K172050 · Vitrolife Sweden AB · MQE · Nov 21, 2017 · Obstetrics/Gynecology
Device Facts
Record ID
K172050
Device Name
Follicle Aspiration Set, Reduced Single Lumen
Applicant
Vitrolife Sweden AB
Product Code
MQE · Obstetrics/Gynecology
Decision Date
Nov 21, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6100
Device Class
Class 2
Indications for Use
Intended for flushing and/or aspiration of oocytes from ovarian follicles.
Device Story
Follicle Aspiration Set is a single-use, sterile, ultrasound-guided transvaginal needle assembly; used for oocyte retrieval from ovarian follicles. Components include stainless steel needle, silicone cork, TPE-O aspiration/flushing tubing, and flushing adapter. Operated by clinicians in clinical settings; needle inserted transvaginally under ultrasound guidance to access follicles; vacuum pump connected via cork/adapter to aspirate or flush oocytes. Output is collected oocytes for laboratory processing. Device facilitates oocyte collection for assisted reproduction; benefits include reliable follicle access and fluid management.
Clinical Evidence
Bench testing only. Studies included sterilization validation (ISO 11137-2:2013), endotoxin testing (USP<85>), needle breakage/elasticity (ISO 9626:2016), and accelerated shelf-life testing. Biocompatibility/toxicity assessed via Mouse Embryo Assay (MEA), achieving ≥80% blastocyst development. Mechanical performance verified via vacuum testing (-500 mmHg), flow rate testing, tensile strength testing, and packaging integrity (ASTM F1886, F1929, F88).
Technological Characteristics
Stainless steel needle; TPE-O aspiration tubing; silicone cork; ABS and MABS adapters; polypropylene sheath. Sensing/actuation: manual ultrasound-guided aspiration/flushing. Dimensions: 1.4mm OD body, 0.9mm OD tip. Sterilization: validated per ISO 11137-2:2013. Connectivity: none (mechanical).
Indications for Use
Indicated for flushing and/or aspiration of oocytes from ovarian follicles in patients undergoing assisted reproduction procedures.
Regulatory Classification
Identification
Assisted reproduction needles are devices used in in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures to obtain gametes from the body or introduce gametes, zygote(s), preembryo(s) and/or embryo(s) into the body. This generic type of device may include a single or double lumen needle and component parts, including needle guides, such as those used with ultrasound.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing).
Predicate Devices
Follicle Aspiration Set, Reduced Single Lumen (K082727)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a symbol of the agency's mission to protect and promote public health through the regulation and supervision of food safety, tobacco, pharmaceuticals, medical devices, and other related products.
November 21, 2017
Vitrolife Sweden AB % Greg Holland Regulatory Specialist Regulatory Specialist, Inc. 3722 Ave. Sausalito Irvine, CA 92606
K172050 Re:
Trade/Device Name: Follicle Aspiration Set, Reduced Single Lumen Regulation Number: 21 CFR§ 884.6100 Regulation Name: Assisted Reproduction Needles Regulatory Class: II Product Code: MOE Dated: October 16, 2017 Received: October 19, 2017
Dear Greg Holland:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/6 description: The image shows the text "Benjamin R. Fisher -S" in a large, bold font. The text is black and is set against a white background. The letters are clear and easy to read. The text appears to be a name, followed by an initial.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K172050
Device Name
Follicle Aspiration Set, Reduced Single Lumen
Indications for Use (Describe) Intended for flushing and/or aspiration of oocytes from ovarian follicles.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Vitrolife" in blue font, with a blue swoosh-like design to the right of the word. The font is a serif font, and the word is written in a slightly bolded style. The swoosh design is a simple, curved line that starts thick and tapers to a point.
# 510(k) Summary - K172050
| Submitter | Vitrolife Sweden AB<br>Gustaf Werners gata 2<br>SE - 421 32 Västra Frölunda<br>SWEDEN<br>Nina Arvidsson<br>TEL: (46) 31 721 80 00<br>FAX: (46) 31 721 80 99 |
|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Greg Holland<br>Regulatory Consultant to Vitrolife Sweden AB<br>Regulatory Specialists, Inc.<br>3722 Ave. Sausalito<br>Irvine, CA 92606<br>TEL: 949.422.3853<br>FAX: 949.552.2821<br>EMAIL: greg@regulatoryspecialists.com |
| Date Prepared | November 17, 2017 |
| Trade Name<br>Common Name<br>Classification Name<br>Regulation Number<br>Regulatory Class<br>Product Code | Follicle Aspiration Set, Reduced Single Lumen<br>Oocyte Retrieval Needle<br>Assisted Reproduction Needles<br>884.6100<br>II<br>MQE (Needle, Assisted Reproduction) |
| Predicate | Follicle Aspiration Set, Reduced Single Lumen (K082727).<br>This predicate device has not been subjected to a design-<br>related recall. |
### Device Description:
The Follicle Aspiration Set. Reduced Single Lumen proposed in this submission is intended for ultrasound-guided transvaginal aspiration and flushing of oocytes from ovarian follicles. This product is comprised of the following components:
- A stainless steel needle .
- A silicone cork to be pushed into the opening of a sampling tube .
- A length of TPE-O tubing for aspiration and flushing ●
- An adapter for flushing .
The subject device is provided sterile and is for single use only. It comes in four versions with the same design but different dimensions, as described below.
{4}------------------------------------------------
| REF | Needle Body [mm] | | | Needle Tip [mm] | | | Aspiration Tubing [mm] | | |
|-----------------------------------------------------------------------------------------------|------------------|-----|-----|-----------------|-----|----|------------------------|------|-----|
| | OD | ID | L | OD | ID | L | OD | ID | L |
| 17175 | 1.4 | 1.2 | 250 | 0.9 | 0.6 | 50 | 2.22 | 1.33 | 600 |
| 17176 | 1.4 | 1.2 | 250 | 0.9 | 0.6 | 50 | 2.22 | 1.33 | 900 |
| 17177 | 1.4 | 1.2 | 300 | 0.9 | 0.6 | 50 | 2.22 | 1.33 | 600 |
| 17178 | 1.4 | 1.2 | 300 | 0.9 | 0.6 | 50 | 2.22 | 1.33 | 900 |
| The cork (REF 21014) has OD of 1.7 mm on the top and 13 mm in the bottom and length of 22 mm. | | | | | | | | | |
#### Indications for Use:
Intended for flushing and/or aspiration of oocytes from ovarian follicles.
## Comparison of Intended Use and Technological Characteristics of Subject and Predicate Devices:
| Devices | K172075 (subject device) | K082727 (predicate device) | | | | |
|-------------------------------------------|--------------------------------------------------------------|-------------------------------------------------------------------------------|--|--|--|--|
| Intended use | | | | | | |
| Indications for Use | Same as predicate | Intended for flushing and/or aspiration<br>of oocytes from ovarian follicles. | | | | |
| Technological characteristics - Design | | | | | | |
| Lumen type | Same as predicate | Single lumen | | | | |
| Needle length | Tip - 50 mm; Body - 250/300 mm | Tip - 40 mm; Body - 260/310 mm | | | | |
| Needle Outer Diameter | Same as predicate | Tip - 0.9 mm; Body - 1.4 mm | | | | |
| Needle Inner Diameter | Same as predicate | Tip - 0.6 mm; Body - 1.2 mm | | | | |
| Aspiration tube length | 600/900 mm | 600 mm | | | | |
| Tubing Outer diameter | 2.22 mm | 1.95 mm | | | | |
| Tubing Inner diameter | 1.33 mm | 1.35 mm | | | | |
| Cannula | Bent cannula on cork | Straight cannula mounted on cork | | | | |
| Cork and cannula geometry | Allows connection of a vacuum pump | Requires a connection/cork adapter | | | | |
| Technological characteristics - Materials | | | | | | |
| Aspiration needle | Same as predicate | Stainless steel | | | | |
| Needle grip | Same as predicate | Stainless steel | | | | |
| Aspiration tubing | Polyolefin based TPE (TPE-0) | Teflon | | | | |
| Cork | Same as predicate | Silicone | | | | |
| Cork adapter | Acrylonitrile Butadiene Styrene (ABS) Sinkral | Stainless steel | | | | |
| Cannula Luer | Methylmethacrylate Acrylonitrile Butadiene<br>Styrene (MABS) | Polypropylene | | | | |
| Cannula for flushing | Same as predicate | Stainless steel | | | | |
| Protective sheath | Polypropylene | N/A | | | | |
| Bonding material | Same as predicate | Epoxy resin | | | | |
The subject and predicate devices have the same intended use – flushing and/or aspiration of oocytes from ovarian follicles.
There are differences in design and materials used in different components between the subject and predicate devices. However, these differences do not raise different questions of safety and effectiveness. Similar features are also common in other cleared oocyte retrieval needles.
### Performance Testing:
{5}------------------------------------------------
The following studies have been performed to support substantial equivalence to the predicate device:
- Sterilization Validation Study per ISO 11137-2:2013 .
- Endotoxin Testing using LAL method per USP<85> to demonstrate that the subject device . meets acceptance criterion of ≤1.2 EU/ device
- Breakage and Elasticity Testing to demonstrate that the stainless needle of the subject . device meets the requirements in ISO 9626:2016
- Accelerated shelf-life testing to demonstrate that the subject device meets the following ● performance criteria at the end of three-year shelf-life:
- * Mouse embryo assay (MEA) - 1-Cell MEA: ≥80% embryos expanded to blastocysts within 96 hours
One-cell mouse embryos were exposed to subject device extracts and cultured at 37°C in an atmosphere containing 5% CO2. The percent of embryos developed to the expanded blastocyst stage within 96 hours were assessed in comparison with the control group.
- * Visual Inspection per ASTM F1886/F1886M-09
- * Dye Penetration Testing per ASTM F1929-12
- * Seal Strength Testing per ASTM F88/F88M-09
- * Vacuum Testing - Assessment of device function under vacuum pressure at -500 mmHg. Devices were shown to operate as intended without signs of damage
- * Flow Testing - Testing was conducted to assess that flow rates were within anticipated ranges under intended use conditions
- * Tensile Strength Testing - Testing was conducted to assess that tensile strength at joints were within anticipated ranges under intended use conditions
### Conclusions
The results of non-clinical performance testing described above demonstrate that the subject device is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.