MaxZero Extension Sets with Needless Connector

K171957 · Carefusion 303, Inc. · FPA · Jul 19, 2017 · General Hospital

Device Facts

Record IDK171957
Device NameMaxZero Extension Sets with Needless Connector
ApplicantCarefusion 303, Inc.
Product CodeFPA · General Hospital
Decision DateJul 19, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5440
Device ClassClass 2
AttributesPediatric, 3rd-Party Reviewed

Indications for Use

The MaxZero™ Extension Set with Needle-Free Connector(s) is a sterile single patient use device intended to be used for the delivery and/or aspiration of fluids to/from an IV catheter.

Device Story

Intravascular extension set with integrated MZ1000 zero-reflux needleless connector; used in hospital environments for fluid delivery/aspiration via IV catheter. Device includes PVC tubing, slide clamps, male/female luers, and Y/T-site connectors. Operated by clinicians; provides needleless access point to IV line; prevents needle stick injuries and microbial ingress. Connector surface is flat/smooth for disinfection with 70% IPA. Supports 7-day use and 200 activations. Pressure-rated configurations allow use with power injectors. Benefits include reduced infection risk and safe, needle-free fluid administration.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including microbial ingress, high-pressure testing, bond pull, priming volume, flow rate, and harsh infusates testing. Biocompatibility evaluated per ISO 10993; sterilization validated per ISO 11137.

Technological Characteristics

Intravascular extension set; PVC tubing; MZ1000 needleless connector. Materials non-DEHP, non-latex. Standards: ISO 594-1/2 (luer), ISO 8536-4/8/9/10 (infusion equipment), ISO 10993 (biocompatibility), ISO 11137 (radiation sterilization). Pressure rated to 325 psi. Connectivity: mechanical luer interface.

Indications for Use

Indicated for single use in pediatrics (neonates, infants, children, adolescents) and immunocompromised patients for direct injection, intermittent infusion, continuous infusion, or aspiration of drugs, blood, and fluids. Pressure rated for use with power injectors up to 325 psi at 10mL/sec.

Regulatory Classification

Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the most prominent. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 # July 19, 2017 CareFusion, Inc. % Mark Job Regulatory Technology Services. LLC 1394 25th Street North West Buffalo, Minnesota 55313 11021 Re: K171957 Trade/Device Name: MaxZero™ Extension Sets with Needleless Connector Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: June 28, 2017 Received: June 29, 2017 Dear Mr. Mark Job: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR {1}------------------------------------------------ Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. # Tara A. Ryan -S for Lori Wiggins Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) Device Name MaxZero™ Extension Set with Needle-Free Connector(s) Indications for Use (Describe) Pressure Rated: The MaxZero™ Extension Set with Needle-Free Connector(s) is for single use only. The MaxZero™ Extension Set with Needle-Free Connector(s) may be used for direct injection, intermittent infusion or aspiration. This set may be used with power injector procedures to a maximum pressure of 325 psi at a flow rate of 10mL per second. Type of Use (Select one or both, as applicable) | <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | |-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------| |-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------| #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside. The shield has a yellow accent on the top left corner. To the right of the circle is the text "CareFusion" in a dark gray sans-serif font. # 510(k) K171957 Summary #### Submitter Information | Submitter: | CareFusion, Inc.<br>10020 Pacific Mesa Blvd.<br>San Diego, CA 92121, USA | |-----------------|--------------------------------------------------------------------------| | Contact Person: | Viviana Lai | | Phone: | 714-922-7679 | | Email: | Viviana.lai@carefusion.com | | Date Prepared: | June 23, 2017 | #### Subject Device Identification | Trade Name: | MaxZero™ Extension Set with Needle-Free Connector(s) | |-----------------------|------------------------------------------------------| | Common Name: | Intravascular Administration Set | | Classification Name: | Intravascular Administration Set | | Classification Panel: | General Hospital | | Regulation Number: | 21 CFR 880.5440 | | Requlatory Class: | Class II | | Product Code: | FPA | #### Predicate Device Identification | Trade Name: | MaxZero™ Extension Sets with Needleless Connector | |----------------------|---------------------------------------------------| | Manufacturer: | Carefusion, Inc. | | 510k Number: | K140831 | | 510K Clearance Date: | April 15 2014 | #### Reason for Submission The objective of this submission is to introduce new components such as Needless Access Connector-Y (NAC-Y also known as Y-site) and Needless Access Connector-T (NAC-T also known as T-connector) connectors to the MaxZero™ Extension Set with Needle-Free Connector(s) product line configurations. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside. The top left corner of the shield is yellow. To the right of the circle is the word "CareFusion" in gray. #### Device Description The MaxZero™ Extension Set with Needle-Free Connector(s) are intravascular extension sets intended for single patient use, including pediatrics (neonates, infants, children, adolescents) and immunocompromised patients, for direct injection, intermittent infusion continuous infusion or aspiration of drugs, blood and fluids, All MaxZero™ Extension Set with Needle-Free Connector(s) include the previously cleared zero reflux MZ1000 needleless Connector bonded to the extension set tubing. The MZ1000 needleless connector allows thorough and easy disinfection due to a solid, flat smooth surface and eliminates the risk of needle stick injuries. The MaxZero™ needleless connectors are sterile single patient devices that can be used for seven (7) days and 200 activations. All extension sets included in this submission are not made from natural rubber latex or DEHP. #### Indication for Use Pressure Rated: The MaxZero™ Extension Set with Needle-Free Connector(s) is for single use only. The MaxZero™ Extension Set with Needle-Free Connector(s) may be used for direct injection, intermittent infusion, continuous infusion or aspiration. This set may be used with power injector procedures to a maximum pressure of 325 psi at a flow rate of 10mL per second. #### Technological Characteristics The information provided in the premarket notification demonstrates that the subject device, MaxZero™ Extension Set with Needle-Free Connector(s), is substantially equivalent to the legally marketed predicated devices. The subject device and the predicate device are intended to be used for the delivery and/or aspiration of fluids to/from an IV catheter in a hospital environment. The subject and the predicate devices are similar in physical properties, materials, and configuration. Each device includes connectors that allow for needleless access to the IV line during IV therapy eliminating the risk of needle injury. The subject device incorporates MaxZero™ Needleless Connector attached to IV tubing. Components of the subject devices are made of materials that are substantially equivalent to the predicate devices listed above and this submission includes comprehensive biocompatibility testing for all device materials included in this submission. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for CareFusion. The logo consists of an orange circle with a white shield inside, and the text "CareFusion" in gray. The shield has a yellow accent on the top left. # Substantial Equivalence Table | | MaxZero™ Extension Set with<br>Needle-Free Connector(s)<br>(Subject Device) | CareFusion MaxZero™<br>Extension Sets with Needleless<br>Connector<br>(Predicate device- K140831) | |------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | FDA Reg. Number | 21 CFR 880.5440 | 21 CFR 880.5440 | | FDA Regulation<br>Name | Intravascular Administration Set | Intravascular Administration Set | | FDA Class | Class II | Class II | | FDA Product Code | FPA | FPA | | Product Description | MaxZero™ Extension Set with<br>Needle-Free Connector(s) and the<br>predicate devices are sterile, single<br>patient use, including pediatrics<br>(neonates, infants, children,<br>adolescents) and<br>immunocompromised patients for<br>direct inject, intermittent infusion,<br>continuous infusion or aspiration of<br>drugs, blood and fluids | MaxZero™ Extension Sets with<br>Needleless Connector and the<br>predicate devices are sterile, single<br>patient use, including pediatrics<br>and immunocompromised patients<br>for direct inject, intermittent<br>infusion, continuous infusion or<br>aspiration of drugs, blood and<br>fluids | | Intended Use | The MaxZero™ Extension Set with<br>Needle-Free Connector(s) is a<br>sterile single patient use device<br>intended to be used for the delivery<br>and/or aspiration of fluids to/from<br>an IV catheter. | The MaxZero™ Extension Sets<br>With Needleless Connector is a<br>sterile single patient use device<br>intended to be used for the<br>delivery and/or aspiration of fluids<br>to/from an IV catheter. | | Indications for Use | Pressure Rated: The MaxZero™<br>Extension Set with Needle-Free<br>Connector(s) is for single use only.<br>The MaxZero™ Extension Set with<br>Needle-Free Connector(s) may be<br>used for direct injection,<br>intermittent infusion, continuous<br>infusion or aspiration. This set may<br>be used with power injector<br>procedures to a maximum pressure<br>of 325 psi at a flow rate of 10mL<br>per second. | Pressure Rated: The MaxZero™<br>multi fuse extension set with<br>needleless connector(s) is for<br>single use only. The extension set<br>may be used for direct injection,<br>intermittent infusion, continuous<br>infusion or aspiration. This set may<br>be used with power injector<br>procedures to a maximum<br>pressure of 325 psi at a flow rate<br>of 10mL per second.<br><br>Non Pressure Rated: The<br>MaxZero™ multi fuse extension set<br>with needleless connector(s) is for<br>single use only. The extension set<br>may be used for direct injection,<br>intermittent infusion, continuous<br>infusion or aspiration. | | Needless Connector | CareFusion MZ1000 | CareFusion MZ1000 | | | MaxZero™ Extension Set with<br>Needle-Free Connector(s)<br>(Subject Device) | CareFusion MaxZero™<br>Extension Sets with Needleless<br>Connector<br>(Predicate device- K140831) | | Device Components | MZ1000 Needleless Connector, PVC<br>tubing with various lengths, slide<br>clamps, male luer (spin lock),<br>female luer (female wing adapter),<br>NAC- Y connectors (Y-sites), NAC-T<br>connectors (T-connectors) | MZ1000 Needleless Connector,<br>PVC tubing with various length and<br>ID/OD, check valves, anti-siphon<br>valves, slide clamps, Y connectors,<br>T-connectors, stopcocks, male spin<br>lock and female wing adapter | | Provided Sterile | Yes | Yes | | Single Use | Yes | Yes | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside. The shield has a yellow accent on the top left corner. To the right of the circle is the word "CareFusion" in gray sans-serif font. ## Explanation of Similarities and Differences technological Characteristics compared to Predicate Device The Subject MaxZero™ Extension Set with Needle-Free Connector(s) have the following similarities to the predicate devices: - Same Intended Use and Indication for Use ● - Principle of operation ● - Device Design ● - Zero Reflux Needleless Connector ● - Designed to prevent microbial ingress ● - Needleless connector can be disinfected with 3 sec scrub with 70% IPA ● - Maximum clinical use of 7 days 200 activations for the needleless ● connector (single patient use) - . Non-hemolytic and Non-pyrogenic - Can be used with low power injectors with maximum of 325 psi & flow rate ● of 10ml/second (pressure rated sets) - . Not made with DEHP and not made with natural latex rubber - Safe for use in MRI environment ● - Sets can be used with harsh infusates ● The following are technical characteristics differences between the subject and predicate devices: - . The subject device includes Needleless Access Connector-Y (NAC-Y also known as Y site) and Needless Access Connector-T (NAC-T also known as T connector) connectors that provide additional access ports. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for CareFusion. The logo consists of an orange circle with a white shield inside. The shield has a yellow accent on the top left corner. To the right of the circle is the word "CareFusion" in gray sans-serif font. ### Discussion of Performance Data: #### Non-Clinical Data CareFusion performed design verification performance testing to verify, demonstrate and support the claim of substantial equivalence to the predicate devices. All test results met their acceptance criteria and support that the MaxZero™ Extension Set with Needle-Free Connector(s) are appropriately designed for their intended use. Carefusion performed design verification performance testing according to the FDA recognized performance standards and guidelines. - . ISO 594-1:1986 Conical fittings with a 6% (luer) taper of syringes, needles and certain other medical equipment - Part 1: General requirements" - . ISO 594-2:1998 Conical fittings with 6%(luer) taper for syringes, needles, and certain other medical equipment - Part 2 Locking fittings" - . ISO 8536-4:2010 "Infusion equipment for medical use- Part 4: Infusion set for single use, gravity feed" - . ISO 8536-8:2004 "Infusion equipment for medical use – Part 8: Infusion equipment for medical use. Infusion equipment for use with pressure infusion apparatus" - . ISO 8536-9:2004 *Infusion equipment for medical use – Part 9: Fluid lines for single use with pressure infusion equipment" - . ISO 8536-10:2004 "Infusion equipment for medical use – Part 10: Accessories for fluid lines for single use with pressure infusion equipment" - . ISO 14971:2007 devices" - . Guidance for Industry and FDA Staff - Intravascular Administration Sets Premarket Notification Submission[510(k)], July 11,2008 The following additional performance data were collected. - . Microbial Ingress and Barrier testing - High-Pressure Testing - Air Water Interface Visibility - Set Internal Excess Pressure Testing . - . Clamps - Internal Excess Pressure and Tubing Open Fluid Path Testing - Bond Pull Testing ● - . Priming Volume and Flow Rate Testing - . Harsh Infusates Testing {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside. The shield has a yellow accent on the top left corner. To the right of the circle is the word "CareFusion" in a sans-serif font. # Biocompatibility Biocompatibility assessments were conducted in accordance with ISO-10993-1:2009, "Biological evaluation of medical devices – part 1: Evaluation and testing within a risk management process," Carefusion performed the biocompatibility testing of the components and finish product according to the following parts of the ISO 10993 standard. - . ISO 10993-2:2006: "Biological evaluation of medical devices – part 2: Animal welfare requirements" - . ISO 10993-4:2002: "Biological evaluation of medical devices - part 4: Selection of tests for interactions with blood" - . ISO 10993-5:2009: "Biological evaluation of medical devices - part 5: Tests for in vitro cytotoxicity" - . ISO 10993-10:2010: "Biological evaluation of medical devices – part 10: Tests for irritation and delayed-type hypersensitivity" - . ISO 10993-11:2006: "Biological evaluation of medical devices – part 11: Test for systemic Toxicity" - . ISO 10993-12:2012: "Biological evaluation of medical devices - part 12: Sample preparation and reference materials" #### Sterilization and Shelf life The subject device, MaxZero™ Extension Set with Needle-Free Connector(s), is radiation sterilized and the shelf life data supports a shelf life claim of 3 years. Sterilization and shelf life testing were completed according to the following FDA recognized quidelines: - . ISO 11137-1:2006 "Sterilization of health care products - Radiation- Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices" - ISO 11137-2:2006 "Sterilization of health care products Radiation Part 2 -● Establishing the sterilization dose" - . ISO 11607:2003 "Packaging for terminally sterilized medical devices" - ASTM F1980-07: 2002 "Standard Guide for Accelerated Aging of Sterile . Barrier Systems for Medical Devices" - ASTM F1140:2000 Standard Test Method for Burst Testing of Flexible Package Seals Using Internal Air Pressurization within Restraining Plates" - ASTM D4169: 1998 " Standard Practice for Performance Testing of Shipping . Containers and Systems" - ASTM-F1929-98(04): 1998 "Standard Test Method for Detecting Seal Leaks in ● Porous Medical Packaging by Dye Penetration" {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the CareFusion logo. The logo consists of an orange circle with a white shield inside. The shield has a yellow accent on the top left corner. To the right of the circle is the word "CareFusion" in gray sans-serif font. # Clinical Data There is no clinical data included in this submission. #### Conclusion The device met the acceptance criteria for all functional, microbial ingress, sterility, biocompatibility and other performance criteria. The conclusions drawn from the nonclinical tests that demonstrate that the new device is as safe, as effective, and performs as well as the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...