← Product Code [MHX](/productcode/MHX) · K171801

# IntelliVue Patient Monitor MX100 and Multi-Measurement Module X3; intelliVue Hemodynamic Extension, Capnography Extension, Microstream Extension;Transpac IV Dual IBP Cable (K171801)

_Philips Medizin Systeme Boeblingen GmbH · MHX · Oct 6, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K171801

## Device Facts

- **Applicant:** Philips Medizin Systeme Boeblingen GmbH
- **Product Code:** [MHX](/productcode/MHX.md)
- **Decision Date:** Oct 6, 2017
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 870.1025
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The monitor is indicated for use by healthcare professionals whenever there is a need for monitoring the physiological parameters of patients. The monitor is intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitor is intended for use by trained healthcare professionals in a hospital environment. The monitor is also intended for use during patient transport inside the hospital environment. The monitor is only for use on one patient at a time. It is not intended for home use. Not a therapeutic device. The monitor is for prescription use only. The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC 11). ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients. The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring. The Masimo rainbow SET measurement is indicated for the nonitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRac). The Masimo rainbow SET measurement is indicated for use during both no motion conditions, and for patients who are well or poorly perfused. The measurement extension adds physiological measurements to a dedicated host device. It is intended for use by trained healthcare professionals for adult, pediatric, and neonatal patients in a hospital environment and for transport inside hospitals. The measurement extension can only function when it is connected to a dedicated host device.

## Device Story

IntelliVue MX100, X3, and Hemodynamic Extension 867039 are multiparameter patient monitors; acquire physiological signals (ECG, SpO2, NBP, IBP, temp, CO2, cardiac output); display waves, trends, and alarms. Used in hospital settings (bedside or transport) by trained healthcare professionals. MX100/X3 feature 6.1-inch touchscreen; X3 functions as standalone or module for host monitors. Hemodynamic extension adds dual IBP, temp, and cardiac output to host. Acceleration sensor detects monitor orientation. Data transmitted via LAN/wireless to Central Station. Provides real-time monitoring to support clinical decision-making; enhances patient safety through continuous surveillance and alarm generation.

## Clinical Evidence

Bench testing only. No clinical data. Verification and validation included electrical/mechanical safety (IEC 60601 series), software life cycle (IEC 62304), usability testing, and performance testing for respiration, cardiac output, and NIBP/SpO2 accuracy per FDA guidance. All pass/fail criteria met.

## Technological Characteristics

Compact monitors with CPU, measurement boards, and 6.1-inch TFT touchscreen. Housing: PBT+PC plastics. Connectivity: LAN/wireless. Power: External supply, battery, or host. Software: AAMI ANSI IEC 62304 compliant. Standards: IEC 60601-1, 60601-1-2, 60601-1-8, 60601-2-25/27/34, ISO 80601-2-55/56/61.

## Regulatory Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

## Predicate Devices

- IntelliVue Patient Monitor MP2 ([K161531](/device/K161531.md))
- Multi-Measurement Module X2 ([K161531](/device/K161531.md))
- IntelliVue Hemodynamic Extension M3012A ([K033444](/device/K033444.md))
- Transpac 4 Reusable Transducer Cable ([K061573](/device/K061573.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 6, 2017

Philips Medizin Systeme Boeblingen GmbH Markus Stacha Sr. Regulatory Affairs Engineer Hewlett-Packard-Str.-2 Boeblingen, DE 71034 BW

# Re: K171801

Trade/Device Name: IntelliVue Patient Monitor MX100. Multi-Measurement Module X3 and IntelliVue Hemodynamic Extension 867039 Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm) Regulatory Class: Class II Product Code: MHX, DSI, MLD, DOA, DXN, FLL, DRT, DSJ Dated: September 4, 2017 Received: September 6, 2017

# Dear Markus Stacha:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.qov

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number: K171801

Device Name: IntelliVue Patient Monitor MX100, Multi-Measurement Module X3, and IntelliVue Hemodynamic Extension 867039

Indications for Use (Describe)

Indications for Use statement for the MX100 and X3:

The monitor is indicated for use by healthcare professionals whenever there is a need for monitoring the physiological parameters of patients.

The monitor is intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitor is intended for use by trained healthcare professionals in a hospital environment.

The monitor is also intended for use during patient transport inside the hospital environment.

The monitor is only for use on one patient at a time. It is not intended for home use. Not a therapeutic device. The monitor is for prescription use only.

The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC 11).

ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.

The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.

| Type of Use (Select one or both, as applicable) |                                               |
|-------------------------------------------------|-----------------------------------------------|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D)  | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

{3}------------------------------------------------

# Indications for Use (continued)

510(k) Number: K171801

The Masimo rainbow SET measurement is indicated for the nonitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRac). The Masimo rainbow SET measurement is indicated for use during both no motion conditions, and for patients who are well or poorly perfused.

Indications for Use statement for the Hemodynamic Extension 867039:

The measurement extension adds physiological measurements to a dedicated host device. It is intended for use by trained healthcare professionals for adult, pediatric, and neonatal patients in a hospital environment and for transport inside hospitals.

The measurement extension can only function when it is connected to a dedicated host device.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, providing a strong contrast that makes the word stand out.

## 510(k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.

- 1. The submitter of this Premarket Notification is:
Markus Stacha Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany Fax: +49 7031 463-2442 Tel: +49 7031 463-2840 e-mail: markus.stacha@philips.com

This summary was prepared on October 04, 2017.

- 2. The name and classification of the devices:
Trade names: IntelliVue Patient Monitor MX100 and Multi-Measurement Module X3. Common name: Multiparameter Patient Monitor.

Trade name: IntelliVue Hemodynamic Extension 867039. Common name: Multifunction Patient Monitor Module.

| Device Panel              | Classification | ProCode | Description                                                                 |
|---------------------------|----------------|---------|-----------------------------------------------------------------------------|
| Cardiovascular<br>Devices | §870.1025, II  | MHX     | Monitor, Physiological, Patient<br>(with arrhythmia detection or<br>alarms) |
| Cardiovascular<br>Devices | §870.1025, II  | DSI     | Detector and alarm, arrhythmia                                              |
| Cardiovascular<br>Devices | §870.1025, II  | MLD     | Monitor, ST Segment with Alarm                                              |
| Cardiovascular<br>Devices | §870.1100, II  | DSJ     | Alarm, Blood Pressure                                                       |
| Cardiovascular<br>Devices | §870.1110, II  | DSK     | Computer, Blood Pressure                                                    |
| Cardiovascular<br>Devices | §870.1130, II  | DXN     | System, Measurement, Blood-<br>Pressure, Non-Invasive                       |
| Cardiovascular<br>Devices | §870.1435, II  | DXG     | Computer, Diagnostic, Pre-<br>Programmed, Single-Function                   |
| Cardiovascular<br>Devices | §870.1915, II  | KRB     | Probe, Thermodilution                                                       |
| Cardiovascular<br>Devices | §870.2300, II  | DRT     | Monitor, Cardiac (incl.<br>Cardiotachometer & Rate Alarm)                   |
| Cardiovascular<br>Devices | §870.2060, II  | DRQ     | Amplifier and Signal Conditioner,<br>Transducer Signal                      |
| Cardiovascular<br>Devices | §870.2340, II  | DPS     | Electrocardiograph                                                          |
| Cardiovascular<br>Devices | §870.2340, II  | MLC     | Monitor, ST Segment                                                         |
| Cardiovascular<br>Devices | §870.2600, I   | DRJ     | System, Signal Isolation                                                    |
| Cardiovascular<br>Devices | §870.2700, II  | DQA     | Oximeter                                                                    |
| Cardiovascular<br>Devices | §870.2770, II  | DSB     | Plethysmograph, Impedance                                                   |
| Cardiovascular<br>Devices | §870.2900, II  | DSA     | Cable, Transducer and Electrode,<br>incl Patient Connector                  |

### Classification:

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out.

| K171801                                         |                |         |                                                                        |
|-------------------------------------------------|----------------|---------|------------------------------------------------------------------------|
| Device Panel                                    | Classification | ProCode | Description                                                            |
|                                                 | -              | MSX     | System, Network and<br>Communication, Physiological<br>Monitors        |
|                                                 | §870.2910, II  | DRG     | Transmitters and Receivers,<br>Physiological Signal,<br>Radiofrequency |
| Anesthesiology<br>Devices                       | §868.1400, II  | CCK     | Analyzer, Gas, Carbon Dioxide,<br>Gaseous-Phase                        |
|                                                 | §868.2375, II  | BZQ     | Monitor, Breathing Frequency                                           |
| General Hospital<br>and Personal Use<br>Devices | §880.2910, II  | FLL     | Thermometer, Electronic, Clinical                                      |

3. The new IntelliVue Patient Monitor MX100 and Multi-Measurement Module X3 are substantially equivalent to the legally marketed IntelliVue Patient Monitor MP2 and Multi-Measurement Module X2, marketed pursuant to K161531. The new IntelliVue Hemodynamic Extension 867039 is substantially equivalent to the legally marketed IntelliVue Hemodynamic Extension M3012A, marketed pursuant to

K033444. The new Transpac IV Dual IBP Cable 453564588501 is substantially equivalent to the legally marketed Transpac® 4 Reusable Transducer Cable for Philips Monitors 989803177921 manufactured by ICU Medical, marketed pursuant to K061573.

### 4. Description of the new devices

The new IntelliVue Patient Monitor MX100, Multi-Measurement Module X3, and Hemodynamic Extension 867039 acquire multiple physiological patient signals, display measurement values, waves and trends, generate physiological and technical alarms, provide data recording and support patient data management. The devices offer a monitoring solution optimized for the surgical, cardiac, medical and neonatal care environments. They can be located in the patient vicinity at the bedside or can be used mobile, during patient transport inside hospitals. The measurement sensors are applied at diverse bodily locations, depending on the actual physiological parameters monitored.

The new devices have the same range of functions as the legally marketed predicate devices. They use the same measurement parameters as the predicates.

Whereas the predicate devices optionally provide one invasive pressure, the new devices optionally provide two invasive pressures (dual pressure) that can be measured with one pressure connector.

The MX100 and X3 provide multiple non-invasive and invasive measurements: ECG (including arrhythmia and ST), respiration, SpO2, NBP, dual invasive pressure, temperature, and CO2.

The Hemodynamic Extension 867039 adds optional measurements: dual invasive pressure, temperature, and cardiac output /continuous cardiac output to the host device. The measurement extension can only function when it is connected to a dedicated host device.

{6}------------------------------------------------

Specifications of all measurement characteristics, including measurement principles, methods, algorithms, and all detailed performance specifications and measurement alarm specifications, are the same as those in the predicate devices. The new devices reuse unchanged existing accessories of the predicates.

The new IntelliVue Patient Monitor MX100 and Multi-Measurement Module X3 are compact, rugged, lightweight monitors with built-in measurements. The X3 can be used in two ways: as a multi-measurement module for the Philips IntelliVue family of patient monitors and as a stand-alone monitor.

The MX100 and X3 have a 6.1' TFT flat panel color display with a multi-finger touchscreen as input device.

The MX100 and X3 can interact with the Central Station via LAN or wireless link.

The new Transpac IV Dual IBP Cable 453564588501 is a reusable cable intended to connect two ICU Transpac pressure transducers with one of the new Philips measuring devices (Patient Monitor MX100, Multi-Measurement Module X3, or Hemodynamic Extension 867039). The cable transmits analog voltage signals from the pressure transducers to the Philips device and provides electrical power from the Philips device to the transducers.

### 5. Intended Use

The Intended Use of the new Philips IntelliVue Patient Monitor MX100, Multi-Measurement Module X3 and Hemodynamic Extension 867039 has not changed as a result of the device modification. The indications for use statement of the new devices is the same as that of the predicate devices, except for the environment of use. Whereas the predicate IntelliVue models MP2 and X2 can be used during patient transport inside and outside of the hospital environment, the use of the new Intelli Vue MX100 and X3 is limited to the environment inside hospitals.

Indications for Use statement for the MX100 and X3:

The monitor is indicated for use by healthcare professionals whenever there is a need for monitoring the physiological parameters of patients.

The monitor is intended to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates. The monitor is intended for use by trained healthcare professionals in a hospital environment.

The monitor is also intended for use during patient transport inside the hospital environment.

The monitor is only for use on one patient at a time. It is not intended for home use. Not a therapeutic device. The monitor is for prescription use only.

The ECG measurement is intended to be used for diagnostic recording of rhythm and detailed morphology of complex cardiac complexes (according to AAMI EC 11). ST segment monitoring is intended for use with adult patients only and is not clinically validated for use with neonatal and pediatric patients.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is plain white.

The Integrated Pulmonary Index (IPI) is intended for use with adult and pediatric (1 to 12 years) patients only. The IPI is an adjunct to and not intended to replace vital sign monitoring.

The Masimo rainbow SET measurement is indicated for the noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRac). The Masimo rainbow SET measurement is indicated for use during both no motion and motion conditions, and for patients who are well or poorly perfused.

Indications for Use statement for the Hemodynamic Extension 867039:

The measurement extension adds physiological measurements to a dedicated host device. It is intended for use by trained healthcare professionals for adult, pediatric, and neonatal patients in a hospital environment and for transport inside hospitals. The measurement extension can only function when it is connected to a dedicated host device.

#### 6. Technological Characteristics

The new IntelliVue Patient Monitor MX100, Multi-Measurement Module X3 and Hemodynamic Extension 867039 have the same relevant-technological characteristics as their predicate devices (as stated in Item 3 above) with regard to operating principles, mechanism of action, energy sources, portability, user interface, robustness. They all have very similar dimensions, weight and environmental specifications.

The new MX100 and X3 and their predicates MP2 and X2 have the same hardware and software architecture. Both the new and the predicate devices are compact monitoring devices with a built-in central processing unit, measurement boards, and flat panel display with touch-screen.

The new measurement extension and the predicate measurement extension have the same hardware architecture. Both the new and the predicate extension are compact multi-parameter measuring devices without display and control elements, with built-in power board and measurement boards.

The new MX100 and X3 and their predicates MP2 and X2 can be powered from the same sources: dedicated external power supply, internal battery, or connected host monitor (new X3 and predicate X2 only).

The new measurement extension and the predicate extension are powered from the connected measuring host device via proprietary Philips MSL interface.

Whereas the housings of the predicate devices mainly consist of ABS + PC-FR plastics, the housing of all new devices mainly consist of PBT + PC plastics. The new housing materials enhance the ruggedness of the new devices, without compromising the electrical characteristics, by making them more resistant against mechanical and chemical stress (cleaning detergents).

The new MX100 and X3 have a 6.1 inch state-of-the-art TFT flat panel display with 1024x480 resolution and wide viewing angle. The display is covered by the very

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, providing a strong contrast that makes the word stand out.

resistant and easy to clean Corning® Gorilla® Glass, compared to the polyester film that covers the display of the predicates MP2 and X2. The new MX100 and X3 are supplied with a capacitive multi-touch screen as input device. The multi-touch screen allows use of operating controls such as swiping and multi-finger gestures.

The new devices use modern internal electronic components such as CPU and memory system. A state-of-the-art acceleration sensor is used to determine the monitor position for the adaption of the display orientation.

- 7. Summary of V&V activities
The new devices have been subject to a series of V&V activities:

- Testing according to the recognized consensus standards: ।
	- IEC 60601-1-2: 2007 (Ed. 3) (Electromagnetic Compatibility) .
	- AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012 (Ed. 3.1) (Basic safety and ● essential performance)
	- IEC 60601-2-25: 2011 (Ed. 2) (Electrocardiographs) ●
	- IEC 60601-2-27: 2011 (Ed. 3) (ECG monitoring equipment) ●
	- IEC 80601-2-30: 2013 (Ed. 1.1) (Automated noninvasive sphygmomanometers) ●
	- IEC 60601-2-34: 2011 (Ed. 3) (Invasive blood pressure monitoring equipment)
	- ISO 80601-2-55: 2011 (Ed. 1) (Respiratory gas monitors)
	- ISO 80601-2-56: 2009 (Ed. 1) (Clinical thermometers)
	- . ISO 80601-2-61: 2011 (Ed. 1) (Pulse Oximeters)
	- AAMI ANSI IEC 62304:2006 (Ed. 1) (Software life cycle processes) ●
	- IEC 60601-1-6: 2013 (Ed. 3.1) (Usability) ●
	- IEC 60601-1-8: 2012 (Ed. 2.1) (Alarms) .

All applicable requirements have been met.

- Additional relevant testing according to the device-specific FDA guidance documents:
	- "Non-Invasive Blood Pressure (NIBP) Monitor Guidance": ●
		- o The intra-device variability between a minimum of three devices

Comparison to the intra-arterial reference standard for mean blood pressure.

All applicable requirements have been met.

- "Pulse Oximeters Premarket Notification Submissions [510(k)s] Guidance for . Industry and Food and Drug Administration Staff":
	- o Accuracy of Pulse Oximeters testing that demonstrates that SpO2 and pulse rate values calculated by the OEM system are not corrupted during communication to the host device
	- o Display values, outputs, and indicators.

All applicable requirements have been met.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, providing a strong contrast that makes the word stand out.

- Testing as required by the Hazard Analysis.
All specified pass/fail criteria have been met. The test results have confirmed the effectiveness of the implemented design risk mitigation measures.

- Additional environmental testing (temperature, humidity) and mechanical testing -(mechanical classes 7M1, 7M2, 7M3, and disinfectant resistance) All specified test requirements have been met. The tests have confirmed that the new devices work safely and according to their specifications and indicated claims during tests simulating general hospital conditions, handling and transport in hospital environments, disinfection, and storage.
- Unit level tests, integration tests, functional system level tests, and regression system level tests.

The new devices perform according to the specified criteria:

- Alarm, Measurement, Operating System, Display and Operation Unit Tests ●
- Integration tests
- . Functional tests for Dual Pressure and Operating Controls
- Alarm, Measurement, and Interface Regression Tests .

All specified test requirements have been met. The test results demonstrate that modified and previously available device functions work correctly according to the specifications and labeling claims.

- -Usability and User Preference Testing on the new MX100 and X3. All specified test requirements have been met. The new devices have been found to be safe and effective for the intended users, uses, and use environments.
- -Performance testing of Respiration rate and Cardiac Output (C.O.) All specified test requirements have been met. The new devices perform according to the specified criteria that are the same as those for the predicate devices:
	- Respiration rate accuracy and resolution, respiration wave bandwidth and respiration alarm delay
	- Accuracy and repeatability specifications for right heart and transpulmonary mode
	- . Visibility of thermodilution curve in the cardiac output window on the host monitor where the new X3 is connected
	- Temperature drift limits and small signal warnings.
- Performance testing of the new Transpac IV Dual IBP Cable. -All specified test requirements have been met. The new cable has passed all durability tests in simulated test scenarios:
	- Cleaning and Disinfection ●
	- Mechanical / tensile strength
	- Monitor connector insertion / removal force
	- Operating Temperature ●
	- . Connector insertion cycle.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is white.

# 8. Conclusion

Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the new devices with respect to the predicates.

Testing comprised electrical and mechanical safety tests, environmental tests, performance tests, usability tests, functional and regression tests.

Pass/Fail criteria were based on the specifications cleared for the predicate devices and test results showed substantial equivalence.

The results demonstrate that the new devices meet all defined reliability requirements and performance claims.

---

**Source:** [https://fda-staging.innolitics.com/device/K171801](https://fda-staging.innolitics.com/device/K171801)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
