K171986 · CAO Group, Inc. · GEX · Sep 22, 2017 · General, Plastic Surgery
Device Facts
Record ID
K171986
Device Name
Sterling Supreme Diode Laser
Applicant
CAO Group, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 22, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sterling Supreme Diode Laser is indicated for dentistry and oral soft tissue procedures of: 1) The removal of lesions, excision, incision, vaporization, hemostasis, and photocoagulation on soft tissues including abscess treatment, contouring, curettage, sulcular debridement, pulpotomy, gingivectomy, troughing, and removal of inflamed edematous tissue; 2) Temporary relief of minor muscle and joint pain, stiffness, minor arthritis pain, temporary increase in local circulation, and temporary relaxation of muscles by means of topical elevated tissue temperature from infrared spectral emissions; 3) Light activation of bleaching materials for teeth whitening and laser-assisted whitening/bleaching of teeth.
Device Story
Sterling Supreme Diode Laser delivers 810nm infrared laser energy to soft tissue for dental/oral procedures. Solid-state AlGaAs diodes generate energy; optical fiber transmits energy through a handpiece assembly to a single-use disposable tip. Target tissue absorbs energy, converting it to heat for hemostasis, ablation, or vaporization. Device features color LCD touchscreen, 650nm aiming beam, and wireless foot-actuated switch (2.40-2.48 GHz). Used in dental clinics by practitioners. Factory-established presets available for common procedures; user-definable settings allow saving preferences. Benefits include precise tissue management and therapeutic heating for pain relief.
Clinical Evidence
Bench testing only. Evaluation included final assembly inspection (QAC-OP0298) and main/safety specification verification testing. Results demonstrate the device meets all labeled performance specifications.
Indicated for dental/oral soft tissue procedures (lesion removal, excision, incision, vaporization, hemostasis, photocoagulation, abscess treatment, contouring, curettage, sulcular debridement, pulpotomy, gingivectomy, troughing, removal of inflamed edematous tissue); temporary relief of minor muscle/joint pain, stiffness, arthritis pain, muscle spasm, and increased local circulation; and light activation for teeth whitening/bleaching.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 22, 2017
CAO Group, Inc. Robert Larsen Regulatory Affairs Manager 4628 West Skyhawk Drive West Jordan, Utah 84084
Re: K171986
Trade/Device Name: Sterling Supreme Diode Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology Regulatory Class: Class II Product Code: GEX, ILY Dated: June 26, 2017 Received: July 3, 2017
Dear Mr. Larsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K171986
Device Name Sterling Supreme Diode Laser
Indications for Use (Describe)
The Sterling Supreme Diode Laser is indicated for dentistry and oral soft tissue procedures of:
1) The removal of lesions, excision, incision, vaporization, hemostasis, and photocoagulation on soft tissues including abscess treatment, contouring, curettage, sulcular debridement, pulpotomy, gingivectomy, troughing, and removal of inflamed edematous tissue;
2) Temporary relief of minor muscle and joint pain, stiffness, minor arthritis pain, temporary increase in local circulation, and temporary relaxation of muscles by means of topical elevated tissue temperature from infrared spectral emissions;
3) Light activation of bleaching materials for teeth whitening and laser-assisted whitening/bleaching of teeth.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for CAO Group, Inc. The logo consists of a red square with a white Chinese character inside, followed by the text "CAO GROUP, INC." in black. Below the company name, there is a thin red line, and below that, the text "Easier · Faster · Better™" is displayed in a smaller font size.
West Inrdrin III 8408
301.256.9282 [phone]
877.877.9778 [toll free]
301.256.9287 [fax]
## 510(k) Summary
Prepared By: Robert K. Larsen Preparation Date: September 21, 2017
### Contact Information:
| Contact Name: | Robert K. Larsen |
|------------------|--------------------------------------------------|
| Contact Title: | Regulatory Affairs Manager |
| Address: | 4628 West Skyhawk Drive<br>West Jordan, UT 84084 |
| Contact Phone #: | 801-256-9282 |
| Contact Email: | rob.larsen@caogroup.com |
#### Device Name:
| Trade Name: | Sterling Supreme Diode Laser |
|---------------------------|-------------------------------------------------|
| Common Name: | Soft Tissue Diode Laser |
| Product Code 1: | GEX |
| Regulation 1: | 878.4810 |
| Product Classification 1: | Powered Laser Surgical Instrument<br>Class II |
| Product Code 2: | ILY |
| Regulation 2: | 890.5500 |
| Product Classification 2: | Lamp, Infrared, Therapeutic Heating<br>Class II |
#### Legally Marketed Predicate Devices for Substantial Equivalence:
Precise SHP Diode Laser, manufactured by CAO Group, Inc. (K123443)
#### Rationale for Substantial Equivalence:
The submitted device and identified predicate device share exactly identical indications for use in dentistry and oral soft tissue procedures:
- 1) The removal of lesions, excision, vaporization, ablation, hemostasis, and photocoagulation on soft tissues including abscess treatment, contouring, curettage, sulcular debridement, pulpotomy, frenectomy, gingivectomy, troughing, and removal of inflamed edematous tissue;
- 2) Temporary relief of minor muscle and joint pain, stiffness, minor arthritis pain, muscle spasm, temporary increase in local circulation, and temporary relaxation of muscles by means of topical elevated tissue temperature from infrared spectral emissions;
- 3) Light activation of bleaching materials for teeth whitening and laser-assisted whitening/bleaching of teeth.
{4}------------------------------------------------
4628 W. Skyhawk Drive West Jordan, UT 84084 www.caogroup.com
The submitted device and predicate device shares including emission sources, operating controls, key constructional components, and materials of construction. The devices share similar methods of control systems, safety features, and performance monitoring. The devices share similar performance specifications including power output, emission wavelength, and energy type.
#### Description of Submitted Device:
The Sterling Supreme Diode Laser is a device for delivering laser energy to human soft tissue for a variety of dental and oral procedures and treatments. This energy is generated by solid-state diodes, which provide a consistent and reliable generation of laser energy at 810 ± 20nm for a maximum of 3 watts of energy output. The laser energy is delivered to the surgical site by means of an optical fiber fully ensheathed within a coated steel coil terminated in a handpiece assembly, which allows for the safe transmission of laser energy to the site without creating undue risk to the patient or operatory staff by errant or collateral laser emissions. The target tissues absorb the laser energy converting it to heat. Depending on the intensity or power output of the laser, the heat so generated can cause hemostasis, ablation, or vaporization. The device features some user definable settings, including a selectable 650nm aiming beam, adjustable power output for both the working beam and aiming beam, and continuous delivery or pulse delivery options. The device allows for selection of factoryestablished presets for common dental and oral care procedures, and allows for the operator to save desired settings for quick-reference.
Laser energy is transmitted through the fixed length of optical fiber to the distal end of the handpiece assembly. The handpiece assembly consists of a reusable, removable, sterilizable handpiece sleeve, and for surgical procedures is terminated with a single-use disposable tip. The operator uses the handpiece assembly to position and direct the laser energy to the intended treatment site. The activation of the working beam diodes is accomplished by use of a wireless foot-actuated switch.
### Description of Predicate Device:
The Precise SHP Diode Laser is a device for delivering laser energy to human soft tissue for a variety of dental and oral procedures and treatments. This energy is generated by solid-state diodes, which provide a consistent and reliable generation of laser energy at 810 ± 20nm for a maximum of 3 watts of energy output. The laser energy is delivered to the surgical site by means of an optical fiber fully ensheathed within a coated steel coil terminated in a handpiece assembly, which allows for the safe transmission of laser energy to the site without creating undue risk to the patient or operatory staff by errant or collateral laser emissions. The target tissues absorb the laser energy converting it to heat. Depending on the intensity or power output of the laser, the heat. so generated can cause hemostasis, ablation, or vaporization. The device features some user definable settings, including a selectable 650nm aiming beam, adjustable power output for both the working beam and aiming beam, and continuous delivery or pulse delivery options. The device allows for selection of factory-established presets for common dental and oral care procedures, and allows for the operator to save desired settings for quick-reference.
Laser energy is transmitted through the fixed length of optical fiber to the distal end of the handpiece assembly. The handpiece assembly consists of a reusable, removable, sterilizable handpiece sleeve, and for surgical procedures is terminated with a single-use disposable tip. The operator uses the handpiece assembly to position and direct the laser energy to the intended treatment site. The activation of the working beam diodes is accomplished by use of a wireless foot-actuated switch.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for CAO GROUP, INC. The logo consists of a red square with a white Chinese character inside, followed by the text "CAO GROUP, INC." in black. Below the company name, there is a tagline that reads "Easier · Faster · Better™".
| 4628 W. Skyhawk Drive<br>West Jordan, UT 84084 | 801.256.9282 [phone]<br>877.877.9778 [toll free]<br>801.256.9287 [fax] |
|------------------------------------------------|------------------------------------------------------------------------|
| www.caogroup.com | |
#### Indications for Use of the Submitted Device:
The Sterling Supreme Diode Laser is indicated for use for -
- 1) The removal of lesions, excision, vaporization, ablation, hemostasis, and photocoagulation on soft tissues including abscess treatment, contouring, curettage, sulcular debridement, pulpotomy, frenectomy, gingivectomy, troughing, and removal of inflamed edematous tissue;
- 2) Temporary relief of minor muscle and joint pain, stiffness, minor arthritis pain, muscle spasm, temporary increase in local circulation, and temporary relaxation of muscles by means of topical elevated tissue temperature from infrared spectral emissions;
- 3) Light activation of bleaching materials for teeth whitening and laser-assisted whitening/bleaching of teeth.
| Technological Characteristics and Substantial Equivalence: | | |
|------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| | CAO Group, Inc. Sterling Supreme | CAO Group, Inc. Precise SHP |
| Working Beam<br>Output<br>Wavelength | 810±10 nm | 810±10 nm |
| Working Beam<br>Output Power | 0.1 - 3.0 watts | 0.5 - 3.0 watts |
| Aiming Beam<br>Output Power | < 3mW | < 3mW |
| Aiming Beam<br>Output<br>Wavelength | 650±20 nm | 650±20 nm |
| Laser Source | AlGaAs Diode | AlGaAs Diode |
| Laser<br>Activation | Wireless foot pedal<br>2.40 - 2.48 GHz, IEEE 802.15.4 Protocol | Wireless foot pedal<br>2.40 - 2.48 GHz, IEEE 802.15.4 Protocol |
| Pulse Control | Digital emission control<br>Fixed pulse duration of 0.05 seconds, 10Hz | Digital emission control<br>Selectable fixed pulse duration of 0.03, 0.05,<br>or 0.075 seconds, all at 10Hz |
| Cooling<br>Method | Heatsink / Fan air cooled | Heatsink / Fan air cooled |
| Electrical<br>Power Input<br>(System) | 100-240 VAC @ 47-63 Hz, 1.6A (switchable) | 100-240 VAC @ 47-63 Hz, 1.6A (switchable) |
| User Interface | Color LCD Touch Screen | Color LCD Touch Screen |
| Laser Delivery<br>System | Sheathed quartz optical fiber | Sheathed quartz optical fiber |
| Handpiece<br>design | Inner stem composed of stainless steel<br>permanently attached to the fiber cable;<br>Removable, sterilizable anodized aluminum<br>outer sleeve | Inner stem composed of stainless steel<br>permanently attached to the fiber cable;<br>Removable, sterilizable anodized aluminum<br>outer sleeve |
| Patient Contact<br>Articles | Single use disposable tip composed of silica<br>quartz, stainless steel, and ABS plastic | Single use disposable tip composed of silica<br>quartz, stainless steel, and ABS plastic |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for CAO Group, Inc. The logo consists of a red square with a white Chinese character inside, followed by the text "CAO GROUP, INC." in black. Below the company name is the tagline "Easier · Faster · Better™" in a smaller font size.
| 4628 W. Skyhawk Drive | 801.256.9282 [phone] |
|-----------------------|--------------------------|
| West Jordan, UT 84084 | 877.877.9778 [toll free] |
| www.caogroup.com | 801.256.9287 [fax] |
| | CAO Group, Inc. Sterling Supreme | CAO Group, Inc. Precise SHP |
|--------------------------------------------|-------------------------------------------------|-------------------------------------------------|
| Energy emissions for select device presets | | |
| Curettage | 1.4 watts, Pulse, 30-second timer | 1.4 watts, 0.05 Pulse |
| Gingivectomy | 1.0 watts, continuous emission | 1.0 watts, continuous emission |
| Phototherapy - | 321 J/cm² | 321 J/cm² |
| Pain Relief | (2.0 watts, continuous emission, therapy tip, 5 | (2.0 watts, continuous emission, therapy tip, 5 |
| | minute exposure) | minute exposure) |
| Whitening | 900 1 | 900 1 |
| | (3 watts, Continuous, 5 minute total exposure) | (3 watts, Continuous, 5 minute total exposure) |
| Dimensions | 7.5″ x 5.5″ x 4″ | 5.5″ x 4.75″ x 2.375″ |
| Weight | 3.5 lbs | 4.0 lbs |
| 510(k) Number | Pending this application | K123443 |
| Indications for | The Sterling Supreme Diode Laser is indicated | The Pioneer Elite Diode Laser is indicated for |
| Use | for dentistry and oral soft tissue procedures | dentistry and oral soft tissue procedures of: |
| | of: | |
| | 1) The removal of lesions, excision, incision, | 1) The removal of lesions, excision, incision, |
| | vaporization, ablation, hemostasis, and | vaporization, ablation, hemostasis, and |
| | photocoagulation on soft tissues including | photocoagulation on soft tissues including |
| | abscess treatment, contouring, curettage, | abscess treatment, contouring, curettage, |
| | sulcular debridement, pulpotomy, | sulcular debridement, pulpotomy, |
| | frenectomy, gingivectomy, troughing, and | frenectomy, gingivectomy, troughing, and |
| | removal of inflamed edematous tissue; | removal of inflamed edematous tissue; |
| | 2) Temporary relief of minor muscle and joint | 2) Temporary relief of minor muscle and joint |
| | pain, stiffness, minor arthritis pain, muscle | pain, stiffness, minor arthritis pain, muscle |
| | spasm, temporary increase in local | spasm, temporary increase in local |
| | circulation, and temporary relaxation of | circulation, and temporary relaxation of |
| | muscles by means of topical elevated tissue | muscles by means of topical elevated tissue |
| | temperature from infrared spectral | temperature from infrared spectral |
| | emissions; | emissions; |
| | 3) Light activation of bleaching materials for | 3) Light activation of bleaching materials for |
| | teeth whitening and laser-assisted | teeth whitening and laser-assisted |
| | whitening/bleaching of teeth. | whitening/bleaching of teeth. |
To summarize the technological characteristics, the submitted device and the predicate device share the same technology of generating 810nm infrared laser energy via solid state diodes. The submitted and predicate devices share the same methods of delivering the laser energy through a silica quartz fiber to an affixed handpiece assembly. The devices share the same method of using a single-use disposable tip composed of the same materials for delivering the laser energy to the treatment site. The devices share functionality of a wireless footswitch to activate the 810nm laser energy, such wireless footswitches using the same wireless frequency and transmission technology. The devices share the same method of digital pulse control, with similar manner of fixed pulse width. The devices share the same indications for use. The devices share similar usage of color touchscreen technology as the user interface.
The devices differ in the physical shape of the devices differ in that the predicate uses an off-theshelf component as the touchscreen interface, while the submitted device uses an embedded color touchscreen interface.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for CAO Group, Inc. The logo consists of a red square with a white Chinese character inside, followed by the text "CAO GROUP, INC." in black. Below the company name, the text "Easier · Faster · Better" is written in a smaller font size.
4628 W. Skyhawk Drive
West Jordan, UT 84084
**www.caogroup.com**
801.256.9282 [phone]
877.877.9778 [toll free]
801.256.9287 [fax]
#### Conformity to Standar
The Sterling Supreme Diode Laser is designed to comply with the performance requirements of ANSI/AAMI ES60601-1:2005+A1 (Edition 3.0), AAMI/ANSI/ IEC 60601-1-2:2007 (Edition 3.0), IEC 60601-1-6:2013 (Edition 3.1), IEC 60601-2-22:2007 (Edition 3.1), and IEC 60825-1:2007 (Edition 2.0).
#### Performance Data:
Bench testing on an evaluation sample of the submitted device was performed consistent with internal requirements:
- . QAC-OP0298 - Final assembly inspection of Sterling Supreme Diode Laser
- Sterling Supreme Main and Safety Specification Verification Test ●
Results of these tests demonstrate the product meets the device's labeled specifications.
#### Conclusion:
The Sterling Supreme Diode Laser is substantially equivalent to the listed predicate. This device shares identical intended uses, identical operating principles, similar design features, and identical functional and performance characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.