K172083 · Clariance, Sas · OVD · Nov 8, 2017 · Orthopedic
Device Facts
Record ID
K172083
Device Name
Idys™ ALIF System
Applicant
Clariance, Sas
Product Code
OVD · Orthopedic
Decision Date
Nov 8, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Idys™ ALIF (Anterior Lumbar Integrated Fusion) System is intended for use in patients with degenerative disc disease (DDD) at one (1) or two (2) contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved levels. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. The Idys™ ALIF System should be used with the integrated fixation screws provided. The Idys™ ALIF System is intended to be used with autograft.
Device Story
Intervertebral body fusion device; used for anterior lumbar interbody fusion (ALIF). Implanted by surgeons in clinical settings to restore intervertebral height and lordosis. Device consists of PEEK cage with tantalum markers, integrated titanium alloy plate, and screws. Cage features teeth on superior/inferior surfaces to grip vertebral endplates and resist expulsion; internal slots hold autograft to promote fusion. Standalone fixation achieved via integrated plate and screws. Benefits include stabilization of spinal segments in DDD patients. Device is purely mechanical; no software or electronic components.
Clinical Evidence
Bench testing only. Testing included static and dynamic axial compression, compression shear, subsidence, expulsion, torsion, and wear particle characterization per ASTM F2077 and ASTM F2267. Compliance with sterilization and packaging standards (ISO 11137-1, 11137-2, 11607-1, 11607-2) confirmed.
Technological Characteristics
Materials: PEEK (ASTM F2026), Tantalum markers (ASTM F560), Titanium alloy (ASTM F136). Design: Interbody spacer with integrated plate and screw fixation. Features: Teeth for endplate engagement, open geometry for autograft. Sterilization: Not specified (standards cited). Connectivity: None.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1), including those with up to Grade I spondylolisthesis or retrolisthesis, who have failed at least 6 months of non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
November 8, 2017
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized caduceus symbol. To the right of it is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Clariance SAS % Janice M. Hogan Regulatory Counsel Hogan Lovells US LLP 1735 Market Street, Suite 2300 Philadelphia, Pennsylvania 19103
Re: K172083
Trade/Device Name: Idys™ ALIF System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVD Dated: October 12, 2017 Received: October 12, 2017
Dear Ms. Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.qov
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
> Sincerely, Katherine D. Kavlock -S for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page
510(k) Number (if known) K172083
Device Name
ldys™ ALIF System
Indications for Use (Describe)
The Idys™ ALIF (Anterior Lumbar Integrated Fusion) System is intended for use in patients with degenerative disc disease (DDD) at one (1) or two (2) contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved levels. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. The Idys™ ALIF System should be used with the integrated fixation screws provided.
The Idys™ ALIF System is intended to be used with autograft.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number
FORM FDA 3881 (8/14)
{3}------------------------------------------------
K172083 Page 1 of 3
## K172083
## 510(k) SUMMARY
## CLARIANCE's Idys™ ALIF System
## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
CLARIANCE, SAS 18 Rue Robespierre F-62217 Beaurains, FRANCE Phone: +33 (0)3 21 16 12 15 Facsimile: +33 (0)3 21 15 50 73
Contact Person: Pascal Rokegem, Chief Technology Officer
Date Prepared: October 12, 2017
## Name of Device and Name
ldys™ ALIF System
## Common or Usual Name
Lumbar Intervertebral Body Fusion Device
#### Classification Name
888.3080 - Intervertebral body fusion
#### Product Code
OVD
#### Predicate Devices
Medtronic, DIVERGENCE-L, K150135 (primary)
CLARIANCE, Idys LIF Cage, K131178 (additional)
## Intended Use / Indications for Use
The Idys™ ALIF (Anterior Lumbar Fusion Integrated Fusion) System is intended for use in patients with degenerative Disc Disease (DDD) at one (1) or two (2) contiguous levels of the lumbosacral spine (L2 to S1). DDDis defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved levels.
{4}------------------------------------------------
K172083 Page 2 of 3
These patients should be skeletally mature and have had six (6) months of non-operative treatment. The Idys™ ALIF System should be used with the integrated fixation screws provided. The Idys™ ALIF System is intended to be used with autograft.
## Device Description
The Idys™ ALIF System is manufactured from medical grade polyetheretherketone (INVIBIO PEEK OPTIMA LT1) and is to be used with autograft. The device has a shape which restores the intervertebral height and lordosis. The device contains two slots to receive the autologous bone graft to promote the fusion process between the endplates. The superior and inferior surfaces of the implant are designed with teeth which interact with the surface of the vertebral endplates and helps resist back out.
The Idys™ ALIF System is connected to a plate with four screws for a standalone implantation. The Idys™-ALIF System are made of ASTM F2026 compliant polyetheretherketone (PEEK) and markers made of Tantalum according to ASTM F560, the plate and screws are made of ASTM F136 titanium alloy.
## Comparison of Technological Characteristics
The Idys™ ALIF System and the predicate DIVERGENCE-L™ are designed for use as lumbar intervertebral body fusion devices. All devices are composed of PEEK cages or "interbody spacers" available with various configurations to accommodate patient anatomy. These cages incorporate Tantalum radiographic markers to allow the verification of the cage position during surgery and post-operatively.
The Idys™ ALIF System and the DIVERGENCE-L™ cages feature a similar outer shape and design. Both devices are rectangular with rounded corners, and feature open spaces to allow the incorporation of bone graft which is essential to promote the fusion process. Each of the superior and inferior surfaces of the Idys™ ALIF System and the DIVERGENCE-L™ devices are designed with teeth to grip the surface of the vertebral endplates and help in resisting expulsion.
In terms of configurations, the Idys™ ALIF System cages are very similar to those of the cleared DIVERGENCE-L™ predicate (K150135). The Idys™ ALIF System cages are available with four footprints with different heights ranges and lordotic angles. The differences in the dimensions of the Idys™ ALIF System cages compared to those cleared for the DIVERGENCE-L™ predicate do not raise new issues of safety and effectiveness because the available heights, lordotic options and footprints fall within the range of dimensions offered by the predicate device. Mechanical testing has shown that the addition of a smaller footprint does not have an adverse impact on the mechanical performance of the Idys™ ALIF System cages .
The shape and size offerings differ slightly from the Idys LIF (K13178), which has a flatter and longer shape, but is made from the same materials. The differences in shape do not raise different types of safety or effectiveness questions and mechanical testing shows at least equivalent performance.
{5}------------------------------------------------
The plates intended to be used with the Idys™ ALIF System have similar dimensions as compared to those cleared for DIVERGENCE-L™ predicate device (K150135). The width and the profile of the predicate DIVERGENCE-L is greater than the Idys™ ALIF System plate, but these differences do not raise new or different issues of safety and effectiveness and are supported by testing.
The bone screws proposed with the Idys™ ALIF System have similar dimensions compared to those cleared for the DIVERGENCE-L™ predicate device (K150135) in terms of lengths and diameters. Additionally, the screws of the Idys™ ALIF System are made from the same ASTM F136 medical grade material titanium alloy (Ti6Al4V ELI) as the one used for the cleared DIVERGENCE-L™ predicate screws.
## Performance Data
Performance testing was conducted per Modified ASTM F2077 for inclusion of the plate and screw in the testing protocol and ASTM F2267. Specifically, CLARIANCE performed static and dynamic axial compression testing, static and dynamic compression shear testing, subsidence testing, expulsion testing, torsion testing, and wear particle characterization. The results of these studies were determined to be substantially equivalent to legally marketed devices.
The subject device also complies with 11137-1, 11137-2, 11607-1 and 11607-2.
## Conclusion
The Idys™ ALIF System is as safe and effective as the Divergence-L (K150135) and Idys LIF (K131178). The Idys™ ALIF System has the same intended uses, similar indications, principles of operation, and technological characteristics as the Divergence-L and the Idys LIF. Bench testing has shown equivalent performance. Thus. the Idys™ ALIF System is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.