Alevicyn SG Antimicrobial Gel
K171935 · Sonoma Pharmaceuticals, Inc. · FRO · Dec 1, 2017 · SU
Device Facts
| Record ID | K171935 |
| Device Name | Alevicyn SG Antimicrobial Gel |
| Applicant | Sonoma Pharmaceuticals, Inc. |
| Product Code | FRO · SU |
| Decision Date | Dec 1, 2017 |
| Decision | SESE |
| Submission Type | Traditional |
| Device Class | Class U |
| Attributes | Therapeutic |
Indications for Use
Rx Indications: Alevicyn SG Antimicrobial Gel is intended to moisten the wound bed and facilitate autolytic debridement of acute and chronic dermal lesions. Alevicyn SG Antimicrobial Gel is indicated to manage and relieve the burning, itching, erythema, scaling, and pain experienced with various types of dermatoses, including atopic dermatitis, radiation dermatitis, seborthea and sebortheic dermal irritation, sores, injuries, and ulcers of dermal tissue in addition to moistening and lubricating absorbent wound dressings. The Alevicyn SG Antimicrobial Gel may also be used to manage partial or full thickness wounds such as first and second degree burns, stage I - IV pressure ulcers, diabetic and stasis ulcers, abrasions and skin irritations, surgical wounds (donor and graft sites, incisions), and trauma wounds. Alevicyn SG Antimicrobial Gel helps to relieve dry waxy skin by maintaining a moist wound & skin environment, which is beneficial to the healing process. OTC Indications for Use: Alevicyn SG Antimicrobial Gel is intended for OTC use to relieve itch and pain from minor skin irritations, minor cuts, exit sites, minor abrasions and minor burns, including sunburns. Hydrogel is also intended to moisten and lubricate absorbent wound dressings and moisten the wound bed. A moist wound and skin environment facilitates autolytic debridement and is beneficial to wound management and the healing process.
Device Story
Alevicyn SG Antimicrobial Gel is an opaque, low-viscosity, topical spray gel. It functions as a skin emollient and moisturizer to maintain a moist wound environment, facilitating autolytic debridement and relieving symptoms like pain, itching, and burning. The device is applied topically to the wound bed or skin; it can also be used to moisten absorbent wound dressings. It is intended for use by healthcare professionals (Rx) or patients (OTC) in clinical or home settings. The gel contains hypochlorous acid and sodium hypochlorite as preservatives. By maintaining moisture, it supports the natural healing process of various dermal conditions and wounds.
Clinical Evidence
Bench testing only. Testing included biocompatibility (cytotoxicity, irritation, sensitization per ISO 10993-1), toxicological risk assessment, and antimicrobial preservative effectiveness testing against Proteus mirabilis, Serratia marcescens, MRSA, and Acinetobacter baumannii. Physical characterization included pH, viscosity, and Free Available Chlorine (FAC) analysis.
Technological Characteristics
Opaque, semi-viscous spray gel. Ingredients: water, sodium lithium magnesium fluorosilicate, dimethicone, sodium phosphate, sodium bicarbonate, sodium chloride, sodium sulfate, hypochlorous acid, and sodium hypochlorite. Non-sterile, preserved per USP <51>. Delivered via PET bottle with finger pump sprayer.
Indications for Use
Indicated for patients with acute/chronic dermal lesions, dermatoses (atopic/radiation dermatitis, seborrhea), partial/full thickness wounds (burns, pressure/diabetic/stasis ulcers, surgical/trauma wounds), and minor skin irritations/burns. Suitable for Rx and OTC use.
Predicate Devices
- Alevicyn SG Antipruritic Gel (K152945)
- Puracyn Plus Antimicrobial Gel (K150799)
Reference Devices
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 1, 2017
Sonoma Pharmaceuticals, Inc. Brian Martin Director of Regulatory Affairs and Quality Control 1129 N. McDowell Boulevard Petaluma, California 94954
Re: K171935
Trade/Device Name: Alevicyn SG Antimicrobial Gel Regulatory Class: Unclassified Product Code: FRO Dated: November 2, 2017 Received: November 3, 2017
Dear Mr. Martin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171935
Device Name Alevicyn SG Antimicrobial Gel
### Indications for Use (Describe)
Rx Indications: Alevicyn SG Antimicrobial Gel is intended to moisten the wound bed and facilitate autolytic debridement of acute and chronic dermal lesions. Alevicyn SG Antimicrobial Gel is indicated to manage and relieve the burning, itching, erythema, scaling, and pain experienced with various types of dermatoses, including atopic dermatitis, radiation dermatitis, seborthea and sebortheic dermal irritation, sores, injuries, and ulcers of dermal tissue in addition to moistening and lubricating absorbent wound dressings. The Alevicyn SG Antimicrobial Gel may also be used to manage partial or full thickness wounds such as first and second degree burns, stage I - IV pressure ulcers, diabetic and stasis ulcers, abrasions and skin irritations, surgical wounds (donor and graft sites, incisions), and trauma wounds. Alevicyn SG Antimicrobial Gel helps to relieve dry waxy skin by maintaining a moist wound & skin environment, which is beneficial to the healing process.
OTC Indications for Use: Alevicyn SG Antimicrobial Gel is intended for OTC use to relieve itch and pain from minor skin irritations, minor cuts, exit sites, minor abrasions and minor burns, including sunburns. Hydrogel is also intended to moisten and lubricate absorbent wound dressings and moisten the wound bed. A moist wound and skin environment facilitates autolytic debridement and is beneficial to wound management and the healing process.
| Type of Use (Select one or both, as applicable) |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"> <span> <div style="width:20px; height:20px; border:1px solid black; display:flex; justify-content:center; align-items:center;"> <span> <svg height="10" width="10"> <line style="stroke:rgb(0,0,0);stroke-width:2" x1="0" x2="10" y1="0" y2="10"></line> <line style="stroke:rgb(0,0,0);stroke-width:2" x1="0" x2="10" y1="10" y2="0"></line> </svg> </span> </div> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> |
| <div style="display:flex; align-items:center;"> <span> <div style="width:20px; height:20px; border:1px solid black; display:flex; justify-content:center; align-items:center;"> <span> <svg height="10" width="10"> <line style="stroke:rgb(0,0,0);stroke-width:2" x1="0" x2="10" y1="0" y2="10"></line> <line style="stroke:rgb(0,0,0);stroke-width:2" x1="0" x2="10" y1="10" y2="0"></line> </svg> </span> </div> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Sonoma Pharmaceuticals. The logo consists of a geometric shape on the left and the company name on the right. The geometric shape is a stylized flower or star-like design with teal and gray colors. The company name "Sonoma" is written in gray, and the word "PHARMACEUTICALS" is written in teal below it.
# 510(k) SUMMARY
The following is a summary of 510(k) safety and effectiveness information in accordance with 21 CFR 807.92.
#### I. SUBMITTER
Sonoma Pharmaceuticals. Inc. 1129 North McDowell Blvd. Petaluma, CA 94954 Phone: (707) 283-0550 Fax: (707) 283-0551
Contact Person: Brian W. Martin, Director of Regulatory Affairs and Quality Control Date Prepared: November 28, 2017
#### II. DEVICE
Name of Device: Alevicyn SG Antimicrobial Gel Common or Usual Name: Wound Gel Classification Name: Dressing, Wound, Drug Regulatory Class: Unclassified Product Code: FRO
#### III. PREDICATE DEVICE
Alevicyn SG Antipruritic Gel (K152945) manufactured by Sonoma Pharmaceuticals, Inc. Puracyn Plus Antimicrobial Gel (K150799), manufactured by Innovacyn, Inc.
IV. REFERENCE DEVICE Alevicyn SG Gel (K143590) manufactured by Sonoma Pharmaceuticals, Inc.
#### DEVICE DESCRIPTION V.
The Alevicyn SG Antimicrobial Gel is an opaque gel, slightly chlorinated odor, low viscosity spray gel. The product will be supplied in polyethylene terephthalate (PET) round-bottles with polypropylene (PP) screw-top closure and a finger pump sprayer. Ingredients: Water, sodium lithium magnesium fluorosilicate, dimethicone, sodium phosphate, sodium bicarbonate, sodium chloride, sodium sulfate, with hypochlorous acid and sodium hypochlorite as gel preservatives.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Sonoma Pharmaceuticals. The logo consists of a circular design made up of interconnected teal and gray shapes on the left. To the right of the design is the word "Sonoma" in gray, with the word "PHARMACEUTICALS" in teal underneath.
#### VI. INDICATIONS FOR USE
Rx Indication: Alevicyn SG Antimicrobial Gel is intended to moisten the wound bed and facilitate autolytic debridement of acute and chronic dermal lesions. Alevicyn SG Antimicrobial Gel is indicated to manage and relieve the burning, itching, erythema, scaling, and pain experienced with various types of dermatoses, including atopic dermatitis, radiation dermatitis, seborrhea and seborrheic dermal irritation, sores, injuries, and ulcers of dermal tissue in addition to moistening and lubricating absorbent wound dressings. The Alevicyn SG Antimicrobial Gel may also be used to manage partial or full thickness wounds such as first and second degree burns, stage I -IV pressure ulcers, diabetic and stasis ulcers, abrasions and skin irritations, surgical wounds (donor and graft sites, incisions), and trauma wounds. Alevicyn SG Antimicrobial Gel helps to relieve dry waxy skin by maintaining a moist wound & skin environment, which is beneficial to the healing process.
OTC Indications: Alevicyn SG Antimicrobial Gel is intended for OTC use to relieve itch and pain from minor skin irritations, minor cuts, exit sites, minor lacerations, minor abrasions and minor burns, including sunburns. Hydrogel is also intended to moisten and lubricate absorbent wound dressings and moisten the wound bed. A moist wound and skin environment facilitates autolytic debridement and is beneficial to wound management and the healing process.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Sonoma Pharmaceuticals. The logo consists of a geometric shape on the left and the company name on the right. The geometric shape is made up of several curved lines that form a flower-like pattern. The company name is written in a sans-serif font, with the word "Sonoma" in a larger font than the word "Pharmaceuticals."
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE VII.
| Submitter/<br>Holder | Proposed Device:<br>Alevicyn SG Antimicrobial Gel,<br>Sonoma Pharmaceuticals | Predicate Device:<br>K152945 - Rx<br>Alevicyn SG Gel,<br>Sonoma Pharmaceuticals | Predicate Device:<br>K150799<br>Puracyn Plus Antimicrobial<br>HydroGel | Reference Device:<br>K143590 - OTC<br>Alevicyn SG Gel,<br>Sonoma Pharmaceuticals | |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Prescription<br>Indications<br>for<br>Use | Rx Indication: Alevicyn SG<br>Antimicrobial Gel is intended to<br>moisten the wound bed and facilitate<br>autolytic debridement of acute and<br>chronic dermal lesions. Alevicyn SG<br>Antimicrobial Gel is indicated to<br>manage and relieve the burning,<br>itching, erythema, scaling, and pain<br>experienced with various types of<br>dermatoses, including atopic<br>dermatitis, radiation dermatitis,<br>seborrhea and seborrheic dermatitis,<br>dermal irritation, sores, injuries, and<br>ulcers of dermal tissue in addition to<br>moistening and lubricating absorbent<br>wound dressings. The Alevicyn SG<br>Antimicrobial Gel may also be used to<br>manage partial or full thickness<br>wounds such as first and second<br>degree burns, stage I - IV pressure<br>ulcers, diabetic and stasis ulcers,<br>abrasions and skin irritations, surgical<br>wounds (donor and graft sites,<br>incisions), and trauma wounds.<br>Alevicyn SG Antimicrobial Gel helps<br>to relieve dry waxy skin by<br>maintaining a moist wound & skin<br>environment, which is beneficial to the<br>healing process. | Rx Indication: Under the<br>supervision of a healthcare<br>professional, Alevicyn SG Gel is<br>intended to manage and relieve<br>the burning, itching, erythema,<br>scaling, and pain experienced<br>with various types of dermatoses,<br>including atopic dermatitis,<br>radiation dermatitis, and<br>seborrhea and seborrheic<br>dermatitis. The Alevicyn SG Gel<br>may also be used to relieve the<br>pain of first and second degree<br>burns. Alevicyn SG Gel helps to<br>relieve dry, waxy skin by<br>maintaining a moist wound &<br>skin environment, which is<br>beneficial to the healing process. | Rx Indication: Puracyn® Plus<br>Antimicrobial Hydrogel<br>Professional Formula is<br>intended for use by healthcare<br>professionals to moisten the<br>wound bed and facilitate<br>autolytic debridement of acute<br>and chronic dermal lesions, as<br>indicated below. Puracyn®<br>Plus Antimicrobial Hydrogel<br>Professional Formula relieves<br>itch and pain associated with<br>dermal irritation, sores, injuries<br>and ulcers of dermal tissue in<br>addition to moistening and<br>lubricating absorbent wound<br>dressings. Puracyn® Plus<br>Antimicrobial Hydrogel<br>Professional Formula is<br>indicated for the management<br>of partial or full thickness<br>wounds such as 1st and 2nd<br>degree burns, stage I - IV<br>pressure ulcers, diabetic and<br>stasis ulcers, abrasions and skin<br>irritations, surgical wounds<br>(donor and graft sites,<br>incisions), trauma wounds, and<br>various dermatoses including<br>atopic dermatitis | Rx Indication: Alevicyn SG Gel<br>is intended to manage and relieve<br>the burning, itching and pain<br>experienced with various types of<br>dermatoses, including radiation<br>dermatitis and atopic dermatitis.<br>The Alevicyn SG Gel may also be<br>used to relieve the pain of first<br>and second degree burns.<br>Alevicyn SG Gel helps to relieve<br>dry waxy skin by maintaining a<br>moist wound & skin environment.<br>which is beneficial to the healing<br>process. | |
| Submitter/<br>Holder | Proposed Device:<br>Alevicyn SG Antimicrobial Gel, | Predicate Device:<br>K152945 - Rx | Predicate Device:<br>K150799 | Reference Device:<br>K143590 - OTC | |
| | Sonoma Pharmaceuticals | Alevicyn SG Gel,<br>Sonoma Pharmaceuticals | Puracyn Plus Antimicrobial<br>HydroGel | Alevicyn SG Gel,<br>Sonoma Pharmaceuticals | |
| OTC<br>Indications<br>for Use | OTC Indication: Alevicyn SG<br>Antimicrobial Gel is intended for OTC<br>use to relieve itch and pain from minor<br>skin irritations, minor cuts, exit sites,<br>minor lacerations, minor abrasions and<br>minor burns, including sunburns.<br>Hydrogel is also intended to moisten<br>and lubricate absorbent wound<br>dressings and moisten the wound bed.<br>A moist wound and skin environment<br>facilitates autolytic debridement and is<br>beneficial to wound management and<br>the healing process. | None | OTC Indication: Puracyn®<br>Plus Duo-Care™<br>Antimicrobial Wound & Skin<br>Hydrogel is intended for OTC<br>use to relieve itch and pain<br>from minor skin irritations,<br>minor cuts, exit sites, minor<br>lacerations, minor abrasions<br>and minor<br>burns, including sunburns.<br>Hydrogel is also intended to<br>moisten and lubricate<br>absorbent wound dressings and<br>moisten the wound bed. A<br>moist wound and skin<br>environment facilitates<br>autolytic debridement and is<br>beneficial to wound<br>management and the healing<br>process. | OTC Indication: Alevicyn SG Gel<br>is intended to relieve the burning<br>and itching associated with many<br>common types of skin irritation,<br>lacerations, abrasions, and minor<br>burns. Alevicyn SG Antipruritic<br>Gel is also indicated for the<br>management of irritation and pain<br>from minor burns, including<br>sunburn. | |
| Mechanism<br>of Action | Skin emollient and moisturizer | Same | Same | Same | |
| Delivery<br>System | Bottle, Spray Gel | Same | Bottle - Twist Top | Same | |
| Description | Appearance: Opaque Colorless semi<br>viscous gel<br>Odor: Slightly Chlorinated | Same | Same | Same | |
| Submitter/<br>Holder | Proposed Device:<br>Alevicyn SG Antimicrobial Gel,<br>Sonoma Pharmaceuticals | Predicate Device:<br>K152945 – Rx<br>Alevicyn SG Gel,<br>Sonoma Pharmaceuticals | Predicate Device:<br>K150799<br>Puracyn Plus Antimicrobial<br>HydroGel | Reference Device:<br>K143590 – OTC<br>Alevicyn SG Gel,<br>Sonoma Pharmaceuticals | |
| Ingredients | Water, sodium lithium magnesium<br>fluorosilicate, dimethicone, sodium<br>phosphate, sodium<br>bicarbonate, sodium chloride, sodium<br>sulfate, with hypochlorous acid and<br>sodium hypochlorite<br>as gel preservatives. | Same | Electrolyzed water,<br>Lithium Magnesium,<br>Sodium Silicate,<br>Phosphates,<br>Sodium Bicarbonate,<br>Sodium Chloride,<br>Sodium Sulfate,<br>Hypochlorous Acid<br>(0.01%),<br>Sodium Hypochlorite | Same | |
| | Sterility<br>Claim | Non-sterile<br>Preserved/Conforming to USP <51> | Same | Non Sterile | Same |
| | | Shelf Life | 24 months | Same | 2 years |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Sonoma Pharmaceuticals. The logo consists of a stylized flower-like design on the left, with the word "Sonoma" in a gray sans-serif font to the right. Below "Sonoma" is the word "PHARMACEUTICALS" in a smaller, teal-colored font. The flower-like design is made up of teal and gray shapes arranged in a circular pattern.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Sonoma Pharmaceuticals. The logo consists of a circular design on the left and the company name on the right. The circular design is made up of interlocking shapes in shades of gray and teal, forming a flower-like pattern around a central white space. To the right of the circular design, the word "Sonoma" is written in a gray, sans-serif font, with the word "PHARMACEUTICALS" written in a smaller, teal font directly below it.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Sonoma Pharmaceuticals. The logo consists of a geometric design on the left and the company name on the right. The geometric design is a circular shape made up of intersecting curved lines in gray and teal. The company name "Sonoma" is in gray, and the word "PHARMACEUTICALS" is in teal below it.
#### VIII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
# Biocompatibility Testing
The biocompatibility evaluation for the Alevicyn SG Antimicrobial Gel was conducted in accordance with the FDA biocompatibility guidance issued on June 16, 2016 and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1 : Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The Alevicyn SG Antimicrobial Gel was evaluated for cytotoxicity, irritation, sensitization. In addition, a toxicological risk assessment was conducted to support the prolonged use of the device.
# Bench Testing
The following tests were performed to support the performance of Alevicyn SG Antimicrobial Gel: visual inspection, pH, viscosity, Free Available Chlorine (FAC) and Antimicrobial Preservative Effectiveness Testing against the following bacteria in the gel while on the shelf: Proteus mirabilis, Serratia marcescens, antibiotic resistant Methicillin-Resistant Staphylococcus aureus (MRSA), and Acinetobacter baumannii. The Alevicyn SG Antimicrobial Gel meets specification and performance characteristics and is substantially equivalent to the predicate devices.
#### IX. CONCLUSION
Alevicyn SG Antimicrobial Gel is substantially equivalent in intended use, technological characteristics, safety and effectiveness to Alevicyn SG Gel (K152945), and Puracyn Plus Antimicrobial Gel (K150799). Therefore, the Alevicyn SG Antimicrobial Gel is substantially equivalent to the predicate devices.