The uMEC Series Patient Monitor family consists of the uMEC7, uMEC10, uMEC12, uMEC15 and uMEC15S patient monitors. All of the devices in the family are multiparameter monitors indicated for monitoring, displaying, reviewing, and transferring multiple physiological parameters. The specific indications for use are provided below. The uMEC series patient monitors are intended for monitoring, displaying, and transferring of multiple physiological parameters including ECG (3-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, QT analysis, and heart rate (HR)), respiration rate (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure(IBP), cardiac output (C.O.), carbon dioxide (CO2). All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following: The arrhythmia detection and SpO2 are intended for adult and pediatric patients only; ST Segment analysis is intended for adult patients only; and C.O. monitoring is restricted to adult patients only. The monitor is to be used in healthcare facilities by clinical professionals or under their guidance.
Device Story
Multiparameter patient monitors for healthcare facilities; operated by clinical professionals. Inputs: ECG leads, SpO2 sensors, temperature probes, NIBP cuffs, IBP transducers, C.O. sensors, CO2 sampling lines. Transforms physiological signals into real-time waveforms, numeric values, and alarms. Outputs: Integrated touchscreen display, thermal recorder, and audible/visual alarms. Assists clinicians in patient status assessment and decision-making. Benefits: Continuous monitoring of vital signs to detect physiological changes.
Clinical Evidence
Bench testing only. Functional and system-level testing performed to verify performance against specifications and consensus standards (e.g., IEC 60601 series, AAMI/ANSI EC57). No clinical data provided.
Technological Characteristics
Multiparameter monitors with integrated touchscreen (10.4" to 15"). Power: AC or rechargeable Li-ion battery. Connectivity: Data transfer capabilities. Standards: ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-49, IEC 60601-2-27, IEC 60601-2-34, ISO 80601-2-56, ISO 80601-2-61, ISO 80601-2-55, AAMI/ANSI EC57. Software: Mindray proprietary algorithms for ECG, arrhythmia, ST, QT, and physiological parameter processing.
Indications for Use
Indicated for adult, pediatric, and neonatal patients for monitoring, displaying, and transferring physiological parameters including ECG, Resp, Temp, SpO2, PR, NIBP, IBP, C.O., and CO2. Arrhythmia detection and SpO2 limited to adult/pediatric; ST segment analysis and C.O. limited to adult patients only.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
November 15, 2017
Shenzhen Mindray Bio-Medical Electronics Co., Ltd Yanhong Bai Manager Regulatory Affairs Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan Shenzhen, 518057 CN
Re: K171901
Trade/Device Name: uMEC Series Patient Monitors (including uMEC6, uMEC7, uMEC12, uMEC15, uMEC15S) Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including St-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MHX, DSI, MLD, DRT, DXN, DSK, FLL, DQA, CCK, DXG, DSJ, DRS Dated: October 11, 2017 Received: October 16, 2017
Dear Yanhong Bai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
M& Millelreme
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K171901
Device Name
uMEC Series Patient Monitors (including uMEC6, uMEC7, uMEC12, uMEC12, uMEC15, uMEC15S)
Indications for Use (Describe)
The uMEC Series Patient Monitor family consists of the uMEC7, uMEC10, uMEC12, uMEC15 and uMEC15S patient monitors. All of the devices in the family are multiparameter monitors indicated for monitoring, displaying, reviewing, and transferring multiple physiological parameters. The specific indications for use are provided below.
The uMEC series patient monitors are intended for monitoring, displaying, and transferring of multiple physiological parameters including ECG (3-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, QT analysis, and heart rate (HR)), respiration rate (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure(IBP), cardiac output (C.O.), carbon dioxide (CO2).
All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following: The arrhythmia detection and SpO2 are intended for adult and pediatric patients only;
ST Segment analysis is intended for adult patients only; and
C.O. monitoring is restricted to adult patients only.
The monitor is to be used in healthcare facilities by clinical professionals or under their guidance.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the uMEC series patient monitors is provided below.
| Device Common Name: | Patient Monitor |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade Name: | uMEC Series Patient Monitors(including uMEC6, uMEC7, uMEC10, uMEC12, uMEC15, uMEC15S) |
| Applicant: | SHENZHEN MINDRAY BIO-MEDICAL<br>ELECTRONICS CO., LTD.<br>Mindray Building, Keji 12th Road South<br>High-tech Industrial Park, Nanshan<br>Shenzhen 518057, P.R. China<br>Tel: +86 755 81888998<br>Fax: +86 755 26582680 |
| Contact: | Yanhong Bai<br>Manager Regulatory Affairs<br>SHENZHEN MINDRAY BIO-MEDICAL<br>ELECTRONICS CO., LTD.<br>Mindray Building, Keji 12th Road South<br>High-tech Industrial Park, Nanshan 518057, P.R. China<br>Tel: +86 755 81885635<br>Fax: +86 755 26582680<br>E-mail: baiyanhong@mindray.com |
| Date Prepared: | June 23, 2017 |
| Classification Regulation: | 870.1025 - Arrhythmia detector and alarm (including<br>ST-segment measurement and alarm) |
| | Class II |
| Panel: | Cardiovascular |
| Product Code: | MHX - monitor, physiological, patient (with arrhythmia<br>detection or alarms) |
{4}------------------------------------------------
### Subsequent Product Codes:
| Product<br>Code | Regulation<br>Number | Panel | Regulation description | Device Common Name |
|-----------------|----------------------|----------------|----------------------------------------------------------------------------|--------------------------------------------------------------|
| DSI | 21 CFR<br>870.1025 | Cardiovascular | Arrhythmia detector and alarm (including ST-segment measurement and alarm) | detector and alarm,<br>arrhythmia |
| MLD | 21 CFR<br>870.1025 | Cardiovascular | Arrhythmia detector and alarm (including ST-segment measurement and alarm) | monitor, st segment with<br>alarm |
| DRT | 21 CFR<br>870.2300 | Cardiovascular | Cardiac Monitor (including cardiotachometer and rate alarm) | monitor, cardiac (incl.<br>cardiotachometer & rate<br>alarm) |
| DXN | 21 CFR<br>870.1130 | Cardiovascular | Noninvasive blood pressure measurement system | system, measurement<br>, blood-pressure, |
| DSK | 21 CFR<br>870.1110 | Cardiovascular | Blood pressure computer | computer, blood-pressure |
| FLL | 21 CFR<br>880.2910 | Cardiovascular | Clinical electronic thermometer | thermometer, electronic,<br>clinical |
| DQA | 21 CFR<br>870.2700 | Cardiovascular | Oximeter | oximeter |
| CCK | 21 CFR<br>868.1400 | Anesthesiology | Carbon dioxide gas analyzer | analyzer, gas,<br>carbon-dioxide, |
| DXG | 21 CFR<br>870.1435 | Cardiovascular | Single-function, preprogrammed diagnostic | computer, diagnostic,<br>pre-programmed, |
| DSJ | 21 CFR<br>870.1100 | Cardiovascular | Blood pressure alarm | alarm, blood-pressure |
| DRS | 21 CFR<br>870.2850 | Cardiovascular | transducer, blood-pressure, extravascular | Extravascular blood<br>pressure transducer. |
Predicate Device: K152902 –Passport Series Patient Monitors
# Indication for Use:
The uMEC Series Patient Monitor family consists of the uMEC6, uMEC7, uMEC10, uMEC12, uMEC15 and uMEC15S patient monitors. All of the devices in the family are multiparameter monitors indicated for monitoring, displaying, reviewing, and transferring multiple physiological parameters. The specific indications for use are provided below.
The uMEC series patient monitors are intended for monitoring, displaying, reviewing, and K171901
{5}------------------------------------------------
transferring of multiple physiological parameters including ECG (3-lead , 5-lead or 12-lead selectable, arrhythmia detection, ST segment analysis, and heart rate (HR)), respiration rate (Resp), temperature (Temp), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure(IBP), cardiac output (C.O.), carbon dioxide (CO2).
All the parameters can be monitored on single adult, pediatric, and neonatal patients with the exception of the following:
The arrhythmia detection and SpO2 are intended for adult and pediatric patients only; ST Segment analysis is intended for adult patients only;
C.O. monitoring is restricted to adult patients only;
The monitor is to be used in healthcare facilities by clinical professionals or under their guidance.
### Device Description:
The subject uMEC Series Patient Monitors include six monitors:
- uMEC6 Patient Monitor ●
- . uMEC7 Patient Monitor
- . uMEC10 Patient Monitor
- uMEC12 Patient Monitor
- uMEC15 Patient Monitor
- uMEC15S Patient Monitor ●
The uMEC Series Patient Monitors are Mindray's new generation monitoring product family with ergonomic and flexible design in platform of both software and hardware to meet the clinical needs of monitoring.
#### Performance Data:
- To establish the substantial equivalence of the uMEC Series Patient Monitors, Mindray conducted functional and system level testing on the subject device. The testing provided an evaluation of the performance of the device relevant to each of the differences between the subject device and the predicate device. The functional and system level testing showed that the devices continue to meet specifications and the performance of the device is equivalent to the predicate.
- Mindray has followed the following FDA Guidance Documents relevant to this device: .
- Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm Guidance for Industry and FDA (10/28/2003)
- Cardiac Monitor Guidance (Including Cardiotachometer and Rate Alarm)
- . Mindray has conducted testing to ensure the subject device meets relevant consensus standards.
{6}------------------------------------------------
ANSVAAMI ES60601-1:2005/(R)2012 and A1:2012,C1:2009/(R)2012 and A2:2010/(R)2012 Medical electrical equipment--Part 1:General requirements for basic safety and essential performance
IEC 60601-1-2:2007 Medical electrical equipment--Part 1-2: General requirements for basic safety and essential performance-- Collateral standard: Electromagnetic compatibility--Requirements and tests
IEC 60601-1-8:2012 Medical electrical equipment - part 1-8: general requirements for basic safety and essential performance - collateral standard: general requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems
IEC 60601-1-6: 2013 Medical electrical equipment-part 1-6: general requirements for basic safety and essential performance--collateral standard: usability
IEC 60601-2-49:2011 Medical electrical equipment --Part 2-49: Particular requirements for the basic safety and essential performance of multifunction patient monitoring equipment
IEC 60601-2-27:2011 Medical electrical equipment--Part 2-27: Particular requirements for the basic safety and essential performance of electrocardiographic monitoring equipment
IEC 60601-2-34:2011 Medical electrical equipment - part 2-34: particular requirements for the basic safety, including essential performance, of invasive blood pressure monitoring equipment
IEC 80601-2-30:2013 Medical electrical equipment - part 2-30: particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
ISO 80601-2-56: 2009 Medical electrical equipment - part 2-56: particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
ISO 80601-2-61: 2011 Medical electrical equipment - part 2-61: particular requirements for basic safety and essential performance of pulse oximeter equipment
ISO 80601-2-55: 2011 Medical electrical equipment - part 2-55: particular requirements for the basic safety and essential performance of respiratory gas monitors
AAMI / ANSI EC57:2012 testing and reporting performance results of cardiac rhythm and stsegment measurement algorithms
IEC 62133 : 2012 secondary cells and batteries containing alkaline or other non-acid electrolytes safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications
# Comparison of Indications:
Both the predicate device and the subject uMEC Series Patient Monitors are multiparameter patient monitors intended to be used in healthcare facilities under the direction of clinical professionals. Some features present in the cleared Passport Series Patient Monitors are not present in uMEC Series Patient Monitors. In conclusion, the minor differences to the indications K171901
{7}------------------------------------------------
for use do not change the fundamental intended use of the uMEC series Patient Monitors as multiparameter monitors.
# Comparison of Technological Characteristics:
The technological characteristics of the subject device are compared to the predicate device in the table below.
| | Predicate Device (K152902) | Subject Devices | | | | | |
|------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|------------------------|
| Feature | Passport 17m | uMEC6 | uMEC10 | uMEC7 | uMEC12 | uMEC15 | uMEC15S |
| Integrated<br>display and<br>touchscreen | 17" 1280*1024 pixels | 10.4"<br>800*600<br>pixels | 10.4"<br>800*600 pixels | 12.1"<br>800*600 pixels | 12.1"<br>800*600 pixels | 15"<br>1024*768<br>pixels | 15"<br>1024*768 pixels |
| Power<br>supply | Two rechargeable Lithium-ion battery<br>or AC power supply | One standard Rechargeable Lithium-ion battery or One optional<br>Rechargeable Lithium-ion battery or<br>AC power supply | | | | | |
| Battery | Chargeable Lithium-Ion, 11.1 VDC,<br>4500 mAh | Chargeable Lithium-Ion, 11.1 VDC, 2500 mAh (Standard)<br>Chargeable Lithium-Ion, 11.1 VDC, 5000 mAh (Optional ) | | | | | |
| Data<br>Recorder | The thermal recorder records patient<br>information, measurement numerics,<br>up to three waveforms, etc. | Same | | | | | |
| Speaker | Give alarm tones (45 to 85 dB), key<br>tones, QRS tones; support PITCH<br>TONE and multi-level tone<br>modulation | Same | | | | | |
| ECG | 3-lead , 5-lead or 12-lead selectable,<br>arrhythmia detection, ST segment<br>analysis, QT analysis, an interpretation<br>of resting 12-lead ECG, and heart rate<br>(HR);<br>Support Mindray and Mortara<br>algorithm. | 3-lead , 5-lead or 12-lead selectable, arrhythmia detection, ST<br>segment analysis, QT analysis andheart rate (HR);<br>Only support Mindray algorithm and don't support interpretation<br>of resting 12-lead ECG | | | | | |
| ECG<br>measurement<br>range<br>and<br>Accuracy | HR: Neonate: 15 to 350 bpm;<br>Pediatric: 15 to 350 bpm;<br>Adult: 15 to 300 bpm;<br>±1 bpm or ±1%, whichever is greater.<br><br>ST: -2.0 to 2.0 mV RTI;<br>-0.8 to 0.8 mV: ±0.02 mV or ±10%,<br>whichever is greater.<br>Beyond this range: Not specified.<br><br>QT: [200, 800]ms; ±30ms | Same | | | | | |
| Arrhythmia<br>Analysis | Asystole, VFib/VTac, Vtac, Vent.<br>Brady, ExtremeTechy, Extreme<br>Brady, PVC, Couplet, Bigeminy,<br>Trigeminy, R on T, RunPVCs,PVCs,<br>Tachy, Brady, Missed Beats, Vent.<br>Rhythm, PNP, PNC, Multif.<br>PVC,Nonsus. Vtac, Pause, Irr.<br>Rhythm, Afib | Same | | | | | |
| | Predicate Device (K152902) | Subject Devices | | | | | |
| Feature | Passport 17m | uMEC6 | uMEC1<br>0 | uMEC<br>7 | uMEC1<br>2 | uMEC1<br>5 | uMEC15<br>S |
| Respiration<br>rate (Resp) | Measurement range: Adult: 0 to 120<br>rpm; Pediatric, neonate: 0 to 150 rpm.<br>Accuracy: 7 to 150 rpm: ±2 rpm or<br>±2%, whichever is greater;<br>0 to 6 rpm: Not specified. | Same | | | | | |
| Temperatur<br>e (Temp) | Measurement range: 0 to 50 °C (32 to<br>122 F)<br>Accuracy: ±0.1 °C or ±0.2 F (without<br>probe) | Same | | | | | |
| Pulse<br>oxygen<br>saturation<br>(SpO2) | MindraySpO2 Module<br>Measurement range: 0 to 100%<br>Accuracy:70 to 100%: ±2%<br>(adult/pediatric mode)<br>70 to 100%: ±3% (neonate mode)<br>0% to 69%: Not specified.<br>MasimoSpO2 Module<br>Measurement range: 1 to 100%<br>Accuracy:70 to 100%: ±2% (measured<br>without motion in adult/pediatric<br>mode)<br>70 to 100%: ±3% (measured without<br>motion in neonate mode)<br>70 to 100%: ±3% (measured with<br>motion)<br>1% to 69%: Not specified.<br>NellcorSpO2 Module<br>Measurement range: 0 to 100%<br>Accuracy:70 to 100%: ±2%<br>(adult/pediatric)<br>70 to 100%: ±3% (neonate)<br>0% to 69%: Not specified. | MindraySpO2 Module<br>Measurement range: 0 to 100%<br>Accuracy:70 to 100%: ±2% (adult/pediatric mode)<br>0% to 69%: Not specified.<br>MindraySpO2 Module do not support neonate mode and the<br>specifications for Adult/pediatric are same<br>Not Support Masimo and NellcorSpO2 Module | | | | | |
| Pulse rate<br>(PR) | PR from MindraySpO2 Module<br>Measurement range: 20 to 254 bpm<br>Accuracy: ±3 bpm<br>PR from MasimoSpO2 Module<br>Measurement range: 25 to 240 bpm<br>Accuracy: ±3 bpm (measured without<br>motion)<br>±5 bpm (measured with motion)<br>PR from NellcorSpO2 Module<br>Measurement range: 20 to 300 bpm<br>Accuracy:20 to 250 bpm: ±3 bpm<br>251 to 300 bpm, not specified<br>PR from IBP Module<br>Measurement range: 25 to 350 bpm<br>Accuracy: ±1 bpm or ±1%, whichever<br>is greater | PR from<br>MindraySpO2<br>Module<br>Measurement<br>range: 20 to 254<br>bpm<br>Accuracy:±3 bpm<br>PR specifications<br>from MindraySpO2<br>module are same<br>Not Support<br>Masimo and<br>NellcorSpO2<br>Module | | PR from MindraySpO2 Module<br>Measurement range: 20 to 254 bpm<br>Accuracy: ±3 bpm | PR from IBP Module<br>Measurement range: 25 to 350 bpm<br>Accuracy: ±1 bpm or ±1%, whichever is<br>greater | PR specifications from MindraySpO2<br>module and IBP Module are same<br>Not Support Masimo and NellcorSpO2<br>Module | |
{8}------------------------------------------------
{9}------------------------------------------------
| | Predicate Device (K152902) | | | Subject Devices | | | | | |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|-----------|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|------------|------------|-------------|--|
| Feature | Passport 17m | | | uMEC6<br>0 | uMEC1<br>7 | uMEC1<br>2 | uMEC1<br>5 | uMEC15<br>S | |
| Non-<br>invasive<br>blood<br>pressure<br>(NIBP) | Measurement range: | | | Measurement range: | | | | | |
| | | Adult | Pediatric | Neonate | | Adult | Pediatric | Neonate | |
| | Systolic | 40-270 | 40-200 | 40-135 | | 25-290 | 25-240 | 25-140 | |
| | Diastolic | 10-210 | 10-150 | 10-100 | | 10-250 | 10-200 | 10-115 | |
| | Mean | 20-230 | 20-165 | 20-110 | | 15-260 | 15-215 | 15-125 | |
| | Accuracy:<br>Max mean error: ±5 mmHg<br>Max standard deviation: 8 mmHg | | | | Accuracy:<br>Max mean error: ±5 mmHg<br>Max standard deviation: 8 mmHg<br>The measurement range of subject devices is wider than the<br>predicate device | | | | |
| Invasive<br>blood<br>pressure<br>(IBP) | Measurement range: -50 to 300 mmHg<br>Accuracy: ±2% or ±1 mmHg,<br>whichever is greater (without sensor) | | | Not Supported | Same as the predicate device | | | | |
| Cardiac<br>output<br>(C.O.) | Measurement range:<br>C.O: 0.1 to 20 L/min<br>TB: 23 to 43 °C<br>TI: 0 to 27 °C<br>Accuracy<br>C.O: ±5% or ±0.1 L /min, whichever<br>is greater<br>TB, TI: ±0.1 °C (without sensor) | | | Not Supported | Same as the predicate device | | | | |
{10}------------------------------------------------
| | Predicate Device (K152902) | Subject Devices | | | | | |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|------------|------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|-------------|
| Feature | Passport 17m | uMEC6 | uMEC1<br>0 | uMEC1<br>7 | uMEC1<br>2 | uMEC1<br>5 | uMEC15<br>S |
| Carbon<br>dioxide<br>(CO2) | Sidestream CO2 Module:<br>Measurement range0 - 13% Accuracy:<br>0-5%: ±0.26% abs<br>5-10%: ±5% of the reading<br>10-13%: ±10% of the reading<br>±0.3% ABS increase in error for ISO<br>accuracy mode<br>awRR measurement:<br>awRR measurement range: 0 to<br>120rpm<br>awRR measurement accuracy: ±2rpm | Not Supported | | | Sidestream CO2 Module:<br>Measurement range:0 to 20%<br>Accuracy:<br>CO2 concentration<1%:±0.1%<br>1%≤CO2 concentration<5%: ±0.2%<br>5%≤CO2<br>concentration<7%±0.3%7%≤CO2<br>concentration<12% ±0.4%<br>12%≤CO2 concentration≤13%±0.5%<br>13%<CO2<br>concentration≤20%±(0.43%+8%rel)<br><br>±0.3% ABS increase in error for ISO<br>accuracy mode<br><br>awRR measurement:<br>awRR measurement range: 0 to 150rpm<br>awRR measurement accuracy: <60rpm:<br>±1rpm<br>60 to 150 rpm: ±2rpm<br><br>The CO2 measurement range of subject<br>devices is wider than thepredicate device<br>The awRR measurement range of subject<br>devices is wider than thepredicate device.<br>The Accuracy of awRR is also improved | | |
### Substantial Equivalence Conclusion:
Based on the detailed comparison between the previously cleared Passport Series devices (K152902) and the subject uMEC series devices, the performance testing and conformance with applicable standards, the uMEC Series Patient Monitors (including uMEC6, uMEC7, uMEC10, uMEC12, uMEC15, uMEC15S) can be found substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.