K172061 · Nuvasive Specialized Orthopedics, Inc. · HSB · Aug 24, 2017 · Orthopedic
Device Facts
Record ID
K172061
Device Name
PRECICE UNYTE System
Applicant
Nuvasive Specialized Orthopedics, Inc.
Product Code
HSB · Orthopedic
Decision Date
Aug 24, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PRECICE UNYTE System is indicated for open and closed fracture fixation, corrective osteotomies, pseudoarthrosis, mal-unions, limb lengthening, or bone transport of long bones.
Device Story
The PRECICE UNYTE System is an intramedullary fixation device for long bones. It consists of a nail containing an enclosed rare earth magnet, a telescoping lead screw/nut assembly, and planetary gearing, along with locking screws, an end cap, and an External Remote Controller (ERC). The device is inserted into the intramedullary canal and secured with locking screws. The ERC is used non-invasively to adjust the nail, facilitating fracture reduction, limb lengthening, or bone transport. The system is intended for use by surgeons in clinical settings. The device benefits patients by providing a non-invasive method for bone adjustment and fixation.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by the identical design and technological characteristics to the predicate device, with previous bench testing for the nail and ERC remaining applicable.
Technological Characteristics
Intramedullary fixation rod system. Components: nail with rare earth magnet, telescoping lead screw/nut assembly, planetary gearing, locking screws, end cap, and External Remote Controller (ERC). Nail and end cap sterilized via gamma radiation; screws and instruments supplied non-sterile. Mechanical adjustment via external magnetic control.
Indications for Use
Indicated for patients requiring open and closed fracture fixation, corrective osteotomies, pseudoarthrosis, mal-unions, limb lengthening, or bone transport of long bones.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
Predicate Devices
PRECICE UNYTE System (formerly PRECICE Trauma Nail System) (K160267)
Reference Devices
PRECICE Intramedullary Limb Lengthening System (K170346)
{0}------------------------------------------------
## Please wait...
If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.
You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.
For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.
Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.
{1}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with three lines above them that resemble the wings of a bird.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
NuVasive Specialized Orthopedics, Inc. Cora Sim Associate Manager, Regulatory Affairs 101 Enterprise. Suite 100 Aliso Viejo, California 92656
August 24, 2017
Re: K172061
Trade/Device Name: PRECICE UNYTE System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary Fixation Rod Regulatory Class: Class II Product Code: HSB Dated: July 6, 2017 Received: July 7, 2017
Dear Cora Sim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{2}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
K172061 page 1 of 2
Image /page/3/Picture/1 description: The image shows the logo for Nuvasive Specialized Orthopedics, Inc. The logo features a stylized leaf-like shape in purple and silver on the left. To the right of the shape is the company name, "NUVASIVE" in silver, with the tagline "SPECIALIZED ORTHOPEDICS, INC." in purple below it.
## PRECICE UNYTE System 510(k) Summarv - K172061 August 7, 2017
- 1. Company: NuVasive Specialized Orthopedics, Inc. 101 Enterprise, Suite 100 Aliso Viejo, CA 92656
- Contact: Cora Sim Associate Manager, Regulatory Affairs Phone: (949) 544-6478 Fax: (949) 837-3664
Date of Submission: August 7, 2017
- 2. Proprietary Trade Name: PRECICE UNYTE System
- Classification Name: Intramedullary Fixation Rod (21 CFR 888.3020) 3.
- 4. Product Code: HSB (Rod, Fixation, Intramedullary and Accessories)
- ട. Product Description: The PRECICE UNYTE System is composed of the PRECICE UNYTE Nail (supplied sterile), locking screws, end cap, surgical instruments and an external remote controller (ERC). The Nail is available in various diameters, lengths and screwhole configurations to accommodate a variety of patient anatomies. The locking screws are also available in a variety of diameters and lengths. The PRECICE UNYTE Nail and end cap is supplied sterile by gamma radiation while the locking screws and reusable instruments are supplied non-sterile and must be sterilized prior to use. The UNYTE Nail contains an enclosed rare earth magnet, telescoping lead screw/nut assembly, and planetary gearing. The UNYTE Nail is supplied pre-distracted by 10 mm (femur and tibia models), 15 mm and 20 mm (humeral model), to allow for compression fracture reduction techniques.
- Indications: The PRECICE UNYTE System is indicated for open and closed fracture fixation, 6. corrective osteotomies, pseudoarthrosis, mal-unions, limb lengthening, or bone transport of long bones.
- 7. Substantial equivalence: A detailed comparison to the predicate device demonstrates that the PRECICE UNYTE System is substantially equivalent to the following 510(k) cleared device:
Trade Name: PRECICE UNYTE System (formerly called PRECICE Trauma Nail System) Common Name: Intramedullary Fixation Rod 510(k) Clearance Number: K160267
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a logo for "NUVASIVE SPECIALIZED ORTHOPEDICS, INC.". The logo features a stylized, abstract shape on the left, which is partially purple and partially silver. To the right of the shape is the word "NUVASIVE" in silver, and below that is the phrase "SPECIALIZED ORTHOPEDICS, INC." in purple.
In addition, the PRECICE UNYTE System designates the following 510(k) cleared reference devices:
| Trade Name: | PRECICE Intramedullary Limb Lengthening System |
|--------------------------|------------------------------------------------|
| Common Name: | Intramedullary Fixation Rod |
| 510(k) Clearance Number: | K170346 |
| Trade Name: | PRECICE UNYTE CoCr System |
| Common Name: | Intramedullary Fixation Rod |
| 510(k) Clearance Number: | K162927 |
| Trade Name: | AOS Tibial Nail System |
| Common Name: | Intramedullary Fixation Rod |
| 510(k) Clearance Number: | K130625 |
Substantial equivalence is based on indications for use, technological characteristics, and principles of operation.
The PRECICE UNYTE System subject to this 510(k) and the predicate device have similar indications for use. Specifically, both systems are indicated for open and closed fracture fixation, pseudoarthrosis, or mal-unions and non-unions of long bones. The PRECICE Intramedullary Limb Lengthening System and PRECICE UNYTE CoCr System reference devices listed above both include limb lengthening with their indications for use. The AOS Tibial Nail System reference device includes bone transport within the indications for use. This 510(k) is intended to expand the indications of the PRECICE UNYTE System to align with the predicate and reference devices in the PRECICE family and in the same product code by including limb lengthening and bone transport.
The PRECICE UNYTE System has the same technological characteristics and the same principle of operation as that of the predicate. The design of both intramedulary nails is identical. Both devices are inserted into the intramedullary canal of the long bone and secured with locking screws. Both devices are adjusted non-invasively by the External Remote Controller (ERC).
There are no technological or performance changes to the PRECICE UNYTE Nail being made as a result of this submission, therefore all testing that was performed on the predicate PRECICE UNYTE Nail are applicable.
There are no changes to the design of the ERC being made as a result of this submission, therefore all testing that was performed on the ERC use with the predicate PRECICE UNYTE Nail System for the ERC are applicable.
Conclusions can be drawn from these tests that the PRECICE UNYTE System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.