K171817 · Cook Biotech Incorporated · FTM · Mar 13, 2018 · General, Plastic Surgery
Device Facts
Record ID
K171817
Device Name
Biodesign Diaphragmatic Hernia Graft
Applicant
Cook Biotech Incorporated
Product Code
FTM · General, Plastic Surgery
Decision Date
Mar 13, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Biodesign® Diaphragmatic Hernia Graft is intended for implantation to reinforce soft tissues where weakness exists including the repair of diaphragmatic/hiatal hernias. The graft is supplied sterile and is intended for one time use only.
Device Story
Biodesign Diaphragmatic Hernia Graft is a surgical mesh implant; composed of porcine small intestinal submucosa (SIS) processed into an acellular, collagen-rich extracellular matrix. Device is multilayered, perforated, and reinforced with polyglycolic acid (PGA) suture to prevent delamination. Supplied sterile in double Tyvek pouches; ethylene oxide sterilized. Intended for surgical implantation by clinicians to reinforce soft tissue in diaphragmatic/hiatal hernia repairs. Prior to implantation, graft is rehydrated with sterile saline or autologous body fluids (e.g., blood, bone marrow aspirate, platelet concentrate). Upon implantation, device provides mechanical reinforcement; over time, it remodels into host tissue, leaving no graft material behind. Subject device is physically identical to predicate; modification is limited to expanded rehydration fluid options in instructions for use.
Clinical Evidence
No clinical data. Bench testing only. Verification testing compared mechanical properties (tensile strength, suture retention strength, hydration time, handling) of the device rehydrated with saline versus autologous body fluids. Results demonstrated no significant differences in performance.
Technological Characteristics
Material: Porcine small intestinal submucosa (SIS) extracellular matrix. Components: PGA suture. Configuration: Multilayered sheets (rectangular or U-cut). Sterilization: Ethylene Oxide (EO). Connectivity: None. Energy source: None (passive implant).
Indications for Use
Indicated for patients requiring soft tissue reinforcement for the repair of diaphragmatic or hiatal hernias.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below.
March 13, 2018
Cook Biotech Incorporated Nick Wang, Ph.D. Regulatory Scientist 1425 Innovation Place West Lafayette, Indiana 47906
Re: K171817
Trade/Device Name: Biodesign Diaphragmatic Hernia Graft Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: OWV, FTM Dated: December 11, 2017 Received: December 13, 2017
Dear Dr. Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may: therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K171817
Device Name Biodesign® Diaphragmatic Hernia Graft
Indications for Use (Describe)
The Biodesign® Diaphragmatic Hernia Graft is intended for implantation to reinforce soft tissues where weakness exists including the repair of diaphragmatic/hiatal hernias.
The graft is supplied sterile and is intended for one time use only.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|----------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### I. SUBMITTER
Cook Biotech Incorporated 1425 Innovation Place West Lafayette, IN 47906
Phone: (765) 497-3355
Contact Person: Perry W. Guinn Date Prepared: 12 February 2018
#### DEVICE II.
| Name of Device: | Biodesign® Diaphragmatic Hernia Graft |
|-----------------------|---------------------------------------|
| 510(k) Number: | K171817 |
| Common or Usual Name: | Surgical Mesh |
| Classification Name: | Surgical Mesh |
| Regulation number: | 21 CFR §878.3300 |
| Product Code: | OWV, FTM |
| Regulatory Class: | Class II |
## III. PREDICATE DEVICE
Biodesign® Diaphragmatic Hernia Graft (K133011, Cook Biotech Incorporated)
## IV. REFERENCE DEVICE
Meso Bilayer Surgical Mesh (K132025, DSM Biomedical)
#### DEVICE DESCRIPTION V.
The Biodesign® Diaphragmatic Hernia Graft, the subject device, has the same indications for use and is physically identical to the predicate device (K133011) with the same name. The only difference between the subject device and the predicate device is the rehydration fluid options described in the devices' instructions for use (IFU).
Both the subject and predicate devices are part of a family of implant devices manufactured from porcine small intestine that have been processed to remove the tunica mucosa from the inner intestinal surface and the serosa and tunica muscularis from the outer intestinal surface. After further disinfection to ensure the material is virally inactive, the resulting layer is a three dimensional, acellular, collagen-rich extracellular matrix (ECM) that is termed small intestinal submucosa (SIS). Both the subject and predicate
{4}------------------------------------------------
devices are multilayered SIS sheets configured in two shapes: rectangular shape (7 x 10 cm) and the U-cut shape (7 x 10 cm with U-shaped notch). Both devices are perforated and have polyglycolic acid (PGA) suture stitched across the device and around the perimeter. The purpose of the suture stitching is to prevent delamination due to device handling during implantation. Both devices are provided in a double Tyvek® pouch and are ethylene oxide (EO) sterilized.
Prior to implantation, the Biodesign® Diaphragmatic Hernia Graft, the subject device, can be rehydrated with sterile saline or autologous body fluids (e.g., blood, bone marrow aspirate, or blood concentrates such as platelet concentrate).
Upon implantation, the Biodesign® Diaphragmatic Hernia Graft (both subject and predicate device) will provide mechanical reinforcement of soft tissue in the repair of diaphragmatic/hiatal hernias. Over time, the device will incorporate (remodel) into the body such that no graft material is left behind.
### VI. INDICATIONS FOR USE
The Biodesign® Diaphragmatic Hernia Graft is intended for implantation to reinforce soft tissues where weakness exists including the repair of diaphragmatic/hiatal hernias.
The graft is supplied sterile and is intended for one time use only.
## VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Biodesign® Diaphragmatic Hernia Graft device is identical to its predicate in that they are both intended to reinforce soft tissues where weakness exists including the repair of diaphragmatic/hiatal hernias. Moreover, the subject device's component materials, fundamental scientific technology, and mode of action are identical to that of the predicate device.
The only modification to the Biodesign® Diaphragmatic Hernia Graft (subject device) is the addition of more rehydration fluid options in the instructions for use (IFU). The subject device includes sterile saline and autologous body fluids (e.g., blood, bone marrow aspirate, or blood concentrates such as platelet concentrate) as possible rehydration fluids whereas the predicate device describes rehydration with saline. The reference device Meso Bilayer Surgical Mesh (K132025) also indicated for hiatal hernia repairs, shares the same rehydration fluid information (rehydration with saline or autologous body fluids) as the subject device. Please see Table 5-1 for a more detailed comparison of the subject, predicate and reference devices.
{5}------------------------------------------------
| Device | BiodesignⓇ<br>Diaphragmatic<br>Hernia Graft<br>(subject device) | Biodesign® Diaphragmatic<br>Hernia Graft<br>(predicate device) | Meso Bilayer Surgical Mesh<br>(reference device) |
|-----------------------------|-----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | CBI | CBI | DSM Biomedical |
| 510 (k)<br>Number | K171817 | K133011 | K132025 |
| Product Code | | FTM | FTM, OXH |
| Intended Use | Unchanged from<br>predicate device | For implantation to<br>reinforce soft tissues where<br>weakness exists including<br>the repair of<br>diaphragmatic/hiatal hernias | For implantation to reinforce<br>soft tissues where weakness<br>exists in patients requiring soft<br>tissue repair and reinforcement<br>in plastic and reconstructive<br>surgery including but not<br>limited to the following<br>procedures: reinforcement of<br>primary closure such as suture<br>line reinforcement and muscle<br>flap reinforcement; hernia repair<br>(e.g. hiatal, femoral,<br>paracolostomy, umbilical) |
| Materials and<br>Components | | Small intestinal submucosa<br>PGA suture | Porcine tissue and synthetic<br>absorbable polymer |
| Device<br>Configuration | | 7x 10 cm size (rectangular<br>and U-cut shape) | Unknown (not specified in<br>510(k) summary) |
| Shelf-Life | | 18 months | 36 months |
| Packaging | | foil pouch | Double peel packages |
| Sterilization | | Ethylene Oxide (EO) | Ethylene Oxide (EO) |
| One-time Use | | Yes | Yes |
| Rehydration<br>Fluid | Saline and<br>autologous body<br>fluids | Saline | Saline and autologous body<br>fluids |
#### Table 5-1. Substantial Equivalence Information
## PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
#### Non-Clinical testing
Product verification testing was performed on sterilized finished devices to evaluate device mechanical properties after rehydration with body fluids. Comparative tests (rehydration with saline vs. with body fluids) were performed to evaluate the device's
{6}------------------------------------------------
tensile strength, suture retention strength, hydration time and simulated use/handling properties.
Results of the testing showed rehydration with body fluids did not significantly change the mechanical and handling properties of the device.
#### VIII. CONCLUSIONS
For purposes of determinations of substantial equivalence under section 513(i) of the FD&C Act (21 U.S.C. § 360c(i)), the Biodesign® Diaphragmatic Hernia Graft, the subject device, has the same intended use and functions under the same mode of action and fundamental scientific technology as the predicate device with the same name. In addition, the subject device is composed of the same materials and is manufactured using the same processes as the predicate device. The only modification is the addition of autologous body fluids as possible rehydration fluids in the device's IFU. The absence of changes in the fundamental scientific technology and intended use of the device, as well as a review of the risk analysis and completion of verification and validation activities, provide evidence to support the conclusion that the modification does not introduce new risks and that the subject device performs comparably to the predicate device that is currently marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.