PICO 7 Single Use Negative Pressure Wound Therapy System
K172005 · Smith & Nephew Medical Limited · OMP · Feb 2, 2018 · General, Plastic Surgery
Device Facts
Record ID
K172005
Device Name
PICO 7 Single Use Negative Pressure Wound Therapy System
Applicant
Smith & Nephew Medical Limited
Product Code
OMP · General, Plastic Surgery
Decision Date
Feb 2, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
PICO 7 is indicated for patients who would benefit from a suction device (Negative Pressure Wound Therapy) as it may promote would healing via removal of low to moderate levels of exudate and infectious materials. Appropriate wound types include: Chronic Acute Traumatic Subacute and dehisced wounds Partial-thickness burns Ulcers (such as diabetic or pressure) Flaps and grafts Closed Surgical incisions PICO 7 Single Use Negative Pressure Wound Therapy System is suitable for use both in a hospital and homecare setting.
Device Story
PICO 7 is a portable, battery-powered, electro-mechanical suction pump system for negative pressure wound therapy (NPWT). It connects via flexible tubing to a super-absorbent adhesive dressing. The system applies negative pressure (-60 to -100 mmHg) to the wound bed to remove exudate and infectious materials, promoting healing. Designed for single-use, the pump and dressings are supplied sterile. It is intended for use in hospital and homecare settings by healthcare professionals or patients/caregivers. The device incorporates a magnet. The system manages low to moderate wound exudate levels. Clinical benefit is derived from the maintenance of a consistent negative pressure environment at the wound site, facilitating fluid management and wound closure.
Clinical Evidence
Bench testing only. Wound model testing evaluated fluid management and pressure maintenance across various dressing sizes (10x20cm, 25x25cm, multisite) at low (0.6 g/cm²/24h) and moderate (1.1 g/cm²/24h) flow rates. Results confirmed dressings maintained negative pressure between -60 mmHg and -100 mmHg for at least 95% of test duration. Human factors summative studies confirmed the device is safe and effective for intended users (healthcare professionals and lay users).
Technological Characteristics
Electro-mechanical suction pump; magnet-containing. Materials evaluated per ISO 10993-1. Electrical safety/EMC compliance: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11. Sterile, single-use. Operates at -60 to -100 mmHg. Flexible tubing and super-absorbent adhesive dressings.
Indications for Use
Indicated for patients requiring negative pressure wound therapy to promote healing by removing low to moderate exudate and infectious materials from chronic, acute, traumatic, subacute, dehisced wounds, partial-thickness burns, ulcers (diabetic/pressure), flaps, grafts, and closed surgical incisions. Suitable for hospital and homecare use.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
Predicate Devices
PICO Single Use Negative Pressure Wound Therapy System (K163387)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
February 2, 2018
Smith & Nephew Medical Limited Susan McLoughlin Regulatory Affairs Director, Advanced Wound Management 101 Hessle Road Hull, hu3 2bn GB
Re: K172005
Trade/Device Name: PICO 7 Single Use Negative Pressure Wound Therapy System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: OMP Dated: December 21, 2017 Received: December 26, 2017
### Dear Susan McLoughlin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
#### 510(k) Number (if known) K172005
#### Device Name
PICO 7 Single Use Negative Pressure Wound Therapy System
Indications for Use (Describe)
PICO 7 is indicated for patients who would benefit from a suction device (Negative Pressure Wound Therapy) as it may promote would healing via removal of low to moderate levels of exudate and infectious materials. Appropriate wound types include:
- Chronic
- Acute
- Traumatic
- Subacute and dehisced wounds
- Partial-thickness burns
- Ulcers (such as diabetic or pressure)
- Flaps and grafts
- Closed Surgical incisions
PICO 7 Single Use Negative Pressure Wound Therapy System is suitable for use both in a hospital and homecare setting.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY
## PICO 7 Single Use Negative Pressure Wound Therapy System
| General Information | |
|------------------------------------|--------------------------------------|
| Submitter Name/Address: | Smith & Nephew Medical Limited |
| | 101 Hessle Road, |
| | Hull |
| | HU3 2BN |
| | United Kingdom |
| Establishment Registration Number: | 8043484 |
| Contact Person: | Susan McLoughlin, Regulatory Affairs |
| | Director |
| Phone Number | +44 1482 673637 |
| Date Prepared: | 1 February 2018 |
| Device Description | |
| Trade Name: | PICO 7 Single Use Negative Pressure |
| | Wound Therapy System |
| Common or Usual Name: | Negative Pressure Wound Therapy |
| | powered suction pump |
| Classification Name: | Powered suction pump (21 CFR |
| | 878.4780) |
| Regulatory Class: | Class II |
| Product Code: | OMP |
| Predicate Device Information | |
| 510(k) Number: | K163387 |
| Device: | PICO Single Use Negative Pressure |
| | Wound Therapy System |
| Clearance Date: | April 18, 2017 |
{4}------------------------------------------------
## Device Description
PICO 7 Single Use Negative Pressure Wound Therapy System is a small, lightweight, portable, magnet containing electro-mechanical pump system connected through a flexible tube to a super-absorbent, gentle adhesive dressing. Secondary fixation strips are also provided to ensure an adequate seal is achieved. The pump, the dressing and fixation strips are supplied sterile and for single use. PICO 7 was designed to deliver negative pressure wound therapy to low and to moderate exudate levels of acute and chronic wound types. PICO 7 Single Use Negative Pressure Wound Therapy System is suitable for use in both a hospital and homecare setting. A carry bag is not provided with the system however will be available to order separately.
## Indications for Use
PICO 7 is indicated for patients who would benefit from a suction device (Negative Pressure Wound Therapy) as it may promote wound healing via removal of low to moderate levels of exudate and infectious materials.
Appropriate wound types include:
- . Chronic
- Acute .
- Traumatic ●
- Subacute and dehisced wounds
- Partial-thickness burns, Ulcers (such as diabetic or pressure)
- Flaps and grafts .
- . Closed surgical incisions
PICO Single Use Negative Pressure Wound Therapy System is suitable for use in both a hospital and homecare setting.
{5}------------------------------------------------
## Comparison between New and Predicate Devices
The Indications for Use statement of the PICO 7 Single Use Negative Pressure Wound Therapy System is identical to the predicate device. The technological principal for delivering the negative pressure wound therapy for both the subject and predicate devices are identical. The main differences between the subject device and the predicate device are:
- (a) the introduction of a new pump to deliver negative pressure,
- (b) introduction of a magnet to the device,
- (c) the introduction of dual dressing kits and single dressings kit labelled as PICO 7 Single Use NPWT System,
- (d) the introduction (addition to existing range) of PICO Fluid Management Pack with 5 individually packaged dressings for use with PICO Single Use Negative Pressure Wound Therapy systems,
- (e) the extension of wear time of the dressings when used on moderately exuding wounds.
## Non-Clinical Tests (Bench)
The following bench tests have successfully been completed:
- Wound Model Testing of PICO 7 Kits 10x20cm, 25x25cm, small multisite . 15x20cm, large multisite 20x25cm dressings at low and moderate flow rate
The results of the tests showed that all of the dressings managed fluid at a flow rate modelling a low (0.6 g/cm²/24 hours) and a moderately exuding wound (1.1 g/cm²/24 hours) respectively for a simulated wound of 25% of the dressing absorbent pad area. The results also showed that the dressings maintained negative pressure within the test requirements of -60 mmHg to -100 mmHg for a minimum of 95% of the test time at the simulated wound base throughout the test. Based on the results of the tests it can be concluded that the PICO 7 systems can be used for their intended purposes.
- Wound Model Testing of PICO 7 kits 10x20cm and 25x25cm dressings at low and moderate flow rate (vertical orientation)
{6}------------------------------------------------
The results of the tests showed that all of the dressings managed fluid at a flow rate modelling a low (0.6 q/cm²/24 hours) and a moderately exuding wound (1.1 g/cm²/24 hours) respectively for a simulated wound of 25% of the dressing area. The results also showed that the dressings maintained negative pressure within the test requirements of -60 mmHq to -100 mmHq for a minimum of 95% of the test time at the simulated wound base throughout the test. Based on the results of the tests it can be concluded that the PICO 7 systems can be used for their intended purposes.
## Human Factors
PICO 7 has been evaluated according to FDA guidance 'Applying Human Factors and Usability Engineering to Medical devices, Feb 3rd 2016' and IEC62366-1:2015. Critical tasks were identified following a risk management process and a series of preliminary testing carried out with the two identified user groups. Healthcare Professionals and Lay Users, in order to optimize the design. Following this a summative study was carried out on the entire device followed by an additional study to assess labelling improvements. The studies concluded that the device labelling is safe and effective for its intended users in the intended environments of use.
## Biocompatibility and Sterilization Testing
PICO 7 Single Use Negative Pressure Wound Therapy System has been evaluated according to the Biological Evaluation of Medical Devices Standard BS EN ISO 10993, with particular reference to Part 1 (2009): Evaluation and testing within a risk management process and FDA guidance 'Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". PICO 7 Single Use NPWT Use System uses the same dressings with soft port as PICO Single Use NPWT System. The biocompatibility of the dressings with soft port used in PICO Single Use NPWT System has been addressed and cleared in 510(k) K166387,
{7}------------------------------------------------
April 18, 2017. The exterior pump components have been assessed through chemical characterisation testing, review of compliance to recognised standards, ethylene oxide and ethylene chlorohydrin residuals testing and supplier biocompatibility test statements.
## Electrical Safety and Electromagnetic Compatibility (EMC)
PICO 7 Single Use Negative Pressure Wound Therapy System has been tested and assessed respectively in accordance with the following electrical safety standards and has been found to be in compliance with the following standards IEC60601-1, IEC60601-1-2, IEC60601-1-6 and IEC60601-1-11.
## Conclusions
In establishing substantial equivalence to the predicate devices, Smith & Nephew Medical Ltd evaluated the indications for use, materials, technology, product specifications and energy requirements of the device. Performance testing, service life testing and electrical safety testing has been completed to demonstrate that the PICO 7 Single Use Negative Pressure Wound Therapy System is substantially equivalent to the predicate for the intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.