K172049 · Cry IQ AB · GEH · Aug 30, 2017 · General, Plastic Surgery
Device Facts
Record ID
K172049
Device Name
CryoIQ DERM, CryoIQ PRO, CryoIQ EquiMED
Applicant
Cry IQ AB
Product Code
GEH · General, Plastic Surgery
Decision Date
Aug 30, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CryoIQ Multi LINE are reusable devices intended to destroy tissue during surgical procedures by applying extreme cold.
Device Story
CryoIQ Multi LINE is a reusable cryosurgical device for tissue destruction. It utilizes a replaceable nitrous oxide (N2O) cartridge (16g, 25g, or 40g) attached to a unit body. Operation involves pressing a lever to dispense a continuous stream of liquid N2O through a capillary tube/applicator tip. The device achieves a constant temperature of -89°C (-128°F) at the treatment site via heat evaporation upon phase transition, causing cellular necrosis. Used by physicians or medical professionals in clinical settings. The device is handheld and mechanical; no software or electronic components are involved. Benefits include precise, controlled application of extreme cold for lesion removal.
Clinical Evidence
Bench testing only. Testing included freezing effect verification, mechanical integrity, and gas consumption analysis. The device complies with ASTM F882-84 for Cryosurgical Medical Instruments. No clinical data was required or provided.
Technological Characteristics
Materials: brass, stainless steel, gold-plated components, borosilicate glass 3.3 (DURAN) tips, PTFE/acrylonitrile-butadiene-rubber seals. Energy source: pressurized N2O gas (50 bar/725 psi). Principle: heat evaporation via phase transition. Form factor: handheld unit with lever actuator and interchangeable tips (1mm-4mm, angled, long). Compliance: ASTM F882-84. No software or electronic connectivity.
Indications for Use
Indicated for the destruction of tissue during surgical procedures by applying extreme cold, including treatment of genital lesions, molluscum contagiosum, verruca plantaris, verruca plana, actinic keratosis, seborrheic keratosis, skin tags, and lentigo.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 30, 2017
Cry IO AB % Ms. Cherita James Regulatory Consultant M Squared Associates. Inc 515 Eight Avenue, St. 1212 New York, New York 10018
Re: K172049
Trade/Device Name: CryoIQ MULTI LINE Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: Class II Product Code: GEH Dated: July 5, 2017 Received: July 6, 2017
Dear Ms. James:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerelv.
# Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K172049
Device Name CryoIQ Multi LINE
Indications for Use (Describe)
The CryoIQ Multi LINE are reusable devices intended to destroy tissue during surgical procedures by applying extreme cold.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K172049
In accordance with 21 CFR 807.92, the following information constitutes the CryolQ AB summary for the CryolQ Multi LINE.
## I. Submission Date
August 28, 2017
## II. Submitter
| Name: | CryoIQ AB |
|----------|---------------------------------------------|
| Address: | Apelröds vägen 1<br>439 32 Onsala<br>Sweden |
| Contact person: | Stefan Skafte, President |
|-----------------|--------------------------|
| Phone number: | +46 31-400 500 |
| Fax number: | +46 300-56 94 99 |
## III. Device
| Name of device: | CryolQ MULTI line |
|-----------------------|-----------------------------------|
| Common or usual name: | Cryosurgical unit & accessories |
| Classification name: | Cryosurgical unit & accessories |
| Regulatory class: | Class II per regulations 878.4350 |
| Classification code: | GEH |
| Panel: | General & plastic surgery |
#### IV. Predicate devices
| | 510(k) | Company | Device |
|----------------------|---------|----------------------------------|-------------|
| Primary<br>predicate | K091721 | New Medical Technologies<br>GmbH | CryoSuccess |
| Reference | K024009 | H&O Equipments NV/SA | CryoProbe-C |
| | K102006 | STC Consulting LLC | CryOmega |
## V. Device Description
The CryolQ devices consist of a unit body with a lever and an applicator (capillary tube, tip). A disposable cartridge filled with nitrous oxide (N₂O, laughing qas) will be attached to the unit body by slightly screwing into it. The cartridge has a build-in valve for safety storage and release of the gas.
The CryolQ designed to dispense a continuous stream of liquid nitrous oxide when actuated by pressing the lever.
{4}------------------------------------------------
The device functions by means of heat evaporation upon phase transition where the nitrous oxide is applied to the treatment area by means of the applicator (capillary tube, tip) at a constant temperature of -89°C/-128°F (cold performance) followed by evaporation. Physicians or medical professionals can use the gas in procedures requiring the destruction of tissue using the extreme cold of nitrous oxide (-89°C/-128°F).
## VI. Intended use
The CryoIQ Multi LINE are reusable devices intended to destroy tissue during surgical procedures by applying extreme cold.
#### VII. Comparisons of the technological characteristics with the predicate device
CryolQ MULTI line exists of 3 devices, CryolQ DERM, CryoIQ PRO, CryolQ EquiMed. They all have the same intended use to destroy tissue during surgical procedures by applying extreme cold, same technical features but they differ in design.
| | Subject device | Predicate Device | Reference devices | | S.E |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|
| 510(k) | No number yet | K091721 | K024009 | K102006 | |
| 510(K)<br>sponsor | CryolQ AB | New Medical<br>Technologies<br>GmbH | H&O<br>Equipments<br>NV/SA | STC<br>Consulting<br>LLC | |
| Device<br>Name | CryolQ MULTI<br>line | CryoSuccess | CryoProbe-C | CryOmega | |
| Models | CryolQ DERM<br>CryolQ PRO<br>CryolQ EquiMed<br>Std tip (ø 1mm)<br>Dermatology tip<br>(ø 2mm)<br>Dermatology tip<br>(ø 3mm)<br>Dermatology tip<br>(ø 4mm)<br>Derm. tip, White<br>(ø 1mm)<br>Derm. tip, Blue<br>(ø 2mm)<br>Derm. tip, Green<br>(ø 3mm)<br>Long tip 13 cm<br>Angled tip 45° | CryoSuccess<br>Std tip (ø 1mm)<br>Std tip (ø 2mm)<br>Std tip (ø 3mm)<br>Std tip (ø 4mm)<br>Gynecology tip | CryoProbe I c<br>CryoProbe I x | CryOmega | |
| | Subject device | Predicate Device | Reference devices | | S.E |
| 510(k) | No number yet | K091721 | K024009 | K102006 | |
| 510(K)<br>sponsor | CryolQ AB | New Medical<br>Technologies<br>GmbH | H&O<br>Equipments<br>NV/SA | STC<br>Consulting<br>LLC | |
| Indication for<br>use | To destroy tissue<br>during surgical<br>procedures by<br>applying extreme<br>cold. | To destroy tissue<br>during surgical<br>procedures by<br>applying extreme<br>cold | To destroy<br>tissue during<br>surgical<br>procedures<br>by applying<br>extreme cold | Surgical<br>destruction of<br>target tissue<br>by applying<br>cryogenic<br>gases at<br>extreme low<br>temperatures.<br>The list below<br>shows<br>examples of<br>the type of<br>lesions that<br>can be<br>treated:<br>-Genital<br>lesions<br>-Molluscum<br>Contagiosum<br>-Verruca<br>Plantaris<br>-Verruca<br>Plana<br>-Actinic<br>Keratosis<br>-Seborrheic<br>Keratosis<br>-Skin Tags<br>-Lentigo | Yes |
| Type of<br>lesions to be<br>treated | -Genital lesions<br>-Molluscum<br>Contagiosum<br>-Verruca<br>Plantaris<br>-Verruca Plana<br>-Actinic Keratosis<br>-Seborrheic<br>Keratosis<br>-Skin Tags<br>-Lentigo | Not stated | Not stated | -Genital<br>lesions<br>-Molluscum<br>Contagiosum<br>-Verruca<br>Plantaris<br>-Verruca<br>Plana<br>-Actinic<br>Keratosis<br>-Seborrheic<br>Keratosis<br>-Skin Tags<br>-Lentigo | Yes |
| Nitrogen | Nitrous Oxide<br>(N2O, laughing<br>gas). Available in<br>16 g, 25 g and 40<br>g cartridge. | Nitrous Oxide<br>(N2O, laughing<br>gas). Available in<br>23.5 g cartridge. | Nitrous Oxide<br>(N2O,<br>laughing<br>gas).<br>Available in 8<br>g or 16 g<br>cartridges. | Nitrous Oxide<br>(N2O,<br>laughing<br>gas).<br>Available in<br>16 g<br>cartridge. | Yes |
| | Cartridge<br>pressure: 50 bar<br>(725 psi) | Cartridge<br>pressure: 50 bar<br>(725 psi) | Cartridge<br>pressure: 50<br>bar (725 psi) | Cartridge<br>pressure: 50<br>bar (725 psi) | Yes |
| | Subject device | Predicate Device | Reference devices | | S.E |
| 510(k) | No number yet | K091721 | K024009 | K102006 | |
| 510(K)<br>sponsor | CryolQ AB | New Medical<br>Technologies<br>GmbH | H&O<br>Equipments<br>NV/SA | STC<br>Consulting<br>LLC | |
| | Shelf life: 2 years | Shelf life: 2 years | Shelf life: 2<br>years | Shelf life: 2<br>years | Yes |
| Shelf life<br>Device | 5 years | 5 years (with 2<br>years warranty) | 2 years | Disposable<br>unit, is<br>discarded<br>after liquefied<br>gas is<br>emptied | Yes |
| Treatment<br>Temperature | Liquid freezing:<br>-89°C/-128°F<br>constant temp. | Liquid freezing:<br>-89°C/-128°F<br>constant temp. | Liquid<br>freezing:<br>-89°C/-128°F<br>constant<br>temp. | Liquid<br>freezing:<br>-89°C/-128°F<br>constant<br>temp. | Yes |
| Material | Housing:<br>Cryo device has<br>no housing.<br>The device<br>material is brass<br>and stainless<br>steel,<br>gold plated | Housing:<br>Cryo device has<br>no housing.<br>The device<br>material is brass<br>and stainless<br>steel,<br>gold plated | Housing:<br>Aluminum | Housing:<br>plastic | Yes |
| | Cryo tip: metal,<br>gold plated, cryo<br>tip in borosilicate<br>glass 3.3 (such<br>as DURAN) | Cryotip: metal,<br>gold plated,<br>cryotip in<br>borosilicate glass<br>3.3 (such as<br>DURAN) | Micro-<br>Applicator:<br>Unknown | Micro-<br>Applicator:<br>Unknown | Yes |
| | Lock cap:<br>Plastic | Protective Cap:<br>Thermoplastic<br>rubber | Lock cap:<br>Plastic | Lock cap:<br>Unknown | Yes |
| | Filter:<br>Stainless steel | Filter:<br>Stainless steel | Filter:<br>Unknown | Filter:<br>Plastic | Yes |
| | O-rings/Seals<br>PTFE and<br>acrylonitrile-<br>butadiene-rubber | O-rings/Seals<br>PTFE and<br>acrylonitrile-<br>butadiene-rubber | O-<br>rings/Seals:<br>Unknown | O-rings/Seals<br>Butadiene-<br>rubber | Yes |
| | Nitrous oxide<br>cartridges:<br>Metal | Nitrous oxide<br>cartridges:<br>Metal | Nitrous oxide<br>cartridges:<br>Metal | Nitrous oxide<br>cartridges:<br>Metal | Yes |
| Mode of Use | Apply spray<br>topically | Apply spray<br>topically | Apply spray<br>topically | Apply spray<br>topically | Yes |
| Storage<br>Conditions | Protect the unit<br>against heat and<br>exposure to<br>direct sunlight.<br>The storage<br>temperature is<br>between - | Protect the unit<br>against heat and<br>exposure to direct<br>sunlight. The<br>storage<br>temperature is<br>between - | Store in a<br>cool dry<br>place and<br>keep out of<br>reach of<br>children,<br><50°C/122°F | <50°C/122°F | Yes |
| | Subject device | Predicate Device | Reference devices | | S.E |
| 510(k) | No number yet | K091721 | K024009 | K102006 | |
| 510(K)<br>sponsor | CryolQ AB | New Medical<br>Technologies<br>GmbH | H&O<br>Equipments<br>NV/SA | STC<br>Consulting<br>LLC | |
| | max.<br>50°C/122°F | 10°C/14°F and<br>max. 45°C/113°F. | | | |
| Cleaning | The unit body<br>and cryo tip can<br>be cleaned and<br>disinfected with<br>an alcohol-based<br>disinfectant or<br>alcohol. | The unit body and<br>cryo tip can be<br>cleaned and<br>disinfected with<br>an alcohol-based<br>disinfectant or<br>alcohol. | All external<br>parts can be<br>wiped with a<br>cloth soaked<br>in any non-<br>corrosive<br>sterilization<br>solution or<br>alcohol. | Unknown | Yes |
| Sterilization | N/A | Steam sterilize<br>applicators at<br>134°C/273°F,<br>according to the<br>instruction in the<br>manual of your<br>steam sterilizer<br>and according to<br>the country<br>specific law | N/A | N/A | Yes |
| Safety | Complies with<br>ASTM: F882-84<br>for Cryosurgical<br>Medical<br>Instrument | Complies with<br>ASTM: F882-84<br>for Cryosurgical<br>Medical<br>Instrument | Complies<br>with ASTM:<br>F882-84 for<br>Cryosurgical<br>Medical<br>Instrument | Complies<br>with ASTM:<br>F882-84 for<br>Cryosurgical<br>Medical<br>Instrument | Yes |
| Operating<br>Principle | CryolQ functions<br>by means of heat<br>evaporation upon<br>phase transition<br>where liquid<br>nitrous oxide<br>(N₂O, laughing<br>gas) is applied to<br>the treatment<br>area by means of<br>a capillary tube<br>(applicator, tip) at<br>a constant<br>temperature of -<br>89°C/-128°F<br>(cold<br>performance)<br>followed by<br>evaporation.<br>This results in<br>cellular<br>destruction<br>(necrosis) in the<br>treatment area.<br>The amount of<br>gas dispensed is<br>controlled by the<br>medical<br>professional | Cryosuccess<br>functions by<br>means of heat<br>evaporation upon<br>phase transition<br>where liquid<br>nitrous oxide<br>(N₂O, laughing<br>gas) is applied to<br>the treatment<br>area by means of<br>a capillary tube<br>(applicator, tip) at<br>a constant<br>temperature of -<br>89°C/-128°F (cold<br>performance)<br>followed by<br>evaporation. This<br>results in cellular<br>destruction<br>(necrosis) in the<br>treatment area.<br>The amount of<br>gas dispensed is<br>controlled by the<br>medical<br>professional<br>Pressing the lever<br>of the device. | The patented<br>principle of<br>the<br>CryoProbe is<br>based upon<br>the direct<br>flow of N₂O in<br>liquid phase<br>for the<br>purpose of<br>freezing with<br>resulting in<br>necrosis of<br>tissue in the<br>practice of<br>medicine. To<br>achieve this<br>phenomenon<br>an<br>economical<br>gas cartridge<br>is used. The<br>cartridge is<br>filled with<br>liquid N₂O<br>(83%) and<br>the rest with<br>N₂O gas. The<br>liquid N₂O is<br>the | Gas<br>dispensed<br>using<br>actuator<br>lever. Spray<br>controlled by<br>on/off<br>actuator. | Yes |
| | Subject device | Predicate Device | Reference devices | | S.E |
| 510(k) | No number yet | K091721 | K024009 | K102006 | |
| 510(K)<br>sponsor…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.