TRYPTIK2C-Plate® Anterior Cervical Plate System is intended to be used for temporary stabilization of the cervical spine (C2-C7) during the development of solid spinal fusion in patients with instability caused by the following degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumor, spondylolisthesis, deformity (i.e., scoliosis, kyphosis, lordosis), pseudarthrosis, and failed previous fusions
Device Story
TRYPTIK2C-Plate® is an anterior cervical plate system for spinal stabilization. System comprises 1- to 4-level plates (20mm-92mm length) and fixed/variable-angle bone screws (Ø 4.0mm, Ø 4.5mm). Plates attach to anterior cervical vertebral bodies (C2-C7) to provide immobilization as an adjunct to fusion. Features integrated Nitinol expansive rings as a one-step anti-back out locking mechanism for screws. Used in surgical settings by orthopedic or neurosurgeons. Provides mechanical stabilization to facilitate bone healing; benefits patient by maintaining alignment during fusion process. Supplied sterile with dedicated surgical instrumentation.
Clinical Evidence
Bench testing only. Includes static push-out testing of screws and corrosion testing per ASTM F2129-17. No clinical data provided.
Technological Characteristics
Plates and screws: Titanium alloy Ti6Al4V ELI (ISO 5832.3, ASTM F136). Anti-back out ring: Nitinol (ASTM F2063). Dimensions: 2.4mm thick, 16.5mm wide. Mechanical locking: Integrated expansive ring. Sterilization: Gamma. Bacterial endotoxin limit: 20 EU/device.
Indications for Use
Indicated for temporary stabilization of cervical spine (C2-C7) during solid spinal fusion in patients with instability due to degenerative disc disease, trauma (fractures), tumor, spondylolisthesis, spinal stenosis, deformity (scoliosis, kyphosis, lordosis), pseudarthrosis, or failed previous fusions.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
February 26, 2018
SPINEART Franck Pennesi Chief Technical Officer 3 Chemin du Pré Fleuri 1228 Plan Les Ouates Geneva, Switzerland
Re: K171797
Trade/Device Name: TRYPTIK2c-Plate® Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: KWQ Dated: January 26, 2018 Received: January 29, 2018
Dear Franck Pennesi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Vincent J. Devlin -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
Device Name
TRYPTIK2C-Plate® Anterior Cervical Plate System
### Indications for Use (Describe)
TRYPTIK2C-Plate® Anterior Cervical Plate System is intended to be used for temporary stabilization of the cervical spine (C2-C7) during the development of solid spinal fusion in patients with instability caused by the following degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumor, spondylolisthesis, deformity (i.e., scoliosis, kyphosis, lordosis), pseudarthrosis, and failed previous fusions
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) [ ] Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
# TRADITIONAL 510k TRYPTIK2c-Plate Anterior Cervical Plate System
Image /page/3/Picture/2 description: The image shows the logo for Spineart. The logo features the word "spineart" in a stylized font, with the "a" and "r" in a lighter purple color. Above the text is a purple butterfly-like design.
## 510(k) SUMMARY
| 510k | TRADITIONAL |
|-----------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Basis for submission | This submission is intended to address a line extension of the Tryptik®2 Anterior<br>Cervical Plate system and introduce a design change of the clip that constitutes<br>the anti-back out mechanism of the cervical plate. |
| Submitted by | SPINEART<br>3 Chemin du Pré Fleuri<br>1228 PLAN LES OUATES<br>GENEVA SWITZERLAND |
| Contacts | Franck PENNESI Chief Technical Officer<br>Phone : +41 22 570 1200 Fax : +41 22 594 8306<br>Mail : fpennesi@spineart.com<br>Regulatory contact : Dr Isabelle DRUBAIX (Idée Consulting) idrubaix@nordnet.fr |
| Date Prepared | June 12, 2017 |
| Common Name | Anterior Cervical Plate System |
| Trade Name | TRYPTIK2C-Plate Anterior Cervical Plate System |
| Classification Name | Spinal intervertebral body fixation orthosis |
| Class | II |
| Product Code | KWQ Appliance, Fixation, Spinal Intervertebral Body |
| CFR section | 888.3060 Spinal intervertebral body fixation orthosis |
| Device panel | ORTHOPEDIC |
| Legally marketed<br>predicate devices | Primary predicate: Tryptik2c-Plate Anterior Cervical Plate manufactured by<br>Spineart K153042 |
| | Additional predicates: Diamond Anterior Cervical Plate System manufactured by<br>Amendia, Inc (K100265), Spider Cervical Plating (SCP) System manufactured by X-<br>Spine Systems, Inc (K052292) and Vectra System manufactured by Synthes Spine<br>CO.LP (K050451) |
| Indications for use | TRYPTIK2C-Plate Anterior Cervical Plate System is intended to be used for<br>temporary stabilization of the cervical spine (C2-C7) during the development of<br>solid spinal fusion in patients with instability caused by the following<br>degenerative disc disease (as defined by neck pain of discogenic origin with<br>degeneration of the disc confirmed by patient history and radiographic studies),<br>trauma (including fractures), tumor, spondylolisthesis, spinal stenosis, deformity<br>(i.e., scoliosis, kyphosis, lordosis), pseudarthrosis, and failed previous fusions |
| Description of the device | The TRYPTIK2 <span style="vertical-align: super;">C-Plate</span> Anterior Cervical Plate System consists of 1-level, 2-level, 3-level and 4-level plates with cancellous and cortical fixed-angle and variable-angle bone screws. The plate attaches to the anterior portion of the vertebral body of the cervical spine (C2-C7) and functions as an adjunct to fusion to provide immobilization and stabilization of cervical segments of the spine. The Plate configurations are ranging in total lengths from 20mm (1-level) to 92mm (4-level). The plate is 2.4mm thick and 16.5mm wide and incorporates integrated expansive rings for anti-back out of the bone screws that functions as a one-step locking mechanism. The Screw range comes in two diameters, Ø 4.0 and Ø 4.5, and length is ranging from 12mm to 22mm.<br>The TRYPTIK2C-Plate® Anterior Cervical Plates and screws are all made of Titanium alloy Ti6Al4V ELI conforming to ISO 5832.3 and ASTM F136 with an expansive ring made of Nitinol conforming to ASTM F2063.<br>The TRYPTIK2 <span style="vertical-align: super;">C-Plate</span> Anterior Cervical Plates and screws are delivered sterile (gamma sterilization). The TRYPTIK2 <span style="vertical-align: super;">C-Plate</span> Anterior Cervical Plate System is supplied with all the surgical instruments required for its installation. Bacterial endotoxin testing as specified in USP standard is used for pyrogenicity testing to achieve the Endotoxin limit of 20 EU / device |
| Technological<br>characteristics compared<br>to the predicate devices | As was established in this submission the Tryptik2 ACP is substantially equivalent and has the same technological characteristics to its predicate devices in areas including function, material composition, design, range of sizes and mechanical performance. |
| Discussion of Testing | The following non-clinical tests were conducted:<br>- Static push-out testing of the screw per in-house protocol<br>- Corrosion Testing per ASTM F2129-17<br>The result of these studies shows that the Tryptik2 ACP device shows a very good behavior with regards to the pitting corrosion resistance. Tryptik2 ACP meets or exceeds the performance of the predicate device. Therefore the Tryptik2 ACP device is substantially equivalent to its predicate devices. |
| Conclusion | Based on the design features, technological characteristics, feature comparisons, indications for use, and non-clinical performance testing, the Tryptik2 ACP has demonstrated substantial equivalence to the identified predicate devices. |
{4}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.