Powder Free Nitrile Patient Examination Glove, Blue Colored, Non Sterile, Tested for Use with Chemotherapy Drugs
K171873 · Zhonghong Pulin Medical Products Co., Ltd. · LZA · Nov 15, 2017 · General Hospital
Device Facts
Record ID
K171873
Device Name
Powder Free Nitrile Patient Examination Glove, Blue Colored, Non Sterile, Tested for Use with Chemotherapy Drugs
Applicant
Zhonghong Pulin Medical Products Co., Ltd.
Product Code
LZA · General Hospital
Decision Date
Nov 15, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. These gloves were tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
Device Story
Disposable, non-sterile, blue nitrile patient examination glove; provides barrier protection for examiner's hands; prevents cross-contamination between patient and examiner. Tested for permeation resistance against specific chemotherapy drugs per ASTM D6978-05. Used in clinical settings by healthcare professionals. Output is physical barrier protection; informs clinical decision-making regarding safe handling of hazardous chemotherapy agents. Benefits include reduced risk of exposure to patient pathogens and hazardous drugs.
Clinical Evidence
No clinical data. Bench testing only. Compliance verified via ASTM D6319-10 (physical/dimensional), ASTM D5151-06 (pinholes), ASTM D6124-06 (powder residual), and ASTM D6978-05 (chemotherapy permeation). Biocompatibility testing per ISO 10993-10:2010 confirmed non-irritating and non-sensitizing.
Technological Characteristics
Nitrile compound; powder-free; non-sterile; blue color. Dimensions and physical properties meet ASTM D6319-10. Barrier protection verified via ASTM D5151-11. Biocompatibility per ISO 10993-10:2010. Minimum thickness 0.05mm; minimum length 230mm.
Indications for Use
Indicated for use as a disposable patient examination glove worn on the examiner's hand to prevent contamination between patient and examiner. Tested for use with chemotherapy drugs.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile (K151750)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
November 15, 2017
Zhonghong Pulin Medical Products Co., Ltd. % Da Shi Manager Zhonghong Pulin Medical Products Co.,Ltd. 800 - 10 Four Seasons Place Toronto, M9B 6H7 Ca
Re: K171873
Trade/Device Name: Powder Free Nitrile Patient Examination Glove, Blue Colored, Non Sterile, Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LZA, LZC Dated: October 13, 2017 Received: October 13, 2017
Dear Da Shi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K171873
Device Name
Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs
Indications for Use (Describe)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
These gloves were tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
| Drug and Concentration | Minimum Breakthrough Detection Time in Minute |
|-------------------------------------|-----------------------------------------------|
| Carmustine (BCNU) 3.3 mg/ml | 14.7 |
| Cisplatin 1.0 mq/ml | >240 |
| Cyclophosphamide (Cytoxan) 20 mg/ml | >240 |
| Cytarabine 100 mq/ml | >240 |
| Dacarbazine (DTIC) 10.0 mg/ml | >240 |
| Doxorubicin Hydrochloride 2.0 mq/ml | >240 |
| Etoposide (Toposar) 20.0 mg/ml | >240 |
| Fluorouracil 50.0 mq/ml | >240 |
| Ifosfamide 50.0 mq/ml | >240 |
| Methotrexate 25 mq/ml | >240 |
| Mitomycin C 0.5 mq/ml | >240 |
| Mitoxantrone 2.0 mq/ml | >240 |
| Paclitaxel (Taxol) 6.0 mq/ml | >240 |
| Thiotepa 10.0 mq/ml | 58.8 |
| Vincristine Sulfate 1.0 mq/ml | >240 |
Please note the following drugs have extremely low permeation times: Carmustine (BCNU) (3.3mg/ml) 14.7 minutes and Thiotepa (10 mg/ml) 58.8 minutes.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Zhong Hong Pulin. The logo consists of a red square with the letters "Z" and "Z" separated by a vertical line inside. To the right of the square are the words "ZHONG HONG" in a larger font, with "PULIN" underneath in a smaller font. The text is black.
# 510(k) SUMMARY
Preparation Date: Nov. 8th, 2017
#### Submitter:
| Company Name: | Zhonghong Pulin Medical Products Co., Ltd. |
|-----------------|----------------------------------------------------------------------|
| Address: | Pachigang Industrial Park, Luannan County<br>Tangshan, Hebei, 063502 |
| Country: | China |
| Phone No.: | +86 315-4165760 |
| Fax No.: | +86 315-4167664 |
| E-mail Address: | RA@zhonghongpulin.com |
#### Contact Person:
| Contact Person : | Li, Jinge (Regulatory Affairs Manager) |
|------------------|----------------------------------------|
| Contact Email: | ljg@zhonghongpulin.cn |
#### Name of the Device:
Device trade or proprietary name: Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs Device common or usual name: Patient Examination Glove Device Classification Name: LZA - Polymer Patient Examination Glove Device Classification Name: LZC - Patient Examination Glove, Specialty Regulation Number: 21 CFR 88.6250 FDA Device Class: Class 1 Product Code: LZA, LZC
#### Predicate Device:
Class I patient Examination glove and tested for use with Chemotherapy Drugs, Powder Free, LZC, which meets all the requirement of ASTM D 6319-10 and FDA 21 CFR 880.6250.
Device Name: Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs Copmany Name: Kossan International Sdn. Bhd. 510(K) Number: K151750
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains a logo for Zhong Hong Pulin. The logo consists of a red square with the letters "Z" and "Z" separated by a vertical line. To the right of the square are the words "ZHONG HONG" in a larger font, with the word "PULIN" underneath in a smaller font.
#### Device Description:
The subject device in this 510(k) Notification is a Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile, Tested for Use with Chemotherapy Drugs
The subject device is a patient examination glove made from nitrile compound, blue in color, powder free and non-sterile (as per 21 CFR 880.6250, class I).
The principle operation of the medical device to provide single use barrier protection for the wearer and the device meets all the requirement specifications for Barrier Protection, tensile properties as defined in ASTM D6319-10, Standard specification for Nitrile Examination Gloves.
### Intended use of the Device (Indication for use)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.
These gloves were tested for use with chemotherapy drugs per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.
| Drug and Concentration | Minimum Breakthrough Detection Time in Minutes |
|-------------------------------------|------------------------------------------------|
| Carmustine (BCNU) 3.3 mg/ml | 14.7 |
| Cisplatin 1.0 mg/ml | >240 |
| Cyclophosphamide (Cytoxan) 20 mg/ml | >240 |
| Cytarabine 100 mg/ml | >240 |
| Dacarbazine (DTIC) 10.0 mg/ml | >240 |
| Doxorubicin Hydrochloride 2.0 mg/ml | >240 |
| Etoposide (Toposar) 20.0 mg/ml | >240 |
| Fluorouracil 50.0 mg/ml | >240 |
| Ifosfamide 50.0 mg/ml | >240 |
| Methotrexate 25 mg/ml | >240 |
| Mitomycin C 0.5 mg/ml | >240 |
| Mitoxantrone 2.0 mg/ml | >240 |
| Paclitaxel (Taxol) 6.0 mg/ml | >240 |
| Thiotepa 10.0 mg/ml | 58.8 |
| Vincristine Sulfate 1.0 mg/ml | >240 |
Please note that the following drugs have extremely low permeation times: Carmustine (BCNU) 14.7 minutes and Thiotepa 58.8 minutes.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains a logo for "ZHONG HONG PULIN". The logo consists of a red square with the letters "Z" and "Z" inside, separated by a vertical line. To the right of the square, the words "ZHONG HONG" are written in a larger font, and below that, the word "PULIN" is written in a smaller font.
#### Summary of the Technological Characteristics of the Device:
The subject device is summarized with the following technological characteristics compared to ASTM or equivalent standard.
| Characteristics | Standard | Device performance |
|-----------------------|-------------------------------------------------------------|------------------------------------------------------------------|
| Dimension | ASTM standard D 6319-10 | Meets |
| Physical Properties | ASTM standard D 6319-10 | Meets |
| Freedom from pinholes | 21 CFR 800.20 | Meets |
| | ASTM D5151-11 | |
| Powder Residual | ASTM D6319-10 and D6124-<br>06(Reapproved 2011) | Meets |
| | | |
| Biocompatibility | Primary Skin Irritation ISO<br>10993-10:2010 | Not a primary skin irritant under<br>the conditions of the study |
| | Dermal sensitization in the<br>guinea pig ISO 10993-10:2010 | Not a contact sensitizer under<br>the conditions of the study |
#### Substantial Equivalence Based on Assessment of Non-Clinical Performance Data
Bench tests were conducted to verify that the proposed device met all specifications and the proposed device is Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the subject device complies with the following standards:
ASTM D6319-10 Standard Specification for Nitrile Examination Gloves for Medical application.
ASTM D5151-06 (2011), Standard Test Method for Detection of Holes in Medical Gloves. ASTM D6124-06 (2011). Standard Test Method for Residual Powder on Medical Gloves. ASTM D6978-05 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs
The subject device meets the same test standards conducted by the predicate device. Both device has the same sizes, color, physical and dimensional characteristics.
The minimum breakthrough detection time of Carmustine for the subject device is below 30 minutes, similar with predicate K151750 (Blue).
The minimum breakthrough detection time of Thiotepa for the subject device is at 58.8 minutes. The subject device is having longer permeation time than predicate K151750 (Blue).
Warning statement (Do Not Use with Carmustine and Thiotepa) for subject device is included in Labeling, similar with predicate device.
The subject device has similar thickness with predicate K151750 at palm, and similar length with predicate K151750 (Blue).
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains a logo with the text "ZHONG HONG" on the top line and "PULIN" on the bottom line. To the left of the text is a red square containing two white letters, "Z" and "Z", separated by a vertical white line. The logo appears to be for a company or organization named Zhong Hong Pulin.
The subject device is having identical specification with predicate K151750 (Blue) with thickness at minimum 0.05mm and length at minimum 230mm.
The difference in labeling with additional drugs tested do not affect the safety and effectiveness of the subject device.
The subject device and the predicate device K151750 (Blue) share the same intended use, same Nitrile material, same compliance with ASTM standards. There is no difference between the subject device and the predicate device K151750 (Blue) with respect to intended use, nonclinical performance data and technological characteristics.
Consequently, the gloves that are the subject of this submission are substantially equivalent to a legally marketed glove K151750 (Blue).
#### Substantial Equivalent Based on Assessment of Clinical Performance Data
Clinical data is not needed for this submission.
#### Biocompatibility
Biocompatibility tests indicated that under the conditions of the studies, the gloves were nonsensitizing and non-irritating.
#### Legally Marketed Device to which Substantial Equivalence is Claimed
The legally marketed predicate device in scope is as follows:
K151750 - Powder Free Nitrile Patient Examination Glove, White Colored, Non-Sterile, Powder Free Nitrile Patient Examination Glove, Blue Colored, Non-Sterile
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.