Combo Stimulator MT9000 For TENS/IF/MIC mode 1. Symptomatic relief of chronic intractable pain 2. Post traumatic pain 3. Post surgical pain For EMS mode 1. Relaxation of muscle spasm. 2. Increase of local blood flow circulation 3. Prevention or retardation of disuse atrophy 4. Muscle re-education 5. Maintaining or increasing range of motion. 6. Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis Combo Stimulator LT7102 For TENS/IF/MIC mode 1. Symptomatic relief of chronic intractable pain 2. Post traumatic pain 3. Post surgical pain For EMS/RUSS mode 1. Relaxation of muscle spasm. 2. Increase of local blood flow circulation 3. Prevention or retardation of disuse atrophy 4. Muscle re-education 5. Maintaining or increasing range of motion. 6. Immediate post-surgical stimulation of muscles to prevent venous thrombosis TENS Stimulator Intensity 10 1. Symptomatic relief of chronic intractable pain 2. Post traumatic pain 3. Post surgical pain
Device Story
Device delivers electrical current to nerves/muscles via skin-applied electrodes; used for pain relief and muscle stimulation. User operates device via keypad and LCD display to adjust intensity levels. Powered by battery or adapter. Models include TENS, EMS, Interferential (IF), Microcurrent (MIC), and Russian (RUSS) stimulation modes. Healthcare providers or patients use device to manage pain or rehabilitate muscles. Output affects clinical decision-making by providing non-pharmacological pain management and physical therapy support. Benefits include pain reduction, improved circulation, and muscle function maintenance.
Clinical Evidence
Bench testing only. Includes electrical safety (IEC 60601-1, IEC 60601-2-10), electromagnetic compatibility (IEC 60601-1-2), and software verification/validation (IEC 62304). Biocompatibility of electrodes evaluated per ISO 10993.
Indicated for symptomatic relief of chronic intractable, post-traumatic, or post-surgical pain. EMS/RUSS modes indicated for muscle spasm relaxation, local blood flow increase, disuse atrophy prevention/retardation, muscle re-education, range of motion maintenance/increase, and post-surgical stimulation to prevent venous thrombosis.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 2, 2017
Shenzhen Dongdixin Technology Co., Ltd. Siping Yuan R.A. Specialist No. 3 Building Xilibaimang Xusheng Industrial Estate Nanshan Shenzhen, 518108 CN
Re: K171978
Trade/Device Name: Combo Stimulator MT9000, Combo Stimulator LT7102, TENS Stimulator InTENSity 10 Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF, GZJ, LIH Dated: June 15, 2017 Received: July 5, 2017
Dear Siping Yuan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Vivek J. Pinto -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
### 510(k) Number (if known) K171978
### Device Name
Combo Stimulator MT9000, Combo Stimulator LT7102, TENS Stimulator InTENSity 10
Indications for Use (Describe)
Combo Stimulator MT9000
For TENS/IF/MIC mode
- 1. Symptomatic relief of chronic intractable pain
- 2. Post traumatic pain
- 3. Post surgical pain
- For EMS mode
- Relaxation of muscle spasm. 1.
- 2. Increase of local blood flow circulation
- 3. Prevention or retardation of disuse atrophy
- 4. Muscle re-education
- న. Maintaining or increasing range of motion.
- Immediate post-surgical stimulation of muscles to prevent venous thrombosis 6.
## Combo Stimulator LT7102
For TENS/IF/MIC mode
- Symptomatic relief of chronic intractable pain 1.
- 2. Post traumatic pain
- 3. Post surgical pain
For EMS/RUSS mode
- Relaxation of muscle spasm. 1.
- 2. Increase of local blood flow circulation
- 3. Prevention or retardation of disuse atrophy
- Muscle re-education 4.
- న. Maintaining or increasing range of motion.
- 6. Immediate post-surgical stimulation of muscles to prevent venous thrombosis
## TENS Stimulator InTENSity 10
- 1. Symptomatic relief of chronic intractable pain
- 2. Post traumatic pain
- 3. Post surgical pain
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for DONGDIXIN TECHNOLOGY CO., LTD. The logo is in blue, except for the X in DIXIN, which is in yellow. There are also three Chinese characters to the right of the English text.
001_510 (k) Summary Version: 1.1
# 510(k) SUMMARY
as required by section 21 CFR 807.92
| Date of Submission: | 06/15/2017 |
|---------------------|------------------------------------------------------------------------------------------|
| Submitter's Name: | Shenzhen Dongdixin Technology Co., Ltd |
| Address: | No. 3 Building, Xilibaimang Xusheng Industrial<br>Estate Nanshan, Shenzhen, CHINA 518108 |
| E-mail: | yuansp@dundex.com |
| Tel: | +86(755) 27652471 |
| FAX: | +86(755) 27652674 |
| Contact: | Siping Yuan |
{5}------------------------------------------------
## DO TECHNOLOGY CO., L
001_510 (k) Summary Version: 1.1
- 1. Proposed Device:
1)
2)
3)
| Trade Name: | Combo Stimulator MT9000<br>Combo Stimulator LT7102<br>TENS Stimulator Intensity 10 |
|----------------------|------------------------------------------------------------------------------------|
| Classification Name: | Stimulator, Muscle, Powered |
| Regulation Number: | 890.5850 |
| Product Code: | IPF |
| Device Class: | II |
| Trade Name: | Combo Stimulator MT9000<br>Combo Stimulator LT7102 |
| Classification Name: | Stimulator, Nerve, Transcutaneous. For pain relie- |
| Regulation Number: | 882.5890 |
| Product Code: | GZJ |
| Device Class: | II |
| Trade Name: | Combo Stimulator MT9000<br>Combo Stimulator LT7102 |
|----------------------|----------------------------------------------------|
| Classification Name: | Interferential current therapy |
| Regulation Number: | Unclassified |
| Product Code: | LIH |
| Device Class: | II |
#### 2. Predicate Device:
| Legally Marketed Device: | MT9000 Series Electro-Stimulator |
|--------------------------|----------------------------------------|
| 510(k) Number: | K093138 |
| Manufacturer: | Shenzhen Dondixin Technology Co., Ltd. |
| Legally Marketed Device: | Sonicator® Plus 940, ME940 |
| 510(k) Number: | K071137 |
| Manufacturer: | Mettler Electronics Corp. |
#### 3. Description of Proposed Device:
MT9000, LT7102 and Intensity 10 are Transcutaneous Electrical Nerve Stimulator and muscle stimulator for pain relief and/or Electrical Muscle Stimulator sends gentle electrical current to underlying nerves and muscle group via electrode applied on the skin. The parameters of units are controlled by the press buttons. Its intensity level is adjustable according
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for DONGDIXIN TECHNOLOGY CO., LTD. The logo is in blue, except for a diagonal line in the middle of the word "DONGDIXIN", which is in orange. To the right of the company name are three Chinese characters in blue. The words "TECHNOLOGY CO., LTD." are in a smaller font size and are located below the company name.
to the needs of patients.
The 3 models MT9000, LT7102 and Intensity 10 have similar housing with a viewable LCD display, an accessible keypad, and accessible battery storage compartment. The LCD is located on the upper half of the rectangular face of the device, above the keypad. The LCD is used to display system information to the user.
#### 4. Description of All Device Modification(s)
The Combo Stimulator MT9000 only changes the energy type and added the adapter power supply. Compared with the existing device (MT9000), they have the same control mechanism, IFU, operating principle, performance specification, dimensional specifications, materials and software. The new change part have been evaluated and passed the requirement based on IEC60601-1-2 and IEC60601-2-10.
The Combo Stimulator LT7102 change the 9-volt battery to the recharge Lithium battery, add the electrode accessories like as different part electrode belt and make it more convenient for the user. Meanwhile, we added the RUSSIAN electrical stimulation mode that together with the TENS, IF and MICROCURRENT. Compared with the existing device (MT9000), they have the same control mechanism, IFU, operating principle, materials and performance specification. And this modification does not affect the intended use of the device or alter the fundamental scientific technology of the device. The new change part have been evaluated and passed the requirement based on IEC60601-1, IEC60601-1-2 and IEC60601-2-10.
Compared with the existing device (MT9001), The TENS Stimulator Intensity 10 only has TENS mode and the following parts have changed:
- Changes the energy type and added the adapter power supply. 1)
- 2) Change the dimensional specification.
- Change the LCD UI and make it more convenient for the user. 3)
And these changes have been evaluated and passed the requirement based on IEC60601-1, IEC60601-1-2 and IEC60601-2-10.
#### 5. Proposed Device Intended for Use Statement:
Device Name: Combo Stimulator MT9000, Combo Stimulator LT7102, TENS Stimulator Intensity 10
## Indications for Use:
## Combo Stimulator MT9000
For TENS/IF/MIC mode
- 1. Symptomatic relief of chronic intractable pain
- 2. Post traumatic pain
- 3. Post surgical pain
For EMS mode
{7}------------------------------------------------
## DON TECHNOLOGY CO., L
001 510 (k) Summary Version: 1.1
- 1. Relaxation of muscle spasm.
- 2. Increase of local blood flow circulation
- 3. Prevention or retardation of disuse atrophy
- 4. Muscle re-education
- 5. Maintaining or increasing range of motion.
- 6. Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
## Combo Stimulator LT7102
For TENS/IF/MIC mode
- 1. Symptomatic relief of chronic intractable pain
- 2. Post traumatic pain
- 3. Post surgical pain
- For EMS/RUSS mode
- 1. Relaxation of muscle spasm.
- 2. Increase of local blood flow circulation
- 3. Prevention or retardation of disuse atrophy
- 4. Muscle re-education
- 5. Maintaining or increasing range of motion.
- 6. Immediate post-surgical stimulation of muscles to prevent venous thrombosis
## TENS Stimulator Intensity 10
- 1. Symptomatic relief of chronic intractable pain
- 2. Post traumatic pain
- 3. Post surgical pain
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for DONGDIXIN TECHNOLOGY CO., LTD. The logo is in blue and orange. The text "DONGDIXIN" is in large, bold, blue letters, with an orange "X". Below that, the text "TECHNOLOGY CO., LTD." is in smaller, blue letters.
## 6. Technological Characteristics and Substantial Equivalence
| | | New device | New device | New device | Predicate device | Predicate device | Discussion |
|----|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| 1 | 510K# | To be assigned | To be assigned | To be assigned | K093138 | K071137 | N/A |
| 2 | Device Name | Combo Stimulator<br>LT7102 | Combo<br>Stimulator<br>MT9000 | TENS Stimulator<br>Intensity 10 | MT9000 Combo<br>TENS/EMS/IF/MIC<br>Stimulator | Sonicator Plus940 | N/A |
| 3 | Manufacturer | Shenzhen Dongdixin Technology Co., Ltd. | | | | Mettler Electronics | N/A |
| 4 | Intended for use | Combo Stimulator MT9000<br>For TENS/IF/MIC mode<br>1. Symptomatic relief of chronic intractable pain<br>2. Post traumatic pain<br>3. Post surgical pain<br>For EMS mode<br>1. Relaxation of muscle spasm.<br>2. Increase of local blood flow circulation<br>3. Prevention or retardation of disuse atrophy<br>4. Muscle re-education<br>5. Maintaining or increasing range of motion.<br>6. Immediate post-surgical stimulation of calf muscles to<br>prevent venous thrombosis | | | For TENS/IF/MIC<br>mode<br>1. Symptomatic<br>relief of chronic<br>intractable pain<br>2. Post<br>traumatic pain<br>3. post surgical<br>pain<br>For EMS mode<br>1. Relaxation of<br>muscle spasm.<br>2. Increase of<br>local blood flow<br>circulation<br>3. Prevention<br>or retardation of<br>disuse atrophy<br>4. Muscle<br>re-education<br>5. Maintaining<br>or increasing<br>range of motion | 4-Pole Interferential,<br>2-Pole Interferential,<br>TENS and<br>Microcurrent<br>waveforms<br>1. Symptomatic relief<br>of chronic intractable<br>pain<br>2. Post-traumatic<br>pain<br>3. Post-surgical pain<br>EMS, TENS, Hi Volt<br>and Russian<br>waveforms<br>I. Relaxation of<br>muscle spasms<br>2. Increase local<br>blood circulation<br>3. Prevention or<br>retardation of disuse<br>atrophy<br>4. Muscle<br>re-education | Same |
| | | Combo Stimulator LT7102<br>For TENS/IF/MIC mode<br>1. Symptomatic relief of chronic intractable pain<br>2. Post traumatic pain<br>3. Post surgical pain | | | | | |
| | | | | | | | |
| | | For EMS/RUSS mode<br>1. Relaxation of muscle spasm.<br>2. Increase of local blood flow circulation<br>3. Prevention or retardation of disuse atrophy<br>4. Muscle re-education<br>5. Maintaining or increasing range of motion.<br>6. Immediate post-surgical stimulation of muscles to prevent venous thrombosis | | | 6. Immediate<br>post-surgical<br>stimulation of<br>muscles to<br>prevent venous<br>thrombosis | 5. Maintaining or<br>increasing range of<br>motion<br>6. Immediate post<br>surgical stimulation<br>of calf muscles to<br>prevent venous<br>thrombosis | |
| | | TENS Stimulator Intensity 10<br>1. Symptomatic relief of chronic intractable pain<br>2. Post traumatic pain<br>3. Post surgical pain | | | | | |
| 5 | Power Source | 3.7V Li-ion Battery<br>Charger output: 5.0V<br>DC, 300mA | 9V Battery<br>Adapter Output: 9V<br>DC 0.8A | 9V Battery<br>Adapter Output: 9V<br>DC 0.8A | 9V Battery | AC Line | Different, but they all<br>comply with IEC 60601-1<br>requirements. |
| | Method of Line current<br>isolation | N/A | N/A | N/A | Battery Supply N/A | Reinforced insulation | |
| | - Patient Leakage<br>Current<br>-Normal condition | 5 uA | 4 uA | 1 uA | 0.61uA | >50uA | |
| | -Single fault condition | 5 uA | 4 uA | 9 uA | 0.68uA | >50uA | |
| 6 | Number of Output<br>Modes | 5 | 4 | 1 | MT9000 4<br>MT9001 1<br>MT9002 1<br>MT9003 1<br>MT9004 1 | 8 | Different, the number of<br>output mode is the feature<br>of device, doesn't affect the<br>safety and effectiveness |
| 7 | Number of Output<br>Channels | 2 | 2 | 2 | 2 | 4 | Different, the number of<br>output channels is the<br>feature of device, doesn't |
| | - Synchronous or<br>Alternating? | Synchronous and<br>Alternating | Synchronous and<br>Alternating | Synchronous | Synchronous and<br>Alternating | Synchronous and<br>Alternating | affect the safety and<br>effectiveness |
| | - Method of Channel<br>Isolation? | By enclosure | By enclosure | By enclosure | By enclosure | By transformer | Different, but they all<br>comply with IEC 60601-1<br>requirements. |
| 8 | Constant Current?<br>Constant Voltage? | Yes<br>No | Yes<br>No | Yes<br>No | Yes<br>No | Yes<br>No | Same |
| 9 | Software/Firmware/<br>Microprocessor<br>Control? | Yes | Yes | Yes | Yes | Yes | Same |
| 10 | Automatic Overload<br>Trip?<br>Automatic Over Current<br>Trip? | Yes<br>Yes | Yes<br>Yes | Yes<br>Yes | Yes<br>Yes | Yes<br>Yes | Same |
| 11 | Automatic No Load<br>Trip? | Yes | Yes | Yes | Yes | Yes | Same |
| 12 | Automatic Shut off? | Yes | Yes | Yes | Yes | Yes | Same |
| 13 | Patient Override<br>Control? | No | No | No | No | No | Same |
| 14 | Indicator Display<br>-On/Off Status?<br>-Voltage/Current Level?<br>-Low Battery? | Yes<br>Yes<br>Yes | Yes<br>Yes<br>Yes | Yes<br>Yes<br>Yes | Yes<br>Yes<br>Yes | Yes<br>Yes<br>N/A | Same<br>Same<br>Same |
| 15 | Waveform<br>TENS<br>EMS<br>Interferential<br>Microcurrent<br>Russian | Biphasic<br>Biphasic<br>Biphasic<br>Monophasic<br>Biphasic | Biphasic<br>Biphasic<br>N/A<br>N/A<br>N/A | Biphasic<br>Biphasic<br>N/A<br>N/A<br>N/A | Biphasic<br>Biphasic<br>Biphasic<br>Monophasic<br>N/A | Biphasic<br>Biphasic<br>Biphasic<br>Monophasic or<br>Biphasic | Similar |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for DONGDIXIN TECHNOLOGY CO., LTD. The logo is in blue, except for a yellow line that crosses the "X" in "DIXIN". There are also three Chinese characters to the right of the English text.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for DONGDIXIN Technology Co., LTD. The logo is in blue and yellow. The text "DONGDIXIN" is in large, bold, blue letters, and the text "TECHNOLOGY CO., LTD." is in smaller, blue letters below it. To the right of the English text is the company name in Chinese.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for DONGDIXIN Technology Co., LTD. The logo is in blue and yellow. The text "DONGDIXIN" is in large, bold, blue letters, and the text "TECHNOLOGY CO., LTD." is in smaller, blue letters below it. To the right of the English text is the company name in Chinese.
| 16 | Shape | | | | | Biphasic | |
|----|-----------------------------------------|-----------------------------------------------|-----------------------------------------------|-----------------------------------------------|-----------------------------------------------------------------------------|-----------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| | TENS | Square | Square | Square | Square | Square | Same with unmodified<br>device MT9000, different |
| | EMS | Square | Square | N/A | Square | Sinusidal | from Sonicator Plus940. |
| | Interferential | Square | Square | N/A | Square | Sinusidal | The shape is the feature |
| | Microcurrent | Square | Square | N/A | Square | Square | of device, doesn't affect |
| | Russian | Square | N/A | N/A | N/A | Gated Sinusoidal | the safety and<br>effectiveness |
| 17 | Max Output Voltage (V) ±20% | | | | | | |
| | 500Ω | | | | | | Similar with unmodified |
| | TENS | 50 | 48 | 52.5 | 48 | 50 | device MT9000, different |
| | EMS | 50 | 48 | N/A | 48 | 49 | from Sonicator Plus940, |
| | Interferential | 15.75 | 17.5 | N/A | 17.5 | 49 | they all comply with IEC |
| | Microcurrent | 0.4 | 0.36 | N/A | 0.36 | 0.38 | 60601-2-10. |
| | Russian | 15.75 | N/A | N/A | N/A | 50 | |
| 18 | Max Output Current (mA) ±20% | | | | | | |
| | 500Ω | | | | | | Similar with unmodified |
| | TENS | 100 | 96 | 105 | 96 | 100 | device MT9000, different |
| | EMS | 100 | 96 | N/A | 96 | 98 | from Sonicator Plus940, |
| | Interferential | 31.5 | 35 | N/A | 35 | 98 | they all comply with IEC |
| | Microcurrent | 0.8 | 0.76 | N/A | 0.76 | 0.76 | 60601-2-10. |
| | Russian | 31.5 | N/A | N/A | N/A | 100 | |
| 19 | Pulse Width Range | | | | | | Similar with unmodified |
| | TENS | 50-400us | 50-300us | 100us~260us | 50-300us | 50-300us | device MT9000, different |
| | EMS | 200~400us | 50-300us | N/A | 50-300us | 500,250,200us | from Sonicator Plus940, |
| | Interferential | 100/200/400us | 125us | N/A | 125us | 500,250,200us | they all comply with IEC |
| | Microcurrent | 2-200ms | 2-200ms | N/A | 2-200ms | 1.25ms-1.67s | 60601-2-10. |
| | Russian | 400us | N/A | N/A | N/A | 400us | |
| 20 | Frequency | | | | | | |
| | TENS | 1~150Hz | 0.5-150Hz | 50~150Hz | 0.5-150Hz | 0.5-250Hz | |
| | EMS | 1~100Hz | 1-150Hz | N/A | 1-150Hz | 2kHz,4kHz,5kHz | |
| | Interferential | 2.5K,5K,10K Hz | 4kHz | N/A | 4kHz | 2kHz,4kHz,5kHz | |
| | | | | | | | |
| | Microcurrent<br>Russian | 1-150Hz<br>2.5kHz | 1-150Hz<br>N/A | N/A<br>N/A | 1-150Hz<br>N/A | 0.3-400Hz<br>2.5kHz | |
| 21 | Beat Frequency<br>Interferential | 1-200Hz | 1-150Hz | N/A | 1-150Hz | 1-250Hz | |
| 22 | Timer Range (minutes) | 5-90 minutes | 0-60 minutes | 15-60 minutes | 0-60 minutes | 0-60 minutes | Different, the treatment<br>time doesn't affect the<br>safety and effectiveness |
| 23 | Compliance with<br>Voluntary Standards? | IEC60601-1,<br>IEC60601-1-2,<br>IEC60601-2-10 | IEC60601-1,<br>IEC60601-1-2,<br>IEC60601-2-10 | IEC60601-1,<br>IEC60601-1-2,<br>IEC60601-2-10 | IEC60601-1,<br>IEC60601-1-2,<br>IEC60601-2-10,<br>MDD93/42/EEC,<br>Annex II | IEC60601-1,<br>IEC60601-1-2,<br>IEC60601-2-10,<br>MDD93/42/EEC, Annex<br>II | Similar |
| 24 | Compliance with 21<br>CFR 898? | Yes | Yes | Yes | Yes | Yes | Same |
| 25 | Weight (Ibs.) | 0.28 | 0.28 | 0.28 | 0.28 | 11 | Different, the different |
| 26 | Dimensions (in.) H x W<br>x L | 4.6x2.36x0.9 | 4.5x2.55x0.9 | 4.5x2.55x0.9 | 4.5x2.55x0.9 | 4.9x13.6x10.5 | weight and dimension<br>doesn't affect the safety<br>and effectiveness. |
| 27 | Housing Materials &<br>Construction | Enclosure: ABS | Enclosure: ABS | Enclosure: ABS | Enclosure: ABS | Metal Casing | Different, but they all<br>comply with IEC 60601-1<br>requirements. |
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image is a logo for DONGDIXIN TECHNOLOGY CO., LTD. The logo is in blue and orange. The word "DONGDIXIN" is in large, bold, blue letters, with the "X" in "DIXIN" being orange. Below "DONGDIXIN" is the text "TECHNOLOGY CO., LTD." in smaller, blue letters. To the right of the text is a column of three Chinese characters.
{13}------------------------------------------------
## 7. Performance Date:
The following performance data are provided in support of the substantial equivalence determination:
## 6.1 Biocompatibility testing
Compared with unmodified device, the new product MT9000 and Intensity 10 has the same accessories which has been evaluated based on the ISO10993 standard and submits to FDA (k093138).
The new product LT7102 added the electrode accessories like as different part electrode belt previously had been submitted with Shenzhen Dongdixin Technology Co., Ltd (k141076) which has been evaluated based on the ISO10993 standard and submit to FDA.
## 6.2 Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the LT2061. The system complies with the IEC 60601-1 and IEC 60601-2-10 standards for safety and the IEC 60601-1-2 standard for EMC.
## 6.3 Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern. IEC 62304 was followed.
## 8. Conclusions:
The LT7102, MT9000 and Intensity 10 have the same intended use and technological characteristics as the predicate device of Sonicator Plus 940, Model ME940 device and MT9000 Combo TENS/EMS/IF/MIC Stimulator. Moreover, bench testing and safety report supplied in this submission demonstrates that the difference in the submitted models could maintain the same safety and effectiveness as that of predicate device. In the other words, those engineering difference do not affect the intended use or alter the fundamental scientific technology of the device. Thus, the LT7102, MT9000 and Intensity 10 are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.