← Product Code [PLA](/productcode/PLA) · K172029

# EnsoETM (K172029)

_Attune Medical · PLA · Jan 5, 2018 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K172029

## Device Facts

- **Applicant:** Attune Medical
- **Product Code:** [PLA](/productcode/PLA.md)
- **Decision Date:** Jan 5, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.5910
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

Model # ECD03-A: The EnsoETM is a thermal regulating device, intended to: • connect to a Gaymar Medi-Therm III Conductive Hyper/Hypothermia System or Stryker Altrix Precision Temperature Management System to control patient temperature, • allow enteral administration of fluids, • and provide gastric decompression and suctioning. Model # ECD04-A: The EnsoETM is a thermal regulating device, intended to: • connect to a Cincinnati Sub-Zero Blanketrol II or Blanketrol III Hyper-Hypothermia System to control patient temperature, • allow enteral administration of fluids, • and provide gastric decompression and suctioning.

## Device Story

EnsoETM is a multi-lumen silicone tube inserted into the esophagus like a standard orogastric tube. Two lumens connect to an external heat exchanger (Gaymar Medi-Therm III, Stryker Altrix, or Cincinnati Sub-Zero Blanketrol II/III) to circulate fluid for patient temperature modulation. A third central lumen connects to wall suction for gastric decompression or an enteral system for fluid administration. Used by clinicians in acute care settings. Device provides dual functionality: thermal regulation and gastric access. Benefits include simultaneous temperature management and gastric maintenance. Single-use, disposable, non-implantable; duration of use ≤36 hours.

## Clinical Evidence

Bench testing only. Testing included tensile force, gastric lumen burst strength, leakage, flow rate, vacuum resistance, positive pressure leakage, stress cracking, axial load/torque separation resistance, and dimensional/flexural modulus verification per ISO 80369-3:2016.

## Technological Characteristics

Multi-lumen medical-grade silicone tube. Features male ENFit connector for gastric lumen (compliant with ISO 80369-3:2016). Coolant lumen fittings vary by model (Clik-Tite or Colder PLC series). Non-implantable, single-use. Connects to external hyper/hypothermia heat exchangers. No electronic components or software.

## Regulatory Identification

An esophageal thermal regulation device is a prescription device used to apply a specified temperature to the endoluminal surface of the esophagus via an external controller. This device may incorporate a mechanism for gastric decompression and suctioning. The device is used to regulate patient temperature.

## Special Controls

In combination with the general controls of the FD&C Act, the Esophageal Cooling Device is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) The patient contacting materials must be demonstrated to be biocompatible.
(2) Non-clinical performance evaluation must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Mechanical integrity testing.
(ii) Testing to determine temperature change rate(s).
(iii) Testing to demonstrate compatibility with the indicated external controller.
(iv) Shelf life testing.
(3) Animal testing must demonstrate that the device does not cause esophageal injury and that body temperature remains within appropriate boundaries under anticipated conditions of use.
(4) Labeling must include the following:
(i) Detailed insertion instructions.
(ii) Warning against attaching the device to unintended connections, such as external controllers for which the device is not indicated, or pressurized air outlets instead of vacuum outlets for those devices, including gastric suction.
(iii) The operating parameters, name, and model number of the indicated external controller.
(iv) The intended duration of use.

## Predicate Devices

- Attune Medical EnsoETM (model #: ECD01-A); [K170009](/device/K170009.md)
- Attune Medical EnsoETM (model #: ECD02-A); [K152450](/device/K152450.md)

## Reference Devices

- Maquet Critical Care Ab Edi Catheter ENFit; [K153688](/device/K153688.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 5, 2018

Attune Medical Erik Kulstad President/CEO 3440 S. Dearborn St. #215-South Chicago, Illinois 60616

Re: K172029

Trade/Device Name: EnsoETM, Models ECD03-A and ECD04-A Regulation Number: 21 CFR 870.5910 Regulation Name: Esophageal Thermal Regulation Device Regulatory Class: Class II Product Code: PLA Dated: December 1, 2017 Received: December 5, 2017

Dear Erik Kulstad:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Nicole G. Ibrahim -S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# SECTION 4: Indications for Use Statement

| Form Approved: OMB No.                          | 0910-0120                                                                                                                                                                                                                                                                                                                                                 |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Expiration Date:                                | January 31, 2017                                                                                                                                                                                                                                                                                                                                          |
|                                                 | See PRA Statement below.                                                                                                                                                                                                                                                                                                                                  |
| 510(k) Number (if known)                        | K172029                                                                                                                                                                                                                                                                                                                                                   |
| Device Name                                     | EnsoETM                                                                                                                                                                                                                                                                                                                                                   |
| Indications for Use (Describe)                  |                                                                                                                                                                                                                                                                                                                                                           |
|                                                 | Model # ECD03-A:<br>The EnsoETM is a thermal regulating device, intended to:<br>• connect to a Gaymar Medi-Therm III Conductive Hyper/Hypothermia System or Stryker Altrix Precision Temperature<br>Management System to control patient temperature,<br>• allow enteral administration of fluids,<br>• and provide gastric decompression and suctioning. |
|                                                 | Model # ECD04-A:<br>The EnsoETM is a thermal regulating device, intended to:<br>• connect to a Cincinnati Sub-Zero Blanketrol II or Blanketrol III Hyper-Hypothermia System to control patient temperature,<br>• allow enteral administration of fluids,<br>• and provide gastric decompression and suctioning.                                           |
| Type of Use (Select one or both, as applicable) |                                                                                                                                                                                                                                                                                                                                                           |
|                                                 | <span style="text-decoration: underline;">Prescription Use</span> (Part 21 CFR 801 Subpart D) <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span>                                                                                                                                                                                                    |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services  

 Food and Drug Administration  

 Office of Chief Information Officer  

 Paperwork Reduction Act (PRA) Staff  

 PRAStaff@fda.hhs.gov

*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*

FORM FDA 3881 (8/14)

Page 1 of 1

PSC Publishing Scraces (301) 443-6740 EF

{3}------------------------------------------------

### Summary of Safety and Effectiveness

#### Submitter / 510(k) Holder

| Company:                     | Attune Medical                                          |
|------------------------------|---------------------------------------------------------|
| Address:                     | 3440 S. Dearborn St.<br>#215-South<br>Chicago, IL 60616 |
| Phone:                       | +1-312-725-4756                                         |
| Contact Person:              | Erik Kulstad<br>CEO/President                           |
| Date Prepared:               | June 30, 2017                                           |
| Device Name & Classification |                                                         |
| Trade Name:                  | EnsoETM                                                 |
| Model Number(s):             | ECD03-A, ECD04-A                                        |
| Classification Name:         | Esophageal Thermal Regulation Device (21 CFR 870.5910)  |
| Product Code:                | PLA                                                     |
| Class:                       | II                                                      |

#### Predicate Devices

1. Attune Medical EnsoETM (model #: ECD01-A); K170009

2. Attune Medical EnsoETM (model #: ECD02-A); K152450

#### Reference Device

1. Maquet Critical Care Ab Edi Catheter ENFit; K153688

#### Device Description

The EnsoETM is a multi-lumen silicone tube that is placed in the esophagus in a similar manner to a standard orogastric tube to control a patient's temperature, while simultaneously maintaining access to the stomach to allow gastric decompression, drainage, and enteral administration of fluids, thereby maintaining the functionality of the standard orogastric tube. Modulation and control of patient temperature is achieved by connecting the device to an external heat exchanger. Two lumens connect to the external heat exchanger. A third central lumen connects to wall suction to allow standard gastric decompression or an enteral administration system to allow enteral administration of fluids. The device is made of standard medical-grade silicone. It is a single-use, disposable, non-implantable device with an intended duration of use of 36 hours or less.

{4}------------------------------------------------

#### Indications for Use (Subject Device 1; Model #: ECD03-A)

The EnsoETM is a thermal regulating device, intended to:

- . connect to a Gaymar Medi-Therm III Conductive Hyper/Hypothermia System or Stryker Altrix Precision Temperature Management System to control patient temperature,
- allow enteral administration of fluids,
- and provide gastric decompression and suctioning.

#### Indications for Use (Subject Device 2; Model #: ECD04-A)

The EnsoETM is a thermal regulating device, intended to:

- connect to a Cincinnati Sub-Zero Blanketrol III or Blanketrol III Hyper-Hypothermia System to control patient temperature,
- . allow enteral administration of fluids,
- and provide gastric decompression and suctioning.

#### Technological Characteristics

The EnsoETM product family consists of four distinct models: the ECD01-A (Predicate Device 1 cleared under K170009), ECD02-A (Predicate Device 2 cleared under K152450), ECD03-A (Subject Device 1), and ECD04-A (Subject Device 2). Table 1 describes the differences between the four models. Device characteristics not described in Table 1 are identical for all models.

| Model #                            | Coolant Lumen<br>Fittings       | Gastric Lumen<br>Fittings | Intended heat<br>exchanger                                | Enteral administration<br>of fluids? |
|------------------------------------|---------------------------------|---------------------------|-----------------------------------------------------------|--------------------------------------|
| ECD01-A<br>(Predicate<br>Device 1) | Clik-Tite<br>connectors         | 5° silicone taper         | Stryker/Gaymar Medi-<br>Therm III or Stryker<br>Altrix    | No                                   |
| ECD02-A<br>(Predicate<br>Device 2) | Colder PLC series<br>connectors | 5° silicone taper         | Cincinnati Sub-Zero<br>Blanketrol II or<br>Blanketrol III | No                                   |
| ECD03-A<br>(Subject<br>Device 1)   | Clik-Tite<br>connectors         | Male ENFit<br>connector   | Stryker/Gaymar Medi-<br>Therm III or Stryker<br>Altrix    | Yes                                  |
| ECD04-A<br>(Subject<br>Device 2)   | Colder PLC series<br>connectors | Male ENFit<br>connector   | Cincinnati Sub-Zero<br>Blanketrol II or<br>Blanketrol III | Yes                                  |

#### Table 1: Differences between EnsoETM models

#### Non-Clinical Performance Testing

Non-clinical performance testing was conducted to demonstrate the technological characteristics of the subject devices are substantially equivalent to the predicate devices. The following tests were performed:

{5}------------------------------------------------

| Test performed                                 | Discussion                                                                                                                                                                                                              |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Tensile force                                  | Tensile force testing demonstrated the ENFit connector and ENFit connector joint on the subject devices are suitable for use, including insertion into and removal from the esophagus.                                  |
| Gastric lumen<br>burst strength                | Gastric lumen burst strength testing demonstrated the gastric lumen of the subject devices can withstand the operating pressures supplied by an enteral administration system.                                          |
| Gastric lumen<br>leakage                       | Gastric lumen leakage testing demonstrated the gastric lumen, gastric lumen connector, and gastric lumen connector joint of the subject devices do not allow enterally administered fluids to escape the gastric lumen. |
| Gastric lumen<br>flow rate                     | Gastric lumen flow rate testing demonstrated the gastric lumen of the subject devices is capable of providing flow of fluids representative of those clinically used for enteral feeding in acute care facilities.      |
| Resistance to<br>vacuum                        | Resistance to vacuum testing demonstrated the subject devices are capable of performing gastric decompression and suctioning.                                                                                           |
| Positive pressure<br>leakage                   | Positive pressure liquid leakage testing demonstrated the ENFit connector does not leak fluid while under pressure.                                                                                                     |
| Leakage by<br>pressure decay                   | Leakage by pressure decay testing demonstrated the ENFit connector does not leak fluid while under pressure.                                                                                                            |
| Stress cracking                                | Stress cracking testing demonstrated the ENFit connector does not fail when subjected to typical use-case stresses.                                                                                                     |
| Resistance to<br>separation from<br>axial load | Resistance to separation from axial load testing demonstrated the ENFit connector does not inadvertently separate when subjected to typical use-case axial loads.                                                       |
| Resistance to<br>separation from<br>unscrewing | Resistance to separation from unscrewing testing demonstrate the ENFit connector does not inadvertently separate when subjected to typical use-case torques.                                                            |
| Resistance to<br>overriding                    | Resistance to overriding testing demonstrated the ENFit connector does not override the threads when subjected to typical use-case torques.                                                                             |
| Disconnection by<br>unscrewing                 | Disconnection by unscrewing demonstrated the ENFit connector is capable of being disconnected by the end user with a reasonable amount of torque.                                                                       |
| Dimensional<br>verification                    | Dimensional verification demonstrated the dimensions of the ENFit connector satisfy the dimensional requirements of ISO 80369-3:2016.                                                                                   |
| Flexural modulus                               | Flexural modulus verification demonstrated the material used to manufacture the ENFit connector satisfies the flexural modulus requirements of ISO 80369-3:2016.                                                        |

### Conclusion

The results of all performance testing demonstrate the subject devices have technological characteristics that do not raise new questions of safety or effectiveness when compared to the predicate devices. Therefore, the subject devices are substantially equivalent to the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K172029](https://fda-staging.innolitics.com/device/K172029)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).