K171769 · Straumann USA, LLC (On Behalf of Institut Straumann Ag) · DZE · Nov 14, 2017 · Dental
Device Facts
Record ID
K171769
Device Name
Straumann PURE Ceramic Implants
Applicant
Straumann USA, LLC (On Behalf of Institut Straumann Ag)
Product Code
DZE · Dental
Decision Date
Nov 14, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Straumann® PURE Ceramic Implant (Monotype) is indicated for restoration in single tooth gaps and in an edentulous or partially edentulous jaw. The prosthetic restorations used are single crowns, fixed partial or full dentures, which are connected to the implants through the corresponding components. The ø3.3 mm reduced diameter implants are recommended for central and lateral incisors only.
Device Story
Straumann® PURE Ceramic Implants are endosseous dental implants made of 100% yttrium-stabilized zirconia (Y-TZP). The device features a monotype design with an integrated ceramic abutment and a ZLA (sand-blasted, large-grit, acid-etched) surface to promote osseointegration. Used by dental professionals in clinical settings to support single crowns, fixed partial, or full dentures. The implant is surgically placed into the jawbone to replace missing teeth. The device is provided sterile via Ethylene Oxide (EO) sterilization. The clinician selects the appropriate diameter and length based on the patient's anatomical needs. The output is a stable, biocompatible foundation for prosthetic dental restorations, benefiting patients by restoring oral function and aesthetics.
Clinical Evidence
Bench testing only. No clinical data provided. Validation included sterilization per ISO 11135 (SAL 10-6), packaging integrity (ISTA 2A), shelf-life (ASTM F1980), and biocompatibility (ISO 10993-1). Mechanical testing (static, dynamic fatigue, torque) performed per ISO 14801:2007 confirmed performance consistency after hydrothermal aging. Surface chemistry analysis showed no significant differences between sterilization methods.
Technological Characteristics
Material: 100% yttrium-stabilized zirconia (Y-TZP). Surface: ZLA (sand-blasted, large-grit, acid-etched). Design: Cylindrical monotype with 1.8 mm machined neck. Dimensions: Apical diameters 3.3/4.1 mm; coronal diameters 3.5/4.8 mm; lengths 8-14 mm. Thread pitch: 0.8 mm. Sterilization: Ethylene Oxide (ISO 11135).
Indications for Use
Indicated for patients requiring dental restoration in single tooth gaps or edentulous/partially edentulous jaws. Reduced diameter (ø3.3 mm) implants restricted to central and lateral incisors.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
Straumann USA, LLC (on behalf of Institut Straumann AG) % Jennifer Jackson Director, Regulatory Affairs & Quality Straumann USA, LLC 60 Minuteman Road Andover, Massachusetts 01810
Re: K171769
Trade/Device Name: Straumann® PURE Ceramic Implants Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE Dated: August 15, 2017 Received: August 16, 2017
Dear Jennifer Jackson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
November 14, 2017
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Mary S. Runner -S
for
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K171769
Device Name Straumann® PURE Ceramic Implants
### Indications for Use (Describe)
The Straumann® PURE Ceramic Implant (Monotype) is indicated for restoration in single tooth gaps and in an edentulous or partially edentulous jaw. The prosthetic restorations used are single crowns, fixed partial or full dentures, which are connected to the implants through the corresponding components.
The g3.3 mm reduced diameter implants are recommended for central and lateral incisors only.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10pt">☒ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size:10pt">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
#### 5.1 Submitter
Straumann USA, LLC (on behalf of Institut Straumann AG) 60 Minuteman Road Andover, MA 01810 Phone Number: 1-978-747-2509 Fax Number: 1-978-747-0023 E-mail: jennifer.jackson@straumann.com Contact Person: Jennifer M. Jackson, MS Prepared By: Chanrasmey White Date of Submission: November 14, 2017 5.2 Device Trade Name: Straumann® PURE Ceramic Implants Common Name: Endosseous Dental Implant Classification Name: Implant, Endosseous, Root-Form Regulatory Number: 8872.3640 Classification: Class II Product Code: DZE Subsequent Product Code: NHA 5.3 Predicate Device
Primary Predicate: K151328, Straumann® PURE Ceramic Implants
### 5.4 Device Description
The Straumann® PURE Ceramic Implants are made of 100% yttrium-stabilized zirconia. The endosteal region presents macro- and micro-roughness, a ZLA surface. The implant has a 1.8 mm high machined neck. The implant features a monotype design where the ceramic abutment for final restoration is already built in. Straumann® PURE Ceramic Implant prosthetic components are identified with RD (Regular Diameter) Straumann USA, LLC November 14, 2017 5-1
{4}------------------------------------------------
### 510(k) Summary
corresponding to the neck diameter of 4.8 mm, and ND (Narrow Diameter) corresponding to the neck diameter of 3.5 mm.
#### 5.5 Indications for Use
The Straumann® PURE Ceramic Implant (Monotype) is indicated for restoration in single tooth gaps and in an edentulous or partially edentulous jaw. The prosthetic restorations used are single crowns, fixed partial or full dentures, which are connected to the implants through the corresponding components.
The ø3.3 mm reduced diameter implants are recommended for central and lateral incisors only.
### Technological Characteristics 5.6
The proposed Straumann® Pure Ceramic Implants intended use, material, fundamental operating principles and overall design are identical to the primary predicate device, K151328. The subject device is identical to the primary predicate, K151328 with the exception that the subject device will be delivered to the user sterile via Ethylene Oxide as opposed to sterile via H₂O₂plasma.
The additional components mentioned in the predicate device indications for use statement are not subject to this submission as the change only impacts the sterilization of the endosseous dental implants. The technological characteristics of the subject devices are compared to the predicate device in the following table.
{5}------------------------------------------------
### 510(k) Summary
| FEATURE | SUBJECT DEVICE | PRIMARY PREDICATE DEVICE |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | K171769 | K151328 |
| Indications for Use | The Straumann® PURE<br>Ceramic Implant (Monotype)<br>is indicated for restoration in<br>single tooth gaps and in an<br>edentulous or partially<br>edentulous jaw. The prosthetic<br>restorations used are single<br>crowns, fixed partial or full<br>dentures, which are connected<br>to the implants through the<br>corresponding components.<br>The ø3.3 mm reduced<br>diameter implants are<br>recommended for central<br>and lateral incisors only. | The Straumann® PURE Ceramic<br>lmplant (Monotype) is indicated for<br>restoration in single tooth gaps and<br>in an edentulous or partially<br>edentulous jaw. The prosthetic<br>restorations used are single crowns,<br>fixed partial or full dentures, which<br>are connected to the implants<br>through the corresponding<br>components.<br>The ø3.3 mm reduced diameter<br>implants are recommended for<br>central and lateral incisors only.<br>The Straumann® PURE Ceramic<br>Implant Protective Cap is intended<br>to protect the Straumann® PURE<br>Ceramic Implant (Monotype)<br>during the healing phase after<br>implant placement for up to 6<br>months.<br>Temporary copings are intended to<br>serve as a base for temporary<br>crown or bridge restoration for the<br>Straumann® PURE Ceramic<br>Implant (Monotype) for up to 30<br>days. |
| Material | Y-TZP | Y-TZP |
| Surface<br>Treatment | Sand-blasted, large-<br>grit, acid etched | Sand-blasted, large-grit, acid<br>etched (ZLA®) |
| Apical Diameter | Ø3.3 and Ø4.1 mm | Ø3.3 and Ø4.1 mm |
| Coronal<br>Diameter | Ø3.5 and Ø4.8 mm | Ø3.5 and Ø4.8 mm |
| Length | 8, 10, 12, and 14 mm | 8, 10, 12, and 14 mm |
| Implant<br>Design | Cylindrical Monotype | Cylindrical Monotype |
| Implant Neck | 1.8 mm, machined | 1.8 mm, machined |
| Thread Pitch | 0.8 mm | 0.8 mm |
| Implant/Abutment<br>Connection | None (Monotype) | None (Monotype) |
| Sterilization | End user receives product<br>sterilized per Ethylene Oxide<br>Validated per ISO 11135 to<br>a SAL of 10-6 | End user receives product<br>sterilized per H2O2 - plasma in<br>accordance with DIN EN 556-<br>1:2002 (equivalent to ANSI AAMI<br>ST67:2011) to a SAL of 10-6 |
{6}------------------------------------------------
510(k) Summary
### 5.7 Performance Data
The following performance data were provided in support of the substantial equivalence determination.
### Sterilization and Shelf-Life
A sterilization validation was performed per ISO 11135, Sterilization of healthcare products – Ethylene Oxide – Requirements for development, validation and routine control of a sterilization process for medical devices, using the Half Cycle Overkill Approach. The validation demonstrates that the sterilization process and equipment is capable of reliably and consistently sterilizing the subject device to a minimum Sterility Assurance Level (SAL) of 10-6.
A transportation study has been performed per ISTA 2A to demonstrate the integrity of the sterile barrier system considering the new sterilization process. The subject devices continue to meet product release requirements after being subjected to worst-case environmental and handling changes.
A packaging stability study was performed to demonstrate the integrity of the packaging system and the sterile barrier system after sterilization process via Ethylene Oxide,
handling, distribution, transport and storage up to the defined product shelf life. The subject devices continue to meet product release requirements after being subjected to worst-case environmental and handling changes.
Shelf-Life studies have been conducted in accordance with ASTM F1980.
### Biocompatibility Testing
The biological safety of the Straumann® PURE Ceramic Implants sterilized via Ethylene Oxide was re-evaluated and further biocompatibility testing was required according to ISO and FDA standard (FDA, 2016; ISO 10993-1, 2009). EO and ECH residues were not detected in the chemical analysis and no cytotoxicity was observed. Based on the test results the change in sterilization method did not affect the biological safety profile of the devices.
### Bench Testing
Mechanical testing (static, dynamic fatigue and torque) and surface chemistry analysis was performed to demonstrate the new sterilization method does not affect product performance. Static and dynamic fatigue strength testing was performed per ISO 14801:2007 and the FDA-Guidance to "Root-form endosseous dental implants Straumann USA, LLC November 14, 2017
{7}------------------------------------------------
### 510(k) Summary
and endosseous dental implant abutments." All mechanical testing performed demonstrated the new sterilization method does not affect the product performance after being subjected to hydrothermal aging. Cleanliness analysis demonstrated no significant difference in surface chemistry between H2O2 plasma (sterility method per predicate device) and EO sterilization.
### 5.8 Conclusion
The documentation submitted in this premarket notification demonstrates that the Straumann® PURE Ceramic Implants are substantially equivalent to the predicate device and do not pose new issues of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.