# OverStitch SX Endoscopic Suturing System (ESS) (K171886)

_Apollo Endosurgery, Inc. · OCW · Nov 15, 2017 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K171886

## Device Facts

- **Applicant:** Apollo Endosurgery, Inc.
- **Product Code:** [OCW](/productcode/OCW.md)
- **Decision Date:** Nov 15, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The OverStitch™ Endoscopic Suturing System is intended for endoscopic placement of suture(s) and approximation of soft tissue.

## Device Story

The OverStitch Endoscopic Suturing System is a single-use, sterile endoscopic accessory designed for tissue approximation. It consists of a Needle Driver Assembly, Anchor Exchange Device, Tissue Helix, and Suture Cinch. The system attaches to either dual-channel or single-channel endoscopes. The clinician operates the device by squeezing a handle to actuate a needle body, which exchanges a suture-anchor assembly to perform stitching. The Tissue Helix gathers tissue for approximation, and the Suture Cinch secures the suture using a PEEK component. The device is used in clinical settings by physicians to perform endoscopic suturing, potentially benefiting patients by enabling minimally invasive tissue repair within the gastrointestinal tract. The system is entirely mechanical; it does not involve electronic processing, software, or AI/ML algorithms.

## Clinical Evidence

No clinical data was required. Substantial equivalence was demonstrated through bench testing, including needle passing reliability, needle pull-off strength, Suture Cinch deployment/pull-off strength, and Tissue Helix acquisition reliability. Biocompatibility was verified per ISO 10933-1 (cytotoxicity, sensitization, irritation, toxicity, implantation, genotoxicity). Sterility was validated per ISO 11135:2014.

## Technological Characteristics

Materials: Stainless steel, ABS, silicone, Nitinol, PTFE, VESTAKEE PEEK, Cobalt Chromium MP35. Sensing/Actuation: Mechanical handle-actuated needle driver and anchor exchange. Form factor: Endoscope-mounted catheter system (117-165cm length). Connectivity: None. Sterilization: Ethylene Oxide (EO) to SAL 10-6. Software: None.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- OverStitch™ Endoscopic Suturing System ([K081853](/device/K081853.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

November 15, 2017

Apollo Endosurgery Ms. Maritza Ward Manager, Regulatory Affairs 1120 S. Capital of Texas Hwy. Building 1, Suite 300 Austin, Texas 78746

Re: K171886

Trade/Device Name: OverStitch Endoscopic Suturing System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: HCF Dated: October 12, 2017 Received: October 16, 2017

Dear Ms. Ward:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# David Krause -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K171886

#### Device Name

Overstitch™ Endoscopic Suturing System

Indications for Use (Describe)

The OverStitch™ Endoscopic Suturing System is intended for endoscopic placement of suture(s) and approximation of soft tissue.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|                                                 |  |

Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) Summary

| Owner's Name & Address: | Apollo Endosurgery<br>1120 S. Capital of Texas Hwy.<br>Building 1, Suite 300<br>Austin, TX 78746                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:         | Maritza Ward<br>Manager, Regulatory Affairs<br>Phone: (512)-279-5114<br>Email : Maritza.Ward@apolloendo.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Date:                   | November 9, 2017                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Trade Name:             | OverStitch™ Endoscopic Suturing System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Common Name:            | Endoscopic Tissue Approximation Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Product Code:           | HCF                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Classification:         | Class II (21 CFR 876.1500)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Classification Name     | Endoscope and Accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Predicate Devices:      | K081853 – OverStitch™ Endoscopic Suturing System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Device Description      | The OverStitch™ Endoscopic Suturing System and accessories<br>are intended for endoscopic placement of sutures and<br>approximation of soft tissue within the gastrointestinal tract<br>utilizing either a dual channel or single-channel endoscope. The<br>system is comprised of the Needle Driver Assembly and Anchor<br>Exchange Device (collectively referred to as ESS), and<br>accessories such as the Tissue Helix and Suture Cinch devices.<br>All devices are sterile packaged and designed for single use and<br>are manufactured from various thermoplastic, silicone and<br>stainless steel materials. |

The OverStitch ESS is designed for compatibility with dual channel endoscopes. The endcap of the Needle Driver Assembly attaches to the distal end of the endoscope, and the handle attaches to the proximal end. The handle of the Needle Driver Assembly is squeezed to actuate the needle body and exchange the proprietary suture assembly with the Anchor Exchange to perform stitching operations.

OverStitch Sx ESS

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Apollo Endosurgery. The logo features a stylized human figure with a circular head and a swirling, ribbon-like body. To the right of the figure is the word "apollo" in blue, sans-serif font. Below "apollo" is the word "endosurgery" in a smaller, similar font, also in blue.

The OverStitch Sx ESS is designed for compatibility with single channel endoscopes. The ESS is mounted onto the endoscope using silicone straps distributed along the length of the catheter and endcap assembly. The external catheter sheath has two working channels through which the Anchor Exchange and OverStitch accessories can operate, independent of the endoscope channel. The handle of the Needle Driver Assembly is squeezed to actuate the needle body and exchange the proprietary Suture-Anchor assembly with the Anchor Exchange to perform stitching operations.

#### OverStitch Tissue Helix

The OverStitch Helix is an optional device that is designed to acquire tissue by rotating the device handle into targeted area until tissue is gathered into the exposed helix coil. The acquired tissue is then pull into proximity of the needle body to complete the stitching operation.

#### OverStitch Suture-Anchor Assembly

The OverStitch Suture-Anchor Assembly is comprised of a 510(k) cleared suture product, manufactured from either polydioxanone or polypropylene materials attached to an implantable anchor manufactured from cobalt chrome and stainless steel. The Anchor component is intended to function with the OverStitch ESS device to perform stitching operations and serves as an anchor to secure suture placement, once released from the ESS.

### OverStitch Suture Cinch

The Overstitch Suture Cinch device is comprised of thermoplastic and stainless steel materials and includes an implantable PEEK Cinch component designed to secure the placement of suture as a final step of an OverStitch procedure. The device functions by squeezing the handle and deploying the PEEK components that press-fit onto the tail end of the suture to maintain suture position in situ.

Indications for Use:

The OverStitch Endoscopic Suturing System is intended for endoscopic placement of suture(s) and approximation of soft tissue.

{5}------------------------------------------------

## Technological Characteristics:

| NEEDLE DRIVER<br>ASSEMBLY | Gen 0 Needle Driver<br>Assembly                                                                                                                                                              | Gen 2 Needle Driver<br>Assembly | Sx Needle Driver<br>Assembly                                                                  |                                                        |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|-----------------------------------------------------------------------------------------------|--------------------------------------------------------|
| 510(k) Clearance          | K081853                                                                                                                                                                                      | Proposed                        | Proposed                                                                                      |                                                        |
| Single Use                | Yes                                                                                                                                                                                          |                                 |                                                                                               |                                                        |
| Shelf-Life Claim          | 1-year                                                                                                                                                                                       | 3-years                         | 1-year                                                                                        |                                                        |
| Materials                 | Stainless Steel Delrin/ABS                                                                                                                                                                   | Stainless Steel ABS             | Stainless Steel ABS Silicone                                                                  |                                                        |
| Dimensions                | Length: 130cm<br>O.D.: 2mm (distal end)                                                                                                                                                      |                                 | Length: 117cm<br>O.D.: 2mm (distal end)                                                       |                                                        |
| Suture Mechanism          | The handle of the Needle Driver Assembly is squeezed to actuate the needle body<br>and exchange the proprietary suture assembly with the Anchor Exchange to perform<br>stitching operations. |                                 |                                                                                               |                                                        |
| Compatible Scopes         | Dual channel OD between 12.8 mm and 14.0 mm <170 cm length                                                                                                                                   | Olympus GIF-2TH180 GIF-2T160    | Single Channel Endoscopes that are 8.8 - 9.8 mm in diameter and up to 110cm in working length |                                                        |
| Endcap<br>+ Scope         | Position                                                                                                                                                                                     | Seats on distal scope face      | Seats on distal scope face                                                                    | Side mount to distal end in any rotational orientation |
|                           | Attachment                                                                                                                                                                                   | Back-loaded wire                | Interference fit                                                                              | Two straps                                             |
| Sterilization             | Terminally sterilized to SAL 10-6 using a validated EO method                                                                                                                                |                                 |                                                                                               |                                                        |

| ANCHOR EXCHANGE   | Gen 0 Anchor Exchange                                      | Gen 2 Anchor Exchange                                                                 |
|-------------------|------------------------------------------------------------|---------------------------------------------------------------------------------------|
| 510(k) Clearance  | K081853                                                    | Proposed                                                                              |
| Single Use        | Yes                                                        |                                                                                       |
| Shelf-Life Claim  | 1-year                                                     | 3-years<br>(when packaged as ESS-G02-160)<br>1-year<br>(when packaged as ESS-G02-Sx1) |
| Materials         | Stainless Steel Delrin/ABS                                 | Stainless Steel ABS                                                                   |
| Dimensions        | Length: 130cm<br>O.D.: 3.2mm (distal end)                  |                                                                                       |
| Compatible Scopes | Dual channel OD between 12.8 mm and 14.0 mm <170 cm length | Dual Channel<br>Olympus  GIF-2TH180 GIF-2T160<br>Single Channel                       |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Apollo Endosurgery. The logo features a stylized human figure with a blue circle for a head and a swirling blue line representing the body. To the right of the figure is the word "apollo" in blue, followed by the word "endosurgery" in a smaller font size and also in blue. The logo is clean and modern, with a focus on the company's name and area of expertise.

| ANCHOR EXCHANGE                           | Gen 0 Anchor Exchange                                         | Gen 2 Anchor Exchange                                                               |
|-------------------------------------------|---------------------------------------------------------------|-------------------------------------------------------------------------------------|
|                                           |                                                               | Endoscopes that are 8.8 - 9.8mm in<br>diameter and up to 110cm in working<br>length |
| Method of Actuation                       | Sliding Lever                                                 | Button on top of handle                                                             |
| Integrated With Needle<br>Driver Assembly | Yes                                                           | No                                                                                  |
| Sterilization                             | Terminally sterilized to SAL 10-6 using a validated EO method |                                                                                     |

| SUTURE CINCH        | Gen 0 Suture Cinch                                                                                                   | Gen 2 Suture Cinch                                                                                                                              |
|---------------------|----------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Clearance    | K081853                                                                                                              | Proposed                                                                                                                                        |
| Single Use          | Yes                                                                                                                  |                                                                                                                                                 |
| Shelf-Life Claim    | 1-year                                                                                                               | 3-years                                                                                                                                         |
| Materials           | Stainless Steel ABS Medical grade metallic or plastic materials                                                      | Stainless Steel ABS VESTAKEE PEEK                                                                                                               |
| Cinch design        | PEEK or Stainless Plug and collar components designed to press-fit together onto suture and hold construct in place. | PEEK Plug and collar components designed to press-fit together onto suture and hold construct in place.                                         |
| Dimensions          | Suture Cinch Assembly:<br>Length: 182cm<br>O.D.: 2.4mm (distal end)                                                  | Suture Cinch Assembly:<br>Length: 152cm<br>O.D.: 2.4mm (distal end)                                                                             |
|                     | Deployed Cinch:<br>Length: 8mm<br>O.D. 2.4mm                                                                         | Deployed Cinch:<br>Length: 8mm<br>O.D. 2.4mm                                                                                                    |
| User Sequence       | Helix, Device Handle, Anchor Exchange, Cinch per DFU                                                                 |                                                                                                                                                 |
| Compatible Scopes   | Dual channel OD between 12.8 mm and 14.0 mm <170 cm length                                                           | Dual Channel<br>Olympus  GIF-2TH180 GIF-2T160   Single Channel<br>Endoscopes that are 8.8 - 9.8mm in diameter and up to 110cm in working length |
| Method of Actuation | Squeezing of Handle                                                                                                  |                                                                                                                                                 |
| Sterilization       | Terminally sterilized to SAL 10-6 using a validated EO method                                                        |                                                                                                                                                 |

| TISSUE HELIX      | Gen 0 Tissue Helix                         | Gen 2 Tissue Helix                                |
|-------------------|--------------------------------------------|---------------------------------------------------|
| 510(k) Clearance  | K081853                                    | Proposed                                          |
| Single Use        | Yes                                        |                                                   |
| Shelf-Life Claim  | 1-Year                                     | 3-years                                           |
| Materials         | • Stainless Steel<br>• Derlin/ABS          | • Stainless Steel<br>• Nitinol<br>• PTFE<br>• ABS |
| Dimensions        | Length: 165cm<br>O.D.: 2.0 mm (distal end) | Length: 165cm<br>O.D.: 2.4mm (distal end)         |
| Compatible Scopes | • Dual channel                             | Dual Channel                                      |

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Apollo Endosurgery. The logo features a stylized human figure with a spiral design for the body, and a blue circle for the head. To the right of the figure is the word "apollo" in blue, with the word "endosurgery" underneath in a smaller font, also in blue.

|                                 | • OD between 12.8 mm and 14.0 mm<br>• <170 cm length          | Olympus<br>· GIF-2TH180<br>· GIF-2T160<br>Single Channel<br>Endoscopes that are 8.8 - 9.8mm in<br>diameter and up to 110cm in working<br>length |
|---------------------------------|---------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Outer Sheath<br>Protecting Coil | No                                                            | Yes                                                                                                                                             |
| Sterilization                   | Terminally sterilized to SAL 10-6 using a validated EO method |                                                                                                                                                 |

| ANCHOR-SUTURE<br>ASSEMBLY | Gen 0 Anchor                                                                                                                                                                                                                                                                                          | Gen 2 Anchor                                                                                                                                                                                                                                                                                                                                        |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Clearance          | K081853                                                                                                                                                                                                                                                                                               | Proposed                                                                                                                                                                                                                                                                                                                                            |
| Single Use                | Yes                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                     |
| Shelf-Life Claim          | 1-year                                                                                                                                                                                                                                                                                                | 3-years                                                                                                                                                                                                                                                                                                                                             |
| Materials                 | • Stainless Steel<br>• Medical Grade Plastic                                                                                                                                                                                                                                                          | • Stainless Steel 316L<br>• Cobalt Chromium MP35                                                                                                                                                                                                                                                                                                    |
| Dimensions                | Anchor<br>Length: 8.4 mm<br>O.D.: 1.3 mm<br><br>Suture<br>Length: 185 cm<br>O.D.: 2.0 mm                                                                                                                                                                                                              | Anchor<br>Length: 6.6 mm<br>O.D.: 1.0 mm<br><br>Suture<br>Length: 185 cm<br>O.D.: 2.0 mm                                                                                                                                                                                                                                                            |
| Suture Anchor Design      | Suture-Anchor has recessed features on<br>both ends of the component designed to<br>engage with the Needle Body Assembly<br>and the Anchor Exchange. The suture is<br>swaged onto the center of the needle<br>body to allow for the needle body to<br>double as an anchor for the suture<br>construct | Suture-Anchor has a recessed feature<br>on one end designed to engage with<br>the Anchor Exchange, and a latch<br>feature on the opposite end designed<br>to engage with the Needle Body<br>Assembly. The suture is swaged onto<br>the center of the needle body to allow<br>for the needle body to double as an<br>anchor for the suture construct |
| Sterilization             | Terminally sterilized to SAL 10-6 using a validated EO method                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                     |

Non-Clinical Performance Data:

Appropriate product testing was performed on all subject devices to evaluate conformance to product specifications and equivalence to the predicate designs. Verification and validation for the proposed system of devices was conducted in accordance with protocol. All devices were evaluated to their individual functional and reliability requirements, as well as system compatibility with various endoscopes.

#### Performance Testing

Bench testing included needle passing reliability, needle pull off strength, Suture Cinch deployment and pull-off strength, Tissue Helix acquisition reliability and bond strength.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Apollo Endosurgery. The logo features a stylized human figure with a spiral design in place of the body, colored in blue. To the right of the figure is the word "apollo" in a bold, sans-serif font, also in blue. Below "apollo" is the word "endosurgery" in a smaller, lighter blue font.

#### Sterility

Sterility claims (SAL 10-6) were confirmed by executing a sterilization validation in accordance with ISO 11135:2014.

Shelf-Life Shelf-life claims were confirmed by repeating sterility and functional testing on aged product.

The results of all studies confirmed equivalency between the subject and predicate devices, and that no new issues of safety or efficacy were raised.

- Biocompatibility: Biocompatibility testing and toxicological assessments were performed on all OverStitch devices in accordance with their risk category requirements, as defined in ISO 10933-1. Testing included cytotoxicity, sensitization, irritation, acute systemic toxicity, material mediated pyrogenicity, subacute/subchronic toxicity, implantation, and genotoxicity.
- Clinical Performance Data: Clinical testing was not required to demonstrate substantial equivalence.
- Basis of Substantial Equivalence: Based on a comparison of indications for use and technological characteristics, the proposed devices have demonstrated substantial equivalence to their predicates.

---

**Source:** [https://fda-staging.innolitics.com/device/K171886](https://fda-staging.innolitics.com/device/K171886)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).