K171928 · Med-Hot Thermal Imaging, Inc. · LHQ · Dec 8, 2017 · Obstetrics/Gynecology
Device Facts
Record ID
K171928
Device Name
Med-Hot Thermal Imaging Systems
Applicant
Med-Hot Thermal Imaging, Inc.
Product Code
LHQ · Obstetrics/Gynecology
Decision Date
Dec 8, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2980
Device Class
Class 1
Indications for Use
The Med-Hot Thermal Imaging Systems are intended to review, measure and record skin temperature patterns and variations emitted from the human body. They are intended for use as adjunctive diagnostic imaging for thermally significant indications in the regions of the head and neck, breast, chest, abdomen, back and extremities. The value of the value of these thermal patterns is determined by professional investigation. This device is intended for use by qualified technical personnel trained in its use.
Device Story
Non-contact infrared imaging system; captures skin temperature patterns via microbolometer camera (320x240 or 640x480 array). System includes infrared camera, TotalVision software, and computer. Used in clinical settings by trained technical personnel. Camera transmits data via Gigabit Ethernet to computer; software displays real-time thermal imagery and allows manual analysis of regions of interest (spot, line, square). Software performs no automated diagnostic interpretation; user visually interprets thermal patterns to infer clinical significance. Provides adjunctive data for professional investigation of thermally significant indications. Benefits include non-invasive visualization of thermal variations.
Clinical Evidence
No clinical data. Bench testing only, including software verification and validation, environmental testing (shock/vibration), EMC, and safety testing per ISO/IEC standards.
Technological Characteristics
Non-contact infrared imaging; microbolometer detector; 320x240 or 640x480 resolution; 50-60 Hz frame rate; Gigabit Ethernet interface. Factory calibration with firmware-based temperature conversion. Standards: ISO 14971 (risk management), IEC 60068-2-27/6 (shock/vibration), IEC 60529 (IP code), EN 61000-6-2/3 (EMC), EN 60950 (safety), 47 CFR Part 15 (RF).
Indications for Use
Indicated for adjunctive diagnostic imaging of skin temperature patterns in head, neck, breast, chest, abdomen, back, and extremities. Intended for use by trained technical personnel.
Regulatory Classification
Identification
A telethermographic system for adjunctive diagnostic screening for detection of breast cancer or other uses is an electrically powered device with a detector that is intended to measure, without touching the patient's skin, the self-emanating infrared radiation that reveals the temperature variations of the surface of the body. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized eagle. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Med-Hot Thermal Imaging, Inc. % E. J. Smith Consultant Smith Associates 1468 Harwell Avenue CROFTON MD 21114
December 8, 2017
Re: K171928
Trade/Device Name: Med-Hot Thermal Imaging Systems Regulation Number: 21 CFR 884.2980 Regulation Name: Telethermographic System Regulatory Class: I Product Code: LHO Dated: October 28, 2017 Received: November 2, 2017
Dear E. J. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
## Page 2 - E. J. Smith
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171928
#### Device Name
Med-Hot Thermal Imaging Systems
#### Indications for Use (Describe)
The Med-Hot Thermal Imaging Systems are intended to review, measure and record skin temperature patterns and variations emitted from the human body. They are intended for use as adjunctive diagnostic imaging for thermally significant indications in the regions of the head and neck, breast, chest, abdomen, back and extremities. The value of the value of these thermal patterns is determined by professional investigation. This device is intended for use by qualified technical personnel trained in its use.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement on last page.
{3}------------------------------------------------
# 510(k) Summary
| Sponsor: | Med-Hot Thermal Imaging, Inc. | |
|---------------------------|-------------------------------------------|--|
| Company Address: | 5120 S. Florida Ave., Suite 301 | |
| | Lakeland, FL 33813 | |
| Telephone: | 863-646-1599 | |
| Fax: | 863-646-1544 | |
| Contact Person: | Carol Chandler | |
| Summary Preparation Date: | December 6, 2017 | |
| Device Name: | | |
| Trade Name: | Med-Hot Thermal Imaging Systems | |
| Common/Usual Name: | System, Telethermographic (Adjunctive Use | |
| Classification Name: | System, Telethermographic (Adjunctive Use | |
| Regulation Number: | 21 CFR 884.2980 | |
| Device Class: | Class I | |
### Predicate Device:
| Manufacturer | Product Name | 510(k) Number |
|-------------------------------|--------------|---------------|
| Med-Hot Thermal Imaging, Inc. | MTI 2000 | K063047 |
### Device Description:
The Med-Hot Thermal Imaging Systems are non-contact infrared imaging devices, with all functions controlled at the computer screen. TotalVision is a patented, clinical personnelfriendly software application, validated in the field. The system is delivered with a computer including installed and tested software.
The Med-Hot Thermal Imaging camera is suitable for high quality imaging of the human body:
- . The resolution (320 X 240 / 640 X 480) provides adequate detail to detect and visualize minute thermal details
- . Microbolometer technology provides low power consumption and high temperature reliability
- 50-60 Hz image frequency provides real-time image viewing and capture with no loss of detail due to movement.
- Automatic focus option to provide privacy imaging, a comfort feature for both client and
{4}------------------------------------------------
user
- . Gigabit Ethernet interface-fastest industry standard for data transfer of data dense files
- . Factory calibration - temperature conversion files reside in the camera firmware, not in the user's computer, providing reliable, enhanced accuracy. See explanation below.
- . 25 degree standard lens provides the most practical field of view in a clinical setting with limited space.
The TotalVision Capture Software allows for incoming image data from the camera head and configures that data in a form that can be displayed on the computer screen. This display will include an image of the scene within the camera's field of view allowing the user to visualize thermal patterns or analyze the image in terms relative temperature values. The software makes no determination regarding what the thermal patterns or relative temperature values mean. It will be left to the user to infer areas of interest based on his/her visual interpretation of those patterns or values.
The new device consists of:
- 1. Infrared Camera
- ▲ MAX 076 320 x 240 array detector
- レ MAX 307 640 x 480 array detector
- 2. TotalVision Software
- 3. Laptop or desktop computer
### Indications for Use:
The Med-Hot Thermal Imaging Systems are intended to review, measure and record skin temperature patterns and variations emitted from the human body. They are intended for use as adjunctive diagnostic imaging for thermally significant indications in the regions of the head and neck, breast, chest, abdomen, back and extremities. The significance of the value of these thermal patterns is determined by professional investigation. This device is intended for use by qualified technical personnel trained in its use.
### Predicate Device Comparison:
Synopsis of the comparison analysis:
- . Med-Hot Thermal Imaging Systems and the predicate device (MTI 2000) are intended for use as an adjunct to other clinical diagnostic procedures in the diagnosis, quantifying, screening of differences in skin surface temperature changes.
- . Both view live imagery from the camera, record snap shots to disk, include postprocessing of fast thermal events, provide analysis by region of interest (spot, line, square) and save the information in a file organizer.
- . The Med-Hot TotalVision Capture software is an evolution of the original Med-Hot software program included in K063047. Improvements and new features added were based on user-driven needs and requests over the years.
{5}------------------------------------------------
### Comparison of Intended Use and Principle of Operation:
As thermal sensors, the principle operation of all thermal imaging or infrared cameras is inherently the same, regardless of technical specifications or applications. They capture and record objects within the camera view for evaluation of thermal variations.
For clinical or medical application, the information from and infrared device is strictly limited to thermal findings. The location of these findings on the body, combined with intensity and distribution of thermal patterns may lead the reader/observer to intentionally or incidentally make a thermal correlation with the expected or abnormal patterns to the anatomy in that region of the body.
The Med-Hot Thermal Imaging Systems is substantially equivalent in intended use, principle of operation and similar software features to the FDA cleared Med-Hot Thermal Imaging, Inc. MTI2000.
### Non-Clinical Performance Testing:
- . ISO 14971, Medical devices - application of risk management to medical Devices
- IEC 60068-2-27 Environmental testing - Part 2-27: Tests - Test Ea and guidance: Shock
- IEC 60068-2-6 Environmental testing - Part 2-6: Tests - Test Fc: Vibration
- IEC 60529 Degrees of protection provided by enclosures (IP Code)
- EN 61000-6-3 Electromagnetic compatibility (EMC). Generic standards. Emission standard for residential, commercial and light-industrial environments
- EN 61000-6-2 Electromagnetic compatibility (EMC). Generic standards. Immunity for industrial environments
- EN 60950 Information technology equipment - Safety - Part 1: General requirements
- 47 CFR Part 15 RADIO FREQUENCY DEVICES ●
- . Software Verification and validation according to FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
### Clinical Studies:
No clinical studies were conducted.
### Conclusion:
Based upon the technology characteristics and safety and performance testing, it is the conclusion of Med-Hot Thermal Imaging Inc., that the Med-Hot Thermal lmaging Systems are as safe and effective as the predicate devices and raises no new issues of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.