K171952 · Dimesol, Inc. · FJK · Mar 30, 2018 · Gastroenterology, Urology
Device Facts
Record ID
K171952
Device Name
Dimesol Tubing Sets for Hemodialysis
Applicant
Dimesol, Inc.
Product Code
FJK · Gastroenterology, Urology
Decision Date
Mar 30, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5820
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Dimesol Tubing Sets for Hemodialysis are single-use sterile medical devices intended to connect the patient to the hemodialyzer and the hemodialysis delivery system in hemodialysis treatment. The compatibility of available configurations is the responsibility of the physician / clinician in charge.
Device Story
Dimesol Tubing Sets are single-use extracorporeal blood circuits connecting patients to hemodialyzers and delivery systems. Components include arterial/venous lines, pump segments, drip chambers with filters, clamps, luer lock connectors, and injection ports. Operated by medical professionals in hospital settings; tubing interfaces with electric peristaltic blood pumps. Device facilitates blood transport, anticoagulant infusion, pressure monitoring, and air bubble trapping. Benefits include maintaining sterile extracorporeal circuit integrity during hemodialysis.
Clinical Evidence
Bench testing only. Device performance validated against ISO 8638 (hemodialysis equipment) and ISO 594-2 (conical fittings). Biocompatibility testing performed per ISO 10993 series, including cytotoxicity, sensitization, systemic toxicity, and hemolysis. No clinical data presented.
Technological Characteristics
Materials: PVC, PP, PC, ABS, PE. Components: arterial/venous lines, pump segments, drip chambers, clamps, luer locks. Sterilization: electron beam (SAL 10^-6). Standards: ISO 8638, ISO 594-2, ISO 10993, ISO 9001, ISO 13485. Non-implantable, single-use disposable.
Indications for Use
Indicated for patients with acute and chronic renal failure requiring hemodialysis treatment.
Regulatory Classification
Identification
A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
Predicate Devices
Nipro Blood Tubing Set with Transducer Protector and Priming Set (K072024)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 30, 2018
Dimesol, Inc. % Courtney J. Miller Partner Frost Brown Todd LLC 10 West Broad Street, Suite 2300 Columbus. OH 43215
Re: K171952
Trade/Device Name: Dimesol Tubing Sets for Hemodialysis Regulation Number: 21 CFR§ 876.5820 Regulation Name: Hemodialysis System and Accessories Regulatory Class: II Product Code: FJK, FKB Dated: February 12, 2018 Received: February 13, 2018
Dear Courtney J. Miller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT
510(k) Number: K171952
FORM: FDA 3881
#### Device Name: Dimesol Tubing Sets for Hemodialysis
Indications for Use:
The Dimesol Tubing Sets for Hemodialysis are single-use sterile medical devices intended to connect the patient to the hemodialyzer and the hemodialysis delivery system in hemodialysis treatment. The compatibility of available configurations is the responsibility of the physician / clinician in charge.
X AND/OR Over-The-Counter Use Prescription Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
### K171952-REVISED 510(K) SUMMARY UNDER 21 CFR 807.92
This 510(k) summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content in this 510(k) summary has been provided in conformance with 21 CFR Part 807.92.
#### Submitter's Information A.
| (1) Name: | Frost Brown Todd LLC |
|-----------------------|-------------------------------------------------------|
| (2) Address: | 10 West Broad St., Suite 2300, Columbus, OH 43215 USA |
| (3) Phone: | 614-559-7288 |
| (4) Fax: | 614-464-1737 |
| (5) Contact Person: | (Mr.) Courtney J. Miller, J.D., LL.M. |
| (6) Preparation Date: | June 27, 2017 |
# On Behalf of Applicant Entity (owner of 510(k))
| (1) Applicant Name: | DIMESOL, INC. |
|------------------------|-------------------------------------------------|
| (2) Applicant Address: | 509 Fishing Creek Rd., Lewisberry, PA 17339 USA |
| (3) Applicant Phone: | 717-938-8391 |
| (4) Applicant Fax: | 717-938-3957 |
# Manufactured for Dimesol, Inc. by:
| (1) Manufacturer Name: | Bain Medical Equipment (Guangzhou) Co., Ltd. |
|---------------------------|------------------------------------------------------------------------------------------------------------------------|
| (2) Manufacturer Address: | 10 Juncheng Road, Eastern Area, Guangzhou Economic &<br>Technological Development District, Guangzhou, China<br>510760 |
| (3) Manufacturer Phone: | 0086-20-82265249 |
| (4) Manufacturer Fax: | 0086-20-32067500 |
#### B. Device Name
| (1) Trade Name: | Dimesol Tubing Sets for Hemodialysis |
|---------------------|---------------------------------------------------------------------------------|
| | (2) Common Name: Blood Tubing (or Blood Line or Tubing) Sets for Hemodialysis |
| (3) Classification: | Class II, per 21 CFR 876.5820 (Hemodialysis system and |
| | accessories) |
| (4) Product Code: | FJK / FKB: Gastroenterology/Urology |
| (5) Standard(s): | MDD 93/42/EEC; ISO 8638; ISO9001; ISO13485 |
{4}------------------------------------------------
### K171952-REVISED 510(K) SUMMARY UNDER 21 CFR 807.92
#### C. Legally Marketed Predicate Device(s)
- Nipro Blood Tubing Set with Transducer Protector and Priming Set (K072024) - JMS Blood Tubing Sets (K032975)
#### D. Device Description
[0001] The Dimesol blood tubing sets for hemodialysis are single use arterial and venous blood line (tubing) sets for use with polyethersulfone hollow-fiber hemodialyzers or other dialyzers that are indicated for hemodialysis treatment of acute and chronic renal failure. TABLE 1. below, provides the basic product codes used for the Dimesol (blood tubing sets for hemodialysis. There are five (5) basic product codes that represent the various configurations (i.e., models) in which the device may be manufactured. TABLE 1 also provides the dimensions (in millimeters) for the pump segment and further identifies the hemodialysis system with which the devices are compatible. TABLE 2, below, provides the basic device components and their respective functions for both the arterial and venous lines.
#### TABLE 1. Product Codes and Specifications
| Product Code | Pump Seqment<br>(IDxODxLength in mm) | System/Device<br>Compatibilities |
|--------------|--------------------------------------|----------------------------------|
| EDIM-BL-001E | 8.00x12.00x350 | Fresenius 2008K |
| EDIM-BL-002E | 8.00x12.00x350 | Fresenius 2008K |
| EDIM-BL-003E | 8.00x12.00x350 | Fresenius 2008K |
| EDIM-BL-004E | 8.00x12.00x350 | Fresenius 2008K |
| EDIM-BL-005E | 8.00x12.00x350 | Fresenius 2008K |
#### TABLE 2. Basic Device Components and Functions (Dimesol: Arterial and Venous Lines)
| | Component | Function |
|---|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Vented Dialyzer<br>Connector Caps | Prevents contamination of blood path through Dialyzer Connector (Arterial/Venous Lines). |
| 2 | Dialyzer Connectors | Connects the arterial and venous lines to the appropriate dialyzer blood port (Blue for venous line and Red for arterial line). |
| 3 | Blood (Drip) Chamber<br>(with or w/o filter) | Provides an air space for monitoring pressures and trapping air bubbles. Venous blood chamber includes conical mesh filter to block clots and to trap foam, or air bubbles before entering patient |
| 4 | Blood (Drip) Chamber Cover | Permits monitoring line and level adjustment line to be sited |
| 5 | Large On/Off Clamps | Typically located along the main blood lines for permitting occlusion of blood tubing set |
| 6 | Small On/Off Clamps | Typically located at all heparin lines, monitoring lines, and level |
{5}------------------------------------------------
### K171952-REVISED 510(K) SUMMARY UNDER 21 CFR 807.92
| 7 | Luer Lock Connectors | adjusting lines for occlusion, if necessary. |
|----|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 8 | Vented Caps for Luer Lock Connectors | Female luer lock connectors provide a connection site based on an internationally recognized standard within the entire blood tubing set. |
| 9 | Recirculation Luer Lock Connectors | Vented caps cover the luer lock connectors to prevent contamination. |
| 10 | Transducer Protectors (TP) | Recirculation connectors allow the connection of hemodialysis blood tubing for recirculation within the extracorporeal blood circuit. |
| 11 | Blood Injection Ports | Prevent blood contamination risk at pressure monitoring line to dialysis machine through the hydrophobic membrane |
| 12 | Patient Connectors (Self-Injection) | Provide access sites for blood sampling and/or medications administration (Red and Blue). |
| 13 | Heparin Infusion Line | International connection to vascular access devices (e.g., AV Fistula needle set) in the extracorporeal blood circuit. |
| 14 | Pump Segment and T-Connectors | Provides adequate anticoagulant for blood before entering dialyzer via a syringe pump. Normally includes a female luer lock connector, a small clamp, and small diameter tubing. |
| 15 | Tubing Segments | Delivery of blood at pre-determined pump rate. |
| 16 | Drainage Bag | Transports blood and fluid during hemodialysis; available in various sizes and lengths. |
| 17 | Negative Pressure Pillow | Collects priming saline waste. |
| 18 | IV | Collapses when predetermined negative pressures are detected. |
| 19 | Suspended Spike (Cap) | Infusion |
| | | Priming (plugged into saline bag) |
Dimesol tubing sets for hemodialysis are intended for use as the extracorporeal [0002] blood circuit during hemodialysis and to connect a dialysis patient to a hemodialyzer. These tubing sets are intended for single use only. The major device components of these tubing sets are manufactured from medical-grade polyvinyl chloride (PVC), polypropylene (PP), polycarbonate (PC), acrylonitrile butadiene styrene (ABS), and fire-retardant polyethylene (PE).
#### E. Indications for Use
[0003] The Dimesol Tubing Sets for Hemodialysis are single-use sterile medical devices intended to connect the patient to the hemodialyzer and the hemodialysis delivery system in hemodialysis treatment. The compatibility of available configurations is the responsibility of the physician / clinician in charge.
{6}------------------------------------------------
#### K171952-REVISED 510(K) SUMMARY UNDER 21 CFR 807.92
#### F. Substantial Equivalence
[0004] The Dimesol (DORA) blood tubing sets (or simply "tubing sets") for hemodialysis are considered by the Submitter to be substantially equivalent in construction, design, materials, performance, and intended use to the JMS Blood Tubing Sets (K032975) and/or the Nipro Blood Tubing Set with Transducer Protector and Priming Set (K072024). All of these tubing sets are intended for use as a portion of the extracorporeal blood circuit during hemodialysis and to connect a dialysis patient to a hemodialyzer. The comparison to the JMS Blood Tubing Sets (K032975) is based primarily on structure, components, materials, and other features common to both the Dimesol blood tubing sets and JMS Blood Tubing Sets (K032975). However, the comparison to the Nipro Blood Tubing Set with Transducer Protector and Priming Set (K072024) is based on structure, components, and materials, as well as performance and biocompatibility.
[0005] As detailed in TABLE 3, below, the Dimesol (DORA) blood tubing sets for hemodialysis and the predicate JMS Blood Tubing Sets (K032975) and/or the Nipro Blood Tubing Set with Transducer Protector and Priming Set (K072024) have equivalent structural and materials characteristics. As detailed in TABLE 4, below, the Dimesol (DORA) blood tubing sets for hemodialysis and the predicate JMS Blood Tubing Sets (K032975) and/or the Nipro Blood Tubing Set with Transducer Protector and Priming Set (K072024) have equivalent general characteristics. As illustrated in TABLE 5, below, the Dimesol (DORA) blood tubing sets for hemodialysis were tested for various biocompatibility factors and found to be within the accepted test parameters of ISO 10993. The Nipro Blood Tubing Set with Transducer Protector and Priming Set (K072024) was also determined to meet the requirements of ISO 10993; thus, the two devices are substantially equivalent with regard to biocompatibility. As illustrated in TABLE 6, below, the Dimesol (DORA) blood tubing sets for hemodialysis were tested for various performance characteristics and found to be within the accepted test parameters of ISO 8638 and ISO 594-2. The Nipro Blood Tubing Set with Transducer Protector and Priming Set (K072024) was also determined to meet the requirements of ISO 8638 and ISO 594-2; thus, the two devices are substantially equivalent with regard to performance characteristics.
| | Component | Material | Blood<br>Contacting | Supplier | JMS<br>Blood Tubing Sets<br>(K032975) | Nipro<br>BloodTubing Sets<br>(K072024) |
|---|-----------------------------------|----------|---------------------|--------------|---------------------------------------|----------------------------------------|
| 1 | Vented Dialyzer<br>Connector Caps | PP | No | Bain Medical | component or<br>equivalent present | component or<br>equivalent present |
| 2 | Dialyzer<br>Connectors | PVC | Yes | Bain Medical | component or<br>equivalent present | component or<br>equivalent present |
TABLE 3. Substantial Equivalence Comparison Table: Components and Materials (A/V)
<sup>510(</sup>k) Premarket Notification for Dimesol, Inc. Blood Tubing Sets for Hemodialysis
{7}------------------------------------------------
### K171952-REVISED 510(K) SUMMARY UNDER 21 CFR 807.92
| 3 | Blood (Drip)<br>Chamber<br>(with or w/o filter) | PVC | Yes | Bain Medical | component or equivalent present | component or equivalent present |
|----|-------------------------------------------------|-----|-----|--------------|---------------------------------|---------------------------------|
| 4 | Blood (Drip) Chamber Cover | PVC | Yes | Bain Medical | component or equivalent present | component or equivalent present |
| 5 | Large On/Off Clamps | PP | No | Bain Medical | component or equivalent present | component or equivalent present |
| 6 | Small On/Off Clamps | PP | No | Bain Medical | component or equivalent present | component or equivalent present |
| 7 | Luer Lock Connectors | PC | Yes | Bain Medical | component or equivalent present | component or equivalent present |
| 8 | Vented Caps for Luer Lock Connectors | PE | No | Bain Medical | component or equivalent present | component or equivalent present |
| 9 | Recirculation Luer Lock Connectors | PP | Yes | Bain Medical | component or equivalent present | component or equivalent present |
| 10 | Transducer Protectors (TP) | SBC | Yes | Bain Medical | component or equivalent present | component or equivalent present |
| 11 | Blood Injection Ports | PVC | Yes | Bain Medical | component or equivalent present | component or equivalent present |
| 12 | Patient Connectors (Self-Injection) | PC | Yes | Bain Medical | component or equivalent present | component or equivalent present |
| 13 | Heparin Infusion Line | PP | No | Bain Medical | component or equivalent present | component or equivalent present |
| 14 | Pump Segment and T-Connectors | PVC | Yes | Bain Medical | component or equivalent present | component or equivalent present |
| 15 | Tubing Segments | PVC | Yes | Bain Medical | component or equivalent present | component or equivalent present |
#### TABLE 4. Substantial Equivalence Comparison Table: General Factors (A/V)
| | Factor | Dimesol (DORA) AVF Needle Set | JMS Blood Tubing Sets (K032975) | Nipro Blood Tubing Sets (K072024) |
|---|-------------------------|----------------------------------------------------------------------------------------------|---------------------------------|-----------------------------------|
| 1 | Device Class | II | SE | SE |
| 2 | Product Code | FJK | SE | SE |
| 3 | Regulation No. | 21 CFR 876.5820 | SE | SE |
| 4 | intended use | hemodialysis | SE | SE |
| 5 | indications for use | acute and chronic<br>renal failure | SE | SE |
| 6 | instructions for use | - remove cap<br>- insert needle into<br>patient's vein<br>- connect with blood<br>tubing set | SE | SE |
| 7 | configuration of device | See Device Description | SE | SE |
| 8 | implantable device | non-implantable | SE | SE |
{8}------------------------------------------------
# K171952-REVISED 510(K) SUMMARY UNDER 21 CFR 807.92
| 9 | target population | dialysis patients | SE | SE |
|----|--------------------------------|---------------------------------------------------|-----------------------------------|------------------|
| 10 | anatomical sites | arterial / venous access | SE | SE |
| 11 | location of use | hospital use | SE | SE |
| 12 | energy used and/or | electric peristaltic<br>blood pump | SE | SE |
| 13 | human factors | used by medical<br>professionals | SE | SE |
| 14 | design factors | hemodialysis circuit<br>arterial and venous lines | SE | SE |
| 15 | performance | single use | SE | SE |
| 16 | materials | PVC, PP, PC, ABS, PE | PVC, PE, PP, ABS,<br>PC | SE |
| 17 | biocompatibility | Compatible with ISO<br>10993 series standards | Biocompatible<br>10993 compatible | biocompatible |
| 18 | environmental<br>compatibility | disposable | SE | SE |
| 19 | sterilization method | electron beam<br>(SAL 10-6) | SE<br>(SAL 10-6) | SE<br>(SAL 10-6) |
| 20 | Non-pyrogenic | Non-pyrogenic | SE | SE |
## TABLE 5. Biocompatibility Comparison with K072024
| Dimesol Blood Tubing Sets (K171952) | Predicate Device<br>(K072024) |
|-------------------------------------|-------------------------------|
| Cytoxicity | Biocompatible |
| Sensitization | Conforms to ISO<br>10993-1 |
| Intracutaneous Reactivation | |
| Acute systemic Toxicity | |
| Hemolysis | |
| Partial Thromboplastin Time | |
| Complement System | |
| In vitro Chromosomal Aberration | |
| Bacterial Reverse Mutation | |
| Mouse Bone Marrow Micronucleus | |
| Pyrogen | |
| Mechanical Hemolysis | |
{9}------------------------------------------------
### K171952-REVISED 510(K) SUMMARY UNDER 21 CFR 807.92
### TABLE 6. Performance Comparison with K072024
| Dimesol Blood Tubing Sets (K171952) | | Predicate Device<br>(K072024) |
|-------------------------------------|--------------------------------|--------------------------------|
| Specific Tests Performed | Conforms to ISO<br>8638: 2010 | Conforms to ISO<br>8638: 2010 |
| Structural Integrity | Conforms to ISO<br>594-2: 1998 | Conforms to ISO<br>594-2: 1998 |
| Connectors to Hemodialyzer | | |
| Needle Access Ports | | |
| Transducer Protectors | | |
| Tubing Compliance | | |
| Priming Volume | | |
| Pump Segment Performance | | |
| Conical Fitting | | |
| Tensile Strength | | |
| Simulated Operation | | |
| Repeated Closing | | |
| Endurance Testing | | |
#### G. Conclusion
[0006] The Dimesol blood tubing sets for hemodialysis are believed to be substantially equivalent in construction, design, materials, performance, and intended use to the JMS Blood Tubing Sets (K032975) and/or the Nipro Blood Tubing Set with Transducer Protector and Priming Set (K072024). All of these tubing sets are intended for use as a portion of the extracorporeal blood circuit during hemodialysis and to connect a dialysis patient to a hemodialyzer. In addition, testing of the Dimesol Blood Tubing Sets for Hemodialysis indicates that they are safe and effective for their intended use. The device labeling contains or includes a Package Insert, which includes indications for use, cautions and warnings, as well as the general operating instructions required for proper use of the device. This information promotes safe and effective use of the blood tubing set. Dimesol, Inc. believes that the information provided in this submission clearly describes the Dimesol Blood Tubing Sets for Hemodialysis and demonstrates that they are believed to be substantially equivalent to the aforementioned predicate device(s).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.