ARIX Humerus System

K172008 · Jeil Medical Corporation · HRS · Dec 1, 2017 · Orthopedic

Device Facts

Record IDK172008
Device NameARIX Humerus System
ApplicantJeil Medical Corporation
Product CodeHRS · Orthopedic
Decision DateDec 1, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

ARIX Humerus System is indicated for fractures, fracture dislocations, osteotomies and nonunions of the proximal humerus, particularly in osteopenic bone.

Device Story

The ARIX Humerus System is a rigid internal fixation system consisting of titanium plates and screws. It is used by orthopedic surgeons in a clinical/surgical setting to stabilize bone segments during the healing of proximal humerus fractures, dislocations, osteotomies, or nonunions. The system includes various manual surgical instruments (e.g., drill bits, driver shafts, depth gauges, drill guides) to facilitate implantation. The device is provided non-sterile and must be autoclaved by the hospital prior to use. By providing mechanical stabilization to the bone, the system facilitates proper healing and restores structural integrity to the humerus.

Clinical Evidence

No clinical studies were performed. Substantial equivalence is supported by bench testing, including bending strength and bending fatigue tests per ASTM F382, and compliance with ASTM F543 for bone screws.

Technological Characteristics

Materials: Unalloyed Titanium (ASTM F67) and Ti-6Al-4V alloy (ASTM F136). Components: Plates and screws of varying lengths/diameters. Principle: Rigid internal fixation via locking/cortical screws and plates. Sterilization: Moist heat (autoclave) per ISO 17665-1:2006 to SAL 10^-6. Connectivity: None (mechanical device).

Indications for Use

Indicated for patients with fractures, fracture dislocations, osteotomies, and nonunions of the proximal humerus, including those with osteopenic bone.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text. Jeil Medical Corporation Sejin Ryu RA Specialist 702.703.704.705.706.804.805.807.812-ho. 55. Digital-ro34-gil. Guro-gu, Seoul, 152-728, Korea Re: K172008 Trade/Device Name: ARIX Humerus System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: October 19, 2017 Received: October 26, 2017 Dear Sejin Ryu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. November 30, 2017 {1}------------------------------------------------ Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the text 'K172008 Page 1 of 1'. This text likely represents a document identifier and page number. The document identifier is 'K172008', and the page number indicates that it is page 1 of 1, suggesting it is a single-page document. 702-703-704-705-706-804-805-807 Digital-ro34-gil, Guro-gu, 08378 Tel : +82 2 850 3533 / Fax : +82 2 850 3536 ## Indication for Use 510(k) Number (if known): K172008 Device Name: ARIX Humerus System Indication for Use: ARIX Humerus System is indicated for fractures, fracture dislocations, osteotomies and nonunions of the proximal humerus, particularly in osteopenic bone. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) {3}------------------------------------------------ 702-703-704-705-706-804-805-807-812-ho, 55, Digital-ro34-gil, Guro-gu, 08378, Korea Tel : +82 2 850 3533 / Fax : +82 2 850 3536 ### 510(k) Summary [As required by 21 CRF 807.92] ### 1. Date Prepared [21 CFR 807.92(a)(1)] 29 November 17 - #### 2. Submitter's Information [21 CFR 807.92(a)(1)] - Name of Sponsor: Jeil Medical Corporation 702·703·704·705·706·804·805·807·812-ho,55, Digital-ro34-gil, Address: -Guro-gu, Seoul, 152-728, Korea - Contact Name: Sejin RYU / RA Specialist - Telephone No.: +82 2 850 3500 - -Fax No.: +82 2 850 3536 - Email Address: rsj@jeilmed.co.kr - Registration No.: 3004049923 - Manufacturer: Same as Sponsor O - । Address: Same as Sponsor #### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)] - Trade Name: ARIX Humerus System O - Common Name: Bone Plate and Bone Screw - Classification Name: Single/multiple component metallic bone fixation appliances and accessories - Orthopedic Classification Panel: - Classification Regulation: 21 CFR 880.3030 O - 0 Product Code: HRS, HWC - Device Class: ll #### 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)] The identified predicate devices within submission are shown as follows: - 510(k) Number: K011815 - Applicant: Synthes (USA) - SYNTHES LCP PROXIMAL HUMERUS PLATES ● Device Name: {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo and contact information for JEIL MEDICAL CORPORATION. The logo is on the top left, and the address is listed below. The address is 702-703-704-705-706-804-805-807-812-ho, 55, Digital-ro34-gil, Guro-gu, 08378, Korea. The telephone number is +82 2 850 3533, and the fax number is +82 2 850 3536. - 510(k) Number: K171285 ● - Applicant: Jeil Medical Corporation - Common Name: Bone Plate. Bone Screw - 0 Device Name: ARIX Diaphysis System #### Reference Predicate Device - - 510(k) Number: K151468 - Applicant: Jeil Medical Corporation - 0 Common Name: Bone Plate. Bone Screw - O Device Name: ARIX Wrist System - 510(k) Number: K152158 - O Applicant: Jeil Medical Corporation - 0 Common Name: Bone Plate, Bone Screw - Device Name: ARIX Ankle System There are no significant differences between the Model ARIX Humerus System and the predicate devices that would adversely affect the use of the product. It is substantially equivalent to these devices in design, function, materials, and operational principles as internal fixation components. #### 5. Description of the Device [21 CFR 807.92(a)(4)] The ARIX Humerus System is rigid fixation consisting of plates and screws in various configurations, shapes and sizes, as follows: | | Plate | Cortical Screw | Locking Screw | |------------------------|----------------------------------|----------------------------------------------|---------------------------------------------------| | Type/<br>Configuration | 35-PLHU series | 35-SO-L series | 35L-SO-L series<br>40L-SA series<br>45L-CO series | | Material | ASTM F 67,<br>Unalloyed Titanium | ASTM F 136,<br>Titanium Alloy<br>(Ti-6Al-4V) | ASTM F 136,<br>Titanium Alloy<br>(Ti-6Al-4V) | The ARIX Humerus System is consists of plates and screws. The ARIX Humerus System is made of Pure Titanium and Titanium Alloy (Ti-6AL-4V), which meet ASTM F67, Standard Specification for Unalloyed Titanium for Surgical Implant Applications, and ASTM F136, Standard Specification for Wrought Titanium- 6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications, which h are widely used for surgical implants with well-known biocompatibility. The plates vary essentially through different lengths and number of plate holes. The screws vary essentially through different lengths and diameters. It also includes various manual surgical instruments, as drill bits, driver shafts and depth gauge. It also includes various manual surgical instruments, such as hand body, driver, drill bits, depth gauge, drill sleeve, forceps, drill guide block, sleeve handle and drill guide. The ARIX Humerus System not provided sterile. It is required to be sterilized via autoclave method to reach a SAL of 10-6 by the hospital prior to surgery. The sterilization method is presented in the instruction, which was validated per ISO 17665-1: 2006 Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices. {5}------------------------------------------------ 702-703-704-705-706-804-805-807-812-ho, 55, Digital-ro34-gil. Guro-gu. 08378. Korea Tel : +82 2 850 3533 / Fax : +82 2 850 3536 #### 6. Indication for Use [21 CFR 807.92(a)(5)] ARIX Humerus System is indicated for fractures, fracture dislocations, osteomies, and non-unions of the proximal humerus, particularly in osteopenic bone. #### 7. Technological Characteristics [21 CFR 807.92(a)(6)] ARIX Humerus System Bone Plate: Based on a technical feature comparison, the subject device was found to be similar to predicate devices with regard to design and materials. The subject plates also have a locking feature, similar to the design used in the predicate device (K011815) ARIX Humerus System Bone Screw: They share similar head, neck, and thread designs as the screws that are currently cleared under the predicate device (K011815 and K171285). #### Non-Clinical Test Summary: Bench tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards: - ASTM F 382, Standard Specification and Test Method for Metallic Bone Plates - ASTM F 543, Standard Specification and Test Methods for Metallic Medical Bone Screws ● The following tests were performed with the predicate device: - Plate - Bending strength test per ASTM F382 - Bending fatigue test per ASTM F382 The results of this testing indicate that the ARIX Humerus System is equivalent to predicate device. #### Clinical Test Summary No clinical studies were considered necessary and performed #### 8. Substantial Equivalence [21 CFR 807.92(b)(a) and 807.92] When compared to the predicate device (K011815, K171285) the ARIX Ankle Distal Tibia System presented in this submission has the same: - Intended Use - - -Technological characteristics - -Operating principle - -Design features - -Performance - -Biocompatibility - -Materials - -Method of sterilization and sterility assurance level #### 9. Conclusion [21 CFR 807.92(b)(3)1 In all respects, the ARIX Humerus System is the equivalent of currently marketed devices. This device is made of the same materials and has similar dimensions and characteristics. This device is manufactured from titanium that is used generally in this kind of bone plate/screw system. This device. ARIX Humerus, is substantially equivalent in design, material, and function to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...