c. Traceability, Stability, Expected values (controls, calibrators, or methods):
**Inoculum Density Check.** Inoculum density checks were performed for all quality control and reproducibility organism suspensions and for $10\%$ of the suspensions prepared for susceptibility testing of the fresh clinical isolates. The mean inoculum density was acceptable at $4.40 \times 10^{5}$ CFU/mL.
**Purity Check.** All clinical, challenge and reproducibility test suspensions were subcultured to assure purity.
**Growth or Device Failure.** There were no growth or device failures during the course of the study.
**Quality Control Testing.** Organisms recommended by both FDA and the CLSI were tested with ceftazidime/avibactam at three sites. According to the CLSI M100 S27 document, *K. pneumoniae* ATCC 700603 must be used for routine QC testing of ceftazidime/avibactam. The QC organisms tested included: *E. coli* ATCC 25922, *E. coli* ATCC 35218, *K. pneumoniae* ATCC 700603 and *P. aeruginosa* ATCC 27853. These QC strains were tested a minimum of 20 times per site by both the Etest and the reference method. One quality control strain (*Staphylococcus aureus* ATCC 29213) was tested with the reference method only to provide additional control for this method; $100\%$ of the results obtained with *S. aureus* ATCC 29213 were in the acceptable range. The results demonstrate that the ceftazidime/avibactam Etest can
6
{6}
produce quality control results in the recommended range $>95\%$ of the time. See Table 2 below for a summary of the QC results.
Per the recommendations in the CLSI document M100, additional testing was performed using the reference method with $K$ pneumoniae 700603 with ceftazidime alone and $K$ pneumoniae ATCC 700603 and $E$ coli ATCC 35218 with ampicillin to assure the integrity of the $\beta$ -lactamase encoding plasmid. The testing demonstrated that the QC strains maintained their resistance profiles throughout the testing as evidenced by obtaining MIC values of $16 - 64~\mu \mathrm{g / mL}$ with ceftazidime for $K$ pneumoniae 700603 and obtaining MIC values of $>128~\mu \mathrm{g / mL}$ with ampicillin for the same organism. MIC values of $>32~\mu \mathrm{g / mL}$ were obtained with ampicillin for $E$ coli ATCC 35218.
The sponsor included the following comment in the device labeling: K. pneumoniae ATCC 700603 should be used for routine QC of ceftazidime/avibactam. This QC strain should also be tested against ceftazidime alone to ensure that the plasmid encoding the beta-lactamase has not been lost in this strain.
Table 2. Quality Control Data for Ceftazidime/Avibactam
| QC organism | Ceftazidime/ Avibactam expected range (μg/mL) | Concentration (μg/mL) | Reference | Etest |
| --- | --- | --- | --- | --- |
| E. coliATCC 25922 | 0.06 – 0.5 | <0.06 | - | - |
| | | 0.06 | 1 | - |
| | | 0.125 | 74 | 71 |
| | | 0.25 | 10 | 15 |
| | | 0.5 | 1 | - |
| | | >0.5 | - | - |
| | | | | |
| P. aeruginosaATCC 27853 | 0.5 - 4 | <0.5 | - | - |
| | | 0.5 | - | - |
| | | 1 | 14 | 46 |
| | | 2 | 71 | 40 |
| | | 4 | 1 | - |
| | | >4 | - | - |
| | | | | |
| E. coliATCC 35218 | 0.03 – 0.12 | <0.03 | - | - |
| | | 0.03 | 1 | - |
| | | 0.06 | 77 | 76 |
| | | 0.125 | 8 | 10 |
| | | >0.125 | - | - |
| | | | | |
| K. pneumoniaeATCC 700603 | 0.25 - 2 | <0.25 | - | - |
| | | 0.25 | - | - |
| | | 0.5 | 50 | 1 |
| | | 1 | 35 | 79 |
{7}
| QC organism | Ceftazidime/Avibactam expected range (μg/mL) | Concentration (μg/mL) | Reference | Etest |
| --- | --- | --- | --- | --- |
| | | 2 | 1 | 6 |
| | | >2 | - | - |
d. Detection limit:
Not applicable
e. Analytical specificity:
Not applicable
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Results obtained with Etest Ceftazidime/Avibactam were compared to results obtained with the CLSI broth microdilution reference panel. The CLSI panel was prepared and interpreted according to CLSI recommendations outlined in the CLSI Standard: CLSI Document M07-A10, Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria that Grow Aerobically; Approved Standard – Tenth Edition, Vol. 35, No. 2; January 2015. The testing conditions for the reference method consisted of the following:
- Medium – Cation-adjusted Mueller Hinton Broth with appropriate dilutions of antimicrobial solution added
- Inoculum – Direct colony suspension to achieve a suspension equivalent to a 0.5 McFarland standard suspension
- Incubation 35 °C in ambient air; 16-20 hours for all organisms
Clinical testing was performed at three sites using a total of 1153 clinical isolates (1033 isolates of Enterobacteriaceae and 120 isolates of P. aeruginosa). Of the clinical isolates, 699 were fresh isolates (60.6 %) and 454 were stock isolates (39.4%). Clinical isolates were tested using both Etest and the reference method.
A total of 86 challenge isolates were tested at a single site using Etest and the reference method (54 isolates of Enterobacteriaceae and 32 isolates of P. aeruginosa).
{8}
Combined clinical and challenge isolates included: C. freundii (93 isolates), C. koseri (91 isolates), E. aerogenes (96 isolates), E. cloacae (99 isolates), E. coli (109 isolates), K. oxytoca (102 isolates), K. pneumoniae (107 isolates), M. morganii (65 isolates), P. mirabilis (95 isolates), P. vulgaris (45 isolates), P. rettgeri (57 isolates), P. stuartii (59 isolates), S. marcescens (69 isolates) and P. aeruginosa (152 isolates).
For P. aeruginosa, the combined results from clinical and challenge testing demonstrated a combined EA of 99.3% and CA of 99.3% (Table 3). A total of 145 isolates were determined to have evaluable results; the EA of evaluable results was 99.3%.
For the Enterobacteriaceae, the combined results from clinical and challenge testing demonstrated a combined EA of 99.1% and CA of 99.6% (Table 3). A total of 1060 isolates were determined to have evaluable results; the EA of evaluable results was 99.2%.
Table 3. Performance of Clinical and Challenge Isolates, Ceftazidime/Avibactam Etest.
| | Tot | No. EA | EA % | Eval Tot | Eval EA | EA Eval % | No. CA | CA % | No. R | min | maj | vmj |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| P. aeruginosa | | | | | | | | | | | | |
| Clinical | 120 | 119 | 99.2 | 114 | 113 | 99.1 | 119 | 99.2 | 16 | NA^{a} | 1 | 0 |
| Challenge | 32 | 32 | 100.0 | 31 | 31 | 100.0 | 32 | 100.0 | 18 | NA^{a} | 0 | 0 |
| Combined | 152 | 151 | 99.3 | 145 | 144 | 99.3 | 151 | 99.3 | 34 | NA^{a} | 1 | 0 |
| | | | | | | | | | | | | |
| Enterobacteriaceae^{b} | | | | | | | | | | | | |
| Clinical | 1033 | 1023 | 99.0 | 1015 | 1006 | 99.1 | 1029 | 99.6 | 33 | NA^{a} | 2 | 2 |
| Challenge | 54 | 54 | 100.0 | 45 | 45 | 100.0 | 54 | 100.0 | 12 | NA^{a} | 0 | 0 |
| Combined | 1087 | 1077 | 99.1 | 1060 | 1051 | 99.2 | 1083 | 99.6 | 45 | NA^{a} | 2^{c} | 2^{c,d} |
a NA, Not Applicable due to a lack of an intermediate breakpoint for ceftazidime/avibactam
b Enterobacteriaceae included: Citrobacter freundii, Enterobacter cloacae, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Proteus mirabilis, Citrobacter koseri, Enterobacter aerogenes, Morganella morganii, Providencia rettgeri, Providencia stuartii, and Serratia marcescens
c Etest results are within essential agreement of the reference method
Due to the lack of an intermediate interpretive category for ceftazidime/avibactam, overall very major error rate for Enterobacteriaceae of 4.4% (2/45) was adjusted to 0% when considering that all error results were within EA of the reference method
EA - Essential Agreement (+/- 1 dilution)
CA - Category Agreement
EVAL - Evaluable isolates
R - Resistant isolates
min - minor discrepancies
maj - major discrepancies
vmj - very major discrepancies
Essential Agreement (EA) occurs when the result of the reference method and that of Etest Ceftazidime/Avibactam are within plus or minus one serial two-fold dilution of the antibiotic. Evaluable results are those that are on scale for both Etest Ceftazidime/Avibactam and the reference method. Category Agreement (CA) occurs when the interpretation of the result of the reference method agrees exactly with the interpretation of Etest Ceftazidime/Avibactam.
{9}
A total of 79 resistant isolates were evaluated; however, no resistant strains of Citrobacter koseri were evaluated. The sponsor added the following limitation to the device labeling:
The ability of Etest to detect resistance with the following microorganisms is unknown because resistant strains were not available at the time of comparative testing. If such a strain is observed, it should be submitted to a reference laboratory for further testing.
Ceftazidime/Avibactam: Citrobacter koseri
Three clinical isolates (one isolate each of P. stuartii, E. aerogenes and P. aeruginosa) were determined to be susceptible by the reference method but resistant with ceftazidime/avibactam Etest. Both the P. stuartii and E. aerogenes MIC results were within essential agreement of the results obtained with the reference method. For the single isolate of P. aeruginosa Etest results were not within essential agreement with the reference method. However, the major error rates for Enterobacteriaceae and for P. aeruginosa were determined to be acceptable at 0.2% and 0.8%, respectively.
Two isolates of M. morganii were determined to be resistant by the reference method but susceptible with Etest Ceftazidime/Avibactam resulting in an overall very major error rate for Enterobacteriaceae of 4.4%. Because there is no intermediate breakpoint for ceftazidime/avibactam and because the Etest MIC values obtained with these two isolates were within essential agreement of the reference method, the overall very major error rate for Enterobacteriaceae was adjusted to 0.0% and considered to be acceptable. The following footnote was added to the performance table in the device labeling:
The overall very major error rate for Enterobacteriaceae was 4.4%. The two very major errors were one dilution apart from the reference method and as such fall within essential agreement. Based on the essential agreement and the lack of an intermediate breakpoint, the adjusted very major error rate is 0.0%
In addition, because both of the very major errors occurred with M. morganii (for a very major error rate of 100% for this species) the following limitation was added to the device labeling to address the possibility of very major errors when testing M. morganii:
Due to the lack of an intermediate interpretive category for ceftazidime/avibactam, testing of Morganella morganii with this drug has resulted in very major discrepancies for isolates that are otherwise within essential agreement of the reference method. Testing should be repeated using an alternative testing/reference method prior to reporting results for M. morganii when the Etest MIC is 8 µg/mL.
10
{10}
Enzyme Group Molecular Characterization:
Isolates of Enterobacteriaceae and *P. aeruginosa* harboring various molecular mechanisms of resistance noted in the FDA drug label were tested with ceftazidime/avibactam. The following resistance mechanisms were evaluated:
- Enterobacteriaceae – ESBL, not specified, CTX-M, TEM, SHV, KPC
- *P. aeruginosa* – chromosomal AmpC, OprD loss, KPC, IMP, VIM, other metallo-β-lactamase
Enterobacteriaceae isolates harboring AmpC and OXA were not evaluated. The sponsor included the following footnote to the performance table in the device labeling:
Enzyme characterization was not available for the following organisms at the time of comparative testing, and therefore the performance of Etest Ceftazidime/Avibactam is unknown for Pseudomonas aeruginosa (up-regulation of MexXY and MexAB) and Enterobacteriaceae (OXA and AmpC).
The sponsor also included the following footnote regarding additional resistance mechanisms:
Ceftazidime/avibactam is not active against bacteria that produce metallo-beta-lactamase enzymes and may not have activity against gram-negative bacteria that overexpress efflux pumps or have porin mutations.
MIC Trending
An analysis of trending was conducted using the combined clinical and challenge data for *P. aeruginosa* and for Enterobacteriaceae. This trending calculation takes into account MIC values that are determined to be one or more doubling dilution lower or higher compared to the reference method irrespective of whether the device MIC values are on-scale or not.
The data for 152 *P. aeruginosa* results determined to be evaluable for trending analysis showed no evidence of trending.
The data for 1061 Enterobacteriaceae results determined to be evaluable for trending analysis is presented in Table 4.
11
{11}
12
Table 4. Trending in Enterobacteriaceae, Clinical and Challenge Isolates
| Organism | Difference in MIC as Compared to the CLSI Reference Method | | | |
| --- | --- | --- | --- | --- |
| | No. results evaluable for trending analysis | No. results ≥1 dilution lower | No. results Exact | No. results ≥1 dilution higher |
| Enterobacteriaceae | 1061 | 80 (7.54%)* | 628 (59.19%) | 353 (33.27%)* |
*Difference: 25.73%, 95% CI (22.46 – 28.96%)
A trend towards higher MIC readings was observed in the overall performance of Enterobacteriaceae compared to the CLSI broth microdilution method which raises concerns for the potential occurrence of major errors. The sponsor included the following footnote to the performance table to address the trending observed for ceftazidime/avibactam:
Etest Ceftazidime/Avibactam MIC values tended to be in exact agreement or at least one doubling dilution higher when testing Enterobacteriaceae compared to the reference broth microdilution method.
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
No applicable
4. Clinical cut-off:
Not applicable
{12}
5. Expected values/Reference range:
Table 5. Breakpoints and Interpretive Categories for Ceftazidime/Avibactam (FDA Drug Label)
| Organism | FDA Interpretive Criteria for Ceftazidime/Avibactam MIC (μg/mL) | | |
| --- | --- | --- | --- |
| | S | I | R |
| Enterobacteriaceae and P. aeruginosa | ≤ 8/4 | NA | ≥ 16/4 |
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.