K172092 · Renovo, Inc. · HRX · Oct 4, 2017 · Orthopedic
Device Facts
Record ID
K172092
Device Name
Reprocessed Shavers/Burs
Applicant
Renovo, Inc.
Product Code
HRX · Orthopedic
Decision Date
Oct 4, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.1100
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Reprocessed Linvatec shavers and burs are intended for use with the Linvatec Integrated Drive/Pump System to perform resection of soft tissue and bone. The fields of application include Otolaryngology, Neurological Surgical, Orthopedic, and Spinal surgical procedures. The Reprocessed Smith & Nephew shavers and burs are intended for use with the Smith & Nephew DYONICS Power II Shaver System control unit, for resection of soft and osseous tissues including but not limited to, use in large articular cavities, small articular cavities, and Functional Endoscopic Sinus Surgery (FESS). The FESS application is limited to those small blades which are appropriate for the procedure. The Reprocessed Styker Shavers and Burs are intended for use with the Styker Total Performance System Shaver handpiece and the Stryker Total Performance System by surgeons in the following orthopedic joints: Knee Shoulder Ankle Elbow Wrist Hip Temporamandibular The Reprocessed Styker Shavers and Burs will be used to resect tissue and bur bone. Examples of uses of the product would include resection of torn knee cartilage, subacromial decompression and resection of synovial tissue in other joints.
Device Story
Reprocessed single-use arthroscopic/laparoscopic shavers and burs; used with OEM-specific motorized handpieces (Linvatec, Smith & Nephew, Stryker). Shavers feature rotating inner blades in outer shafts to cut tissue; burs feature abrasive surfaces to grind bone. Housing includes distal window for targeted cutting/protection. Handpiece provides suction for tissue removal. Used in OR by surgeons for joint/sinus procedures. Reprocessing includes cleaning, sterilization, and performance verification to ensure equivalence to new OEM devices. Benefits include cost-effective reuse of surgical instruments while maintaining clinical performance and safety standards.
Clinical Evidence
No clinical data. Substantial equivalence based on benchtop performance testing, including functional testing (sharpness, bending, drop tests), cleaning validation (residual protein/carbohydrates), biocompatibility (cytotoxicity, irritation, toxicity, pyrogenicity, sensitization), and sterilization validation.
Technological Characteristics
Metal shafts with polymer hubs; rotating inner/outer component design. Lubricants used to reduce friction/heat. Powered by user-supplied OEM handpieces. Sterilization via EtO. No electrical/EMC components. Dimensions/configurations vary by model.
Indications for Use
Indicated for resection of soft tissue and bone in orthopedic joints (knee, shoulder, ankle, elbow, wrist, hip, temporomandibular), otolaryngology, neurological surgery, spinal procedures, and FESS (small blades only).
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
Smith & Nephew DYONICS Power II Shaver System (K062849)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image is a logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is an image of three faces in profile, one behind the other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 4, 2017
Renovo. Inc. % Mr. Robert Packard Medical Device Academy, Inc. 345 Lincoln Hill Rd. Shrewsbury, Vermont 05738
Re: K172092
Trade/Device Name: Reprocessed Shavers/Burs Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX Dated: September 21, 2017 Received: September 21, 2017
Dear Mr. Packard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm_for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely, Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172092
Device Name Reprocessed Linvatec Shavers/Burs
Indications for Use (Describe)
The Reprocessed Linvatec shavers and burs are intended for use with the Linvatec Integrated Drive/Pump System to perform resection of soft tissue and bone. The fields of application include Otolaryngology, Neurological Surgical, Orthopedic, and Spinal surgical procedures.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Reprocessed Single-Use Linvatec Shavers & Burs Device Models Subject to Clearance (120 Linvatec devices in total)
| OEM Model # | Description | Diameter | Picture | OEM |
|-------------|----------------------------------------------------------------------------------------------------------|----------|--------------------------------------|-----------------|
| DPS-C002 | Large Hub Shaver Blades,<br>UltraGator | 4.2mm | Image: Shaver blade | Conmed/Linvatec |
| DPS-C003 | Large Hub Shaver Blades,<br>UltraGator | 5.5mm | Image: Shaver blade | Conmed/Linvatec |
| DPS-C010 | Large Hub Shaver Blades, UltraFRR | 4.2mm | Image: Shaver blade | Conmed/Linvatec |
| DPS-C011 | Large Hub Shaver Blades, UltraFRR | 5.5mm | Image: Shaver blade | Conmed/Linvatec |
| DPS-C005 | Large Hub Shaver Blades,<br>UltraTiger | 4.2mm | Image: Shaver blade | Conmed/Linvatec |
| 9399A | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Great White<br>Blade | 3.5mm | Image: Shaver blade | Conmed/Linvatec |
| 9299A | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Great White<br>Blade | 4.2mm | Image: Shaver blade | Conmed/Linvatec |
| 9297A | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Great White<br>Blade | 4.8mm | Image: Shaver blade | Conmed/Linvatec |
| 9599A | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Great White<br>Blade | 5.5mm | Image: Shaver blade | Conmed/Linvatec |
| EL9299 | Large Hub Shaver Blades, Large<br>Hub, Merlin, Shaver, Great White<br>Blade | 4.2mm | Image: Shaver blade | Conmed/Linvatec |
| 35ULT-ST-ZZ | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Ultracut | 3.5mm | Image: Shaver blade | Conmed/Linvatec |
| C9405 | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Ultracut | 4.2mm | Image: Shaver blade | Conmed/Linvatec |
| C9415 | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Ultracut | 5.5mm | Image: Shaver blade | Conmed/Linvatec |
| EL9405 | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Ultracut | 4.2mm | Image: Shaver blade | Conmed/Linvatec |
| 9298A | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Mako Blade | 4.2mm | Image: Shaver blade | Conmed/Linvatec |
| C9299A | Large Hub Shaver Blades, Pre-Bent<br>15°, Great White Blade | 4.2mm | Image: Shaver blade | Conmed/Linvatec |
| C9248 | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Full-Radius<br>Resector | 3.5mm | Image: Shaver blade | Conmed/Linvatec |
| 9247A | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Full-Radius<br>Resector | 4.2mm | Image: Shaver blade | Conmed/Linvatec |
| C9245 | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Full-Radius<br>Resector | 4.8mm | Image: Shaver blade | Conmed/Linvatec |
| 9246A | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Full-Radius<br>Resector | 5.5mm | Image: Shaver blade | Conmed/Linvatec |
| 42FRR-RA-ZZ | Large Hub Shaver Blades, Large<br>Hub, ReAct, Shaver, Great White<br>Blade | 3.5mm | Image: blade | Conmed/Linvatec |
| 42FRR-RA-ZZ | Large Hub Shaver Blades, Large<br>Hub, ReAct, Shaver, Great White<br>Blade | 4.2mm | Image: blade | Conmed/Linvatec |
| C9943A | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Full-Radisu<br>Resector | 2.0mm | Image: blade | Conmed/Linvatec |
| C9944A | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Full-Radius<br>Resector Blade | 2.0mm | Image: blade | Conmed/Linvatec |
| C9264 | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Gator Blade | 3.5mm | Image: blade | Conmed/Linvatec |
| 9263A | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Gator Blade | 4.2mm | Image: blade | Conmed/Linvatec |
| C9262 | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Gator Blade | 4.8mm | Image: blade | Conmed/Linvatec |
| 9260A | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Gator Blade | 5.5mm | Image: blade | Conmed/Linvatec |
| 35GTR-RA-ZZ | Large Hub Shaver Blades, Large<br>Hub, ReAct, Shaver, Gator | 3.5mm | Image: blade | Conmed/Linvatec |
| 42GTR-RA-ZZ | Large Hub Shaver Blades, Large<br>Hub, ReAct, Shaver, Gator | 4.2mm | Image: blade | Conmed/Linvatec |
| EL9263A | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Gator | 4.2mm | Image: blade | Conmed/Linvatec |
| EL9260A | Large Hub Shaver Blades , Large<br>Hub, Sterling, Shaver, Gator | 5.5mm | Image: blade | Conmed/Linvatec |
| C9960A | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Gator | 2.0mm | Image: blade | Conmed/Linvatec |
| C9961A | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Gator | 2.9mm | Image: blade | Conmed/Linvatec |
| C9253 | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Cuda | 3.5mm | Image: blade | Conmed/Linvatec |
| C9254 | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Cuda | 4.2mm | Image: blade | Conmed/Linvatec |
| C9258 | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Cuda | 4.8mm | Image: blade | Conmed/Linvatec |
| C9255 | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Cuda | 5.5mm | Image: blade | Conmed/Linvatec |
| P9355 | Large Hub Shaver Blades, Large<br>Hub, Merlin, Shaver, Cuda | 3.7mm | Image: blade | Conmed/Linvatec |
| 35CUD-RA-ZZ | Large Hub Shaver Blades, Large<br>Hub, ReAct®, Shaver, Cuda | 3.5mm | Image: blade | Conmed/Linvatec |
| 42CUD-RA-ZZ | Large Hub Shaver Blades, Large<br>Hub, ReAct, Shaver, Cuda | 4.2mm | Image: blade | Conmed/Linvatec |
| C9951A | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Cuda | 2.0mm | Image: blade | Conmed/Linvatec |
| C9952A | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Cuda | 2.9mm | Image: blade | Conmed/Linvatec |
| C9240 | Large Hub Shaver Blades, Large<br>Hub, Sterling®, Shaver, Tiger® | 3.5mm | Image: blade | Conmed/Linvatec |
| C9242 | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Tiger | 4.2mm | Image: blade | Conmed/Linvatec |
| 35TGR-RA-ZZ | Large Hub Shaver Blades, Large<br>Hub, ReAct, Shaver, Tiger | 3.5mm | Image: blade | Conmed/Linvatec |
| | | | | |
| 42TGR-RA-ZZ | Large Hub Shaver Blades, Large<br>Hub, ReAct, Shaver, Tiger | 4.2mm | Conmed/Linvatec | |
| C9283 | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, End Cutter | 3.5mm | Image: [C9283] Conmed/Linvatec | |
| C9284 | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, End Cutter | 4.2mm | Image: [C9284] Conmed/Linvatec | |
| C9290 | Large Hub Burs and Routers, Large<br>Hub, Sterling®, Shaver, Dragon End<br>Cutter | 3.5mm | Image: [C9290] Conmed/Linvatec | |
| C9291 | Large Hub Burs and Routers, Large<br>Hub, Sterling, Shaver, Dragon End<br>Cutter | 4.2mm | Image: [C9291] Conmed/Linvatec | |
| C9292 | Large Hub Burs and Routers, Large<br>Hub, Sterling, Shaver, Dragon End<br>Cutter | 5.5mm | Image: [C9292] Conmed/Linvatec | |
| C9760 | Large Hub Burs and Routers, Large<br>Hub, Sterling, Shaver, Meniscus<br>Cutter | 3.5mm | Image: [C9760] Conmed/Linvatec | |
| 9761A | Large Hub Burs and Routers, Large<br>Hub, Sterling, Shaver, Meniscus<br>Cutter | 4.2mm | Image: [9761A] Conmed/Linvatec | |
| C9762 | Large Hub Burs and Routers, Large<br>Hub, Sterling, Shaver, Meniscus<br>Cutter | 5.5mm | Image: [C9762] Conmed/Linvatec | |
| C9274 | Large Hub Burs and Routers, Large<br>Hub, Sterling, Shaver, Slotted<br>Whisker | 4.2mm | Image: [C9274] Conmed/Linvatec | |
| C9275 | Large Hub Burs and Routers, Large<br>Hub, Sterling, Shaver, Slotted<br>Whisker | 5.5mm | Image: [C9275] Conmed/Linvatec | |
| H9110 | Large Hub Burs and Routers, Large<br>Hub, Sterling®, Bur, Spherical | 3.5mm | Image: [H9110] Conmed/Linvatec | |
| H9111 | Large Hub Burs and Routers, Large<br>Hub, Sterling, Bur, Spherical | 4.5mm | Image: [H9111] Conmed/Linvatec | |
| H9112 | Large Hub Burs and Routers, Large<br>Hub, Sterling, Bur, Spherical | 5.5mm | Image: [H9112] Conmed/Linvatec | |
| H9112RH | Large Hub Burs and Routers, Large<br>Hub, Sterling, Bur, Spherical, 20°<br>Reduced Hood | 5.5mm | Image: [H9112RH] Conmed/Linvatec | |
| EL9111RH | Large Hub Burs and Routers, Large<br>Hub, Extended Length, Sterling,<br>Bur, Spherical, 20° Reduced Hood | 4.5mm | Image: [EL9111RH] Conmed/Linvatec | |
| 55SPH-EL-RH | Large Hub Burs and Routers, Large<br>Hub, Extended Length, Sterling,<br>Bur, Spherical 20° Reduced Hood | 5.5mm | Image: [55SPH-EL-RH] Conmed/Linvatec | |
| C9911A | Large Hub Burs and Routers, Large<br>Hub, Small Joint, Sterling, Bur,<br>Spherical | 2.9mm | Image: [C9911A] Conmed/Linvatec | |
| H9121 | Large Hub Burs and Routers, Large<br>Hub, Sterling, Bur, Tapered | 4.0mm | Image: [H9121] Conmed/Linvatec | |
| H9121RH | Large Hub Burs and Routers, Large<br>Hub, Sterling, Bur, Tapered, 20°<br>Reduced Hood | 4.0mm | Image: [H9121RH] Conmed/Linvatec | |
| H9101 | Large Hub Burs and Routers, Large<br>Hub, Sterling, Bur, Oval | 4.0mm | Image: bur | Conmed/Linvatec |
| 40VAL-ST-FH | Large Hub Burs and Routers, Large<br>Hub, Sterling, Bur, Oval, Fully<br>Hooded | 4.0mm | Image: bur | Conmed/Linvatec |
| H9101RH | Large Hub Burs and Routers, Large<br>Hub, Sterling, Bur, Oval, 20°<br>Reduced Hood | 4.0mm | Image: bur | Conmed/Linvatec |
| H9101LH | Large Hub Burs and Routers, Large<br>Hub, Sterling, Bur, Oval, Left Helix | 4.0mm | Image: bur | Conmed/Linvatec |
| H9103RH | Large Hub Burs and Routers, Large<br>Hub, Sterling, Bur, Oval, 20º<br>Reduced Hood | 5.0mm | Image: bur | Conmed/Linvatec |
| H9102 | Large Hub Burs and Routers, Large<br>Hub, Sterling, Bur, Oval | 6.0mm | Image: bur | Conmed/Linvatec |
| H9201LH | Large Hub Burs and Routers, Large<br>Hub, Sterling, Bur, Oval, Left Helix | 6.0mm | Image: bur | Conmed/Linvatec |
| EL9101 | Large Hub Burs and Routers, Large<br>Hub, Extended Length, Sterling,<br>Bur, Oval | 4.0mm | Image: bur | Conmed/Linvatec |
| 60VAL-EL-RH | Large Hub Burs and Routers, Large<br>Hub, Sterling, Bur, Oval, 20°<br>Reduced Hood | 6.0mm | Image: bur | Conmed/Linvatec |
| H9118 | Large Hub Burs and Routers, Large<br>Hub, Sterling®, Bur, Cyclone | 4.5mm | Image: bur | Conmed/Linvatec |
| H9119 | Large Hub Burs and Routers, Large<br>Hub, Sterling, Bur, Cyclone | 5.5mm | Image: bur | Conmed/Linvatec |
| H9131 | Large Hub Burs and Routers, Large<br>Hub, Sterling, Router, Vortex, 5<br>Flutes, Hooded | 4.5mm | Image: bur | Conmed/Linvatec |
| H9134 | Large Hub Burs and Routers, Large<br>Hub, Sterling, Router, Vortex, 5<br>Flutes, Unhooded | 4.5mm | Image: bur | Conmed/Linvatec |
| H9132 | Large Hub Burs and Routers, Large<br>Hub, Sterling, Router, Vortex, 5<br>Flutes, Hooded | 6.0mm | Image: bur | Conmed/Linvatec |
| 29VTX-SJ-ZZ | Large Hub Burs and Routers, Large<br>Hub, Small Joint, Sterling, Bur,<br>Vortex Router | 2.9mm | Image: bur | Conmed/Linvatec |
| 35VTX-SJ-ZZ | Large Hub Burs and Routers, Large<br>Hub, Small Joint, Sterling, Bur,<br>Vortex Router | 3.5mm | | Conmed/Linvatec |
| H9133 | Large Hub Burs and Routers, Large<br>Hub, Sterling, Lighting | 4.5mm | Image: bur | Conmed/Linvatec |
| M9299A | Micro Hub Blades, Micro Hub,<br>Sterling®, Shaver, Great White | 4.2mm | Image: bur | Conmed/Linvatec |
| | | | | |
| M9298A | Micro Hub Blades, Micro Hub, Sterling, Shaver, Mako | 4.2mm | | Conmed/Linvatec |
| C9940 | Micro Hub Blades, Micro Hub, Sterling®, Blade, Full-Radius Resector | 2.0mm | | Conmed/Linvatec |
| C9941 | Micro Hub Blades, Micro Hub, Sterling, Blade, Full-Radius Resector | 2.9mm | | Conmed/Linvatec |
| C9942 | Micro Hub Blades, Micro Hub, Sterling, Blade, Full-Radius Resector | 3.5mm | | Conmed/Linvatec |
| MC9248 | Micro Hub Blades, Micro Hub, Sterling, Blade, Full-Radius Resector | 3.5mm | | Conmed/Linvatec |
| M9247A | Micro Hub Blades, Micro Hub, Sterling, Blade, Full-Radius Resector | 4.2mm | | Conmed/Linvatec |
| C9960 | Micro Hub Blades, Micro Hub, Sterling, Blade, Gator | 2.0mm | | Conmed/Linvatec |
| C9961 | Micro Hub Blades, Micro Hub, Sterling, Blade, Gator | 2.9mm | | Conmed/Linvatec |
| C9962 | Micro Hub Blades, Micro Hub, Sterling, Blade, Gator | 3.5mm | | Conmed/Linvatec |
| MC9264 | Micro Hub Blades, Micro Hub, Sterling, Blade, Gator | 3.5mm | | Conmed/Linvatec |
| MC9263A | Micro Hub Blades, Micro Hub, Sterling, Blade, Gator | 4.2mm | | Conmed/Linvatec |
| C9950 | Micro Hub Blades, Micro Hub, Sterling®, Blade, Cuda | 2.0mm | | Conmed/Linvatec |
| C9951 | Micro Hub Blades, Micro Hub, Sterling, Blade, Cuda | 2.9mm | | Conmed/Linvatec |
| C9952 | Micro Hub Blades, Micro Hub, Sterling, Blade, Cuda | 3.5mm | | Conmed/Linvatec |
| MC9253 | Micro Hub Blades, Micro Hub, Sterling, Blade, Cuda | 3.5mm | | Conmed/Linvatec |
| MC9254 | Micro Hub Blades, Micro Hub, Sterling, Blade, Cuda | 4.2mm | | Conmed/Linvatec |
| MC9258 | Micro Hub Blades, Micro Hub, Sterling, Blade, Cuda | 4.8mm | | Conmed/Linvatec |
| MC9290 | Micro Hub Blade, Micro Hub, Sterling, Blade, Dragon, End Cutters | 3.5mm | | Conmed/Linvatec |
| MC9291 | Micro Hub Blades, Micro Hub, Sterling, Blade, Dragon, End Cutter | 4.2mm | | Conmed/Linvatec |
| MC9292 | Micro Hub Blades, Micro Hub, Sterling, Blade, Dragon, End Cutter | 5.5mm | | Conmed/Linvatec |
| C9981 | Micro Hub Blades, Micro Hub, Sterling, Blade, Slotted Whisker | 2.9mm | | Conmed/Linvatec |
| MC9274 | Micro Hub Blades, Micro Hub, Sterling, Blade, Slotted Whisker | 4.2mm | | Conmed/Linvatec |
| | | | | |
| C9911A | Large Hub Burs and Routers, Large<br>Hub, Sterling, Bur, Spherical | 2.9mm | Image: bur | Conmed/Linvatec |
| C9912 | Micro Hub Burs and Routers,<br>Micro Hub, Sterling, Bur, Spherical | 3.5mm | Image: bur | Conmed/Linvatec |
| MH9111 | Micro Hub Burs and Routers,<br>Micro Hub, Sterling, Bur, Spherical | 4.5mm | Image: bur | Conmed/Linvatec |
| MH9112 | Micro Hub Burs and Routers,<br>Micro Hub, Sterling, Bur, Spherical | 5.5mm | Image: bur | Conmed/Linvatec |
| MH9102 | Micro Hub Burs and Routers,<br>Micro Hub, Sterling, Bur, Oval | 6.0mm | Image: bur | Conmed/Linvatec |
| MC9106 | Micro Hub Burs and Routers,<br>Micro Hub, Sterling, Bur, Oval, Left<br>Helix | 6.0mm | Image: bur | Conmed/Linvatec |
| C9990 | Micro Hub Burs and Routers,<br>Micro Hub, Sterling, Router, Vortex<br>Two Flutes | 2.9mm | Image: bur | Conmed/Linvatec |
| C9991 | Micro Hub Burs and Routers,<br>Micro Hub, Sterling, Router, Vortex<br>Two Flutes | 3.5mm | Image: bur | Conmed/Linvatec |
| MH9131 | Micro Hub Burs and Routers,<br>Micro Hub, Sterling, Router, Vortex<br>Two Flutes | 4.5mm | Image: bur | Conmed/Linvatec |
| MH9132 | Micro Hub Burs and Routers,<br>Micro Hub, Sterling, Router, Vortex<br>Two Flutes | 6.0mm | Image: bur | Conmed/Linvatec |
| C9405A | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Ultracut | 4.2mm | Image: shaver | Conmed/Linvatec |
| C9298A | Large Hub Shaver Blades, Large<br>Hub, Sterling, Shaver, Mako | 5.5mm | Image: shaver | Conmed/Linvatec |
| 37GRW-ME-ZZ | Large Hub Shaver Blades, Large<br>Hub, Merlin, Shaver, Great White | 3.7mm | Image: shaver | Conmed/Linvatec |
| C9299P | Large Hub Shaver Blades, Large<br>Hub, Merlin, Shaver, Great<br>White,Pre-Bent 15° | 4.2mm | Image: shaver | Conmed/Linvatec |
{4}------------------------------------------------
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172092
Device Name
Reprocessed Smith & Nephew Shavers/Burs
### Indications for Use (Describe)
The Reprocessed Smith & Nephew shavers and burs are intended for use with the Smith & Nephew DYONICS Power II Shaver System control unit, for resection of soft and osseous tissues including but not limited to, use in large articular cavities, small articular cavities, and Functional Endoscopic Sinus Surgery (FESS). The FESS application is limited to those small blades which are appropriate for the procedure.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| Prescription Use (Part 21 CFR 801 Subpart D) | |
| Over-The-Counter Use (21 CFR 801 Subpart C) | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{10}------------------------------------------------
| Reprocessed Single-Use Smith & Nephew Shavers & Burs Device Models Subject to Clearance |
|-----------------------------------------------------------------------------------------|
| (63 Smith & Nephew devices in total) |
| OEM Model # | Description | Diameter | Picture | OEM |
|-------------|-------------------------------------------------------------------------------------------------|----------|--------------------------------------------------------------------------|----------------|
| 7205316 | DYONICS Straight Blades,<br>TURBOWHISKER, Navy Blue Latch | 4.5mm | Image: DYONICS Straight Blades, TURBOWHISKER, Navy Blue Latch | Smith & Nephew |
| 7205315 | DYONICS Straight Blades<br>TURBOTRIMMER, Powder Blue<br>Latch | 4.5mm | Image: DYONICS Straight Blades TURBOTRIMMER, Powder Blue Latch | Smith & Nephew |
| 7205310 | DYONICS Straight Blades<br>SYNOVATOR, Forest Green Latch | 4.5mm | Image: DYONICS Straight Blades SYNOVATOR, Forest Green Latch | Smith & Nephew |
| 7205311 | DYONICS Straight Blades<br>SYNOVATOR, Melon Yellow Latch | 5.5mm | Image: DYONICS Straight Blades SYNOVATOR, Melon Yellow Latch | Smith & Nephew |
| 7205308 | DYONICS Straight Blades Cutter,<br>Red Latch | 3.5mm | Image: DYONICS Straight Blades Cutter, Red Latch | Smith & Nephew |
| 7205309 | DYONICS Straight Blades Cutter,<br>Blue Latch | 4.5mm | Image: DYONICS Straight Blades Cutter, Blue Latch | Smith & Nephew |
| 7205305 | DYONICS Straight Blades Full<br>Radius, Beige Latch | 3.5mm | Image: DYONICS Straight Blades Full Radius, Beige Latch | Smith & Nephew |
| 7205306 | DYONICS Straight Blades Full<br>Radius, Yellow Latch | 4.5mm | Image: DYONICS Straight Blades Full Radius, Yellow Latch | Smith & Nephew |
| 7205341 | DYONICS POWER Shaver Blades,<br>18cm working length Long Full<br>Radius | 4.5mm | Image: DYONICS POWER Shaver Blades, 18cm working length Long Full Radius | Smith & Nephew |
| 7205307 | DYONICS Straight Blades Full<br>Radius, Orange Latch | 5.5mm | Image: DYONICS Straight Blades Full Radius, Orange Latch | Smith & Nephew |
| 7210751 | DYONICS Straight Blades Full<br>Radius ELITE, Beige Latch | 3.5mm | Image: DYONICS Straight Blades Full Radius ELITE, Beige Latch | Smith & Nephew |
| 7210499 | DYONICS Straight Blades Full<br>Radius ELITE, Maroon Latch | 4.5mm | Image: DYONICS Straight Blades Full Radius ELITE, Maroon Latch | Smith & Nephew |
| 7205317 | DYONICS Straight Blades<br>RAZORCUT, Purple Latch | 3.5mm | Image: DYONICS Straight Blades RAZORCUT, Purple Latch | Smith & Nephew |
| 7205318 | DYONICS Straight Blades<br>RAZORCUT, Pink Latch | 4.5mm | Image: DYONICS Straight Blades RAZORCUT, Pink Latch | Smith & Nephew |
| 7205319 | DYONICS Straight Blades<br>RAZORCUT, Light Pink | 5.5mm | Image: DYONICS Straight Blades RAZORCUT, Light Pink | Smith & Nephew |
| 7205312 | DYONICS Straight Blades INCISOR,<br>Medium Grey Latch | 3.5mm | Image: DYONICS Straight Blades INCISOR, Medium Grey Latch | Smith & Nephew |
| 7205313 | DYONICS Straight Blades INCISOR,<br>Lime Green Latch | 4.5mm | Image | Smith & Nephew |
| 7205343 | DYONICS POWER Shaver Blades,<br>18cm working length Long<br>INCISOR | 4.5mm | Image | Smith & Nephew |
| 7205314 | DYONICS Straight Blades INCISOR,<br>Mocha Latch | 5.5mm | Image | Smith & Nephew |
| 7205345 | DYONICS Straight Blades INCISOR<br>Plus, Violet Latch | 4.5mm | Image | Smith & Nephew |
| 7205459 | DYONICS Straight Blades INCISOR<br>Plus, Teal Latch | 5.5mm | Image | Smith & Nephew |
| 7209303 | DYONICS Straight Blades Ultra<br>Series TURBOWHISKER, Navy<br>Blue Latch | 4.5mm | Image | Smith & Nephew |
| 7210976 | DYONICS Straight Blades INCISOR<br>Plus ELITE, Slate Latch | 4.5mm | Image | Smith & Nephew |
| 72200081 | DYONICS Straight Blades INCISOR<br>Plus ELITE, Lilac Latch | 5.5mm | Image | Smith & Nephew |
| 72200414 | DYONICS POWER Shaver Blades,<br>18cm working length Long<br>INCISOR Plus ELITE | 4.5mm | Image | Smith & Nephew |
| 72200095 | DYONICS Straight Blades INCISOR<br>Plus ELITE, White Latch | 3.5mm | Image | Smith & Nephew |
| 72104486 | DYONICS BONECUTTER Blades<br>Full Radius BONECUTTER, Beige<br>Latch | 3.5mm | Image | Smith & Nephew |
| 7206011 | DYONICS BONECUTTER Blades<br>Full Radius BONECUTTER, Yellow<br>Latch | 4.5mm | Image | Smith & Nephew |
| 7206010 | DYONICS BONECUTTER Blades<br>Full Radius BONECUTTER, Orange<br>Latch | 5.5mm | Image | Smith & Nephew |
| 7206013 | DYONICS BONECUTTER Blades<br>SYNOVATOR w/Full Radius inner<br>BONECUTTER, Forest Green Latch | 4.5mm | Image | Smith & Nephew |
| 7206012 | DYONICS BONECUTTER Blades<br>SYNOVATOR w/Full Radius inner<br>BONECUTTER, Melon Yellow<br>Latch | 5.5mm | Image | Smith & Nephew |
| 7205324 | DYONICS Straight Burs Abrader,<br>Aqua Latch | 4.0mm | Image | Smith & Nephew |
| 7205325 | DYONICS Straight Burs Abrader,<br>Black Latch | 5.5mm | Image | Smith & Nephew |
| 7209335 | DYONICS Straight Burs Shielded,<br>White Latch | 4.5mm | Image | Smith & Nephew |
| | | | | |
| 7205326 | DYONICS Straight Burs<br>Acromionizer, Mauve Latch | 4.0mm | Smith & Nephew | |
| 7205327 | DYONICS Straight Burs<br>Acromionizer, Brown Latch | 5.5mm | Smith & Nephew | |
| 7205727 | DYONICS Straight Burs HELICUT,<br>Slate Latch | 4.5mm | Smith & Nephew | |
| 7205328 | DYONICS Straight Burs<br>NOTCHBLASTER, Lilac Latch | 4.0mm | Smith & Nephew | |
| 7205329 | DYONICS Straight Burs<br>NOTCHBLASTER, Peach Latch | 5.5mm | Smith & Nephew | |
| 7205668 | DYONICS Straight Burs<br>ACROMIOBLASTER, Sage Green<br>Latch | 4.0mm | Smith & Nephew | |
| 7205669 | DYONICS Straight Burs<br>ACROMIOBLASTER, Brick Red<br>Latch | 5.5mm | Smith & Nephew | |
| 7205330 | DYONICS Straight Burs<br>STONECUTTER, Maroon Latch | 4.0mm | Smith & Nephew | |
| 72200082 | DYONICS POWER Shaver Burrs,<br>18cm working length, partial<br>hood Long Abrader | 5.5mm | Smith & Nephew | |
| 72200729 | DYONICS Straight Burs ELITE<br>STONECUTTER, Maroon Latch | 5.5mm | Smith & Nephew | |
| 72200730 | DYONICS Straight Burs ELITE<br>ACROMIOBLASTER, Sage Green<br>Latch | 4.0mm | Smith & Nephew | |
| 72200724 | DYONICS Straight Burs ELITE<br>Acromionizer, Mauve Latch | 5.5mm | Smith & Nephew | |
| 72203127 | DYONICS HIGH VISIBILITY Hip<br>Burrs HIGH VISIBILITY Sheath<br>Abrader Burr | 5.5mm | Smith & Nephew | |
| 72203521 | DYONICS Full Radius PLATINUM<br>Blades Full Radius PLATINUM<br>Blade | 4.5mm | Smith & Nephew | |
| 72200723 | DYONICS Straight Burs ELITE<br>Abrader, Black Latch | 5.5mm | Smith & Nephew | |
| 72200728 | DYONICS Straight Burs ELITE<br>STONECUTTER, Maroon Latch | 4.0mm | Smith & Nephew | |
| 72203128 | HIGH VISIBILITY Sheath Abrader<br>Burr | 4.0mm | Smith & Nephew | |
| 72202530 | HIGH VISIBILITY Sheath Flat Top<br>Burr | 4.0mm | Smith & Nephew | |
| 7205320 | ORBIT INCISOR™ | 4.5mm | Image | Smith & Nephew |
| 72200731 | ELITE ACROMIOBLASTER | 5.5mm | Image | Smith & Nephew |
| 72200727 | ELITE NOTCHBLASTER | 5.5mm | Image | Smith & Nephew |
| 72200726 | ELITE NOTCHBLASTER | 4.0mm | Image | Smith & Nephew |
| 7205331 | ELITE ACROMIOBLASTER | 5.5mm | Image | Smith & Nephew |
| 72202531 | BONECUTTER PLATINUM Blade | 4.5mm | Image | Smith & Nephew |
| 72203013 | DYONICS INCISOR Plus PLATINUM | 4.5mm | Image | Smith & Nephew |
| 72200725 | ELITE Acromionizer | 5.5mm | Image | Smith & Nephew |
| 7210486 | Full Radius BONECUTTER | 3.5mm | Image | Smith & Nephew |
| 72200722 | ELITE Abrader | 4.0mm | Image | Smith & Nephew |
| 3419 | Full Radius | 2.9mm | Image | Smith & Nephew |
{11}------------------------------------------------
{12}------------------------------------------------
{13}------------------------------------------------
{14}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172092
Device Name Reprocessed Stryker Shavers/Burs
Indications for Use (Describe)
The Reprocessed Styker Shavers and Burs are intended for use with the Styker Total Performance System Shaver handpiece and the Stryker Total Performance System by surgeons in the following orthopedic joints:
> Knee
- > Shoulder
- > Ankle
- > Elbow
- > Wrist
- > Hip
- > Temporamandibular
The Reprocessed Styker Shavers and Burs will be used to resect tissue and bur bone.
Examples of uses of the product would include resection of torn knee cartilage, subacromial decompression and resection of synovial tissue in other joints.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{15}------------------------------------------------
## Reprocessed Single-Use Stryker Shavers & Burs Device Models Subject to Clearance (72 Stryker devices in total)
| OEM<br>Model # | Description | Diameter | Picture | OEM |
|----------------|------------------------------------------------------------------------------------------|----------|--------------------------------------|---------|
| 375-535-000 | Formula Series Cutters, Tomcat Cutters | 3.5mm | Image: Tomcat Cutter | Stryker |
| 375-545-000 | Formula Series Cutters, Tomcat Cutters | 4.0mm | Image: Tomcat Cutter | Stryker |
| 375-555-000 | Formula Series Cutters, Tomcat Cutters | 5.0mm | Image: Tomcat Cutter | Stryker |
| 375-565-000 | Formula Series Cutters, Tomcat Cutters | 5.5mm | Image: Tomcat Cutter | Stryker |
| 375-542-000 | Formula Series Cutters, Resector Cutters | 4.0mm | Image: Resector Cutter | Stryker |
| 375-552-000 | Formula Series Cutters, Resector Cutters | 5.0mm | Image: Resector Cutter | Stryker |
| 375-562-000 | Formula Series Cutters, Resector Cutters | 5.5mm | Image: Resector Cutter | Stryker |
| 375-534-000 | Formula Series Cutters, Aggressive Plus Cutters | 3.5mm | Image: Aggressive Plus Cutter | Stryker |
| 375-544-000 | Formula Series Cutters, Aggressive Plus Cutters | 4.0mm | Image: Aggressive Plus Cutter | Stryker |
| 375-554-000 | Formula Series Cutters, Aggressive Plus Cutters | 5.0mm | Image: Aggressive Plus Cutter | Stryker |
| 375-564-000 | Formula Series Cutters, Aggressive Plus Cutters | 5.5mm | Image: Aggressive Plus Cutter | Stryker |
| 375-546-000 | Formula Series Cutters, Scalloped Cutter | 4.0mm | Image: Scalloped Cutter | Stryker |
| 375-566-000 | Formula Series Cutters, Scalloped Cutter | 5.5mm | Image: Scalloped Cutter | Stryker |
| 375-531-000 | Formula Series Cutters, Cougar End Cutter | 3.5mm | Image: Cougar End Cutter | Stryker |
| 375-541-000 | Formula Series Cutters, Cougar End Cutter | 4.0mm | Image: Cougar End Cutter | Stryker |
| 375-551-000 | Formula Series Cutters, Cougar End Cutter | 5.0mm | Image: Cougar End Cutter | Stryker |
| 375-737-000 | Formula Series Cutters, End Cutter | 3.5mm | Image: End Cutter | Stryker |
| 375-747-000 | Formula Series Cutters, End Cutter | 4.0mm | Image: End Cutter | Stryker |
| 375-757-000 | Formula Series Cutters, End Cutter | 5.0mm | Image: End Cutter | Stryker |
| 375-884-000 | Formula Series Cutters, True End Cutter | 4.0mm | Image: True End Cutter | Stryker |
| 375-543-000 | Formula Series Cutters, Double Bite Cutter | 4.0mm | Image: Double Bite Cutter | Stryker |
| 375-548-000 | Formula Series Cutters, Slotted Whisker Cutter | 4.0mm | Image: Slotted Whisker Cutter | Stryker |
| 375-558-000 | Formula Series Cutters, Slotted Whisker Cutter | 5.0mm | Image: Slotted Whisker Cutter | Stryker |
| 375-549-000 | Formula Series Cutters, Whisker Cutter | 4.0mm | Image: Whisker Cutter | Stryker |
| 375-941-012 | Standard 12 Flute Barrel Bur | 4.0mm | Image: Standard 12 Flute Barrel Bur | Stryker |
| 375-951-012 | Standard 12 Flute Barrel Bur | 5.5mm | Image: Standard 12 Flute Barrel Bur | Stryker |
| 375-941-000 | Aggressive 6 Flute Barrel Bur | 4.0mm | Image: Aggressive 6 Flute Barrel Bur | Stryker |
| 375-951-100 | Aggressive 6 Flute Barrel Bur | 5.0mm | Image: Aggressive 6 Flute Barrel Bur | Stryker |
| 375-951-000 | Aggressive 6 Flute Barrel Bur | 5.5mm | Image: Aggressive 6 Flute Barrel Bur | Stryker |
| 375-941-200 | Unhooded 6 Flute Barrel Bur | 4.0mm | Image: Unhooded 6 Flute Barrel Bur | Stryker |
| 375-951-200 | Unhooded 6 Flute Barrel Bur | 5.5mm | Image: Unhooded 6 Flute Barrel Bur | Stryker |
| 375-940-012 | Standard 12 Flute Round Bur | 4.0mm | Image: Standard 12 Flute Round Bur | Stryker |
| 375-950-112 | Standard 12 Flute Round Bur | 5.0mm | Image: Standard 12 Flute Round Bur | Stryker |
| 375-950-012 | Standard 12 Flute Round Bur | 5.5mm | Image: Standard 12 Flute Round Bur | Stryker |
| 375-930-000 | Aggressive 6 Flute Round Bur | 3.5mm | Image: Aggressive 6 Flute Round Bur | Stryker |
| 375-940-000 | Aggressive 6 Flute Round Bur | 4.0mm | Image: Aggressive 6 Flute Round Bur | Stryker |
| 375-950-100 | Aggressive 6 Flute Round Bur | 5.0mm | Image: Aggressive 6 Flute Round Bur | Stryker |
| 375-950-000 | Aggressive 6 Flute Round Bur | 5.5mm | Image: Aggressive 6 Flute Round Bur | Stryker |
| 375-940-200 | Unhooded 6 Flute Round Bur | 4.0mm | Image: Unhooded 6 Flute Round Bur | Stryker |
| 375-950-200 | Unhooded 6 Flute Round Bur | 5.5mm | Image: Unhooded 6 Flute Round Bur | Stryker |
| 375-941-200 | Unhooded 6 Flute Barrel Bur | 4.0mm | Image: Unhooded 6 Flute Barrel Bur | Stryker |
| 375-951-200 | Unhooded 6 Flute Barrel Bur | 5.5mm | Image: Unhooded 6 Flute Barrel Bur | Stryker |
| 375-952-000 | Egg Bur 6 Flute | 5.5mm | Image: bur | Stryker |
| 375-941-500 | Formula Series Specialty Bur, 6 Flute SLAP Bur | 4.0mm | Image: bur | Stryker |
| 375-450-500 | Formula Series Specialty Bur, Auger Bur | 5.0mm | Image: bur | Stryker |
| 375-832-000 | Formula Series Specialty Bur, Subchondral Drill<br>(1.5mm Drill Tip) | 4.0mm | Image: bur | Stryker |
| 375-953-000 | Formula Series Specialty Burs, 6 Flute Left Helix<br>Barrel Bur | 5.5mm | Image: bur | Stryker |
| 375-953-012 | Formula Series Specialty Burs, 12 Flute Left Helix<br>Barrel Bur | 5.5mm | Image: bur | Stryker |
| 375-627-000 | Formula Series Cutters and Burs, Small-Joint, Full<br>Radius Cutter | 2.5mm | Image: bur | Stryker |
| 375-637-000 | Formula Series Cutters and Burs, Small-Joint, Full<br>Radius Cutter | 3.5mm | Image: bur | Stryker |
| 375-628-000 | Formula Series Cutters and Burs, Small-Joint,<br>Aggressive Plus Cutter | 2.5mm | Image: bur | Stryker |
| 375-638-000 | Formula Series Cutters and Burs, Small-Joint,<br>Aggressive Plus Cutter | 3.5mm | Image: bur | Stryker |
| 375-636-000 | Formula Series Cutters and Burs, Small-Joint,<br>Tomcat Cutter | 3.5mm | Image: bur | Stryker |
| 375-641-000 | Formula Series Cutters and Burs, Small-Joint<br>Hooded Abrasion Bur | 2.0mm | Image: bur | Stryker |
| 375-647-000 | Formula Series Cutters and Burs, Small-Joint<br>Hooded Abrasion Bur | 3.0mm | Image: bur | Stryker |
| 385-544-000 | Formula Series<br>Hip Length Cutters & Burs, Aggressive Plus, Long<br>Hip Blade | 4.0mm | Image: bur | Stryker |
| 385-554-000 | Formula Series<br>Hip Length Cutters & Burs, Straight Aggressive<br>Plus, Long Hip Blade | 5.0mm | Image: bur | Stryker |
| 385-545-000 | Formula Series<br>Hip Length Cutters & Burs, Tomcat, Long Hip Blade | 4.0mm | Image: bur | Stryker |
| 385-552-000 | Formula Series<br>Hip Length Cutters & Burs, Straight Resector, Long<br>Hip Blade | 5.0mm | Image: bur | Stryker |
| 385-943-000 | Formula Series<br>Hip Length Cutters & Burs, Long Hip 8 Flute Pear<br>Bur | 4.0mm | Image: bur | Stryker |
| 385-953-008 | Formula Series<br>Hip Length Cutters & Burs, Long Hip 8 Flute Pear<br>Bur | 5.5mm | Image: bur | Stryker |
| 385-951-008 | Formula Series<br>Hip Length Cutters & Burs, Long Hip 8 Flute Round<br>Bur | 5.5mm | Image: bur | Stryker |
| 385-950-008 | Formula Series<br>Hip Length Cutters & Burs, Long Hip 8 Flute Barrel<br>Bur | 5.5mm | Image: Stryker | Stryker |
| 275-628-000 | Small Joint Cutters & Burs, Small Joint Aggressive<br>Cutter | 2.5mm | Image: Stryker | Stryker |
| 275-627-000 | Small Joint Cutters & Burs, Small Joint Full Radius<br>Cutter | 2.5mm | Image: Stryker | Stryker |
| 275-637-000 | Small Joint Cutters & Burs, Small Joint Full Radius<br>Cutter | 3.5mm | Image: Stryker | Stryker |
| 275-638-000 | Small Joint Cutters & Burs, Small Joint Aggressive<br>Cutter | 2.5mm | Image: Stryker | Stryker |
| 275-641-000 | Small Joint Cutters & Burs, Small Joint Hooded<br>Abrasion Bur | 2.0mm | Image: Stryker | Stryker |
| 275-647-000 | Small Joint Cutters & Burs, Small Joint Hooded<br>Abrasion Bur | 3.0mm | Image: Stryker | Stryker |
| 275-831-000 | Small Joint Cutters & Burs, Small Joint Round Bur | 3.0mm | Image: Stryker | Stryker |
| 375-951-112 | Standard 12 Flute Barrel Bur | 5.0mm | Image: Stryker | Stryker |
| 375-532-000 | Resector Cutter | 3.5mm | Image: Stryker | Stryker |
{16}------------------------------------------------
{17}------------------------------------------------
{18}------------------------------------------------
1.
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:
SUBMITTER ReNovo, Inc. Mark K. Wells 340 SW Columbia St. Bend OR, 97702 USA Tel: +1.541.422.8880 Fax: +1.541.422.8881
#### II. CONTACT PERSON
Robert Packard, President of Medical Device Academy 345 Lincoln Hill Rd. Shrewsbury, VT 05738 USA Tel: +1.802.281.4318 Email: rob@13485cert.com
Date Prepared: October 2, 2017
#### III. DEVICE
| Name of Device: | Reprocessed Shavers/Burs |
|------------------------------|--------------------------|
| Classification Name: | Arthroscope |
| Regulation: | 21 CFR §888.1100 |
| Regulatory Class: | Class II |
| Product Classification Code: | HRX |
#### OEM/PREDICATE DEVICES IV.
| Predicate Manufacturer: | Stryker Endoscopy |
|-------------------------|------------------------------------------|
| Predicate Trade Name: | Total Performance System Shaver |
| Predicate 510(k): | K973195 |
| Predicate Manufacturer: | Linvatec Corporation |
| Predicate Trade Name: | Integrated Drive/Pump System |
| Predicate 510(k): | K981636 |
| Predicate Manufacturer: | Smith & Nephew, Inc., Endoscopy Division |
| Predicate Trade Name: | DYONICS Power II Shaver System |
| Predicate 510(k): | K062849 |
#### V. DEVICE DESCRIPTION
The Reprocessed Shavers and Burs are intended to be used for arthroscopic and laparoscopic surgical techniques. Shavers have sharp rotating inner blades housed in an outer shaft designed to cut tissue. Burs have an abrasive wearing surface that rotate in an outer sheath to grind down bone. The housing has a window cut out on one side of the distal end, allowing the bur to cut one part of the intended anatomy, while the adjacent one is still protected by the housing on the opposite side of the abrader, bur or blade.
{19}------------------------------------------------
There are various configurations of diameter, length, and cutting or grinding surfaces. The materials of construction are generally: metal shafts, lubricants (between the inner and outer components to reduce friction and heat), and polymer hubs bonded to the shafts. The polymer hub connects to a user-supplied motorized hand piece, which drives the internal abrader, bur or blade inside the outer housing. The handpiece provides suction to remove tissue which has been cut during the surgical procedure.
There are 39 Stryker shavers and 33 Stryker burs included in this submission. The Stryker shavers and burs are used with the Total Performance System Shaver handpiece is made of a motor housing with a coupling mechanism to secure the shavers/burs to the handpiece. The Stryker shavers and burs are intended to be used to resect tissue and bur bone in orthopedic joints such as the knee, shoulder, ankle, elbow, wrist, hip, etc.
There are 31 Smith & Nephew shavers and 32 Smith & Nephew burs included in this submission. The Smith & Nephew shavers and burs are intended to be used with the Smith & Nephew DYONICS Power II Shaver System control unit. This system is compatible with DYONICS POWER handpieces. It can be used in large or small joint cavities or in sinus surgery.
There are 94 ConMed/Linvatec shavers and 26 ConMed/Linvatec burs included in this submission. The Linvatec shavers and burs are intended to be used with the Linvatec Integrated Drive/Pump system. The system has three different handpieces that ca be used for otolaryngology, neurological surgical, orthopedic, and spinal procedures.
#### VI. INDICATIONS FOR USE
The Reprocessed Linvatec shavers and burs are intended for use with the Linvatec Integrated Drive/Pump System to perform resection of soft tissue and bone. The fields of application include Otolaryngology, Neurological Surgical, Orthopedic, and Spinal surgical procedures.
The Reprocessed Smith & Nephew shavers and burs are intended for use with the Smith & Nephew DYONICS Power II Shaver System control unit, for resection of soft and osseous tissues including but not limited to, use in large articular cavities, small articular cavities, and Functional Endoscopic Sinus Surgery (FESS). The FESS application is limited to those small blades which are appropriate for the procedure.
The Reprocessed Styker Shavers and Burs are intended for use with the Styker Total Performance System Shaver handpiece and the Stryker Total Performance System by surgeons in the following orthopedic joints:
> Knee
- > Shoulder
- > Ankle
- > Elbow
- > Wrist
- > Hip
- > Temporamandibular
The Reprocessed Styker Shavers and Burs will be used to resect tissue and bur bone. Examples of uses of the product would include resection of torn knee cartilage, subacromial decompression and resection of synovial tissue in other joints.
## VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE OEM PREDICATE DEVCE
The following characteristics were compared between the subject device and the predicate device in order to demonstrate substantial equivalence:
{20}------------------------------------------------
- Indications for Use - The predicates and subject devices have equivalent indications for use; all are indicated for the resection of soft tissue and bone.
- . Materials - The predicate and subject device materials are equivalent. The lubricants are not identical, but are equivalent in functional and biocompatible performance. The predicate device and subject device materials are identical to their respective predicate (OEM) devices, with the exception of the lubricant.
- Design The predicate and subject devices are identical in design. .
- Energy Source - The subject devices are not in themselves powered, but are powered by a userprovided handpiece. For the subject devices, the handpiece is the same handpiece used by the predicate (OEM).
- Performance Testing - The subject devices are reprocessed versions of the predicate (OEM) devices. The OEM devices are performance-tested cleared medical devices; therefore the performance testing consisted of side-by-side testing of the reprocessed devices to an identical new OEM device. The subject device was tested for sharpness and bending after three reprocessing cycles, and performed equivalently to the OEM device.
- . Cleaning Validation - While it is very important to test the performance of the reprocessed device to ensure it performs as new; an important part of the risk management of a reprocessed device is the need to ensure cleanliness. A cleaning validation was performed to ensure that the reprocessing steps thoroughly cleaned and sanitized the devices so that they were as clean as a new OEM device.
## VIII. SAFTEY AND PERFORMANCE DATA
The following safety and performance data is provided in support of the substantial equivalence determination.
## Performance Testing
- . Functional Testing
- Performance testing comparison against new OEM device (sharpness and bending) O
- Drop Testing comparison against new OEM device O
- Product Stability
- Shelf Life Testing 1-Year Accelerated Aging O
## Biocompatibility Testing
- · Cytotoxicity
- · Irritation
- · Acute Systemic Toxicity
- · Material Mediated Pyrogenicity
- · Sensitization
### Cleaning Validation
- Residual Protein and Carbohydrates .
- Visual Inspection ●
- . Cleaning Performance Qualification
### Sterilization and Packaging
- EtO Sterilization Validation ●
- EtO Residuals Testing ●
- Simulated Shipment Testing
{21}------------------------------------------------
Electrical Safety and EMC Testing
Electrical safety and EMC testing were not applicable.
## Animal Testing
Animal performance testing was not required to demonstrate safety and effectiveness.
## Clinical Testing
Clinical testing was not required to demonstrate the safety and effectiveness of the Reprocessed Shavers/Burs. Substantial equivalence is based upon benchtop performance testing.
#### CONCLUSIONS IX.
Side by side testing ensured that the shavers and burs performed equivalently to the same device new from the OEM. Cleaning validations ensured that the devices are reprocessed and cleaned to the condition of a new OEM device. Therefore, based on a comparison of technological characteristics, indications for use, and performance comparison data, it can be concluded that the proposed Reprocessed Shavers/Burs are substantially equivalent to the OEM predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.