K172144 · Zeltiq Aesthetics · OOK · Nov 1, 2017 · General, Plastic Surgery
Device Facts
Record ID
K172144
Device Name
CoolSculpting System
Applicant
Zeltiq Aesthetics
Product Code
OOK · General, Plastic Surgery
Decision Date
Nov 1, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4340
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K172144 · Nov 1, 2017
CoolSculpting System
Zeltiq Aesthetics
Retrospective analysis of clinical photographs from a prior IDE study
A retrospective photographic analysis was conducted to evaluate the effectiveness of the device in affecting the appearance of lax tissue in the submental area, using data from a previously cleared IDE study.
Subjects from a prior IDE study (n=57 for analysis); Sample Size: 57
Not applicable for this study
Reduction in submental area (≥ 20 mm²), independent panel review of photographs, subject satisfaction
Indications for Use
The CoolSculpting System is a skin cooling or heating device is indicated for cold-assisted lipolysis (breakdown of fat) of the upper arm, bra fat, banana roll, submental area, thigh, abdomen, and flank, or "love handles" in individuals with a Body Mass Index (BMI) of 30 or less. The device is intended to affect the appearance of visible fat bulges in the upper arm, bra fat, banana roll, submental area, thigh, abdomen and flank. When used for cold-assisted lipolysis of the submental area, the device can also affect the appearance of lax tissue in the submental area. Cooling with the device may also be used to minimize pain and thermal injury during laser and dermatological treatments and act as a local anesthetic for procedures that induce minor local discomfort. The CoolSculpting System is also indicated for use to provide localized thermal therapy (hot or cold) to minimize pain post-trauma and post-surgery, and for temporary relief of minor aches, pains, and muscle spasms. The optional massage function can also be used for the temporary relief of minor muscle aches, pain, and spasm and for temporary improvement in local circulation and temporary reduction in the appearance of cellulite. The ZELTIQ Pretreatment Skin Wipe and Gel/Gelpad facilitate thermal contact of the device with a patient's skin by mitigating minor variances in device-to-skin contact.
Device Story
Portable thermoelectric cooling/heating device; applies controlled thermal energy to treatment sites. Comprised of control unit, detachable vacuum/surface applicators, and supplies (liners, gelpads, cycle cards). Used in clinical settings by trained operators. Inputs: manual selection of thermal parameters. Outputs: localized cooling/heating to induce lipolysis or provide thermal therapy. Optional massage function for circulation/cellulite. Healthcare providers use device to reduce fat bulges and improve skin laxity appearance. Benefits: non-invasive fat reduction and symptom relief for minor aches/pains.
Clinical Evidence
Retrospective photo analysis study (n=57) evaluating submental laxity improvement. Primary endpoint: independent panel review of pre- and 12-week post-treatment photos. 77.2% of subjects showed ≥ 20 mm² area reduction; 95.5% of these were identified as responders by physician panel. 75% of subjects reported improved chin toning. Safety profile unchanged from previous clearances.
Technological Characteristics
Thermoelectric cooling/heating system. Components: control unit, vacuum/surface applicators, gelpads. Connectivity: standalone. Software: embedded, unchanged from predicate. Sterilization: not specified (supplies are single-use).
Indications for Use
Indicated for cold-assisted lipolysis of upper arm, bra fat, banana roll, submental area, thigh, abdomen, and flank in individuals with BMI ≤ 30. Intended to affect appearance of visible fat bulges and lax tissue in submental area. Also indicated for localized thermal therapy to minimize pain post-trauma/surgery, relief of minor aches/pains/spasms, and optional massage for circulation improvement and cellulite appearance reduction.
Regulatory Classification
Identification
A contact cooling system for aesthetic use is a device that is a combination of a cooling pad associated with a vacuum or mechanical massager intended for the disruption of adipocyte cells intended for non-invasive aesthetic use.
Special Controls
*Classification.* Class II (special controls). The special controls for this device is FDA's “Guidance for Industry and FDA Staff; Class II Special Controls Guidance Document: Contact Cooling System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration."
November 1, 2017
ZELTIQ Aesthetics, Inc. Ewald Riechert Director of Regulatory Affairs 4410 Rosewood Road Pleasanton, California 94588
Re: K172144
Trade/Device Name: ZELTIQ CoolSculpting System Regulation Number: 21 CFR 878.4340 Regulation Name: Contact Cooling System for Aesthetic Use Regulatory Class: Class II Product Code: OOK Dated: September 28, 2017 Received: September 29, 2017
Dear Ewald Riechert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce pror to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
> Sincerely, Jennifer R. Stevenson -S3 For Binita S. Ashar, M.D., M.B.A., F.A.C.S.
Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172144
Device Name Zeltiq CoolSculpting System
#### Indications for Use (Describe)
The CoolSculpting System is a skin cooling or heating device is indicated for cold-assisted lipolysis (breakdown of fat) of the upper arm, bra fat, banana roll, submental area, thigh, abdomen, and flank, or "love handles" in individuals with a Body Mass Index (BMI) of 30 or less. The device is intended to affect the appearance of visible fat bulges in the upper arm, bra fat, banana roll, submental area, thigh, abdomen and flank. When used for cold-assisted lipolysis of the submental area, the device can also affect the appearance of lax tissue in the submental area.
Cooling with the device may also be used to minimize pain and thermal injury during laser and dermatological treatments and act as a local anesthetic for procedures that induce minor local discomfort.
The CoolSculpting System is also indicated for use to provide localized thermal therapy (hot or cold) to minimize pain post-trauma and post-surgery, and for temporary relief of minor aches, pains, and muscle spasms. The optional massage function can also be used for the temporary relief of minor muscle aches, pain, and spasm and for temporary improvement in local circulation and temporary reduction in the appearance of cellulite.
The ZELTIQ Pretreatment Skin Wipe and Gel/Gelpad facilitate thermal contact of the device with a patient's skin by mitigating minor variances in device-to-skin contact.
| Type of Use ( <i>Select one or both, as applicable</i> ) |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5. 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| I. SUBMITTER: | ZELTIQ™ Aesthetics, Inc.<br>4410 Rosewood Drive<br>Pleasanton, CA 94588 |
|-------------------------|----------------------------------------------------------------------------------------------------------|
| CONTACT: | Ewald Riechert<br>Director of Regulatory Affairs<br>ZELTIQ Aesthetics, Inc.<br>Phone: 925-568-2977 |
| DATE PREPARED: | July 14, 2017 |
| II. DEVICE: | |
| TRADE NAME: | ZELTIQ CoolSculpting System |
| COMMON NAME: | Skin Cooling Device |
| CLASSIFICATION NAME: | Contact Cooling System for Aesthetic Use |
| DEVICE CLASSIFICATION: | Class II, 21 CFR §878.4340 |
| PRODUCT CODE: | OOK |
| III. PREDICATE DEVICES: | Predicate Device: CoolSculpting System (K171069, OOK)<br>Reference Device: Ulthera System (K121700, OHV) |
## IV. DEVICE DESCRIPTION:
The CoolSculpting System is a portable thermoelectric cooling and heating device that applies controlled cooling or heating to a treatment site. The CoolSculpting System is comprised of a control unit, detachable vacuum and surface applicators and supplies such as liners, gel/gelpads, cycle cards, geltraps, gaskets, foam borders and securement system.
## V: INDICATION FOR USE:
The CoolSculpting System is a skin cooling or heating device. The device is indicated for cold-assisted lipolysis (breakdown of fat) of the upper arm, bra fat, back fat, banana roll, submental area, thigh, abdomen, and flank, or "love handles" in individuals with a Body Mass Index (BMI) of 30 or less. The device is intended to affect the appearance of visible fat bulges in the upper arm, bra fat, back fat, banana roll, submental area, thigh, abdomen and
{4}------------------------------------------------
flank. When used for cold-assisted lipolysis of the submental area, the device can also affect the appearance of lax tissue in the submental area.
Cooling with the device may also be used to minimize pain and thermal injury during laser and dermatological treatments and act as a local anesthetic for procedures that induce minor local discomfort.
The CoolSculpting System is also indicated for use to provide localized thermal therapy (hot or cold) to minimize pain post-trauma and post-surgery, and for temporary relief of minor aches, pains, and muscle spasms. The optional massage function can also be used for the temporary relief of minor muscle aches, pain, and spasm and for temporary improvement in local circulation and temporary reduction in the appearance of cellulite.
The ZELTIQ Pretreatment Skin Wipe and Gel/Gelpad facilitate thermal contact of the device with a patient's skin by mitigating minor variances in device-to-skin contact.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE:
The ZELTIQ CoolSculpting System is the same as the device cleared in K171069. No changes have been made to the device to accommodate this new indication.
The CoolSculpting System has been established as safe and effective through many prior clearances (the most recent is K171069). It was demonstrated to be safe and effective for the submental area in K151179.
#### VII. PERFORMANCE DATA:
## Biocompatibility testing
The ZELTIQ CoolSculpting System is the same as the previously cleared predicate device (K171069).
## Electrical safety and electromagnetic compatibility (EMC)
The ZELTIQ CoolSculpting System is the same as the previously cleared predicate device (K171069).
#### Software Verification and Validation Testing
The ZELTIQ CoolSculpting System is the same as the previously cleared predicate device (K171069).
## Performance testing
The ZELTIQ CoolSculpting System is the same as the previously cleared predicate device (K171069).
{5}------------------------------------------------
## Clinical Study
ZELTIQ conducted a retrospective photo analysis study examining the effectiveness of the cryolipolysis system to affect the appearance of lax tissue in the submental area. The photographic analysis was based on the methodology used to obtain clearance of the reference device (Ulthera System, K121700).1
Sixty subjects were enrolled in an original IDE approved study that was used for the submental clearance in K151179. Two of the 60 subjects did not receive full treatments and were excluded from the per-protocol population. Subjects received treatment with the ZELTIQ CoolSculpting System on the submental area, and were followed through a 12-week post-treatment visit.
The primary safety endpoint in the original IDE approved study was met and remains unchanged since the close of the study and was previously submitted in K151179.
The primary effectiveness endpoint in the original IDE involved independent panel review of pre- and 12-week post-treatment photographs of the treatment area. For the per protocol population, the correct baseline photograph identification rate by the independent panel reviewers was 91.4%. This data remains unchanged since the close of the study and was previously submitted in K151179.
The secondary effectiveness endpoint for subject satisfaction was assessed by an IRBapproved questionnaire administered at 12-week post-treatment. Seventy-five percent of the subjects reported that they agree or strongly agree that the treatment made their chin look more toned. This data remains unchanged since the close of the study and was previously submitted in K151179.
This retrospective study used the per-protocol population (n=58) for analysis. One subject was excluded from the analysis due to excessive hair in the submental region, therefore the number of subjects used for analysis was 57. Right and left lateral photographic views of the submental area taken at baseline and at the 12-week post final treatment visit were included in the analysis. Each photo was cropped and masked prior to evaluation. A board certified plastic surgeon identified the following anatomical points on each photograph: the lateral canthus, the anterior most point where the nostril meets the columella, and the point where the chin meets the neck (submental crease). A series of lines were drawn on the photo using AutoCAD software and areas in the submental region were measured. A responder analysis was performed with the criteria being ≥ 20 mm² decrease in area as measured on both the right lateral and left lateral views of the region. A second analysis was performed comparing the responders with results from the independent physician review panel of photos conducted for the previous study.
<sup>1</sup> Oni, G, Hoxworth R, Teotia, S, et al. Evaluation of a Microfocused Ultrasound System for Improving Skin Laxity and Tightening in the Lower Face. Aesthetic Surgery Journal 2014;34(7):1099-1110.
{6}------------------------------------------------
Further evidence of treatment effectiveness found the data from photographic analysis of the treated areas indicates 77.2% (44/57) of subjects exhibited a ≥ 20 mm² area reduction in the submental area and neck. Of those 44 subjects, 42 (95.5%) were correctly identified by the physician panel as having a visible response. The results of this photographic analysis indicate that the CoolSculpting procedure on the submental area can also affect the appearance of lax tissue in the submental area.
The study design and results are summarized in the table below:
| Study Design | Retrospective |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sample Size for<br>analysis | 57 |
| Effectiveness Results | The per protocol population consisted of all the treated<br>subjects followed for 12 weeks with weight change of no<br>more than 5% of total body weight at the time the 12 week<br>images were taken. For the per protocol population, the<br>correct baseline photograph identification rate by the<br>independent panel reviewers was 91.4%.<br><br>Data from photographic analysis showed effect on the<br>appearance of lax tissue (≥ 20 mm²) from baseline to 12<br>weeks post-treatment in 77.2% (44/57) of subjects. Of those<br>44 subjects, 42 (95.5%) were correctly identified by the<br>physician panel as having a visible response.<br><br>Subject satisfaction data from original IDE: 75% of the<br>subjects reported that they agreed that the treatment made<br>their chin look more toned. |
| Safety Results | The primary safety endpoint of the original IDE approved<br>study was met and remains unchanged since the close of the<br>study. All device- and/or procedure-related adverse events<br>have resolved spontaneously. |
Table 5.1. Summary of Retrospective Study for effect on the appearance of lax tissue in the Submental area
The data provides reasonable assurance of safety and effectiveness to demonstrate substantial equivalence of the CoolSculpting System for the additional indication for use of affecting the appearance of lax tissue in the submental area.
{7}------------------------------------------------
## VIII. CONCLUSION:
The ZELTIQ CoolSculpting System is the same as the device cleared in K171069. No changes have been made to the device to accommodate this indication of affecting the appearance of lax tissue in the submental area.
The CoolSculpting System is substantially equivalent to the predicate device for the revised indications for use. The clinical data shows visible effect in the appearance of lax tissue meeting both the safety and effectiveness endpoints.
The CoolSculpting System has been established as safe and effective through many prior clearances (the most recent is K171069). It was demonstrated to be safe and effective for the submental area in K151179.
The CoolSculpting System has not changed to accommodate this expanded indication. The clinical data indicates the same safety and effectiveness for the CoolSculpting System for this expanded intended use. As such, the CoolSculpting System is substantially equivalent to the predicate device (previously cleared CoolSculpting System, K171069).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.