← Product Code [LNS](/productcode/LNS) · K172084

# Delta III Lithotripter (K172084)

_Dornier Medtech America, Inc. · LNS · Aug 8, 2017 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K172084

## Device Facts

- **Applicant:** Dornier Medtech America, Inc.
- **Product Code:** [LNS](/productcode/LNS.md)
- **Decision Date:** Aug 8, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.5990
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Delta III Lithotripter is indicated for the fragmentation of urinary tract stones, i.e., renal calyceal stones, renal pelvic stones, and upper ureteral stones.

## Device Story

Delta III Lithotripter is a modular urological workstation for extracorporeal shock wave lithotripsy (ESWL) and diagnostic/therapeutic urological procedures. System comprises a basic unit with integrated X-ray C-arm and therapy arm, patient table, and control desk with image processing. Device uses shock wave source (EMSE) to fragment stones; X-ray unit provides imaging for stone localization. Operated by clinicians in a clinical setting. Image processing system (UIMS) supports DICOM 3 and PACS connectivity. Modifications include updated cart wheels, electronic component upgrades, power rating label adjustment, collision system logic fix, software update for fluoroscopy delay, additional ultrasound base unit compatibility, and improved hand controller coating. Device assists physicians in stone fragmentation, potentially benefiting patients by providing non-invasive treatment for urinary calculi.

## Clinical Evidence

Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-36, IEC 61846, IEC 60601-2-54, and IEC 62366. Software verification testing performed for collision protection and hand controller function. Engineering analysis confirmed compatibility of additional ultrasound systems.

## Technological Characteristics

Modular system: Basic Unit (power, control, shockwave generation, X-ray C-arm), Patient Table, Control Desk. X-ray imaging chain. Shock wave source (EMSE). Connectivity: DICOM 3, PACS. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-36, IEC 61846, IEC 60601-2-54, IEC 62366.

## Regulatory Identification

An extracorporeal shock wave lithotripter is a device that focuses ultrasonic shock waves into the body to noninvasively fragment urinary calculi within the kidney or ureter. The primary components of the device are a shock wave generator, high voltage generator, control console, imaging/localization system, and patient table. Prior to treatment, the urinary stone is targeted using either an integral or stand-alone localization/imaging system. Shock waves are typically generated using electrostatic spark discharge (spark gap), electromagnetically repelled membranes, or piezoelectric crystal arrays, and focused onto the stone with either a specially designed reflector, dish, or acoustic lens. The shock waves are created under water within the shock wave generator, and are transferred to the patient's body using an appropriate acoustic interface. After the stone has been fragmented by the focused shock waves, the fragments pass out of the body with the patient's urine.

## Special Controls

*Classification.* Class II (special controls) (FDA guidance document: “Guidance for the Content of Premarket Notifications (510(k)'s) for Extracorporeal Shock Wave Lithotripters Indicated for the Fragmentation of Kidney and Ureteral Calculi.”)

## Predicate Devices

- Dornier Delta III Lithotripter ([K170122](/device/K170122.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the seal is a stylized image of three human profiles facing to the right, stacked on top of each other, with flowing lines suggesting movement or connection.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 8, 2017

Dornier MedTech America, Inc. John Hoffer VP Ouality. Regulatory, Clinical 1155 Roberts Blvd Suite 100 Kennesaw, Georgia 30144

K172084 Re:

Trade/Device Name: Delta III Lithotripter Regulation Number: 21 CFR 876.5990 Regulation Name: Extracorporeal Shock Wave Lithotripter Regulatory Class: Class II Product Code: LNS Dated: July 10, 2017 Received: July 10, 2017

Dear John Hoffer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act 

{1}------------------------------------------------

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

# Charles Viviano -S

For

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page

510(k) Number (if known) K172084

R172084

Device Name

Delta III Lithotripter Indications for Use (Describe)

The Delta III Lithotripter is indicated for the fragmentation of urinary tract stones, i.e., renal calyceal stones, renal pelvic stones, and upper ureteral stones.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

区 Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

Page 1 of 1 FDA

PSC Publishing Services (301) 443-6740 EP

{3}------------------------------------------------

### 510(k) SUMMARY

# Delta III Lithotripter

#### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Dornier MedTech America, Inc. 1155 Roberts Blvd. Kennesaw, GA 30144

Date Prepared: July 10, 2017

Contact Person: John Hoffer

Phone: 770-514-6163

#### Name of Device and Name/Address of Sponsor

Delta III Lithotripter

Dornier MedTech America, Inc. 1155 Roberts Blvd. Kennesaw, GA 30144

#### Common or Usual Name

Shock Wave Lithotripter

#### Classification Name

According to 21 C.F.R. § 876.5990, FDA has classified extracorporeal shock wave lithotripters as Class II devices with special controls. The Product Code for these lithotripters is LNS.

#### Predicate Device

Dornier Delta III Lithotripter (K170122)

#### Purpose of the 510(k) Notice

The Delta III Lithotripter that is the subject of this submission is a modification to the Dornier Delta III Lithotripter (K170122) to include the following minor changes:

- The UIMS cart wheels were required to be replaced as a result of their obsolescence, O
- Several electronic components were required to be modified or upgraded as O a result of additional IEC electrical testing,
- In order to affix the CE Mark, the power rating label of the Delta III X-ray device had o to be changed to reflect the effective power as measured instead of the actual power input level.

{4}------------------------------------------------

- A modification to the collision system was required due to instances when the o restoring force relating to the collision system is too low, the collision system output signal remains in "collision mode" even though the collision situation has been corrected.
- O For user convenience and preference, a software change was made that allows the user to utilize fluoroscopy without significant delay directly after a snap shot is taken,
- Additional Ultrasound base units are now being offered to use with the identical o ultrasound transducer, and
- A easier applied inner coating lining of the hand controller has been implemented. o

The modified Delta III has the same intended use, technological characteristics, and principles of operation as the predicate device. The difference between the modified Delta III and the predicate device does not raise any new or different questions of performance, safety or effectiveness. Thus, Dornier believes that the modified Delta III is substantially equivalent.

# Intended Use/Indications for Use

The Delta III Lithotripter is indicated for the fragmentation of urinary tract stones, i.e., renal calyceal stones, renal pelvic stones, and upper ureteral stones.

# Device Description

The Delta III Lithotripter is a modular urological work station designed for extracorporeal shock wave lithotripsy ("ESWL") and for diagnostic and therapeutic procedures usual in urology.

The Delta III is composed of the following modules:

- . Basic Unit with integrated X-ray C-arm and Therapy Arm with camera for Shockwave Treatment;
- Patient Table;
- . Control Desk - Image Processing.

The basic unit contains the power supplies, control unit, power electronics for motor drives, components for shockwave generation, and an integrated Therapy C-arm and an X-Ray C-Arm. The housing can be positioned with its back close to the room wall and has wide side doors for easy service.

The therapy and X-Ray C-arm house the shock wave source ("EMSE") and the complete Xray unit. The X-ray unit consists of the X-ray generator, the X-ray tube, an image receptor system, and a high resolution imaging chain. This provides the imaging to perform the procedures. The C-arms allow for a wide range of movement to facilitate performing urological procedures. The shock wave circuit supplies the shock wave energy needed for the treatment of kidney stones.

The Delta III's urological patient table provides longitudinal, lateral and vertical travel range to allow easy positioning of the stone in the shock wave focus for lithotripsy and urological

{5}------------------------------------------------

procedures.

The image processing system (UIMS) with DICOM 3 capability supports PACS connection and offers complete X-ray control and image handling.

All of the characteristics described above of the Delta III subject to this submission are identical to the predicate device.

# Technological Characteristics

As described in the section above, the Delta III device has the same technical characteristics as the predicate. This includes function and operation of the three main modules that comprise the system, the Basic Unit with integrated X-ray C-arm and Therapy Arm with camera for Shockwave Treatment, the Patient Table and the Control Desk and Image Processing ability. The minor changes that are associated with this submission do not change the essential function and use of the Delta III as compared to the predicate.

# Performance Data

The company has complied with all of the requirements described in FDA's Guidance for the Content of Premarket Notifications (510(k)s) for Extracorporeal Shock Wave Lithotripters Indicated for the Fragmentation of Kidnev and Ureteral Calculi.

The device is in compliance with the following standards:

- IEC 60601-1:2005,Electrical safety of medical devices;
- . IEC 60601-1-2:2007, Electomagnetic compatibility:
- IEC 60601-1-3: 2008, Radiation protection in diagnostic X-ray equipment; ●
- IEC 60601-2-36: 1997, Safety of equipment for extracorporeally induced lithotripsy; ●
- . IEC 61846, Ultrasonics - pressure pulse lithotripters characteristics of fields;
- IEC 60601-2-54, Medical electrical equipment particular requirements for the . basic safety and essential performance of x-ray equipment; and
- . Usability: IEC 62366:2007, Application of usability engineering to medical devices.

In summary, during the verification testing, the electrical safety of the system and the electromagnetic compatibility were fully addressed by demonstrating compliance with the applicable standards. In addition, appropriate software verification testing to sections of IEC 60601-1 for wheel performance and collision protection, and verification of hand controller function testing were also performed to address the modifications made that are subject to this submission.

Also, an engineering analysis concluded that the prior ultrasound localization accuracy testing supported the compatibility of the additional ultrasound systems that were added.

There were no unanticipated new risks identified

# Conclusions

The Delta III is as safe and effective as the cited predicate device. The Delta III has the same intended uses and indications, technological characteristics, and principles of

{6}------------------------------------------------

operation as its predicate device. The minor differences do not alter the intended the use of the device and do not affect its safety and effectiveness when used as labeled. In addition, the minor technological differences between the Delta III and its predicate devices raise no new or different questions of safety or effectiveness. Design controls demonstrate that the Delta III is as safe and effective as the predicate device. Thus, the Delta III is substantially equivalent.

---

**Source:** [https://fda-staging.innolitics.com/device/K172084](https://fda-staging.innolitics.com/device/K172084)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
