This product is used for temporary internal drainage from the ureteropelvic junction to the bladder. They are not intended to remain indwelling more than twelve months.
Device Story
Sterile, single-use ureteral stent set for temporary internal drainage from ureteropelvic junction to bladder; used in clinical settings by physicians. Device consists of radiopaque silicone rubber double pigtail stent with sideports, wire guide, stent positioner with lockable fitting, and T-inserter. Stent features graduation marks for visualization and braided tether for repositioning/removal. Lockable fitting on positioner secures wire guide for one-pass placement. Stent remains in urinary tract via pigtail loops; tether removed after 14 days. Provides drainage to relieve obstruction; benefits patient by maintaining ureteral patency.
Radiopaque silicone rubber stent; 6.0-8.5 Fr diameter; 20-30 cm length. Includes stainless steel or tungsten markers, PTFE-coated stainless steel wire guide, and polyethylene/vinyl positioner. Sterilized via ethylene oxide. Mechanical drainage device; no software or electronic components.
Indications for Use
Indicated for temporary internal drainage from the ureteropelvic junction to the bladder in patients requiring ureteral stenting. Not intended for indwelling use exceeding twelve months.
Regulatory Classification
Identification
A ureteral stent is a tube-like implanted device that is inserted into the ureter to provide ureteral rigidity and allow the passage of urine. The device may have finger-like protrusions or hooked ends to keep the tube in place. It is used in the treatment of ureteral injuries and ureteral obstruction.
Predicate Devices
Universa™ Firm Ureteral Stents and Stent Sets (K161236)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
March 26, 2018
Cook Incorporated Karthik Pillai Regulatory Science Specialist 750 Daniels Wav P.O. Box 489 Bloomington, IN 47402
Re: K172017
Trade/Device Name: Black Silicone Filiform Double Pigtail Stent Set Regulation Number: 21 CFR 876.4620 Regulation Name: Ureteral stent Regulatory Class: II Product Code: FAD Dated: February 16, 2018 Received: February 20, 2018
Dear Karthik Pillai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Joyce M. Whang -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
510(k) Number (if known) K172017
#### Device Name
The Black Silicone Filiform Double Pigtail Ureteral Stent Set
Indications for Use (Describe)
This product is used for temporary internal drainage from the ureteropelvic junction to the bladder. They are not intended to remain indwelling more than twelve months.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is in white, uppercase letters, with a registered trademark symbol next to it. Below "COOK" is the word "MEDICAL" in white, uppercase letters, set against a slightly darker red background.
# 510(k) Summary
COOK INCORPORATED 750 DANIELS WAY, P.O. BOX 489 BLOOMINGTON, IN 47402-0489 V.COOKM
# Black Silicone Filiform Double Pigtail Stent Set 21 CFR §876.4620 Date Prepared: June 30, 2017
#### Submitted By:
| Submission: | Traditional 510(k) Premarket Notification |
|-----------------------|---------------------------------------------------------------|
| Applicant: | Cook Incorporated |
| Applicant Address: | Cook Incorporated<br>750 Daniels Way<br>Bloomington, IN 47404 |
| Contact: | Karthik Pillai |
| Email: | RegSubmissions@CookMedical.com |
| Contact Phone Number: | 812-335-3575 x104929 |
| Contact Fax Number: | 812-332-0281 |
#### Device Information:
| Trade Name: | Black Silicone Filiform Double Pigtail Ureteral Stent Set |
|-----------------------|-----------------------------------------------------------|
| Common Name: | Ureteral Stent |
| Classification Name: | Ureteral Stent |
| Regulation: | 21 CFR §876.4620 |
| Product Code: | FAD |
| Device Class: | II |
| Classification Panel: | Gastroenterology/Urology |
#### Predicate Device:
The predicate device is Universa™ Firm Ureteral Stents and Stent Sets, cleared under 510(k) K161236 on January 27, 2017.
#### Device Description:
The Black Silicone Filiform Double Pigtail Stent Set is a sterile, single-use device. The stents are available with outside diameters of 6.0, 7.0, or 8.5 Fr and specified lengths ranging from 20 to 30 centimeters. The stents are flexible, tubular double pigtail stents constructed of radiopaque silicone rubber. Sideports extend along the body of the stent as
{4}------------------------------------------------
well as on the pigtails to provide drainage. The stents are secured in the urinary tract with pigtail loops on the proximal and distal ends. A braided tether for repositioning and removal of the device is located on the proximal pigtail (bladder end) of the stent. Along the stent are graduation marks to provide visualization during stent advancement and placement. The Black Silicone Filiform Double Pigtail Stent Set includes a stent, wire guide, a stent positioner with a lockable fitting, and a T-inserter. The lockable fitting on the proximal end of the stent positioner secures the position of the stent positioner on the wire guide and allows for one-pass placement of the stent. The Black Silicone Filiform Double Pigtail Stent Set is labeled for a 12-month indwell time; however, the tether should be removed after 14 days.
### Intended Use:
This product is used for temporary internal drainage from the ureteropelvic junction to the bladder. They are not intended to remain indwelling more than twelve months.
### Comparison to Predicate Device:
The Black Silicone Filiform Double Pigtail Ureteral Stent Set is substantially equivalent to the predicate device, the Universa™ Firm Ureteral Stents and Stent Sets, as described in K161236. The table below presents the similarities and differences between the devices for substantial equivalence purposes. The differences between the subject device and the predicate device do not raise any new issues of safety and effectiveness. Performance data are available to support substantial equivalence and were developed using acceptable scientific methods for evaluation.
| | Universa™ Firm Ureteral Stent and<br>Stent Set (K161236) | Black Silicone Filiform Double Pigtail<br>Ureteral Stent Set (Subject of this<br>submission) |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Used for temporary internal drainage<br>from the ureteropelvic junction to the<br>bladder. Ureteral stents have been used to<br>relieve obstruction in a variety of benign,<br>malignant, and post-traumatic conditions.<br>The stents may be placed using<br>endoscopic, percutaneous, or open<br>surgical techniques. | Used for temporary internal drainage<br>from the ureteropelvic junction to the<br>bladder. They are not intended to remain<br>indwelling more than twelve months. |
| Maximum Indwell<br>Time | 12 months | Identical |
| Sterilization | Ethylene oxide | |
| Maximum Shelf<br>Life | 3 years | Identical |
{5}------------------------------------------------
| | UniversaTM Firm Ureteral Stent and<br>Stent Set (K161236) | Black Silicone Filiform Double Pigtail<br>Ureteral Stent Set (Subject of this<br>submission) |
|-----------------------------------|-----------------------------------------------------------|----------------------------------------------------------------------------------------------|
| Principle Operation<br>Components | Double pigtail stent Stent positioner Wire guide | Double pigtail stent Stent positioner with lockable fitting Wire guide T-inserter |
| Dimensions | | |
| Loop Diameter | 13 mm - 20 mm | 15 mm - 18 mm |
| Stent Outer<br>Diameter | 5 Fr - 8 Fr | 6 Fr - 8.5 Fr |
| Stent Length | 18 cm - 30 cm | 20 cm - 30 cm |
| Pusher/Positioner | 5 Fr and 7 Fr | 7 Fr, 8 Fr, and 9 Fr |
| Materials | | |
| Stent | Radiopaque Polycarbonate-based<br>Polyurethane | Radiopaque Silicone Rubber |
| Stent Markings | Black Ink | White Ink |
| Tether | Tevdek® (Polyethylene Terephthalate) | Identical |
| Pusher/Positioner | Polyethylene Non-radiopaque or Vinyl | Identical |
| Pusher/Positioner<br>Marker | Stainless Steel | Stainless Steel or Tungsten |
| Wire Guide | PTFE-coated Stainless Steel | Identical |
#### Technological Characteristics:
The Black Silicone Filiform Double Pigtail Ureteral Stent Set was subjected to the following tests to assure reliable design and performance under the specified testing parameters:
- 1. Sterility
- 2. Packaging
- 3. Biocompatibility
- 4. Flow Rate
- 5. Retention Strength
- 6. Tensile Strength
- 7. Curl Restoration
- 8. Kink Radius
- 9. Radiopacity
- 10. MRI Testing
{6}------------------------------------------------
### Conclusion:
The results of the testing provide reasonable assurance that the Black Silicone Filiform Double Pigtail Ureteral Stent Set has been designed such that it will function as intended. The subject device does not raise new questions of safety or effectiveness as compared to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.