# Laport (K171741)

_Sejong Medical Co., Ltd. · GCJ · Aug 29, 2017 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K171741

## Device Facts

- **Applicant:** Sejong Medical Co., Ltd.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Aug 29, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Laport® Trocars have applications in abdominal, thoracic, and gynecologic minimally invasive procedures to establish a path of entry for endoscopic instruments.

## Device Story

Laport Trocars consist of sleeve and needle assembly (5-12 mm diameter); used in abdominal, thoracic, gynecologic minimally invasive surgery. Three needle types: Safety (sharp flat-bladed tip with protective shield), Optical (transparent tip for endoscope visualization), Bladeless (plastic tip for reduced puncture hazard). Operated by surgeons to establish entry path for endoscopic instruments. Universal seal accommodates wide range of instrument sizes. Device facilitates access to body cavities; enables instrument insertion; maintains pneumoperitoneum via valve system. Benefits include safe, controlled access to surgical sites.

## Clinical Evidence

Bench testing and non-clinical animal study (micropigs) performed. Bench tests: ISO 10993 biocompatibility (cytotoxicity, sensitization, systemic toxicity, irritation), ISO 11737-1 sterilization, shelf-life, leakage, and dimensional measurements. Animal study evaluated insertion/extraction force, fixation, intraperitoneal pressure maintenance, and visualization convenience during laparoscopy. Results support substantial equivalence.

## Technological Characteristics

Sleeve and needle assembly; 5-12 mm diameter. Three tip configurations: Safety (flat-bladed), Optical (transparent), Bladeless (plastic). Universal seal/lip valve. Materials compliant with ISO 10993. Sterilization via Ethylene Oxide. Mechanical device; no energy source or software.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- ENDOPATH® III Trocars ([K032676](/device/K032676.md))
- LAPORT ([K092584](/device/K092584.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the address for the Food and Drug Administration. The address is 10903 New Hampshire Avenue, Document Control Center - WO66-G609, Silver Spring, MD 20993-0002. The address is centered on the page.

August 29, 2017

Sejong Medical Co., Ltd. % Ms. Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA. Inc 690 Roosevelt Irvine, California 92620

Re: K171741

Trade/Device Name: Laport Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: June 5, 2017 Received: June 12, 2017

Dear Ms. Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Jennifer R. Stevenson -S3

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K171741

Device Name

Laport ®

Indications for Use (Describe)

The Laport® Trocars have applications in abdominal, thoracic, and gynecologic minimally invasive procedures to establish a path of entry for endoscopic instruments.

Type of Use (Select one or both, as applicable)

| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D)    </div> | <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)    </div> |
|-----------------------------------------------------------------------------|----------------------------------------------------------------------------|
|-----------------------------------------------------------------------------|----------------------------------------------------------------------------|

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

(K171741)

This summary of 510(K) is being submitted in accordance with requirements of 21 CFR Part 807.92.

### Date:

## 1. Submitter/Applicant

Sejong Medical Co., Ltd. 11, Sinchon 2-ro Paju-si, Gyeonggi-do Republic of Korea 10880

Tel: +82-31-945-8191 Fax: +82-31-945-8190

## 2. U.S Agent/Contact Person

Priscilla Chung LK Consulting Group USA, Inc. 690 Roosevelt, Irvine CA 92620 Phone: 714-202-5789 Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com

## 3. Device

- Trade Name: Laport ®
- Common Name: Sterile Trocars for Endoscopic surgery ●
- Classification: Class II ●
- Classification regulation: 21 CFR 876.1500 ●
- . Product Code: GCJ

#### Predicate Devices: 4.

# Primary Predicate Device: ENDOPATH® III Trocars (K032676) by Ethicon Edo-Surgery, Inc. Reference Predicate Deice: LAPORT (K092584) by MGB EndoskopischeGerate GmbH Berlin

510(k) Submission.

{4}------------------------------------------------

#### 5. Description:

The Laport® Trocars are for use during endoscopic minimally invasive procedures or to gain access potential spaces for endoscopic instruments. The Laport® Trocars are consisting of a sleeve and needle in sizes ranging from 5 to 12 mm in diameter. There are three different needles: Safety, Optical and Bladeless Type. The Safety Type has a sharp flat-bladed tip and shield. The shield on the Safety Type needle is designed to cover the flat-bladed tip to protect internal structures from puncture once the abdominal or thoracic cavity has been entered. The Optical Type has a transparent tip for use with an endoscope to provide visualization for insertions. The Bladeless Type has a plastic tip which reduces hazards during puncture of abdominal cavity.

#### 6. Indication for use:

The Laport® Trocars have applications in abdominal, thoracic, and gynecologic minimally invasive procedures to establish a path of entry for endoscopic instr uments.

#### Performance Data 7.

The following tests were performed on the subject device and the test results support that the subject device is substantially equivalent to the predicate devices.

- · Sterilization Validation Test in accordance with ISO11737-1
- · Shelf Life Validation Test
- · Biocompatibility Tests in accordance with ISO 10993

| Cytotoxicity                              | ISO 10993-5  |
|-------------------------------------------|--------------|
| Ethylene Oxide Sterilization<br>Residuals | ISO 10993-7  |
| Skin Sensitization                        | ISO 10993-10 |
| Systemic Toxicity                         | ISO 10993-11 |
| Irritation                                | ISO 10993-10 |

- · Performance Tests: Appearance, Measurement, Leakage, and Extraction, and Non-Clinical(Animal) test

| Test            | Test Method                                                                                                          |
|-----------------|----------------------------------------------------------------------------------------------------------------------|
| Appearance      | Visual Inspection                                                                                                    |
| Measurement     | Inspecting dimensions using vernier calipers                                                                         |
| Leakage         | When the sample was put air in trocar sleeve<br>with air pressure after close the valve, it<br>should not be leaked. |
| Extraction Test | Appearance, pH, KMnO4, Evaporating                                                                                   |

{5}------------------------------------------------

| Non-Clinical(Animal)<br>test using micropigs | residue, Heavy Metal, UV-vis Spectrum                                                                                                                                                                                                                                                      |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                              | Trocar Insertion & Extraction Test:<br>The operator (expert veterinarian) scored the<br>experience when the trocar was inserted into<br>the abdominal cavity of animals and<br>extracted according to pre-set scoring index.<br>Insertion, Fixation, and Extraction were<br>evaluated.     |
|                                              | Intraperitoneal Pressure Test:<br>The time for intraperitoneal pressure to reach<br>a certain mmHg was recorded. And the<br>bubble test was performed to check the<br>leakage of gas. Time to insufflation was<br>evaluated.                                                               |
|                                              | Convenience of performing laparoscope:<br>The operator inserted the laparoscope<br>thought the trocar needle of optical test<br>devices and scored the convenience of<br>performing laparoscope according to the<br>scoring index demonstrated below.<br>Visualization was also evaluated. |

All the test results demonstrated that the subject device is substantially equivalent to the predicate devices.

#### 8. Basis for Substantial Equivalence

The subject device, Laport® incorporates the same intended use with the predicate devices. The subject device is composed of similar materials with the predicate devices. All the devices are similar in physical & mechanical properties. The differences between the subject device and the predicate device are universal seal and lip valve part. The subject device, Laport® has a universal seal that facilitate using a wide range of instruments from 5mm to 12mm.

We have performed the various performance tests, and the test results support that the subject device is substantially equivalent to the predicate devices.

#### 9. Conclusion

Upon reviewing the information including testing data provided in this submission and comparing intended use, principle of operation and overall technological characteristics, we conclude that the Laport® is substantially equivalent to the predicate devices.

510(k) Submission.

---

**Source:** [https://fda-staging.innolitics.com/device/K171741](https://fda-staging.innolitics.com/device/K171741)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).