K172077 · Cutera, Inc. · GEX · Oct 27, 2017 · General, Plastic Surgery
Device Facts
Record ID
K172077
Device Name
Cutera enlighten III Laser System
Applicant
Cutera, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 27, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The enlighten III laser system is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery. 1064 nm: The 1064 nm wavelength of the enlighten III laser system is indicated for: - treatment of benign pigmented lesions on patients with all skin types (Fitzpatrick I-VI) - tattoo removal for dark colored tattoo inks and for multicolored tattoos containing dark colored tattoo inks on patients with all skin types (Fitzpatrick I-VI) 532 nm: The 532 nm wavelength of the enlighten III laser system is indicated for: - treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-IV - tattoo removal for lighter colored tattoo inks, including red and yellow inks, on patients with Fitzpatrick skin types I-III 670 nm: The 670 nm wavelength of the enlighten III laser system is indicated for: - treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-III - tattoo removal for lighter colored tattoo inks, including green and blue inks, on patients with Fitzpatrick skin types II-IV
Device Story
Multi-wavelength pulsed laser system; console-housed laser, power supplies, control electronics; touchscreen interface; footswitch activation; articulated arm with rotational mirror knuckles; optical zoom handpiece; low-power diode aiming beam. Used in dermatology/plastic surgery clinics by physicians. Inputs: user-selected treatment parameters (wavelength, pulse width, spot size). Outputs: laser energy delivery to skin. Principle: photothermolysis for pigmented lesions and tattoo ink fragmentation. Clinical benefit: removal of unwanted tattoos and benign pigmented lesions. Output affects clinical decision-making by enabling targeted treatment of specific ink colors/lesion types based on wavelength selection.
Clinical Evidence
Prospective, two-center study (n=27 enrolled, 25 analyzed) assessing 670 nm wavelength for green/blue tattoo removal. Primary endpoint: >=50% clearing in >=80% of patients. Results: 84% of subjects achieved >50% clearing. Blinded dermatologists reported significant improvement (mean 3.24/4.0 scale). Adverse events (erythema, edema, frosting, crusting) were expected and resolved without intervention. No serious adverse events.
Indicated for treatment of benign pigmented lesions and tattoo removal. 1064 nm: all skin types (Fitzpatrick I-VI). 532 nm: Fitzpatrick I-IV (lesions) or I-III (tattoos). 670 nm: Fitzpatrick I-III (lesions) or II-IV (tattoos).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
October 27, 2017
Cutera, Inc. Bradley Renton Vice President, Regulatory and Medical Affairs, & Compliance Officer 3240 Bayshore Blvd. Brisbane, California 94005
Re: K172077
Trade/Device Name: Cutera enlighten III Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: October 5, 2017 Received: October 6, 2017
Dear Bradley Renton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172077
Device Name Cutera enlighten III Laser System
#### Indications for Use (Describe)
The enlighten III laser system is intended for use in surgical and aesthetic applications in the medical specialties of dermatology and general and plastic surgery.
1064 nm:
The 1064 nm wavelength of the enlighten III laser system is indicated for:
- · treatment of benign pigmented lesions on patients with all skin types (Fitzpatrick I-VI)
- tattoo removal for dark colored tattoo inks and for multicolored tattoos containing dark colored tattoo inks on patients with all skin types (Fitzpatrick I-VI)
532 nm:
- The 532 nm wavelength of the enlighten III laser system is indicated for:
- · treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-IV
- · tattoo removal for lighter colored tattoo inks, including red and yellow inks, on patients with Fitzpatrick skin types I-III
670 nm:
- The 670 nm wavelength of the enlighten III laser system is indicated for:
- · treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-III
- · tattoo removal for lighter colored tattoo inks, including green and blue inks, on patients with Fitzpatrick skin types II-IV
Type of Use (Select one or both, as applicable)
| <span style="font-size:10pt;"> <span style="font-family:Wingdings;">X</span> Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size:10pt;"><span style="font-family:Wingdings;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|----------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Section 5 510(K) Summary
This 510(K) Summary of safety and effectiveness for the enlighten III laser system is submitted in accordance with the requirements of the SMDA 1990 and following guidance concerning the organization and content of a 510(K) summary.
| Applicant: | Cutera, Inc. |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 3240 Bayshore Blvd., Brisbane, CA 94005 |
| Contact Person: | Bradley Renton |
| | 415-657-5568 - phone<br>415-715-3568 - fax<br>brenton@cutera.com |
| Preparation Date: | October 5, 2017 |
| Device Trade Name: | enlighten III Laser System |
| Common Name: | Dermatology Laser |
| Classification Name: | Instrument, Surgical, Powered, laser<br>79-GEX, 21 CFR 878.4810 |
| Legally Marketed<br>Predicate Devices: | Cutera enlighten III Laser System (K170936)<br>Cynosure PicoSure (K160480)<br>Syneron Candela PicoWay (K170597)<br>Cynosure RevLite (K133254) |
| Device Description: | The enlighten III laser system is a multi-wavelength, pulsed laser<br>system. A key feature of the device is its ability to produce multiple<br>laser wavelengths and pulse widths (1064 and 532 nm: 750 ps and<br>2 ns; 670 nm: 660 ps and 2 ns).<br><br>The laser, power supplies, and control electronics are housed<br>inside a console equipped with a touchscreen control panel. The<br>laser treatment parameters are selected using the control panel.<br>Laser emission is activated by depressing a footswitch. An<br>articulated arm with rotational mirror knuckles delivers the laser<br>beam from a laser head inside the console to a handpiece. The<br>handpiece is equipped with an optical zoom assembly that<br>determines the spot size of the laser beam on the treatment<br>surface. The beam of a low-power diode laser is also transmitted<br>through the arm to provide an aiming beam. |
{4}------------------------------------------------
## Section 5 510(K) Summary
Indications for Use:
#### 1064 nm
The 1064 nm wavelength of the enlighten III laser system is indicated for:
- . treatment of benign pigmented lesions on patients with all skin types (Fitzpatrick I-VI)
- tattoo removal for dark colored tattoo inks and for . multicolored tattoos containing dark colored tattoo inks on patients with all skin types (Fitzpatrick I-VI)
## 532 nm
The 532 nm wavelength of the enlighten III laser system is indicated for:
- . treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-IV
- . tattoo removal for lighter colored tattoo inks, including red and yellow inks, on patients with Fitzpatrick skin types I-III
## 670 nm
The 670 nm wavelength of the enlighten III laser system is indicated for:
- . treatment of benign pigmented lesions on patients with Fitzpatrick skin types I-III
- tattoo removal for lighter colored tattoo inks, including green . and blue inks, on patients with Fitzpatrick skin types II-IV
Performance Data: IEC 60601-1 Medical Electrical Equipment - Part 1: General Requirements for Safety (Edition 3.1, 2012), including:
- . IEC 60601-1-6 Medical Electrical Equipment - Part 1-6: General Requirements for Safety - Collateral Standard: Usability (Edition 3.1, 2013) - Test Report Attachment 4
- IEC 62366 Medical Devices Application of Usability . Engineering to Medical Devices (Edition 1.1, 2014) - Test Report Attachment 5
- IEC 60601-2-22 Medical Electrical Equipment Part 2: . Particular Requirements for Basic Safety and Essential Performance of Surgical, Cosmetic, Therapeutic and Diagnostic Laser Equipment (Edition 3.1, 2012) – Test Report Attachment 6
- IEC 60825-1 Safety of Laser Products Part 1: Equipment ● Classification and Requirements (Edition 2.0, 2007) - Test Report Attachment 7
The product also fulfills the requirements of AAMI/ANSI ES60601-1:2005+A2 (R2012) + A1.
IEC 60601-1-2 Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility (Edition 3, 2007)
{5}------------------------------------------------
## Section 5 510(K) Summary
Results of Clinical An IRB-approved, prospective, two-center clinical study was Study: conducted to assess the safety and efficacy of the 670 nm wavelength of the enlighten III laser system for the clearing of tattoos containing green and/or blue ink.
> Twenty-seven subjects with Fitzpatrick skin types II-V with unwanted tattoos containing green and/or blue ink, either alone or in combination with other ink colors were enrolled. The subjects received nanosecond and/or picosecond treatments with the Cutera enlighten III laser system with the 670 nm wavelength used to target green and blue inks; the 1064 nm wavelength used to target black and gray inks; and the 532 nm wavelength used to target lighter red and yellow inks. All wavelengths of the system could be used as required by the ink colors contained within the tattoo; however, only the clearing of green and blue inks were graded by blinded dermatologists. Standardized photographs were taken prior to and immediately after all study visits. Pain levels during treatment and adverse events were recorded during all visits.
An analysis of data was conducted on twenty-five of the twentyseven subjects originally enrolled (two subjects had voluntarily withdrawn from the study after receiving one and two treatments respectively, but prior to attending the required follow-up visits for photography). Before and after images of the twenty-five subject tattoos were randomized and sent to three blinded dermatologists for correct identification of the before from after image, and grading of the clearing of the green and/or blue ink. All blinded dermatologists correctly identified the before image for all subjects. Greater than 50% clearing was seen in green and blue ink in the tattoos for 84% of subjects indicating that the primary efficacy endpoint for the study (≥ 50% clearing of green and blue inks in ≥ 80% of patients) had been met.
Blinded photographic assessment of the 25 subject tattoos resulted in a clinically and statistically significant improvement score of 3.24 on a 0-4 point scale. The mean and standard deviation were 3.24 ±0.71. Investigator assessment of tattoo clearing at six-week post treatment resulted in a mean improvement score of 3.43 ± 0.7.
Based on subject questionnaires, sixty-nine percent of subjects reported that they were "extremely satisfied" or "satisfied" with the level of clearing and thirty-one percent of subjects reported they were neutral. No patients reported dissatisfaction. Eighty-five percent of subjects would have the investigational treatment aqain.
All subjects tolerated treatments well with few requests for topical anesthetic and without any requests for local infiltration of anesthetic even though allowed by protocol (mean pain score 4.2 ± 2.9 [0-9] on a 10-point scale). As expected, all subjects experienced erythema, edema and frosting immediately following laser treatment, and most experienced crusting in the treated area. All adverse effects resolved with no required intervention. No
{6}------------------------------------------------
## Section 5 510(K) Summary
unexpected device related serious adverse events were noted.
The 670 nm wavelength of the enlighten III laser system was found to be safe and effective, with acceptable adverse effects, for tattoo removal for lighter colored tattoo inks including green and blue inks, in patients with skin types II-IV, allowing the conclusion that the enlighten laser system is substantially equivalent to the predicate devices for the requested indications.
Summary of The Cutera enlighten III laser systems used at each site in the Technological clinical trial had identical device design, components, and Characteristics: performance specifications as the currently cleared enlighten III laser system (K170936). The enlighten III device design and components are also very similar to those of the Cynosure PicoSure, Syneron Candela PicoWay, and Cynosure RevLite predicates. For each of these systems:
- the user interface is located at the front/top of the console;
- . the treatment handpiece is attached to an articulating arm that is connected to the main system console;
- . laser energy is generated within the console and is transmitted through the articulated arm to the handpiece; and
- . the handpiece allows the spot size on the skin to be adjusted according to device specifications.
Each system thus consists of an articulating arm and attached handpiece, as well as an electrically powered system console that produces the laser energy and houses the software and user interface.
The enlighten III laser system provides similar key design aspects, including the same or similar spot sizes, laser wavelengths, pulse widths, and laser types as its predicate devices. The repetition rate of the enlighten III laser system is the same as or within the repetition rate range of the predicate devices. Further, each of the devices presents a range of spot sizes to allow the user to choose the most appropriate spot size for each patient. Therefore, the minor differences do not raise any new safety or effectiveness questions, because the enlighten III parameters are the same as or within the range of the predicates.
Conclusion: Clinical testing of the enlighten III laser system demonstrated that the device performs as intended with a favorable safety profile. Results in the study were similar to those reported for the predicate devices, in support of substantial equivalence. The non-clinical data further support the safety of the device, and software verification and validation testing demonstrates that the enlighten III is expected to perform as intended in the specified use conditions. The Cutera enlighten III laser system is substantially equivalent to the predicate devices.
Summary of
Technological
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.