← Product Code [MAX](/productcode/MAX) · K171893

# Blustone Synergy Lumbar Interbody Fusion System (Basalt, Magma, Obsidian), Blustone Synergy Cervical Interbody Fusion System (Slate) (K171893)

_Bluestone Synergy · MAX · Sep 8, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K171893

## Device Facts

- **Applicant:** Bluestone Synergy
- **Product Code:** [MAX](/productcode/MAX.md)
- **Decision Date:** Sep 8, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Blustone Synergy lumbar implants (Basalt, Magma, Obsidian): The Blustone Synergy lumbar implants are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). These patients should be skeletally mature and have six months of non-operative therapy. Additionally, the Blustone Synergy lumbar implants can be used as an adjunct to fusion in patients diagnosed with degenerative scoliosis. The Blustone Sysergy lumbar implants are intended to be used with supplemental spinal fixation systems that have been cleared for use in the lumbosacral spine. Blustone Synergy cervical implants (Slate): The Blustone Synergy cervical (Slate) implants are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous bone graft when used as an adjunct to fusion in patients with cervical disc disease (DDD) at one level or two contiguous levels from C2 to T1. These patients should be skeletally mature and have had six weeks of non-operative treatment. The Blustone Synergy cervical implants are also to be used with supplemental fixation.

## Device Story

Interbody fusion system consisting of PEEK implants (Basalt, Magma, Obsidian for lumbar; Slate for cervical) with tantalum markers. Implants placed in intervertebral space to facilitate fusion; used with autograft/allogenic bone graft and supplemental fixation. Lumbar devices implanted via posterior, transverse, lateral, or transforaminal approaches; cervical devices via anterior approach. Used by surgeons in clinical settings to treat degenerative disc disease and degenerative scoliosis. Tantalum markers assist in radiographic visualization for proper placement. Benefits include stabilization of spinal segments and promotion of fusion in skeletally mature patients.

## Clinical Evidence

Bench testing only. Mechanical performance evaluated via static/dynamic compression and torsion (ASTM F2077-14), subsidence (ASTM F2267-04(2011)), and expulsion (ASTM Draft Standard F-04.25.02.02). No clinical data provided.

## Technological Characteristics

Materials: Zeniva ZA-500 PEEK (ASTM F2026) and tantalum markers (ASTM F560). Instruments: Stainless steel (ASTM F899). Form factor: Interbody cages for cervical and lumbar spine. Connectivity: None. Sterilization: Not specified. Algorithm: None.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Eminent Spine Interbody Fusion System ([K090064](/device/K090064.md))
- Lumbar I/F Cage ([P960025](/device/P960025.md))
- Aleutian System ([K133614](/device/K133614.md))
- Cascadia System ([K162264](/device/K162264.md))
- Integrity Spine Lumbar Interbody Fusion System ([K151819](/device/K151819.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 8, 2017

Blustone Synergy % Ms. Lisa Peterson Official Correspondent Bluestone Synergy 5520 Ventana Ct. Pueblo, Colorado 81005

Re: K171893

Trade/Device Name: Blustone Synergy Lumbar Interbody Fusion System Basalt, Magma, Obsidian), Blustone Synergy Cervical Interbody Fusion System (Slate) Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX, ODP Dated: June 17, 2017 Received: June 26, 2017

Dear Ms. Peterson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act 

{1}------------------------------------------------

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K171893

#### Device Name

Blustone Synergy Lumbar Interbody Fusion (Basalt, Magma, Obsidian), Blustone Synergy Cervical Interbody Fusion System (Slate)

#### Indications for Use (Describe)

Blustone Synergy lumbar implants (Basalt, Magma, Obsidian):

The Blustone Synergy lumbar implants are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). These patients should be skeletally mature and have six months of non-operative therapy. Additionally, the Blustone Synergy lumbar implants can be used as an adjunct to fusion in patients diagnosed with degenerative scoliosis. The Blustone Sysergy lumbar implants are intended to be used with supplemental spinal fixation systems that have been cleared for use in the lumbosacral spine.

#### Blustone Synergy cervical implants (Slate):

The Blustone Synergy cervical (Slate) implants are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous bone graft when used as an adjunct to fusion in patients with cervical disc disease (DDD) at one level or two contiguous levels from C2 to T1. These patients should be skeletally mature and have had six weeks of non-operative treatment. The Blustone Synergy cervical implants are also to be used with supplemental fixation.

Type of Use (*Select one or both, as applicable*)

| <span style="font-size:10px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------------------------------|-----------------------------------------------|
|------------------------------------------------------------------------------------|-----------------------------------------------|

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary Blustone Synergy İnterbody Fusion System Premarket Notification

| SUBMITTED BY                             | Blustone Synergy<br>5520 Ventana Ct.<br>Pueblo CO 81005                                                                                    |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT PERSON                           | Lisa Peterson<br>Kaedon Consulting, LLC<br>Phone: 512-507-0746<br>lpeterson@kaedonconsulting.com                                           |
| DATE PREPARED                            | September 7, 2017                                                                                                                          |
| CLASSIFICATION NAME                      | Intervertebral Body Fusion Device with Bone Graft, Lumbar<br>(MAX)<br>Intervertebral Body Fusion Device with Bone Graft, Cervical<br>(ODP) |
| DEVICE CLASS                             | Class II                                                                                                                                   |
| REGULATION NUMBER                        | 888.3080 (Product Code MAX)<br>888.3080 (Product Code ODP)                                                                                 |
| COMMON NAME                              | Interbody Fusion Device                                                                                                                    |
| PROPRIETARY NAME                         | Blustone Synergy Lumbar Interbody Fusion System (Basalt,<br>Magma, Obsidian)<br>Blustone Synergy Cervical Interbody Fusion System (Slate)  |
| IDENTIFICATION OF PREDICATE<br>DEVICE(S) | Predicate devices include various cleared interbody fusion<br>systems:                                                                     |
|                                          | Primary<br>Eminent Spine: Eminent Spine Interbody Fusion System<br>(K090064)                                                               |
|                                          | Additional<br>DePuy Acromed: Lumbar I/F Cage (P960025)                                                                                     |

- K2M, Inc: Aleutian System (K133614) and Cascadia System (K162264)
- Integrity Spine: Integrity Spine Lumbar Interbody Fusion -System (K151819)

{4}------------------------------------------------

K171893 Page 2 of 3

## DEVICE DESCRIPTION

The Blustone Synergy Interbody Fusion System will be offered in various device configurations based on surgical approach and patient anatomy, and consist of:

- Blustone Synergy cervical interbody fusion device(s), which may be implanted as a single 1) device via an anterior approach.
- 2) Blustone Synegy lumbar interbody fusion device(s), which may be implanted
	- bi-laterally in pairs via a posterior (PLIF) approach;
	- as a single device via a transverse (T-PLIF) approach;
	- as a single device via a lateral (LLIF) approach;
	- as a single device via a transforaminal (TLIF) approach

The Blustone Synergy System implant components are made of polyether ether ketone (Zeniva ZA-500 PEEK) that conforms to ASTM F2026. Additionally, the devices contain tantalum markers (ASTM F560) to assist the surgeon with proper placement of the device. The System is implanted using a combination of device specific and universal class I instruments manufactured from stainless steel materials that conform to ASTM F899.

#### INDICATIONS

## Blustone Synergy lumbar implants (Basalt, Magma, Obsidian):

The Blustone Synerqy lumbar implants are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when used as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level or two contiquous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). These patients should be skeletally mature and have six months of non-operative therapy. Additionally, the Blustone Synergy lumbar implants can be used as an adjunct to fusion in patients diagnosed with degenerative scoliosis. The Blustone Synergy lumbar implants are intended to be used with supplemental spinal fixation systems that have been cleared for use in the lumbosacral spine.

#### Blustone Synergy cervical implants (Slate):

The Blustone Synerqy cervical (Slate) implants are intervertebral body fusion devices indicated for use with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft when used as an adjunct to fusion in patients with cervical disc disease (DDD) at one level or two contiquous levels from C2 to T1. These patients should be skeletally mature and have had six weeks of non-operative treatment. The Blustone Synergy cervical implants are also to be used with supplemental fixation.

## TECHNOLOGICAL COMPARISON TO PREDICATE(S)

The purpose of this premarket notification is to obtain clearance to market the Blustone Synergy Interbody Fusion System. The Blustone Synergy System implants were compared to the predicate devices identified above, and performance evaluation results, design features, materials and sizes were found to be substantially equivalent to these systems.

{5}------------------------------------------------

K171893 Page 3 of 3

#### DISCUSSION OF NON-CLINICAL TESTING

The following non-clinical tests were conducted:

- Static and dynamic compression testing, conducted in accordance with ASTM F2077-14 -
- -Static and dynamic torsion testing, conducted in accordance with ASTM F2077-14
- -Subsidence testing, conducted in accordance with ASTM F2267-04(2011)
- -Expulsion testing, conducted in accordance with ASTM Draft Standard F-04.25.02.02

#### CONCLUSIONS

The subject and predicate device(s) share the same intended use, primary implant design and equivalent material of manufacture. The non-clinical mechanical test results demonstrate that any minor differences do not impact device performance as compared to the predicates and demonstrate that the Blustone Synergy Interbody Fusion System is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K171893](https://fda-staging.innolitics.com/device/K171893)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
