The subject device intended to measure the diastolic, systolic blood pressures and pulse rate of an adult individual in hospitals, hospital-type facilities and home environments by using a non-invasive oscillometric technique in which an inflatable cuff (size: 12.5~21.5cm(4.9~8.5in) is wrapped around the single wrist. The Subject device is not intended to be diagnostic device.
Device Story
Battery-powered, non-invasive oscillometric blood pressure monitor; wrist-worn inflatable cuff (12.5-21.5cm). Device automatically performs inflation, deflation, and measurement of systolic/diastolic pressure and pulse rate. Used in hospitals, hospital-type facilities, and home environments by adults. Operation involves wrapping cuff around single wrist; device processes pressure oscillations to calculate BP and pulse. Results displayed on LCD for user review. Provides non-diagnostic monitoring of vital signs; aids in tracking blood pressure trends.
Clinical Evidence
Clinical validation study conducted per ISO 81060-2:2013. Study included 87 subjects (36 male, 51 female) using same-arm sequential method. Manual mercury sphygmomanometer used as reference. Results confirmed accuracy within acceptable scope specified by ISO 81060-2. No adverse events reported.
Technological Characteristics
Materials: ABS, PMMA. Power: 2x AAA alkaline batteries. Measurement: Oscillometric technique. Cuff size: 12.5-21.5cm. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 10993-5, ISO 10993-10, IEC 80601-2-30, ISO 81060-2.
Indications for Use
Indicated for adult individuals requiring measurement of diastolic and systolic blood pressure and pulse rate in hospital or home settings. Not intended for diagnostic use.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 10, 2018
Shenzhen Combei Technology Co., Ltd. % Ms. Migo Yang, Consultant Shenzhen Joyantech Consulting Co., Ltd. 1122#. International Mavor Communication Center. Baishizhong Rd 55#, Nanshan District, Shenzhen, Guangdong 518053 CHINA
Re: K171833
Trade/Device Name: Digital Blood Pressure Monitor-Wrist Style, with 14 models: BP800W, BP603W, BP880W, BP885W, BPCBOA-2F, BP850W, BP300W, BP810W, BP602W, BP608W, BP606W, BP660W, BP830W, BP866W Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: November 11, 2017 Received: December 5, 2017
Dear Ms. Migo Yang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
M.A. Wilhelm
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K171833
#### Device Name
Digital Blood Pressure Monitor: Wrist Style: BP800W, BP603W, BP885W, BPCB0A-2F, BP850W, BP300W, BP810W, BP602W, BP608W, BP600W, BP660W, BP830W, BP866W
#### Indications for Use (Describe)
The subject device is intended to measure the diastolic, systolic blood pressures and pulse rate of an adult individual in hospitals, hospital-type facilities and home environments by using a non-invasive oscillometric technique in which an inflatable cuff(size: 12.5-21.5cm(4.9-8.5in) is wrapped around the single wrist.
The Subject device is not intended to be diagnostic device.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> |
| Over-The-Counter Use (21 CFR 801 Subpart C) | <input checked="" type="checkbox"/> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(k) information is submitted as required by requirements of SMDA and 21 CFR §807.92.
## 1 Administrative Information
| Submission Date | Oct 8, 2016 |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer<br>information | Submitter's Name: Shenzhen Combei Technology Co.,Ltd<br>Address: Floor 5, Block B, Building G, Jinxiongda Science<br>Park, South Huanguan Road, Guanlan,<br>Longhua New District, Shenzhen, 518110,<br>Guangdong, China |
| | Contact person: Huaguang.Meng<br>TEL: 0755-29588956<br>FAX: 0755-28588961<br>E-Mail: huaguangmeng@combei.com.cn |
| Submission<br>Correspondent | Contact person: Miss Migo. Yang<br>E-Mail: migo@cefda.com<br>Shenzhen Joyantech Consulting Co., Ltd.<br>1122#,International Mayor Communication Center,<br>Baishizhong Road 55#, Nanshan District, Shenzhen,<br>Guangdong, P.R.China. |
| 历大 17 | Contact person: Mr. Field.Fu<br>E-Mail: cefda13485@163.com<br>Shenzhen Joyantech Consulting Co., Ltd.<br>1122#, International Mayor Communication Center,<br>Baishizhong Road 55#, Nanshan District, Shenzhen,<br>Guangdong, P.R.China |
| Establishment<br>registration number | NA |
## 2 Device Information
| Common name of the device | System, Measurement, Blood-Pressure, Non-Invasive |
|---------------------------|---------------------------------------------------|
| Trade name of the device | Digital Blood Pressure Monitor- Wrist Style; |
| Type/Model of the | Wrist Style: BP800W, BP603W, BP880W, BP885W, |
{4}------------------------------------------------
Shenzhen Combei Technology Co.,Ltd
Version: A/0
| device | BPCB0A-2F, BP850W, BP300W, BP810W, BP602W,<br>BP608W, BP606W, BP660W, BP830W, BP866W; |
|-------------------------------|---------------------------------------------------------------------------------------|
| | Classification panel: Cardiovascular |
| Classification<br>information | Classification name: System, Measurement, Blood-<br>Pressure, Non-Invasive |
| | Regulation Number: 870.1130 |
| | Device Class: II |
| | Product Code: DXN |
| type of 510(k)<br>submission | Traditional |
## 3 Predicate Device Information
Fudakang Industrial Co., Ltd Sponsor: Wrist blood pressure monitor Device: K150430 510(K) Number: |
#### 4 Device Descriptions
The Subject device is battery driven automatic non-invasive Blood Pressure Monitor. It consists of the Main Control Unit, LCD and attachments. The device can automatically complete the inflation, deflation and BP measurement, which intended to measure the diastolic, systolic blood pressures and pulse rate for an adult individual via the oscillometric technique.
The wrist style utilizes an inflatable cuff that is wrapped around the single wrist; the cuff circumference is limited to: 12.5~21.5cm(4.9~8.5in).
## 5 Intended Use
The subject device intended to measure the diastolic, systolic blood pressures and pulse rate of an adult individual in hospitals, hospital-type facilities and
{5}------------------------------------------------
| | VOL_005_510(k) Summary |
|------------------------------------|------------------------|
| Shenzhen Combei Technology Co.,Ltd | 001_510(k) Summary |
| Product: NIBP | Version: A/0 |
home environments by using a non-invasive oscillometric technique in which an inflatable cuff (size: 12.5~21.5cm(4.9~8.5in) is wrapped around the single wrist. The Subject device is not intended to be diagnostic device.
#### 6 Indications for Use
The subject device intended to measure the diastolic, systolic blood pressures and pulse rate of an adult individual in hospital-type facilities and home environments by using a non-invasive oscillometric technique in which an inflatable cuff (size: 12.5~21.5cm(4.9~8.5in) is wrapped around the single wrist. The Subject device is not intended to be diagnostic device.
{6}------------------------------------------------
# 7 SE Comparison
| Characteristics | Subject device | Predicate device<br>(k150430) | Remark |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Device Name | Digital Blood Pressure Monitor-<br>Wrist Style | Wrist Blood Pressure<br>Monitor | NA |
| Device Model | BP800W, BP603W, BP880W,<br>BP885W, BPCB0A-2F,<br>BP850W, BP300W, BP810W,<br>BP602W, BP608W, BP606W,<br>BP660W, BP830W, BP866W | FT-B11W, FT-B12W, FT-<br>B13W, FT-B14W, FTB21Y,<br>FT-B22Y, FT-B11W-V, FT-<br>B12W-V, FT-B13W-V, FT-<br>B14W-V, FT-B21Y-V, FT-<br>B22Y-V, FT-B13W-UR, FT-<br>B13W-BT | NA |
| Manufacturer | Shenzhen Combei Technology<br>Co.,LTD | Fudakang Industrial Co.,<br>Ltd | NA |
| Intended Use/<br>Indication for<br>Use | The subject device intended<br>to measure the diastolic,<br>systolic blood pressures and<br>pulse rate of an adult<br>individual in hospitals,<br>hospital-type facilities and<br>home environments by using<br>a non-invasive oscillometric<br>technique in which an<br>inflatable cuff (available size:<br>12.5~21.5cm(4.9~8.5in) is<br>wrapped around the single<br>wrist.<br>The Subject device is not<br>intended to be diagnostic<br>device. | Fudakang Arm Blood<br>Pressure Monitor and Wrist<br>Blood Pressure Monitor are<br>non-invasive blood<br>measurement system<br>intended to measure the<br>diastolic, systolic blood<br>pressures and pulse rate of<br>an adult individual in<br>hospitals, hospital-type<br>facilities and home<br>environments. | same |
| | The subject device intended<br>to measure the diastolic,<br>systolic blood pressures and<br>pulse rate of an adult<br>individual in hospitals,<br>hospital-type facilities and<br>home environments by using<br>a non-invasive oscillometric<br>technique in which an<br>inflatable cuff (available size:<br>12.5~21.5cm(4.9~8.5in) is<br>wrapped around the single<br>wrist.<br>The Subject device is not | Fudakang Arm Blood<br>Pressure Monitor and Wrist<br>Blood Pressure Monitor are<br>not intended to be<br>diagnostic device | |
| Characteristics | Subject device | Predicate device<br>(k150430) | Remark |
| | intended to be diagnostic<br>device. | | |
| Intended<br>Population | adult | | same |
| Intended<br>Anatomical<br>site | wrist | | same |
| Prescription &<br>OTC | OTC | | same |
| Working<br>Principle | Oscillometric method | | same |
| Internal Power<br>supply | 2- size "AAA" alkaline Batteries | | same |
| Working<br>Current | ≤30mV | | same |
| Contact<br>Material | ABS, PMMA | | same |
| Memory<br>Function | 2×120 memory | 1×60 memory | SE |
| Cuff Size | 12.5cm~21.5cm (4.5n~8.5in) | 135mm~195mm<br>(5.3in~7.7in) | Note01 |
| Measuring<br>range | Pressure: 30 to 280 mmHg (in 1<br>mmHg increment); | Pressure:<br>(40mmHg~280mmHg); | Note02 |
| | Pulse: 40 to 200 beat/minute | Pulse rate range(40-160)<br>hypo/minute | |
| Accuracy | Pressure: ±3mmHg; Pulse: ±5% | Pressure: ±5mmHg; Pulse<br>±5%. | SE |
| Operating<br>Environment | 5~40℃,<br>15%~85%RH | | same |
| Storage &<br>Transport<br>Environment | -10~55℃<br>10%~95%RH | | same |
#### Table 1. Substantial Equivalence Comparison
{7}------------------------------------------------
Shenzhen Combei Technology Co.,Ltd
#### Product: NIBP
Version: A/0
Note01 : Similar : The subject device has larger wrist circumference than predicate device; Both the subject device and predicate device have been tested by ISO81060-2 and clinical test as qualified.
Note02 : Similar : The subject device has a larger measuring range of pressure and pulse than predicate device. Both the subject device and predicate device have been tested by ISO81060-2 and clinical test as qualified.
The subject device is as same as predicate device in Working Principle, Intended use/Indications for Use, Intended patient population, Intended application site, Intended Environments, Working Current, Cuff size, Operating Environment
{8}------------------------------------------------
Shenzhen Combei Technology Co.,Ltd
Storage & Transport Environment. Only their Measuring range are a little bit different. However, the minor difference does not raise any safety or effectiveness issue based on tests.
Thus, the subject device is Substantially Equivalent (SE) to the predicate device which is legally marketed in US.
#### 8 Brief discussions of the non-clinical tests
The subject device conforms to the following guidances and standards:
- Non-Invasive Blood Pressure (NIBP) Monitor Guidance や
- IEC 60601-1:2005+A1:2012: Medical Electrical Equipment Part 1: General や Requirements For Basic Safety And Essential Performance;
- ゃ IEC 60601-1-2:2014 Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests.
- IEC 60601-1-11: 2010 Medical Electrical Equipment Part 1-11: General � Requirements For Basic Safety And Essential Performance - Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Systems Used In The Home Healthcare Environment;
- ISO 10993-5: 2009 /(R)2014 Biological evaluation of medical devices Part 5: ゃ Tests for In Vitro cytotoxicity;
- ゃ ISO 10993-10: 2010 Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization;
- ゃ IEC 80601-2-30: 2013 Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers
## 9 Brief discussions of clinical tests
- ISO 81060-2:2013 Non-invasive sphygmomanometers Part 2: Clinical や validation of automated measurement type;
{9}------------------------------------------------
| Shenzhen Combei Technology Co.,Ltd |
|------------------------------------|
| Product: NIBP |
Version: A/0
In this clinical investigation, eighty seven patients (36 males and 51 females) participated in the clinical study. Same arm sequential method was adopted during the clinical testing. The manual Mercury Sphygmomanometer was used as a reference device. All the subjects were volunteer to take part in the clinical study, all the subjects completed the clinical study without any AE or side-effect. The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2.
## 10 Other information (such as required by FDA guidance)
No other information.
## 11 Conclusions
The subject device: Digital Blood Pressure Monitor-Wrist Style; manufactured by Shenzhen Combei Technology Co.,Ltd is respectively substantially equivalent to the predicate device Wrist Blood Pressure Monitor manufactured by Fudakang Industrial CO.,LTD(K150430).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.