K171738 · Hitachi Healthcare Americas Corporation · JAK · Aug 18, 2017 · Radiology
Device Facts
Record ID
K171738
Device Name
Supria True64
Applicant
Hitachi Healthcare Americas Corporation
Product Code
JAK · Radiology
Decision Date
Aug 18, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Supria True64 system is indicated for head, whole body, and vascular X-ray Computed Tomography applications in patients of all ages. The images can be acquired in either axial, helical, or dynamic modes. The volume datasets acquired by the Supria can be post processed by the system to provide additional information. Post processing capabilities included in the Supria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering. Volume datasets acquired by the Supria can be transferred to external devices via a DICOM standard interface. The guideShot Option adds a remote in-room display and controls to support interventional imaging. The device output can provide an aid to diagnosis when used by a qualified physician.
Device Story
Multi-slice CT system using 64-detector technology; rotates X-ray tube and detector assembly around patient via slip ring. Collects data at multiple angles; high-frequency generator produces fan beam X-ray. High-speed reconstruction sub-system processes data into cross-sectional images. Operated by qualified physicians in clinical settings. Includes guideShot option for interventional imaging via remote in-room display. Post-processing includes CTA, MPR, and volume rendering. Output aids diagnosis; images transferred via DICOM. Benefits include high-quality imaging with potential for low-dose acquisition via iterative reconstruction.
Clinical Evidence
Bench testing only. Performance metrics (dose profile, noise, CT number, uniformity, spatial resolution, MTF) compared to predicate and found equivalent. Clinical evidence limited to 3 clinical image examples for usability assessment and comparative evaluation of FBP vs. Iterative Reconstruction showing noise/artifact reduction.
Technological Characteristics
64-slice CT; solid-state detector; 750mm gantry aperture; 5 MHU anode heat storage; 48kW system max power. PC-based workstation (Windows OS). Connectivity via DICOM. Iterative reconstruction (Intelli IP Advanced/Quick). Automatic Exposure Control (IntelliEC). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-44, NEMA XR 25, NEMA XR 26, IEC 62304.
Indications for Use
Indicated for head, whole body, and vascular X-ray CT applications in patients of all ages.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
HITACHI SUPRIA Whole-body X-ray CT System Phase 3 (K163528)
Submission Summary (Full Text)
{0}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or movement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 18, 2017
Hitachi Healthcare Americas Corporation % Mr. Douglas J. Thistlethwaite Manager of Regulatory Affairs 1959 Summit Commerce Park TWINSBURG OH 44087
Re: K171738
Trade/Device Name: Supria True64 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: July 26, 2017 Received: July 28, 2017
Dear Mr. Thistlethwaite:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
#### K171738
Device Name Supria True64 Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
Indications for Use (Describe)
The Supria True64 system is indicated for head, whole body, and vascular X-ray Computed Tomography applications in patients of all ages. The images can be acquired in either axial, helical, or dynamic modes.
The volume datasets acquired by the Supria can be post processed by the system to provide additional information. Post processing capabilities included in the Supria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering.
Volume datasets acquired by the Supria can be transferred to external devices via a DICOM standard interface. The guideShot Option adds a remote in-room display and controls to support interventional imaging. The device output can provide an aid to diagnosis when used by a qualified physician.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
{4}------------------------------------------------
## Submitter Information
| Submitter: | Hitachi Healthcare Americas<br>1959 Summit Commerce Park<br>Twinsburg, Ohio 44087-2371 |
|-------------------|----------------------------------------------------------------------------------------|
| Contact: | Douglas J. Thistlethwaite |
| Telephone number: | 330-425-1313 |
| Telephone number: | 330-963-0749 |
| E-mail: | thistlethwaited@hitachihealthcare.com |
| Date: | April 18, 2017 |
## Subject Device Name
| Trade/Proprietary Name: | Supria True64 |
|-------------------------|------------------------------------------|
| Regulation Number: | 21 CFR 892.1750 |
| Regulation Name: | Computed tomography x-ray system |
| Product Code | JAK, System, X-Ray, Tomography, Computed |
| Class | II |
| Panel | Radiology |
## Predicate Device Name
| Predicate Device(s): | HITACHI SUPRIA Whole-body X-ray CT System Phase 3 (K163528) |
|----------------------|-------------------------------------------------------------|
| Regulation Number: | 21 CFR 892.1750 |
| Regulation Name: | Computed tomography x-ray system |
| Product Code | JAK, System, X-Ray, Tomography, Computed |
| Class | II |
| Panel | Radiology |
### Device Intended Use
The Supria True64 system is indicated for head, whole body, and vascular X-ray Computed Tomography applications in patients of all ages. The images can be acquired in either axial, helical, or dvnamic modes,
The volume datasets acquired by the Supria can be post processed by the system to provide additional information. Post processing capabilities included in the Supria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering.
Volume datasets acquired by the Supria can be transferred to external devices via a DICOM standard interface.
The guideShot Option adds a remote in-room display and controls to support interventional imaging. The device output can provide an aid to diagnosis when used by a qualified physician.
## Device Description
### Function
The Supria True64 is a multi-slice computed tomography system designed to perform multi-slice CT scanning supported by 64-detector technology. The system allows optimum clinical applications ranging from routine exams in response to the diversified circumstances in imaging whole body regions.
{5}------------------------------------------------
### Scientific Concepts
The Supria True64 system uses 64-slice CT technology, where the X-ray tube and detector assemblies are mounted on a frame that rotates continuously around the patient using slip ring technology. The solid-state detector assembly design collects up to 64 slices of data simultaneously. The X-ray sub-system features a high frequency generator. X-ray tube, and collimation system that produces a fan beam X-ray output. The system can operate in a helical (spiral) scan mode where the patient table moves during scanning. As the X-ray tube/detector assembly rotates around the patient, data is collected at multiple angles.
The collected data is then reconstructed into cross-sectional images by a high-speed reconstruction sub-system. The images are displayed on a Computer Workstation, stored, printed, and archived as required. The workstation is based on current PC technology using the Windows™ operating system.
### Physical and Performance Characteristics
The Supria True64 system consists of a Gantry, Operator's Workstation, Patient Table, High-Frequency X-ray Generator, and accessories. The system performance is similar to the predicate device.
### Performance Comparison
As part of our design validation performance comparison analysis was conducted to demonstrate continued conformance with a special control or recognized standard.
A rationale analysis was then conducted and the results are contained in Table 1:
| Testing Type | Rationale Analysis |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Validation Testing - Bench | There is no change about the following performance.<br>Dose Profile, Noise, Mean CT number and Uniformity, Spatial Resolution, Tomographic Section Thickness and Sensitivity<br>Profile, Tomographic Plane Location, CT dose index<br>Therefore, we judged that Supria is substantially equivalent to the predicate. |
| Validation Testing - Clinical | We provide 3 kinds of clinical image example which we judged to be sufficient to judge a clinical usability. |
Table 1 Performance Comparison Analysis
The analysis confirms the performance characteristics of the Supria True64 are comparable to the predicate device and support our conclusion that the subject system is substantially equivalent.
# Device Technological Characteristics
The technological characteristics of the Supria True64 and the predicate device are listed in Table 2.
| ITEM | Supria True64 | HITACHI Supria PHASE 3 (K163528) | Difference |
|--------------------------------------------------------------|----------------------------------------------------------------------------------------|---------------------------------------------------------|------------|
| Gantry | | | |
| Geometry | Rotate-rotate with offset detector system,<br>slip ring | Rotate-rotate with offset detector system,<br>slip ring | No |
| Scan Time | 0.75, 1.0, 1.5, 2.0 [s] | 0.75, 1.0, 1.5, 2.0 [s] | No |
| X-ray Fan Beam Angle | 51 [deg] | 51 [deg] | No |
| Gantry Tilt | -30 to +30 [deg] | -30 to +30 [deg] | No |
| Gantry Aperture | 750 [mm] | 750 [mm] | No |
| Gantry Dimensions | 1990 x 920 x 1842.5 [mm] | 1990 x 920 x 1842.5 [mm] | No |
| Gantry Weight | 1600 [kg] | 1600 [kg] | No |
| Scan Localizer | Laser | Laser | No |
| ITEM | Supria True64 | HITACHI Supria PHASE 3 (K163528) | Difference |
| Detector | | | |
| Type | Solid state | Solid state | No |
| Number of Channels | 880 [ch] (16 slice)<br>888 [ch] (64 slice) | 880 [ch] (8ch reference) | No |
| Number of Rows | 16 (16 slice)<br>64 (64 slice) | 16 | See 01 |
| Number of Slices | 16 [slice/scan] (Axial) (16 slice)<br>64 [slice/scan] (Axial) (64 slice) | 16 [slice/scan] (Axial) | No |
| X-ray Tube | | | |
| Anode Heat Storage | 5 [MHU] | 5 [MHU] | No |
| Dissipation Rate | 470 [kHU/min] | 470 [kHU/min] | No |
| Tube cooling | Oil/air | Oil/air | No |
| Tube focal spot | Dual 0.7 x 0.8, 1.2 x 1.4 [mm] | Dual 0.7 x 0.8, 1.2 x 1.4 [mm] | No |
| X-ray Generator | | | |
| kW Output | System Maximum 48[kW] /<br>Generator Maximum 51 [kW] | System Maximum 48[kW] /<br>Generator Maximum 51 [kW] | No |
| Max. Power Input | 75 [kVA] | 75 [kVA] | No |
| kVp Range | 80, 100, 120, 140 [kVp] | 80, 100, 120, 140 [kVp] | No |
| mA Range | 10 to 400 [mA] @120kV, 48kW | 10 to 400 [mA] @120kV, 48kW | No |
| Patient Table | | | |
| Range of Movement,<br>Vertical | 450 to 1000 [mm] (CT-WT-21) | 450 to 1000 [mm] (CT-WT-21) | No |
| Range of Movement,<br>Longitudinal | 1910 [mm] (CT-WT-21) | 1910 [mm] (CT-WT-21) | No |
| Range of Movement,<br>Lateral | N/A | N/A | No |
| Scannable Range | 155 cm | 155 cm | No |
| Maximum Load Capacity | 227 [kg] | 227 [kg] | No |
| Display | | | |
| Monitor Type | 24" LCD | 24" LCD | No |
| Matrices, Pixels | 1920 x 1200 | 1920 x 1200 | No |
| Image Enlargements | Up to 9.99x | Up to 9.99x | No |
| Max. Slices Displayed at Once | 25 | 25 | No |
| Image Storage | | | |
| Hard Disk | 110 [GB] (images),<br>200 [GB] (raw data) (16 slice)<br>800 [GB] (raw data) (64 slice) | 110 [GB] (images),<br>200 [GB] (raw data) | See 02 |
| Storage Images | 200,000 | 200,000 | No |
| Archival Storage (Media) | DVD-R/RW, CD-R/RW | DVD-R/RW, CD-R/RW | No |
| ITEM | Supria True64 | HITACHI Supria PHASE 3 (K163528) | Difference |
| Scanning, Reconstruction | | | |
| Localization Scan | Real time | Real time | No |
| Localization Scan Length | 150, 250, 350, 500, 750,<br>1000, 1250, 1500, 1750 [mm] | 150, 250, 350, 500, 750,<br>1000, 1250, 1500, 1750 [mm] | No |
| Max. Scan Time | 100 [s] | 100 [s] | No |
| Helical Beam Pitch | 0.56, 0.81, 1.06, 1.31, 1.56 (16 slice)<br>0.58, 0.83, 1.08, 1.33, 1.58 (64 slice) | 0.56, 0.81, 1.06, 1.31, 1.56 | See 03 |
| Collimation | 1.25, 5, 10, 15, 20 [mm] | 1.25, 5, 10, 15, 20 [mm] | No |
| Reconstruction Matrix | 512 x 512 [pix] | 512 x 512 [pix] | No |
| Reconstruction FOVs | 20 to 500 [mm] | 20 to 500 [mm] | No |
| Slice Thickness | 0.625, 1.0, 1.25, 2.5, 3.75,<br>5.0, 7.5, 10.0 [mm] | 0.625, 1.0, 1.25, 2.5, 3.75,<br>5.0, 7.5, 10.0 [mm] | No |
| Range of CT numbers | -2000 to +4000 (13bit)<br>-32768 to +32767 (16bit) | -2000 to +4000 (13bit)<br>-32768 to +32767 (16bit) | No |
| Reconstruction Time | 0.1 seconds per image or less | 0.1 seconds per image or less | No |
| Performance | | | |
| High-contrast spatial resolution | 0.35 [mm] | 0.35 [mm] | No |
| Low-contrast resolution mm at<br>% at <4 rads | 2.5 [mm] @ 0.25% | 2.5 [mm] @ 0.25% | No |
| 10% MTF | 14.7 [lp/cm] | 14.7 [lp/cm] | No |
| 50% MTF | 12.2 [lp/cm] | 12.2 [lp/cm] | No |
| Dose Controls | | | |
| Bow Tie Filter | Yes. Normal | Yes. Normal | No |
| Automatic Exposure Control | Yes. IntelliEC | Yes. IntelliEC | No |
| Longitudinal Modulation | Yes | Yes | No |
| Angular Modulation | Yes | Yes | No |
| Iterative Reconstruction | Yes. Intelli IP Advanced, Intelli IP Quick | Yes. Intelli IP Advanced, Intelli IP Quick | No |
| Maximum possible pitch with full<br>image quality | 1.56 (16 slice)<br>1.58 (64 slice) | 1.56 | See 03 |
| Dose Displays | | | |
| CTDIv | Yes | Yes | No |
| DLP | Yes | Yes | No |
| Features | | | |
| Axial Scan | Yes | Yes | No |
| Helical Scan | Yes | Yes | No |
| Dynamic Scan | Yes | Yes | No |
| Predict Scan | Yes | Yes | No |
| ECG Retrospective Scan<br>(Helical) | No | No | No |
| ECG Prospective Scan<br>(Axial) | No | No | No |
| guideShot Scan | Yes | Yes | No |
| ITEM | Supria True64 | HITACHI Supria PHASE 3 (K163528) | Difference |
| Automatic Exposure Control | Yes | Yes | No |
| Automatic Exposure Control<br>using Iterative Reconstruction | No | No | No |
| ECG Dose Modulation | No | No | No |
| Adaptive Filter | No | No | No |
| Iterative Reconstruction | Yes. Intelli IP Advanced, Intelli IP Quick | Yes. Intelli IP Advanced, Intelli IP Quick | No |
| Injector Synchronization | Yes | Yes | No |
| Dose Check | Yes | Yes | No |
| Access Control | Yes | Yes | No |
| Automatic Cardiac Phase<br>Search | No | No | No |
| Preview Scan | No | No | No |
| Double Slice at Axial Scan | No | No | No |
| Priority Recon. | No | No | No |
| Dose Report | Yes. Simple Dose Report | Yes. Simple Dose Report | No |
| DICOM | Yes | Yes | No |
| ID Reader | Yes | Yes | No |
| Exam Split | Yes | Yes | No |
| Multi-Planar Reconstruction<br>(MPR) | Yes | Yes | No |
| Volume Rendering | Yes | Yes | No |
| CT Angiography (CTA) | Yes | Yes | No |
| Segmentation | Yes | Yes | No |
| Retouch | Yes | Yes | No |
| Quality Exam | Yes | Yes | No |
| HiMAR | Yes | Yes | No |
| Orbit synchronization scan | Yes | Yes | No |
| Off-time mode | Yes | Yes | No |
| On-time standby | Yes | Yes | No |
| Shutter Scan Reduction | Yes | Yes | No |
#### Table 2 Technological Characteristic Differences
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
The differences from the predicate device to Supria 64 are explained below table.
| Detector | |
|---------------|------------------------------------------------------------------------------------------------------------------------|
| 01 | The 64 lines detector is added. |
| Image Storage | |
| 02 | Raw data disk for 64 slices is added because the data amount for each raw data of 64 slices is .bigger than 16 slices. |
| Dose Controls | |
| 03 | Helical Beam Pitch is fine-tuned for 64 lines. |
Therefore, based on a thorough analysis and comparison of Supra 64 and the predicate device, the technological characteristics do not impact safety and effectiveness.
{9}------------------------------------------------
# Substantial Equivalence
A summary decision was based on analysis of Table 3.
| ITEM | Overall Rationale Analysis |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Gantry | There are no significant changes in technology characteristics, hardware, and software from the predicate device, HITACHI Supria PHASE 3 (K163528). |
| Detector | |
| X-ray Tube | |
| X-ray Generator | |
| Patient Table | |
| Display | |
| Image Storage | |
| Scanning,<br>Reconstruction | |
| Performance | |
| Dose Controls | |
| Dose Displays | |
| Features | |
Table 3 Rationale Analysis: Supria vs. Predicate
Therefore, based on a thorough analysis and comparison of the functions, scientific concepts, physical and performance characteristics, performance comparison and technological characteristics, the proposed Supria True64 Whole-body X-ray CT System is considered substantially equivalent to the currently marketed predicate device (SUPRIA Whole-body X-ray CT System (K163528)) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness.
{10}------------------------------------------------
# Summary of Non-Clinical Testing
This device complies with all applicable requirements for Dose Profile. Noise. Mean CT number and Uniformity, Spatial Resolution, Tomographic Section Thickness and Sensitivity Profile, Tomographic Plane Location, and CT dose index.
Noise, Mean CT number, Uniformity and Spatial Resolution test results for the Supria True64 were compared to results from the Supria PHASE 3 (K163528. Our findings determined the systems to be equivalent. The summary report on these findings is provided in this submission.
In addition, the Supria True64 system is in conformance with the applicable parts of the following standards:
- AAMI ANSI ES60601-1:2005/(R) 2012 and A1:2012, C1:2009/(R)2012 and . A2:2010/(R)2012
Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
- IEC 60601-1-2 Edition 3: 2007 . Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- IEC 60601-1-3 Edition 2.0 2008-01
Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
- IEC 60601-2-44 Edition 3.0 2009-02 Medical electrical equipment Part 2-44: particular requirements for the basic safety and essential performance of x-ray equipment for computed tomography.
- NEMA XR 25 Computed Tomography Dose Check
- NEMA XR26
Access Controls for Computer Tomography: Identification, Interlocks, and Logs
- IEC 62304 First edition 2006-05. Medical device software Software life cvcle processes .
# Summary of Clinical Testing
As part of design validation, Hitachi has provided 3 kinds of clinical image examples which we judged to be sufficient to judge a clinical usability.
In addition, a comparative evaluation was conducted of via FBP and Iterative Reconstruction. It resulted in finding the images from the Supria True64 realized both low dose and high quality through reduction of image noise and artifacts.
# Conclusions
Hitachi believes that, based on the information included in the submission, Supria True64 Whole-body X-ray CT System is substantially equivalent with respect to hardware, base elements of the software, safety, effectiveness, and functionality to the SUPRIA Whole-body Xray System Phase 3 (K163528).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.