K172004 · Cutera, Inc. · PBX · Aug 2, 2017 · General, Plastic Surgery
Device Facts
Record ID
K172004
Device Name
truSculpt
Applicant
Cutera, Inc.
Product Code
PBX · General, Plastic Surgery
Decision Date
Aug 2, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The truSculpt is intended to generate heat within body tissues for the treatment of selected medical conditions, such as the relief of minor aches, pain, and muscle spasms; an increase in local circulation; and temporary reduction in circumference of the abdomen. It is also intended to provide temporary reduction in the appearance of cellulite.
Device Story
Device consists of console, RF handpieces (16 cm2 and 40 cm2), patient belts, comfort switch, and massage roller. Console delivers RF energy to handpieces to heat subcutaneous tissue while monitoring epidermal temperature. Multiplexor directs RF output to up to six handpieces in sequence for hands-free abdominal/flank treatment. Patient-activated switch allows termination of treatment if discomfort occurs. Used in clinical settings by practitioners. Benefits include non-invasive circumferential reduction and pain relief via controlled thermal energy.
Clinical Evidence
Bench testing only. Performance verified through software V&V and compliance with IEC 60601-1, IEC 60601-1-6, IEC 60601-2-2, and IEC 60601-1-2. No new clinical data provided; relies on established safety/effectiveness of predicate devices.
Technological Characteristics
RF-based heating system; console-controlled; multiplexor board for sequential multi-handpiece (up to 6) delivery; 16 cm2 and 40 cm2 handpieces; patient comfort switch; adjustable belts for hands-free operation. Complies with AAMI/ANSI ES60601-1, IEC 60601-1, 60601-1-2, 60601-1-6, and 60601-2-2.
Indications for Use
Indicated for topical heating to elevate tissue temperature for relief of pain, muscle spasms, and increased local circulation. The 2 MHz setting for the 40 cm2 handpiece is indicated for temporary reduction in abdominal circumference. The massage device is indicated for temporary reduction in the appearance of cellulite.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized symbol that resembles three human profiles facing to the right, with flowing lines above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 2, 2017
Cutera, Inc. Mr. Bradley Renton Vice President, Regulatory and Medical Affairs & Compliance Officer 3240 Bayshore Blvd. Brisbane, California 94005
Re: K172004
Trade/Device Name: truSculpt Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: PBX, GEI Dated: June 30, 2017 Received: July 3, 2017
Dear Mr. Renton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
-53
## Jennifer R. Stevenson
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K172004
Device Name truSculpt
Indications for Use (Describe)
The truSculpt RF energy is intended to provide topical heating for the purpose of elevating tor the treatment of selected medical conditions, such as relief of pain and mcrease in local circulation.
Additionally, the 2 MHz setting for the 40 cm2 handpiece is indicated for temporary reduction in circumference of the abdomen.
The truSculpt massage device is intended to provide a temporary reduction in the appearance of cellulite.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K172004
#### Section 5 510(k) Summary
This 510(k) Summary of safety and effectiveness for the truSculpt RF device is submitted in accordance with the requirements of the SMDA 1990 and following guidance concerning the organization and content of a 510(k) summary.
| Applicant: | Cutera, Inc. |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 3240 Bayshore Blvd.<br>Brisbane, CA 94005 |
| Contact Person: | Bradley Renton |
| Telephone:<br>Fax:<br>Email: | 415-657-5568 - phone<br>415-715-3568 - fax<br>brenton@cutera.com |
| Preparation Date: | August 1, 2017 |
| Device Trade Name: | truSculpt |
| Common Name: | Massager, Vacuum, Radio Frequency Induced Heat |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories, PBX,<br>GEI, 21 CFR 878.4400 |
| Legally Marketed<br>Predicate Devices: | Cutera truSculpt RF Device (K162512)<br>Syneron SlimShape System (K163415) |
| Legally Marketed<br>Reference Devices: | Cynosure SculpSure (K160470) |
| Device Description: | The modified truSculpt RF device consists of a console; one 16 $cm^2$ RF<br>handpiece; up to six 40 $cm^2$ puck-style RF handpieces that can attach to<br>belts configured for hands-free abdominal and flank treatments for<br>circumferential reduction; adjustable patient belts; a patient comfort<br>switch; and a truGlide massage roller. All system functions are<br>controlled through the console. The handpieces deliver RF energy to<br>generate a heating profile that produces a moderate temperature rise in<br>the subcutaneous tissue, while monitoring epidermal temperature. In<br>addition, there is a separate mechanical roller that can be used as a<br>massager. |
| Intended Use: | The truSculpt is intended to generate heat within body tissues for the<br>treatment of selected medical conditions, such as the relief of minor<br>aches, pain, and muscle spasms; an increase in local circulation; and<br>temporary reduction in circumference of the abdomen. It is also intended<br>to provide temporary reduction in the appearance of cellulite. |
| Indications for Use: | The truSculpt RF energy is intended to provide topical heating for the<br>purpose of elevating tissue temperature for the treatment of selected<br>medical conditions, such as relief of pain and muscle spasms and<br>increase in local circulation. |
{4}------------------------------------------------
#### K172004
| Attachment 5 |
|----------------|
| 510(K) Summary |
Additionally, the 2 MHz setting for the 40 cm² handpiece is indicated for temporary reduction in circumference of the abdomen.
The truSculpt massage device is intended to provide a temporary reduction in the appearance of cellulite.
In the modified device, the RF output is directed through a multiplexor to Device Modifications: power up to six treatment handpieces in sequence. The modification also adds a patient-activated treatment termination switch; belts configured to enable up to six 40 cm² treatment handpieces to be positioned on the patient's abdomen and flanks prior to treatment for circumferential reduction; and a minor form factor change to the 40 cm² handpiece to enable it to attach to the patient belts. There are no changes to the indications for use.
> When performing abdominal circumferential reduction treatments using the belts to position the form-factor modified 40 cm² handpieces, the identical treatment parameters are available for selection as used in the clinical trial to gain this indication for use, including area treated and RF dose (time and temperature).
Cutera has determined that no new risks have been introduced due to the above changes, as activating multiple handpieces in sequence is equivalent to the operator sequentially moving a single handpiece to a new location and then beginning a new treatment application; and patient feedback was used in the unmodified truSculpt to determine tolerable settings.
#### Medical Electrical Equipment - Part 1: General Performance Data: IEC 60601-1 Requirements for Safety (Edition 3.1, 2012), including:
- . IEC 60601-1-6 Medical Electrical Equipment - Part 1-6: General Requirements for Safety - Collateral Standard: Usability (Edition 3.1, 2013) - Test Report Attachment 4
- IEC 60601-2-2 Medical Electrical Equipment Part 2-2: . Particular Requirements for the Basic Safety and Essential Performance of High Frequency Surgical Equipment and High Frequency Surgical Accessories – Test Report Attachment 5
The product also fulfills the requirements of AAMI/ANSI ES60601-1:2005+A2 (R2012) + A1.
IEC 60601-1-2 Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Disturbances (Edition 4.0, 2014)
A patient comfort switch was added to mitigate risks during hands-free treatment with the modified truSculpt device. The function of the comfort switch is defined in the Software Requirements Specification (SRS) for the modified device. The performance of the patient comfort switch within the modified truSculpt device was verified and validated during software V&V testing. All tests were passed. There were no exceptions.
{5}------------------------------------------------
#### K172004
#
| Summary of<br>Technological<br>Characteristics: | A multiplexor board was designed and added to the truSculpt console to direct RF output to any of 6 output ports to which RF handpieces can be attached. The function of the multiplexor board is defined in the Software Requirements Specification (SRS) for the modified device. The software verifies the state of each switch before enabling RF delivery. The performance of the multiplexor board within the modified truSculpt device was verified and validated during software V&V testing. All tests were passed. There were no exceptions.<br><br>Software verification and validation was performed, and it was demonstrated that the software for the modified truSculpt device performs as intended (V0174 - Software Verification and Validation Report).<br><br>The modified truSculpt RF device has the same intended use and indications for use and the same fundamental scientific technology as the previously cleared unmodified Cutera truSculpt RF device (K162512). It also has the same fundamental scientific technology as the Syneron SlimShape System (K163415) predicate device. The adjustable belt system and patient comfort switch that enable hands-free operation are very similar to the belts and comfort switches employed by the Syneron SlimShape System predicate device and the Cynosure SculpSure (K160470) reference device.<br><br>For each of the devices:<br>the user interface is located at the front/top of the console; the treatment applicators are positioned on the patient's body prior to commencing treatment and held in place by adjustable belts; the devices moderately heat the superficial tissue of the patient to achieve their intended outcome; and the patient can terminate the treatment if the discomfort becomes excessive. The modified truSculpt RF device employs the same fundamental scientific technology, similar key design aspects, and the same or similar treatment areas as its predicate and reference devices. Therefore, the minor differences do not raise any new safety or effectiveness questions. |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion: | Cutera believes that the requested changes are substantially equivalent to the predicate device and do not raise any new issues of safety or effectiveness. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.