NuvoMed Wireless Blood Glucose Monitoring System (Model BGM-6/0352)

K171822 · Biotest Medical Corporation · NBW · Sep 20, 2017 · Clinical Chemistry

Device Facts

Record IDK171822
Device NameNuvoMed Wireless Blood Glucose Monitoring System (Model BGM-6/0352)
ApplicantBiotest Medical Corporation
Product CodeNBW · Clinical Chemistry
Decision DateSep 20, 2017
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2

Indications for Use

The NuvoMed Wireless Blood Glucose Monitoring System (Model BGM-6/0352) is intended for single-patient use only.

Device Story

System measures glucose in capillary whole blood via amperometric test strips; reaction between glucose and reagents produces electrical current proportional to glucose concentration. Meter calculates concentration and displays result on LCD. Used at home by patients with diabetes for self-monitoring. Features Bluetooth 4.0 for wireless transmission of results to iOS app; replaces previous RS232 cable download. Includes error detection (e.g., temperature, battery, insufficient sample, E8 for setup failure). User performs test by inserting strip, applying blood, and viewing result. Data helps patients manage diabetes control programs. Device requires manual cleaning/disinfection with EPA-registered wipes.

Clinical Evidence

Bench testing only. Verification and validation activities included usability testing for Bluetooth data transmission and robustness testing for cleaning/disinfection procedures. Safety, reliability, EMC, and FCC testing were performed per applicable standards.

Technological Characteristics

Electrochemical glucose sensing. Components: PCB, microcontroller, positive LCD, Bluetooth module. Housing: plastic (white). Connectivity: Bluetooth wireless data transmission. Power: battery-operated. Sterilization: N/A (non-sterile).

Indications for Use

Indicated for single-patient use for the quantitative measurement of glucose in capillary whole blood from the finger, palm, or forearm as an aid in monitoring the effectiveness of a diabetes control program.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Submission Summary (Full Text)

{0} # **SPECIAL 510(k): Device Modification** **OIR Decision Summary** **To:** THE FILE **RE:** DOCUMENT NUMBER k171822 This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary): 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. SuperCheck Plus Blood Glucose Monitoring System (Model 5228-S); k141351 2. Submitter's statement that the **INDICATION/INTENDED USE** of the modified device as described in its labeling **HAS NOT CHANGED** along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). 3. A description of the device **MODIFICATION(S)**, including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the **FUNDAMENTAL SCIENTIFIC TECHNOLOGY** of the modified device **has not changed**. **This change was for** a. A change in the name of the device from the SuperCheck Plus Blood Glucose Monitoring System (Model 5228-S) to the NuvoMed Wireless Blood Glucose Monitoring System (Model BGM-6/0352) b. New printed circuit board with new microcontroller unit c. Change to positive display liquid crystal display and reposition of the control test flag (cont) on meter display d. Change to Bluetooth mediated data transmission and removal of RS232 phone jack port for cable data download to personal computer e. An E_8 error message added to indicate that users fail to reset date and time after installing or changing batteries in device's "Start Up" mode f. Change of meter's housing color from black to white 4. **Comparison Information** (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics, and performance characteristics of the NuvoMed Wireless Blood Glucose Monitoring System (Model BGM-6/0352). 5. A **Design Control Activities Summary** which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied 1 {1} The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device. The NuvoMed Wireless Blood Glucose Monitoring System (Model BGM-6/0352) is intended for single-patient use only. Disinfection efficacy studies described for the predicate device (k141351 for SuperCheck Plus Blood Glucose Monitoring System, Model 5228-S) using Clorox Bleach Germicidal Wipes (EPA registration number 67619-12) demonstrated complete inactivation of live Hepatitis B Virus (HBV) on the materials of the meter. Robustness accuracy studies performed for the NuvoMed Wireless Blood Glucose Monitoring System (Model BGM-6/0352) demonstrate that there was no change in performance or in the external materials of the meter after 1100 cleaning and 1100 disinfection cycles designed to simulate 3 years of single-patient use. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures. This device was cleared after the FDA issued final guidance documents for prescription use blood glucose monitoring systems (BGMS) and over-the-counter use blood glucose monitoring systems (SMBG). However, the recommendations in the guidance documents were not followed for this device since the studies were performed prior to the finalization of the guidance documents. 2
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...