The s-Clean TiN Coating Abutments is an endosseous dental implant that is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple-units prosthetic appliance attachment to restore a patient's chewing function. Implants can be placed with a conventional two stage surgical process with an option for transmucosal healing or they can be placed in a single stage surgical process for immediate loading when good primary stability has been achieved and with appropriate occlusal loading.
Device Story
s-Clean TiN Coating Abutments are dental abutments used as an interface between endosseous dental implants and prosthetic appliances. Fabricated from Ti-6Al-4V ELI (Grade 5) or Pure Titanium (Grade 4), the devices feature a Titanium Nitride (TiN) coating applied via Physical Vapor Deposition (PVD). The system includes various configurations (angled, couple, free, free-mill, MOA, sub-octa, sole, ball, and snap abutments) and healing caps. Used in dental clinics by clinicians, the abutments are attached to compatible Dentis implant fixtures to restore chewing function. The device is provided non-sterile and requires end-user steam sterilization. The TiN coating is intended to modify surface characteristics. Clinical decision-making relies on the clinician's assessment of implant stability and prosthetic requirements. Benefits include restoration of masticatory function for edentulous or partially edentulous patients.
Clinical Evidence
No clinical testing was performed for this submission. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993-1, -3, -5, -10, -11) and fatigue testing (ISO 14801).
Technological Characteristics
Materials: Ti-6Al-4V ELI (Grade 5) and Pure Titanium (Grade 4). Surface treatment: TiN coating via PVD. Sterilization: Steam sterilization by end-user (ISO 17665-1, -2; ANSI/AAMI ST79). Mechanical testing: Fatigue testing per ISO 14801.
Indications for Use
Indicated for patients requiring dental restoration via endosseous implants in upper or lower jaw arches; supports single or multiple-unit prosthetic appliances; suitable for two-stage surgical process with transmucosal healing or single-stage process with immediate loading given sufficient primary stability.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 29, 2017
Dentis Co., Ltd. % April Lee Consultant Withus Group Inc. 2531 Pepperdale Drive Rowland Heights, California 91748
Re: K171694
Trade/Device Name: s-Clean TiN Coating Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: October 25, 2017 Received: October 31, 2017
Dear April Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Andrew I. Steen -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K171694
Device Name s-Clean TiN Coating Abutments
Indications for Use (Describe)
The s-Clean TiN Coating Abutments is an endosseous dental implant that is indicated for surgical placement in the upper and lower jaw arches, to provide a root for single or multiple-units prosthetic appliance attachment to restore a patient's chewing function. Implants can be placed with a conventional two stage surgical process with an option for transmucosal healing or they can be placed in a single stage surgical process for immediate loading when good primary stability has been achieved and with appropriate occlusal loading.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(K) Summary
#### Submitter
Dentis Co., Ltd. Sun Chul Shin 99, Seongseoseo-ro, Dalseo-gu Daegu, 42718 South Korea Phone: +82-53-583-2804 Fax: +82-53-583-2806
#### Device Information
Trade Name: s-Clean TiN Coating Abutments Common Name: Dental Abutment Classification Name: Endosseous Dental Abutment Product Code: NHA Panel: Dental Regulation Number: 872.3630 Device Class: Class II Date Prepared: 11/20/2017
# Official Correspondent
Withus Group Inc. April Lee 2531 Pepperdale Drive Rowland Heights, CA 91748 USA Email: withus6664@gmail.com Phone: 909-274-9971 Fax: 909-460-8122
#### Description
The s-Clean TiN Coating Abutments are device made of pure titanium and titanium alloy intended for use as an aid in prosthetic restoration.
The system consists of s-Clean TiN Half Coating Angled Abutment, s-Clean TiN Half and partial Coating Couple Abutment, s-Clean TiN Half Coating Free Abutment, s-Clean TiN Half Coating FreeMill Abutment, s-Clean TiN Half Coating MOA Abutment, s-Clean TiN Partial Sub-Octa Abutment, s-Clean TiN Partial Coating Sole Abutment, s-Clean DOA Ball Abutment, s-Clean DOA Snap Abutment, and s-Clean Sole Abutment Healing Cap.
The dimension of each abutment ranges as below:
- s-Clean TiN Half Coating Angled Abutment: Ø4.5, 5.0, 5.5 and 6.5 mm (D) x 10.18, 10.46, . 11.04, 11.18, 11.46, 11.6, 13.04, 13.18, 13.46 and 13.6 mm (L) x 15° and 25°
- s-Clean TiN Half and Partial Coating Couple Abutment: Ø4.5, 4.8, 5.5 and 6.5 mm (D) x 0.8, 1.3, ● 1.8, 2.3, 3.3, 4.3 and 5.3mm (Cuff H)
- . s-Clean TiN Half Coating Free Abutment: Ø4.5. 5.5 and 6.5mm (D) x 15.36 and 15.5mm (L)
- s-Clean TiN Half Coating FreeMill Abutment: Ø4.0, 4.5, 5.5, 6.5mm (D) and 7.5mm (D) x 1.3, ● 1.8, 2.8, 3.8 mm (Cuff H)
- . s-Clean TiN Half Coating MOA Abutment: Ø4.5 and 5.5mm(D) x 12.64, 14.14, 14.64, 15.14 and 16.14mm(L)
- s-Clean TiN Partial Sub-Octa Abutment: Ø 4.8 mm (D) x 0.8, 1.3, 2.3, 3.3, 4.3 and 5.3 mm (Cuff ● H)
- . s-Clean TiN Partial Coating Sole Abutment: Ø4.5. 5.5. 6.0 and 6.5 mm (D) x 0.8. 1.3. 1.8. 2.3. 4.3, and 5.3mm (Cuff H)
{4}------------------------------------------------
- s-Clean DOA Ball Abutment: Ø3.5 mm (D) x 0.6, 1.6, 2.6, 4.6, 5.6 (Cuff H) ●
- s-Clean DOA Snap Abutment: Ø3.5 mm (D) x 3.35 (Cuff H) ●
- s-Clean Sole Abutment Healing Cap: Ø4.6, 5.1, 5.9, 6.3, 6.8 and 6.9mm (D) x 6, 7, 7.5 and 9 mm (L)
The surface of the abutments is partially or half TiN Coated by using PVD (Physical Vapor Deposition).
The subject device is compatible with:
| Compatible Implant names | Implant Models | Sizes |
|----------------------------------------------------|-----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| Dentis Dental Implant System<br>(K171027) | s-Clean Tapered Fixture<br>s-Clean SAVE Fixture | Diameters: Ø3.7, 4.3, 4.8, 5.5, 6.0<br>Length: 7, 8, 10, 12, 14mm |
| Dentis Dental Implant System<br>(K082843) | s-Clean SAVE Fixture | Diameters: Ø6.5, 7.0<br>Lengths:7, 8, 10, 12mm |
| Haptite Coating Implant System<br>(K111364) | HAPTITE s-Clean Tapered Fixture<br>HAPTITE s-Clean Straight Fixture<br>HAPTITE s-Clean SAVE Fixture | Diameters: Ø3.7, 4.1, 4.3, 4.8, 5.5,<br>6.0, 6.5, 7.0<br>Lengths:7, 8, 10, 12, 14mm |
| Dentis Dental Implant System<br>(K150344) | s-Clean Tapered Fixture | Diameters: Ø3.7, 4.1, 4.8<br>Lengths:7, 8, 10, 12, 14mm |
| oneq-sl s-clean implant system<br>(K153639) | OneQ-SL s-Clean Fixture | Diameters: Ø 3.7, 3.9, 4.2, 5.2, 6.0,<br>7.0, 8.0<br>Lengths:7, 8, 10, 12, 14mm |
| s-clean tapered ii rbm implant<br>system (K160213) | s-Clean Tapered II RBM Fixture | Diameters: Ø3.7, 4.1, 4.3, 4.8<br>Lengths:7, 8, 9, 10, 12, 14mm |
The subject device is provided non-sterile and should be sterilized before use (End user sterilization).
#### Indication for Use
The s-Clean TiN Coating Abutments is an endosseous dental implant that is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple-units prosthetic appliance attachment to restore a patient's chewing function. Implants can be placed with a conventional two stage surgical process with an option for transmucosal healing or they can be placed in a single stage surgical process for immediate loading when good primary stability has been achieved and with appropriate occlusal loading.
{5}------------------------------------------------
#### Predicate Device & Comparison
The subject device is substantially equivalent to the following predicate devices:
#### Primary Predicate
- K171027, Dentis Dental Implant System by Dentis Co., Ltd.
#### Reference Predicates
- K082843, Dentis Dental Implant System by Dentis Co., Ltd. ●
- K111364, Haptite Coating Implant System by Dentis Co., Ltd. .
- K161244, S-Clean Oneq-sl Narrow Implant System by Dentis Co., Ltd. .
- K153521, IH Prosthetic System by Sewon Medix Inc. .
- K142211, OT Equator by Rhein'83 Srl. .
| | Subject device | Primary Predicate | Reference Predicate | Reference Predicate | Reference Predicate | | stage surgical process<br>with an option for<br>transmucosal healing or<br>they can be placed in a<br>single stage surgical<br>process for immediate<br>loading when good<br>primary stability has been<br>achieved and with<br>appropriate occlusal<br>loading. | surgical process with an<br>option for transmucosal<br>healing or they can be<br>placed in a single stage<br>surgical process for<br>immediate loading when<br>good primary stability has<br>been achieved and with<br>appropriate occlusal<br>loading. | two stage surgical<br>procedures and not for<br>immediate load. Also, this<br>system is intended to be<br>used in the molar region. | is dedicated for one and<br>two stage surgical<br>procedures and not<br>dedicated for immediate<br>loading. This system is<br>intended for delayed<br>loading. | bridgework. IH Implant<br>System is for single and<br>two stage surgical<br>procedures. It is intended<br>for delayed loading. |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | NA | K171027 | K082843 | K111364 | K153521 | Product Name | s-Clean TiN Half<br>Coating Angled<br>Abutment | - | Submerged Angled<br>Abutment | - | Angled Abutment |
| Product Name | s-Clean TiN Coating<br>Abutments | Dentis Dental Implant<br>System | Dentis Dental Implant<br>System | Haptite Coating Implant<br>System | IH Prosthetic System | Dimension | Ø4.5, 5.0, 5.5 and 6.5<br>mm (D) x<br>10.18, 10.6, 10.46,<br>11.04, 11.18, 11.46,<br>11.6, 13.04, 13.18,<br>13.46 and 13.6 mm (L)<br>x 15° and 25° | - | Ø4.5, 5.0, 5.5 and 6.5<br>mm (D) x<br>9, 10.5, 12 mm (L) | - | Ø 4.5/5.0/5.5 /6.0<br>X<br>9.5, 10, 10.5,<br>11, 11.5, 12 (L) |
| Manufacturer | Dentis Co., Ltd. | Dentis Co., Ltd. | Dentis Co., Ltd. | Dentis Co., Ltd. | Sewonmedix Inc. | Material | Ti-6Al-4V ELI(Grade<br>5) | - | Ti-6Al-4V ELI(Grade<br>5) | - | Ti-6Al-4V ELI(Grade<br>5) |
| Indications for Use | The s-Clean TiN Coating<br>Abutments is an<br>endosseous dental implant<br>that is indicated for<br>surgical placement in the<br>upper and lower jaw<br>arches, to provide a root<br>form means for single or<br>multiple-units prosthetic<br>appliance attachment to<br>restore a patient's<br>chewing function.<br>Implants can be placed<br>with a conventional two | The Dentis Dental Implant<br>System is an endosseous<br>dental implant that is<br>indicated for surgical<br>placement in the upper<br>and lower jaw arches, to<br>provide a root form means<br>for single or multiple-<br>units prosthetic appliance<br>attachment to restore a<br>patient's chewing<br>function. Implants can be<br>placed with a<br>conventional two stage | The Dentis Dental Implant<br>System is indicated for<br>use in partially or fully<br>edentulous mandibles and<br>maxillae, in support of<br>single or multiple-unit<br>restorations including;<br>cemented retained, screw<br>retained, or overdenture<br>restorations, and terminal<br>or intermediate abutment<br>support for fixed<br>bridgework. This system<br>is dedicated for one and | The HAPTITE Coating<br>Implant System is<br>indicated for use in<br>partially or fully<br>edentulous mandibles and<br>maxillae, in support of<br>single or multiple-unit<br>restorations including;<br>cemented retained, screw<br>retained, or overdenture<br>restorations, and terminal<br>or intermediate abutment<br>support for fixed<br>bridgework. This system | IH Implant System is<br>device made of titanium<br>and titanium alloy<br>indicated for in partially<br>or fully edentulous<br>mandibles and maxillae,<br>in support of single or<br>multiple-unit restorations<br>including; cemented<br>retained or screw retained<br>restorations and terminal<br>or interminal abutment<br>support for fixed | Coating | TiN coating | - | NA | - | TiN coating |
| Angulation | 15°/25° | - | 15°/25° | - | 15°/25° | | | | | | |
| Product Name | s-Clean TiN Half<br>Coating Couple<br>Abutment | Couple Abutment | - | - | Cement Abutment | | | | | | |
{6}------------------------------------------------
{7}------------------------------------------------
| Dimension | Ø4.5, 4.8, 5.5 and 6.5 mm (D) x<br>0.8, 1.3, 1.8, 2.3, 3.3, 4.3 and 5.3mm (Cuff H) | Ø4.0, 4.5, 4.8, 5.5, 6.0 and 6.5 mm (D) x<br>0.8, 1.3, 1.8, 2.3, 3.3, 4.3 and 5.3mm (Cuff H) | - | Ø4.5/5.0/5.5/6.0/6.5 X<br>1.0/1.5/2.0/2.5<br>/3.0/3.5/4.0/4.5/5.0<br>(Cuff H) | |
|------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------|-------------------------------------------------------------------------------|---------------------------|
| Material | Ti-6Al-4V ELI(Grade 5) | Ti-6Al-4V ELI(Grade 5) | - | Ti-6Al-4V ELI(Grade 5) | |
| Coating | TiN coating | NA | - | TiN coating | |
| Product Name | s-Clean TiN Half<br>Coating Free Abutment | Free Abutment | - | - | |
| Dimension | Ø4.5, 5.5 and 6.5mm<br>(D) x 15.36 and<br>15.5mm (L) | Ø4.5 and 5.5 and<br>6.5mm (D) x<br>15.36 and 15.5mm (L) | - | - | |
| Material | Pure Titanium (Grade 4) | Pure Titanium (Grade 4) | - | - | |
| Coating | TiN coating | NA | - | - | |
| Product Name | s-Clean TiN Half<br>Coating FreeMill<br>Abutment | - | - | FreeMilling Abutment | |
| Dimension | Ø4.0, 4.5, 5.5, 6.5mm<br>(D) and 7.5mm (D) x<br>1.3, 1.8, 2.8, 3.8 mm<br>(Cuff H) | - | - | Ø 4.5/5.0<br>/5.5/6.0/6.5mm (D) x<br>1.0/2.0 /3.0mm<br>(Cuff H) | |
| Material | Pure Titanium (Grade 4) | - | - | Pure Titanium<br>(Grade 4) | |
| Coating | TiN coating | - | - | TiN coating | |
| Product Name | s-Clean TiN Half<br>Coating MOA<br>Abutment | - | MOA Abutment | - | |
| Dimension | Ø4.5 and 5.5mm(D) x<br>12.64, 14.14, 14.64,<br>15.14 and 16.14mm(L) | - | Ø4.0 and 5.5mm(D) x<br>12.64, 14.14, 14.64,<br>15.14 and 16.14mm(L) | - | |
| Material | Ti-6Al-4V ELI(Grade<br>5) | - | Ti-6Al-4V ELI(Grade<br>5) | - | |
| Coating | TiN coating | - | - | - | |
| Product Name | s-Clean TiN Partial<br>Coating Sub-Octa<br>Abutment | Sub-Octa Abutment | - | - | |
| Dimension | Ø4.8 mm (D) x<br>0.8, 1.3, 2.3, 3.3, 4.3<br>and 5.3 mm (Cuff H) | Ø4.8mm (D) x<br>0.8, 1.3, 2.3, 3.3, 4.3<br>and 5.3 mm (Cuff H) | - | - | |
| Material | Ti-6Al-4V ELI(Grade<br>5) | Ti-6Al-4V ELI(Grade<br>5) | - | - | |
| Coating | TiN coating | - | - | - | |
| Product Name | s-Clean TiN Partial<br>Coating Sole Abutment | Sole Abutment | - | Solid Abutment | |
| Dimension | Ø4.5, 5.5, 6.0 and 6.5<br>mm (D) x 0.8, 1.3, 1.8, | Ø4.5, 5.5, 6.0 and 6.5<br>mm (D) x 0.8, 1.3, 1.8, | - | Ø4.0/4.5/5.0/5.5/<br>6.0/6.5/7.0<br>X 1.0/1.5/2.0/2.5/ | |
| | 2.3, 4.3, and 5.3mm<br>(Cuff H) | 2.3 4.3 and 5.3mm<br>(Cuff H) | | | 3.0/3.5/4.0/4.5/5.0 (L) |
| Material | Ti-6Al-4V ELI(Grade<br>5) | Ti-6Al-4V ELI(Grade<br>5) | | | Ti-6Al-4V ELI(Grade<br>5) |
| Coating | TiN coating | | | | TiN coating |
| Product Name | s-Clean DOA Ball<br>Abutment | O-Ring Abutment | | | |
| Dimension | Ø3.5 mm (D) x 1.0, 2.0.<br>3.0, 4.0, 5.0(Cuff H) | Ø3.5 mm (D) x 0.5, 2<br>and 4 mm (Cuff H) | | | |
| Material | Ti-6Al-4V ELI(Grade<br>న) | Ti-6Al-4V ELI(Grade<br>5) | | | |
| Coating | TiN coating | | | | |
| Product Name | Sole Healing Cap | | | Sole Healing Cap | |
| Dimension | Ø4.6, 5.1, 5.9, 6.3, 6.8<br>and 6.9mm (D) x 6, 7,<br>7.5 and 9 mm (L) | | | Ø4.9mm (D) x 4.0 mm<br>(L) | |
| Material | Acetal | | | Acetal | |
| Sterilization | Steam sterilization by<br>user | Steam sterilization by<br>user | Steam sterilization by<br>user | Steam sterilization by<br>user | |
| Product Code | NHA | NHA | NHA | NHA | |
| Brief Discussion | The subject device is similar to the predicate devices in shape, angulation, structure, dimension, and material. The<br>difference is TiN coated surface treatment of the abutments. | | | | |
{8}------------------------------------------------
{9}------------------------------------------------
K171694
{10}------------------------------------------------
# s-Clean DOA Snap Abutment
| | Subject device | Primary Predicate | Reference Predicate |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | NA | K142211 | K171027 |
| Product Name | s-Clean TiN Coating Abutments | OT EQUATOR | Dentis Dental Implant System |
| Manufacturer | Dentis Co., Ltd. | Rhein'83 Srl | Dentis Co., Ltd. |
| Indications for Use | The s-Clean TiN Coating Abutments is an<br>endosseous dental implant that is indicated for<br>surgical placement in the upper and lower jaw<br>arches, to provide a root form means for single<br>or multiple-units prosthetic appliance<br>attachment to restore a patient's chewing<br>function. Implants can be placed with a<br>conventional two stage surgical process with an<br>option for transmucosal healing or they can be<br>placed in a single stage surgical process for<br>immediate loading when good primary stability<br>has been achieved and with appropriate<br>occlusal loading. | The OT Equator is designed as<br>an endosseous dental implant<br>retentive component used to<br>retain a complete or partial<br>denture. The OT Equator is<br>screwed into an endosseous<br>implant in the mandible or<br>maxilla.<br>The OT Equator abutments are<br>indicated for use with the<br>implant systems listed in<br>Attachment B. | The Dentis Implant System is an endosseous dental<br>implant that is indicated to use for surgical placement<br>in the upper and lower jaw arches, to provide a root<br>form means for single or multiple units' prosthetic<br>appliance attachment to restore a patient's chewing<br>function. Implants can be placed with a conventional<br>two stage surgical process with an option for<br>transmucosal healing or they can be placed in a single<br>stage surgical process for immediate loading.<br>Immediate loading is restricted to the anterior<br>mandible, based on four splinted interforminal placed<br>implants, and not indicated for single, unsplinted<br>implants. Patients must be subject for dental treatment<br>with endosseous implants. |
| Product Name | s-Clean DOA Snap Abutment | OT Equator Abutment | - |
| Dimension | Ø3.5 mm (D) x 3.35 (Cuff H) | Ø2.5 mm (D) x 1-7 mm(Cuff<br>H) | - |
| Material | Ti-6Al-4V ELI(Grade 5) | Ti-6Al-4V ELI(Grade 5) | - |
| Coating | TiN coating | TiN coating | - |
{11}------------------------------------------------
## Non-Clinical Test Data
The subject device was tested to evaluate its substantial equivalence according to the following standards.
- Biocompatibility tests according to ISO 10993-1:2009, ISO 10993-3:2014, ISO 10993-5:2009, ● ISO 10993-10:2010 and ISO 10993-11:2006.
- Fatigue Testing according to ISO 14801:2007 ●
Below tests were performed for predicate devices and leveraged for the subject device:
- End User Steam Sterilization Test according to ISO 17665-1:2006, -2:2009 and ANSI/AAMI ● ST79 referenced in K161244
The results of the above tests have met the criteria of the standards, and demonstrated the substantial equivalence with the predicate device.
Biocompatibility Test was conducted for the subject devices and it demonstrates that the subject device is biocompatible and substantial equivalence with the predicate device, K153521.
The end user sterilization test was performed for predicate device. K161244 and leveraged for the subject device because the product category, material, manufacturing process, facility, and packaging is the exactly the same as the predicate, K161244.
Fatigue evaluation was performed on the subject device under worst case scenario in accordance with ISO 14801 and "Guidance for Industry and Staff - Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments".
The non-clinical testing results demonstrate that the subject device is substantially equivalent to the predicate device.
#### Summary of clinical testing
No clinical testing was performed for this submission.
#### Substantial Equivalence Discussion
#### Similarities:
The subject device has identical, machining, manufacturing process, angulation and indication for use and similar design, dimension and technological characteristics as the predicate devices.
The subject device has been supposed to performance and product validations prior to release. Testing including biocompatibility tests and fatigue test has been performed to ensure the devices comply with the applicable International and US FDA Guidance.
#### Differences:
The differences between the subject device and predicate devices are detailed shape and dimension of each abutment and area of TiN Coated surface treatment.
Any differences between the subject device and predicate devices do not raise new types of substantially equivalent issues.
{12}------------------------------------------------
# Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification Dentis Co., Ltd. Concludes that the s-Clean TiN Coating Abutments is substantially equivalent to the predicate devices as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.