← Product Code [FGB](/productcode/FGB) · K172098

# Medical Video Endoscope System (K172098)

_Zhuhai Pusen Medical Technology Co., Ltd. · FGB · Aug 31, 2017 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K172098

## Device Facts

- **Applicant:** Zhuhai Pusen Medical Technology Co., Ltd.
- **Product Code:** [FGB](/productcode/FGB.md)
- **Decision Date:** Aug 31, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

This instrument has been designed to be used with endo-therapy accessories such as a biopsy forceps and other ancillary equipment for endoscopy and endoscopic surgery within urinary tract and interior of the kidney.

## Device Story

Digital flexible ureteroscope system comprising Uscope (single-use, sterile) and Eview (video processing unit/touch PC). CMOS sensor at distal tip captures mucosal images; LED in handle provides illumination via fiber optics. Electric signals transmitted to Eview for display. Used in hospitals by trained physicians for diagnostic and therapeutic procedures in urinary tract/kidney. Enables visualization of capillary network and mucosal morphology; facilitates use of ancillary tools like laser fibers and forceps. Single-use design eliminates reprocessing requirements compared to reusable predicates.

## Clinical Evidence

Bench testing only. Performance verified per ISO 8600-1:2015 (dimensions, deflection, FOV, image quality, illumination, leak, flow rate). Electrical safety tested per IEC 60601-1 and IEC 60601-2-18. EMC tested per IEC 60601-1-2. Biocompatibility (cytotoxicity, irritation, sensitization) evaluated per ISO 10993-1. Sterilization validated per ISO 11135:2014.

## Technological Characteristics

Flexible ureteroscope. Materials: SUS 304 stainless steel, PEBAX, silicone rubber, ABS, Fluorinated ethylene propylene. CMOS imaging sensor; LED illumination. Powered by main line or lithium battery. Single-use, EO sterilized. Connectivity: Eview video system/touch PC. Standards: ISO 8600-1, ISO 10993-1, ISO 11607, ISO 11135, IEC 60601-1, IEC 60601-2-18.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Karl Storz Flexible Video-Uretero-Renoscope System ([K131369](/device/K131369.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of three human profiles facing to the right, resembling a bird in flight.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 31, 2017

Zhuhai Pusen Medical Technology Co., Ltd. % Dave Yungvirt CEO Third Party Review Group, LLC The Old Station House 24 Lackawanna Place Millburn, NJ 07041

Re: K172098

> Trade/Device Name: Medical Video Endoscope System Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FGB Dated: August 22, 2017 Received: August 23, 2017

Dear Dave Yungvirt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Charles Viviano -S

For Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K172098

Device Name Medical Video Endoscope System

Indications for Use (Describe)

This instrument has been designed to be used with endo-therapy accessories such as a biopsy forceps and other ancillary equipment for endoscopy and endoscopic surgery within urinary tract and interior of the kidney.

| Type of Use (Select one or both, as applicable)                          |                                                                         |
|--------------------------------------------------------------------------|-------------------------------------------------------------------------|
| <div> <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Traditional 510(k)

# 510(k) SUMMARY

## K172098

#### Submitter's Name, Address and Phone and Fax Number A.

Name & Address of the manufacturer: Zhuhai Pusen Medical Technology Co., Ltd.

5/F, Building 1, No 33, Keji San Road, High-tech Zone, Tangjiawan Town, Zhuhai, Guangdong, PRC.

Phone:

Fax:

86-07566880096

86-07566880401

#### Contact person: B.

Ms. Wang ChangShen

Manager, Regulatory Affairs

Phone: 86-07566880865

#### C. Date prepared:

August 17, 2017

#### Name of the device: D.

Trade name: Medical Video Endoscope system Common name: Ureteroscope and Accessories, Flexible/rigid Model Number of the Handle: PU3022、PU3022A Model Number of the Video system: UTV100 21 CFR 876.1500 Endoscope and accessories. Class II Classification: Product Code: FGB

#### Predicate Device: E.

Trade name: Karl Storz Flexible Video-Uretero-Renoscope System

> Section 5: 510(k) summary 5-1

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Zhuhai Pusen Medical Technology Co., Ltd. The logo consists of the word "PUSEN" in a stylized, sans-serif font, with rounded corners and a light blue color. Below the logo is the company name, "Zhuhai Pusen Medical Technology Co., Ltd.", in a smaller, gray font.

Traditional 510(k)

Common name: Ureteroscope and Accessories, Flexible/rigid

Classification: Endoscope and accessories. 21 CFR 876.1500. Class II

Premarket Notification: Storz K131369

#### Device description: F.

The Medical Video Endoscope system is designed to provide image solution for endoscopy and endoscopic surgery, and perform procedures in the urinary tract and interior of the kidney using appropriate accessory devices (e.g. laser fibers, forceps baskets).

The optical fiber transmits the light emitting from the LED light built in the handle to the coelom. The complementary metal-oxide-semiconductor CMOS imaging sensor in the distal tip of the insertion shaft receives the reflected light from the surface of mucosa and transforms the light into electric signal. The cable transfers the electric signal to the data processing center of the Eview, where the electric signal is processed. As a result the color imaging of the capillary network and mucosal morphology is displayed on the screen of the Eview.

The Medical Video Endoscope system is a digital flexible ureteroscope system that consists of the Eview (the video system, touch PC with data processing center) and the Uscope. The Uscope is provided sterile (sterilized by EO) and intended to be single-use.

The Uscope, as provided sterile, contains an insertion portion, handpiece and the cable. The handpiece is made with polymer plastic, Contacted with users. The insertion portion, as the part connected with patients, includes the sheath which is braided tube made of SUS 304 stainless steel and PEBAX, the controllable portion and the distal tip. The controllable portion is made of SUS 304 stainless steel, covered with silicone rubber. The distal tip is made with ABS. There is a working channel that terminates at the distal tip, which is indirect patient-contacting component, made of Fluorinated ethylene propylene.

The Eview, the video system contains touch PC with data processing center and provide energy to Uscope. The Eview can be powered by the main line or lithium battery.

#### Intended Use رخ

This instrument has been designed to be used with endo-therapy accessories such as

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image contains the logo for Zhuhai Pusen Medical Technology Co., Ltd. The logo consists of the word "PUSEN" in a stylized, sans-serif font, with rounded corners and a light blue color. Below the logo, the company name "Zhuhai Pusen Medical Technology Co., Ltd." is written in a smaller, gray font.

a biopsy forceps and other ancillary equipment for endoscopy and endoscopic surgery within urinary tract and interior of the kidney.

The intended use of subject and predicate devices are the same. Both of the subject and predicate devices are indicated for endoscopic examination in the urinary tract and can be used to examine the interior of the kidney, and using additional accessories to perform various diagnostic and therapeutic procedures.

The Medical Video Endoscope system is for use in a hospital or qualified medical institution. The system is only to be used by skilled physicians trained in clinical endoscopic techniques and procedures.

# H. Technological Characteristics

The technological principle for both the subject and predicate devices is based on the use of an LED light integrated in the handle and fiber light guides to illuminate the cavity under examination, and the CMOS imaging sensor located at the tip of the insertion portion to transfers the video signal to the image system.

The following basic technological elements are the same or similar for the subject and predicate devices:

| Compared items                                 | Subject device                              | Predicate device                            |
|------------------------------------------------|---------------------------------------------|---------------------------------------------|
| Same technological characteristics             |                                             |                                             |
| Scope type                                     | Flexible                                    | Flexible                                    |
| anatomical site                                | Urinary tract and interior of<br>the kidney | Urinary tract and interior<br>of the kidney |
| target population                              | Adults                                      | Adults                                      |
| where used (hospital,<br>home, ambulance, etc) | Hospitals                                   | Hospitals                                   |
| Digital<br>video<br>technology                 | CMOS                                        | CMOS                                        |
| Illumination source                            | LED                                         | LED                                         |

{6}------------------------------------------------

Traditional 510/k)

Image /page/6/Picture/1 description: The image shows the logo for Zhuhai Pusen Medical Technology Co., Ltd. The logo consists of the word "PUSEN" in a stylized, sans-serif font, with the letters in a light blue color. Below the logo is the company name, "Zhuhai Pusen Medical Technology Co., Ltd.", in a smaller, gray font.

|                                       |                            | Traditional 510(k)     |
|---------------------------------------|----------------------------|------------------------|
| Up/down deflection (° )               | Up 270 °<br>Down 270 °     | Up 270 °<br>Down 270 ° |
| Direction of view                     | 0 °                        | 0 °                    |
| Similar technological characteristics |                            |                        |
| Field of view                         | 120 °                      | 90 °                   |
| Maximum insertion portion width (mm)  | 3.2                        | 2.9                    |
| Working length(mm)                    | 650                        | 700                    |
| Minimum instrument channel width (mm) | 1.0                        | 1.2                    |
| Image system                          | UTV100 as the image system | Image 1HD system       |

The main difference between the Medical Video Endoscope system and the predicate devices is reusability. The Uscope of the subject device is single-use and provided sterile, while the predicate devices are reusable. This does not create a difference in intended use.

The technological characteristics of the Medical Video Endoscope system and the predicate device are substantially equivalent.

#### Non-Clinical Performance Data I.

The Medical Video Endoscope system has been verified for its safety and effectivity based on the following performance data.

Electrical safety of the system was evaluated in accordance with IEC 60601-1:2012, and AAMI/ANSI ES60601-1:2005/(R)2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012 and IEC 60601-2-18:2009. Electromagnetic compatibility was evaluated in accordance with IEC 60601-1-2:2007. All evaluation acceptance criteria were met.

Biocompatibility evaluation for the Medical Video Endoscopy system was conducted

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image features the word "PUSEN" in a stylized, sans-serif font, rendered in a light blue color. Below the wordmark, in a smaller, darker font, is the text "Zhuhai Pusen Medical Technology Co., Ltd." The text appears to be the name of a company, possibly a medical technology firm, with the wordmark serving as its logo.

Traditional 510(k)

in accordance with the Guidance document "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" June 16, 2016, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA.

The body contact category is "Surface – Mucosal Membrane" with a contact duration of "Limited" (<24 hours). The Following tests were performed as recommended: Cytotoxicity, Irritation and Sensitization. All evaluation acceptance criteria were met.

Sterile barrier systems were evaluated in accordance with ISO 11607:2006. Sterilization Process has been validated accordance with ISO 11135:2014.

Technological characteristics have been tested for its functions as intended, including verification of performance characteristics per ISO 8600-1:2015(Fourth Edition) (The maximum insertion portion width, the minimum instrument channel width, field of view, direction of view, deflection) and performances characteristics relevant to functions as intended (Image quality, Illumination, Articulation, Leak, Flow rate of water).

The results of Non-Clinical Performance testing demonstrate that the Medical Video Endoscope is considered safe and effective for its intended use.

## Conclusion:

The Medical Video Endoscope is substantially equivalent to the predicate device and does not raise any questions of safety and effectiveness.

---

**Source:** [https://fda-staging.innolitics.com/device/K172098](https://fda-staging.innolitics.com/device/K172098)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).