K171774 · Osteogenics Biomedical, Inc. · NPK · Oct 19, 2017 · Dental
Device Facts
Record ID
K171774
Device Name
RPM Reinforced PTFE Mesh
Applicant
Osteogenics Biomedical, Inc.
Product Code
NPK · Dental
Decision Date
Oct 19, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3930
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
RPM™ Reinforced PTFE Mesh is a temporarily implantable material (non-resorbable) indicated for stabilization and support of bone grafts in alveolar bony defect sites.
Device Story
RPM™ Reinforced PTFE Mesh is a temporarily implantable, non-resorbable barrier used in dental bone grafting. It consists of a titanium frame embedded between two layers of polytetrafluoroethylene (PTFE) sheets. The device is placed between the bone graft and the periosteum to stabilize the graft and maintain space for bone-derived cell repopulation and repair. It features circular 0.66 mm diameter macropores to facilitate direct contact between the graft and periosteum, promoting revascularization and cell infiltration. The mesh is provided pre-shaped in various sizes and can be trimmed by the clinician to conform to tissue contours. It is intended for use by dental professionals in clinical settings. By maintaining space and providing structural support, the device aids in the successful regeneration of bone in alveolar defects.
Clinical Evidence
Bench testing only. No clinical data presented. In vitro characterization included tensile strength (ASTM D638-14), lamination strength, and suture retention force, demonstrating performance at least equivalent to reference devices. Biocompatibility testing confirmed the device is non-cytotoxic (ISO 10993-5) and non-irritant (ISO 10993-10).
Technological Characteristics
Materials: 100% polytetrafluoroethylene (PTFE) and titanium frame. Design: Titanium frame embedded between two PTFE layers with 0.66 mm diameter macropores. Form factor: Pre-shaped sheets in various dimensions. Sterilization: Sterile. Biocompatibility: Compliant with ISO 10993-1.
Indications for Use
Indicated for stabilization and support of bone grafts in alveolar bony defect sites in patients requiring dental bone grafting procedures.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
October 19, 2017
Osteogenics Biomedical, Inc. Shane Shuttlesworth President 4620 71st St. Bldg. 78-79 Lubbock, Texas 79424
Re: K171774
Trade/Device Name: RPM Reinforced PTFE Mesh Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: Class II Product Code: NPK Dated: June 14, 2017 Received: June 14, 2017
Dear Shane Shuttlesworth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Mary S. Runner -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171774
Device Name RPM™ Reinforced PTFE Mesh
Indications for Use (Describe)
RPM™ Reinforced PTFE Mesh is a temporarily implantable material (non-resorbable) indication and support of bone grafts in alveolar bony defect sites.
| Type of Use (Select one or both, as applicable) | | |
|-------------------------------------------------|--|--|
| | | |
|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
#### l. SUBMITTER
| Applicant Name: | Osteogenics Biomedical, Inc. |
|-----------------|------------------------------------------------|
| Address: | 4620 71st St, Bldg. 78<br>Lubbock, Texas 79424 |
| Phone: | (806) 796-1923 |
| Fax: | (806) 796-0059 |
| Contact Person: | Shane Shuttlesworth<br>President |
| Date Prepared: | May 10, 2017 |
#### II. DEVICE
| Trade Name: | RPM™ Reinforced PTFE Mesh |
|--------------------|-------------------------------------|
| Common Name: | PTFE Mesh |
| Regulation Number: | 21 CFR 872.3930 |
| Regulation Name: | Bone Grafting Material |
| Regulatory Class: | II |
| Product Code: | NPK (Barrier, Synthetic, Intraoral) |
#### lll. PREDICATE DEVICE
| Predicate Device: | Osteo-Mesh™ TM-300 (Osteogenics Biomedical, Inc. |
|-------------------|--------------------------------------------------|
| | K984230 |
Cytoplast™ Regentex™ Titanium 250 (K972278) and Bio-Gide® Resorbable Bilayer Membrane (K050466) were used as a reference devices in this submission.
#### IV. DEVICE DESCRIPTION
RPM™ Reinforced PTFE Mesh is placed between bone grafts and the periosteum in dental bone grafting procedures to stabilize and support the bone graft. The PTFE mesh helps create the space needed for bone-derived cells to repopulate and repair the defect.
RPM™ Reinforced PTFE Mesh is composed of proprietary 100% polytetrafluoroethylene sheets reinforced with a titanium frame. The titanium frame is embedded between two layers of PTFE. PTFE is a biologically inert and tissue-compatible material. RPM™ Reinforced PTFE Mesh is manufactured with circular 0.66 mm diameter macropores to allow direct contact between the
{4}------------------------------------------------
bone graft and the periosteum. Direct contact between the periosteum and bone graft allows naturally occurring revascularization and infiltration of cells.
The PTFE mesh is designed to maintain space and conform to tissue contours.
RPM™ Reinforced PTFE Mesh is provided pre-shaped in a variety of shapes and sizes. Outer dimensions include:
- 12 mm x 24 mm
- 14 mm x 24 mm
- 17 mm x 25 mm
- 20 mm x 25 mm
- 25 mm x 36 mm
- 25 mm x 30 mm
- 30 mm x 41 mm
- . 24 mm x 38 mm
- 30 mm x 40 mm
- 13 mm x 18 mm
- . 13 mm x 19 mm
### V. INDICATIONS FOR USE
RPM™ Reinforced PTFE Mesh is a temporarily implantable material (non-resorbable) indicated for stabilization and support of bone grafts in alveolar bony defect sites.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
RPM™ Reinforced PTFE Mesh is substantially equivalent to the predicate device, Osteo-Mesh™ TM-300, and the reference device, Cytoplast™ Regentex™ Titanium 250. RPM™ Reinforced PTFE Mesh is identical in design, function, and intended use to the legally marketed predicate device Osteo-Mesh™ TM-300. RPM™ Reinforced PTFE Mesh is identical in composition,
biocompatibility, sterilization, model sizes, packaging, shelf life, physical properties, and intended use to the legally marketed reference device Cytoplast™ Titanium 250. See comparison table below:
| | RPM™ Reinforced<br>PTFE Mesh | Osteo-Mesh™ TM-300<br>(K984230) | Cytoplast™ Regentex™<br>Titanium 250<br>(K972278) |
|------------------------|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| | Osteogenics Biomedical,<br>Inc. | Osteogenics Biomedical,<br>Inc. | Osteogenics Biomedical,<br>Inc. |
| Product Code | NPK | LEY | LYC |
| Indications for<br>Use | For stabilization and<br>support of bone grafts in<br>alveolar bony defect<br>sites.<br>Rx Only | For stabilization and<br>support of bone grafts in<br>dento-alveolar bony<br>defect sites.<br>Rx Only | A temporarily<br>implantable material<br>(non-resorbable for use<br>as a space-making<br>barrier in the treatment<br>of periodontal defects.<br> |
| | | | Rx Only |
| | | | Rx Only |
| Operational | RPM™ Reinforced PTFE | Osteo-Mesh™ TM-300 is | Cytoplast™ Regentex™ |
| Principles | Mesh is placed between | placed between bone | Titanium 250 is placed |
| | bone grafts and the | grafts and the | between bone grafts and |
| | periosteum in dental | periosteum in dental | the periosteum in dental |
| | bone grafting procedures | bone grafting procedures | bone grafting procedures |
| | to stabilize and support | to stabilize and support | to stabilize and support |
| | the bone graft. The PTFE | the bone graft. The PTFE | the bone graft. The PTFE |
| | mesh helps create the | mesh helps create the | mesh isolates the space |
| | space needed for bone- | space needed for bone- | needed for bone-derived |
| | derived cells to | derived cells to | cells to repopulate and |
| | repopulate and repair | repopulate and repair | repair the defect. |
| | the defect. | the defect. | |
| Design | Titanium frame | Titanium mesh may be | Titanium frame |
| | embedded between two | trimmed and shaped to | embedded between two |
| | layers of PTFE. Titanium | create additional space | layers of PTFE. Titanium |
| | frame may be trimmed | for bone growth. | frame may be trimmed |
| | and shaped to create | Macropores allow direct | and shaped to create |
| | additional space for | contact between the | additional space for |
| | bone growth. | bone graft and the | bone growth. |
| | Macropores allow direct | periosteum. Direct | |
| | contact between the | contact between the | |
| | bone graft and the | periosteum and bone | |
| | periosteum. Direct | graft allows naturally | |
| | contact between the | occurring | |
| | periosteum and bone | revascularization and | |
| | graft allows naturally | infiltration of cells. | |
| | occurring | | |
| | revascularization and | | |
| | infiltration of cells. | | |
| | | | |
| Composition | 100% PTFE, | Titanium | 100% PTFE, |
| | Titanium | | Titanium |
| Pores | Macro | Macro | Micro |
| Use | Single | Single | Single |
| Shelf Life | 4 years | N/A | 4 years |
| Biocompatible | Yes | Yes | Yes |
| Sterilization | Sterile | Non-Sterile | Sterile |
| Model Sizes | Various | Various | Various |
| Maximum | 12 months | Not Stated | Not Stated |
| Duration of | | | |
| Implantation | | | |
{5}------------------------------------------------
#### PERFORMANCE DATA VII.
## Nonclinical Tests Submitted
The substantial equivalence of RPM™ Reinforced PTFE Mesh and its predicate was demonstrated based on in vitro characterization studies, biocompatibility studies, and clinical history of the reference devices.
{6}------------------------------------------------
Non-clinical testing was performed in accordance with FDA recognized consensus standards and FDA guidelines as follows:
In vitro product characterization testing was performed to demonstrate substantial equivalence of the subject device to its reference devices. A series of bench tests were conducted to evaluate material properties, biological properties, chemical and physical properties.
Tensile strength was characterized and compared to the reference device, Bio-Gide® Resorbable Bilayer Membrane, in order to establish a minimum acceptable specification. Lamination strength and suture retention force were characterized and compared to the reference device, Cytoplast™ Regentex™ Titanium 250.
| Test | Test Method | Results |
|------------------------|--------------|-------------------------------------------|
| Tensile Strength | ASTM D638-14 | Tensile strength ≥ reference device |
| Lamination Strength | Internal | Lamination strength ≥ reference device |
| Suture Retention Force | Internal | Suture retention force ≥ reference device |
The comparative bench testing is summarized in the table below.
In vitro biocompatibility testing was performed to assess biocompatibility of the RPM™ Reinforced PTFE Mesh as an implantable material.
The subject device passed the following FDA Blue Book Memorandum G95-1 and ISO 10993-1 testing for the biological evaluation of medical devices.
| Test | Test Method/Model | Results |
|--------------|-------------------------------------------------------------------------|---------------|
| Cytotoxicity | ISO MEM Elution Assay with L-929<br>Mouse Fibroblast Cells, ISO 10993-5 | Non-cytotoxic |
| Irritation | Oral Mucosa Irritation, ISO 10993-10 | Non-irritant |
### VIII. CONCLUSION
The results of in vitro device characterization tests show that the subject device, RPM™ Reinforced PTFE Mesh, is substantially equivalent to the predicate device or reference devices. Tensile strength testing shows that the subject device is at least as strong, statistically, as the reference device, Bio-Gide®. Lamination strength and suture retention force tests show that the subject device is at least as strong, statistically, as the reference device, Cytoplast™ Titanium 250. Biocompatibility testing demonstrated that the subject device is non-cytotoxic and a non-irritant.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.