The KH Ophthalmic Slit-lamp Microscope is an AC-powered slitlamp biomicroscope intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affects the structural properties of the anterior eye segment.
Device Story
The KH Ophthalmic Slit-lamp Microscope is an AC-powered biomicroscope used for clinical eye examinations. It projects a thin, intense beam of light into the patient's eye via a control diaphragm. The device consists of a microscope unit, illumination unit, base unit, chinrest, and power unit. A physician operates the binocular stereoscopic microscope to zoom and observe the anterior eye segment, including the cornea and posterior capsule, in three dimensions. The device aids in diagnosing diseases or trauma affecting the anterior segment. It is used in clinical settings; the output is a direct visual observation of the eye, which the clinician uses to inform diagnostic and treatment decisions. The device benefits patients by enabling detailed structural assessment of the eye.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), biocompatibility (ISO 10993-5, ISO 10993-10), and performance standards (ISO 15004-1, ISO 15004-2, ISO 10939, IEC 62471).
Technological Characteristics
AC-powered slit-lamp biomicroscope. Features binocular stereoscopic microscope (Galilean converging). Illumination: 12V/50W halogen bulb. Includes chinrest, base, and power unit. Complies with IEC 60601-1 (electrical safety), ISO 15004-1/2 (ophthalmic instruments/light hazard), and ISO 10939 (slit-lamp microscopes). Biocompatibility per ISO 10993. No digital camera or image capture modules.
Indications for Use
Indicated for eye examination of the anterior eye segment (cornea epithelium to posterior capsule) to aid in the diagnosis of diseases or trauma affecting the structural properties of the anterior eye segment in patients requiring ophthalmic evaluation.
Regulatory Classification
Identification
An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.
Special Controls
*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 21, 2017
Chongqing Kanghua Ruiming S&T Co., Ltd. Xi You Production Manager No.5, Road1, Tongjiaxi Industrial Park, Beibei Chongqing, 400070 CHINA
Re: K171877
Trade/Device Name: SLM-1ER, SLM-2ER, SLM-3ER Regulation Number: 21 CFR 886.1850 Regulation Name: AC-Powered Slitlamp Biomicroscope Regulatory Class: Class II Product Code: HJO Dated: November 1, 2017 Received: November 1, 2017
Dear Xi You:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bradley S. Cunningham -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below,
51D(k) Number (if known) K171877
Device Name KH Ophthalmic Slit-lamp Microscope Model: SLM-1ER, SLM-2ER
### Indications for Use (Describe)
The KH Oplithalmic Slit-lamp Microscope is an AC-powered slittamp biomicroscope intended for use in eye examination of the anterior eye segment, from the comes epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affects the structural properties of the anterior eve segment.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Offica of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@ida.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
{3}------------------------------------------------
# 510(k) SUMMARY (K 171877)
# KH Ophthalmic Slit-lamp Microscope
#### Submitter's Information 1.
The submitter of this special pre-market notification is:
| Name: | Chongqing Kanghua Ruiming S&T Co., Ltd. |
|------------------------|----------------------------------------------------------------------------------------|
| Address: | No.5, Road1, Tongjiaxi Industrial Park, Beibei,<br>Chongqing, China<br>Zip Code:400070 |
| Company Phone No: | +86-23-63227332 |
| Company Fax No: | +86-23-63227336 |
| Contact Person: | Xi You |
| E-mail: | doumeki2007@163.com |
| Date summary prepared: | April 15, 2017 |
#### 2. Device Identification
| Device Trade Name: | KH Ophthalmic Slit-lamp Microscope<br>Model: SLM-1ER, SLM-2ER |
|---------------------|---------------------------------------------------------------|
| Classification Name | AC- Powered Slit-Lamp Biomicroscope |
| Regulation Number: | 21 CFR 886.1850 |
| Regulation Name: | AC Powered Slit lamp Bio-microscope |
| Regulation Class: | II |
| Product Code: | HJO |
#### Predicate Devices 3.
- Model: SLM-1ER, SLM-2ER, the predicate cited is the Keeler Slit Lamp H-• Series of the same instrument:
| Device Trade Name: | Keeler Slit Lamp H- Series |
|-----------------------|-------------------------------------|
| 510(k) Number; | K131589 |
| Common Name | AC Powered Slit lamp Bio-microscope |
| Class: | II |
| Classification Panel: | 86 |
| Product Code: | HJO |
| Regulation Number: | 886.1850 |
{4}------------------------------------------------
#### Device Description 4.
The KH Ophthalmic Slit-lamp Microscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light. The KH Ophthalmic Slitlamp Microscope is composed of the following components: microscope unit, illumination unit, base unit, chinrest, and table and power unit. The slitlamp biomicroscope is used for the observation of the eye. It has an illumination unit to illuminate the eye, and a binocular stereoscopic microscope to zoom and observe patient's eyes, and also can observe the three-dimensional image.
#### Intended Use / Indications for Use 5.
The KH Ophthalmic Slit-lamp Microscope is an AC-powered slitlamp biomicroscope intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affects the structural properties of the anterior eye segment.
| Feature | H-Series Slit<br>Lamp [K131589] | KH Ophthalmic Slit-<br>lamp Microscope | Discussion |
|-----------------------------------------------------------------|------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intend use | As stated | As stated | No changes |
| Bio-microscope | | | |
| Bio-microscope | Galilean<br>converging<br>binoculars with<br>detachable<br>eyepiece head | • SLM-1ER,<br>Converging<br>binoculars with<br>detachable eyepiece<br>head<br>• SLM-2ER, Galilean<br>converging<br>binoculars with<br>detachable eyepiece<br>head | As head can be separated<br>from the body,<br>intervention of module<br>does not change these<br>elements |
| Digital Camera Module | Not fitted | • SLM-1ER, SLM-<br>2ER: Not fitted | Same |
| Image capture button | Not fitted | • SLM-1ER, SLM-2ER: Not fitted | Same |
| Camera Exposure<br>Buttons | Not fitted | Not fitted | Same |
| Slit Lamp illumination<br>Options | 6VDC 20W<br>halogen bulb /<br>LED | 12V/50W halogen<br>bulb | Required for slit lamp<br>illumination with some<br>light used for background<br>lighting |
| Background<br>Illumination<br>adjuster | Not fitted | • Not fitted | Same |
| Background Light<br>Source Options | Not fitted | • Not fitted | Required for higher<br>background light output<br>but still conforms to<br>requirements of 15004-2<br>Photo toxicity with slightly<br>longer exposure time. |
| Background lighting<br>duty cycle - halogen<br>bulb option only | Not applicable | 50% duty cycle at<br>maximum brightness. | Ensures body of light<br>source does not exceed<br>62.3 °C limit demanded by<br>IEC 60601-1 compliance |
| Duration of illumination | Maximum<br>examination times<br>according to ISO<br>15004-2 and ISO<br>10939 | Maximum examination<br>times according to<br>ISO 15004-2 and ISO<br>10939 | As a result of photo-<br>toxicity testing to ISO<br>15004-2 maximum<br>exposure increased from<br>13 to 17 minutes for bulb<br>option (LED maximum<br>exposure is unchanged at<br>12.5 minutes). |
| Input voltage | AC 100-240 V,<br>50/60 Hz | AC 110 V, 50/60 Hz | Same |
#### Comparison to predicate device 6.
{5}------------------------------------------------
{6}------------------------------------------------
#### Non-Clinical testing 7.
# Safety and EMC:
Safety and EMC test was performed in according to the
- IEC60601-1:2005 + CORR.1(2006)+CORR.2(2007)+AM1(2012) or IEC 60601--1:2012
- -IEC 60601-1-2:2007 Medical electrical equipment - part 1-2: general requirements for basic safety and essential performance - collateral standard: electromagnetic compatibility - requirements and tests.
# Biocompatibility :
Biocompatibility testing was performed to evaluate the biocompatibility of the contact materials in accordance with ISO 10993-1:2009, the device passed each biocompatibility test indentified below:
- Cytotoxicity testing in according to ISO 10993-5:2009 -
- Skin irritation testing in according to ISO 10993-10:2010
- -Vaginal irritation testing in according to ISO 10993-10:2010
- Sensitization testing in according to ISO 10993-10:2010 -
# Performance Data:
The following bench testing was conducted in order to support substantial equivalence:
- -ISO 15004-1:2006 Ophthalmic instruments - Fundamental requirements and test methods - Part 1: General requirements applicable to all ophthalmic instruments. The testing found that the product met the requirements of ISO 15004-1:2006.
- -ISO 15004-2:2007 Ophthalmic Instruments - Fundamental requirements and test methods - Part 2: Light hazard protection. The testing found that the device is a Group 2 instrument.
- -ISO 10939:2007 Ophthalmic Instruments - Slit-lamp microscopes found that the KH OPHTHALMIC SLIT-LAMP MICROSCOPE complies with the requirements of the standard.
- -IEC 62471:2006 Photobiological safety of lamps and lamp systems
#### 8. Conclusions:
The KH Ophthalmic Slit-lamp Microscope (Model: SLM-1ER, SLM-2ER) is substantially equivalent to the predicate device in indications for use and construct, Safety and EMC testing to IEC60601-1 and IEC 60601-1-2, and performance testing to ISO 15004-1, ISO 15004-2 and ISO 10939, Based on the information provided in this submission, the KH Ophthalmic Slit-lamp Microscope(Model: SLM-1ER, SLM-2ER) is substantially equivalent to the predicate devices(K131589).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.