The intended use of the Portable Hyperbaric Chamber Revitalair® 430F is to treat acute mountain sickness under the prescription of a health professional. The medical device is designed for use at physician offices and health institutions.
Device Story
Portable hyperbaric chamber for low-pressure therapy (max 1.4 ATA); encloses patient within pressure boundary. Chamber constructed of polyester-based plastic fabric with aluminum rings; includes 10 transparent windows for patient monitoring. Operated via electric oil-less compressor; manual valves control inflation/pressure. Safety relief valves activate above 1.3 ATA. Used in physician offices and health institutions; operated by healthcare professionals or patients (internal/external controls). Provides hyperbaric environment to treat acute mountain sickness; enables visual monitoring of patient comfort. Benefits patient by providing controlled pressure environment for symptom relief.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including electrical safety (IEC 60601-1), electromagnetic disturbance (IEC 60601-1-2), biocompatibility (ISO 10993, ISO 18562), and pressure testing (ASME PVHO-1).
Technological Characteristics
Portable hyperbaric chamber; polyester-based plastic fabric with aluminum rings; 900mm x 1850mm. Operating pressure max 1.4 ATA. Electric oil-less compressor. Relief valves (metal/plastic) set at 1.3 ATA. 10 transparent windows. Complies with ASME PVHO 1-2012. Biocompatibility per ISO 10993/18562. Electrical safety per IEC 60601-1; EMD per IEC 60601-1-2.
Indications for Use
Indicated for treatment of acute mountain sickness in patients with medical conditions related to altitude sickness. Contraindicated for patients with colds or flu symptoms, recent alcohol consumption, blocked ear canals, blocked sinuses, otic barotraumas, excessive CO2 exposure, pulmonary hyper-expansion, or decompression sickness.
Regulatory Classification
Identification
A hyperbaric chamber is a device that is intended to increase the environmental oxygen pressure to promote the movement of oxygen from the environment to a patient's tissue by means of pressurization that is greater than atmospheric pressure. This device does not include topical oxygen chambers for extremities (§ 878.5650).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Oxavita S.R.L. % Diane Sudduth Sr. RA/QA Consultant Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin, Texas 78746
Re: K171899
Trade/Device Name: Revitalair 430F Regulation Number: 21 CFR 868.5470 Regulation Name: Hyperbaric Chamber Regulatory Class: Class II Product Code: CBF Dated: April 10, 2019 Received: April 11, 2019
Dear Diane Sudduth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171899
Device Name Revitalair® 430F
Indications for Use (Describe)
The intended use of the Portable Hyperbaric Chamber Revitalair® 430F is to treat acute mountain sickness under the prescription of a health professional. The medical device is designed for use at physician offices and health institutions.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# Revitalair® 430F
# K171899
#### 1. Submission Sponsor
Oxavita S.R.L.
Tamborini 5813
Buenos Aires, Argentina
Contact: Eduardo Northing
Title: Exports and Logistics
#### 2. Submission Correspondent
Emergo Global Consulting, LLC
2500 Bee Cave Road
Building 1, Suite 300
Austin, TX 78746
USA Office Phone: (512) 327.9997
Contact: Dr. Diane Sudduth
Title: Senior Consultant, RA
#### 3. Date Prepared
11/22/19
#### 4. Device Identification
| Trade/Proprietary Name: | Revitalair® 430F |
|-------------------------|-----------------------------|
| Common/Usual Name: | Portable Hyperbaric Chamber |
| Classification Name: | Chamber, Hyperbaric |
| Regulation Number: | 21 CFR 868.5470 |
| Product Code: | CBF |
{4}------------------------------------------------
Device Class: Class II
Classification Panel: Anesthesiology Devices
## 5. Legally Marketed Predicate Device(s)
K072757, The Dive HD250 and The Grand Dive HD400, SUMMIT TO SEA
### 6. Indication for Use Statement
The intended use of the Portable Hyperbaric Chamber Revitalair® 430F is to treat acute mountain sickness under the prescription of a health professional. The medical device is designed for use at physician offices and health institutions.
### 7. Device Description
The Portable Hyperbaric Chamber Revitalair® 430F is a hyperbaric chamber for low pressures (operating at pressures of no greater than 1.4 atmospheres absolute (ATA)). The operational design pressure of a hyperbaric chamber that encloses a human within its pressure boundary falls within the scope of the American Society of Mechanical Engineers Pressure Vessels Human Occupancy 1 (ASME PVHO 1-2012).
Revitalair® 430F consists of 2 parts, the cabin or chamber and the compression system or compressor's cabinet. The chamber, weighing 62 pounds, is constructed of an airtight polyester-based plastic fabric joined by aluminum rings, forming a cylinder 900 mm in diameter and 1850 mm in length (Figure 1).
The chamber is inflated with atmospheric air through an electric compressor. Fittings allow the chamber to be connected to compressed air by means of manually controlled valves.
The safety or relief valves are operated at pressures above 1.3 ATM which ensures safe operations. The compressor has an additional safety valve for any obstruction of the supply hose from the compressor to the chamber.
The Revitalair® 430F has 10 transparent windows to let in light, 360º viewing and enable easy verification of the patient's comfort from the outside.
The Revitalair® 430F can be operated from the interior as well as from the exterior.
After folding it up, the Revitalair® 430F is placed in its transportation box or in the optional carrying case.
### 8. Substantial Equivalence Discussion
The following table compares the Portable Hyperbaric Chamber Revitalair® 430F (abbreviated as Revitalair® 430F) to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance testing.
{5}------------------------------------------------
| Manufacturer | SUMMIT TO SEA | SUMMIT TO SEA | Oxavita S.R.L. | Significant<br>Difference |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Trade Name | The Grand Dive<br>HD400 | The Dive HD250 | Revitalair® 430F | |
| 510(k) | K072757 | K072757 | | |
| Product Code | CBF | CBF | CBF | Same |
| Regulation<br>Number | 868.5470 | 868.5470 | 868.5470 | Same |
| Regulation Name | Hyperbaric<br>Chamber | Hyperbaric<br>Chamber | Hyperbaric Chamber | Same |
| Intended Use | Acute Mountain<br>Sickness AMS and<br>its associated mild<br>symptoms | Acute Mountain<br>Sickness AMS and<br>its associated mild<br>symptoms | Acute Mountain<br>Sickness AMS and its<br>associated mild<br>symptoms | Same |
| Prescription -<br>Rx Only | Yes | Yes | Yes | Same |
| Target Population | People with medical<br>conditions by<br>altitude sickness | People with<br>medical conditions<br>by altitude sickness | People with medical<br>conditions by<br>altitude sickness | Same |
| Places of Use | Homes, gyms,<br>offices, outdoor,<br>hospitals and clinics | Homes, gyms,<br>offices, outdoor,<br>hospitals and clinics | Hospitals and clinics | Similar |
| Weight | 32 lb | 15 lb | 62 lb | Different |
| Measures<br>Diameter<br>×<br>Length | 39.37in x 106.29in<br>(1000mm x<br>2700mm) | 26.6in x 82.7in<br>(675mm x<br>2100mm) | 35.4in x 74.8in<br>(900mm x 1900mm) | Similar |
| Windows | 4 | 2 | 10 | Different |
| Transport Belts | 4 | 4 | 2 | Different |
| Relief Valves | Yes | Yes | Yes, Metal | Same |
| Manufacturer | SUMMIT TO SEA | SUMMIT TO SEA | Oxavita S.R.L. | Significant |
| Trade Name | The Grand Dive<br>HD400 | The Dive HD250 | Revitalair® 430F | Difference |
| Drain Valves | Yes, integrated<br>emptying and filling | Yes, integrated<br>emptying and filling | Yes, 2 units only for<br>emptying | Same |
| Operating<br>Pressure | 4psi / 1.3 ATA | 4psi / 1.3 ATA | 4psi / 1.3 ATA | Same |
| Inflation Method | Gast compressor<br>0523 | Gast compressor<br>0523 | Oil-less compressor | Same |
| Locking System | Simple air tight<br>closure | Simple air tight<br>closure | System of metal<br>rings | Different |
| Materials | 33-39 oz uretane<br>coated nylon | 33-39 oz uretane<br>coated nylon | Polyester hatched<br>1100 18x18 Vinyl<br>with phthalate free | Similar |
| Relief Valves | Plastic & Metal | Plastic & Metal | 2 by Plastic & Metal | Similar |
| Compressor | Gast oil-less | Gast oil-less | Oil-less | Similar |
| Air Filter in the<br>Compressor | Yes | Yes | Yes | Same |
| Pressure<br>Measurement | Yes | Yes | Yes | Same |
| Air Filter Chamber | Yes | Yes | Yes | Same |
| Skills of the<br>Person to Leave<br>the Chamber | Yes | Yes | Yes | Same |
| Operating<br>Temperature °C | -50°F to 125°F | -50°F to 125°F | 59 °F / 113 °F | Similar |
| Manufacturer | SUMMIT TO SEA | SUMMIT TO SEA | Oxavita S.R.L. | Significant<br>Difference |
| Trade Name | The Grand Dive<br>HD400 | The Dive HD250 | Revitalair® 430F | Difference |
| Contraindications | Patients with:<br>Colds, flu<br>symptoms,<br>- Recent Alcohol<br>consumption,<br>- Blocked ear<br>canals,<br>- Blocked Sinuses,<br>- Otic barotraumas,<br>- Excessive CO2<br>Exposure,<br>- Pulmonary hyper<br>expansion<br>Decompression<br>sickness | Patients with:<br>Colds, or flu<br>symptoms,<br>- Recent Alcohol<br>consumption<br>- Blocked Ear<br>Canals,<br>- Blocked Sinuses,<br>- Otic barotraumas,<br>- Excessive C02<br> | Patients with: Colds,<br>or flu symptoms,<br>- Recent Alcohol<br>consumption<br>- Blocked Ear Canals,<br>- Blocked Sinuses,<br>- Otic barotraumas,<br>- Excessive C02<br>Exposure,<br>- Pulmonary hyper<br>expansion<br>Decompression<br>Sickness | Same |
| Labeling | Comparable | Comparable | Comparable | Same |
# Table 5A – Comparison of Characteristics
{6}------------------------------------------------
{7}------------------------------------------------
# 9. Non-Clinical Performance Data
As part of demonstrating safety and effectiveness of the Revitalair® 430F Hyperbaric Chamber and in showing substantial equivalence to the predicate device that are subject to this 510(k) submission, Oxavita completed an number of non-clinical performance tests. Testing was performed as determined by the hazard analysis. The Revitalair® 430F Hyperbaric Chamber meets all the requirements for overall design, biocompatibility, and electrical safety results confirming that the design output meets the design inputs and specifications for the device.
The Revitalair® 430F Hyperbaric Chamber passed all the testing in accordance with internal requirements, and the standards shown below to support substantial equivalence of the subject device:
- . Electrical safety testing per IEC 60601-1
- Electromagnetic Disturbance (EMD) testing per IEC 60601-1-2
- Biocompatibility testing - Cytotoxicity, Sensitization, Gas emission VOC, Particulates, Inorganic gases (Ozone, CO, CO2) per ISO 10993 and ISO 18562
{8}------------------------------------------------
In addition to the guidance and standards testing, the following life testing was performed:
- Device Life Report which address's the device's useful life based on reliability analysis of the individual components
- . Pressure Testing based on ASME PVHO-1-2007, sections: 2-7.2 al 2-7.7; 4-8.1.1; 4-8.1.2; 4-8.1.4 which address's the device's response to excessive pressures.
#### 10. Clinical Performance Data
There was no human clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with proven safety and efficacy for the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.
#### 11. Statement of Substantial Equivalence
The Portable Hyperbaric Chamber Revitalair® 430F, as designed and manufactured, is determined to be substantially equivalent to the cited predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.