K171657 · Precision Spine, Inc. · MAX · Aug 11, 2017 · Orthopedic
Device Facts
Record ID
K171657
Device Name
ShurFit 2C Lumbar Interbody Fusion System
Applicant
Precision Spine, Inc.
Product Code
MAX · Orthopedic
Decision Date
Aug 11, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ShurFit 2C Interbody Fusion Devices (LLIF and TLIF) are indicated for intervertebral body fusion of the spine in skeletally mature patients. The device is designed to be used with autograft to facilitate fusion. One device is used per intervertebral space for the LLIF and TLIF. The ShurFit 2C Interbody Fusion Devices (LLIF and TLIF) are intended for use at either one level or two contiguous levels in the lumbar spine, for L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The lumbar device is to be used with supplemental fixation and autograft. Patients should have at least six months of non-operative treatment with a lumbar intervertebral fusion device.
Device Story
Interbody fusion device providing mechanical support to lumbar spine segments (L2-S1) until biological fusion occurs. Implants (LLIF/TLIF) inserted by surgeons during spinal fusion procedures. Device acts as spacer packed with autogenous bone graft. Subject device adds Commercially Pure Titanium (CpTi) and Hydroxyapatite (HA) coating to previously cleared PEEK cages to enhance surface properties. Output is mechanical stabilization of intervertebral space. Benefits include facilitation of spinal fusion in patients with degenerative disc disease.
Clinical Evidence
Bench testing only. Mechanical performance evaluated per ASTM F2077-11 and F2267-04 (static/dynamic axial compression, compression shear, expulsion, subsidence). Coating properties evaluated per ASTM F1147, F1044, F1160, and F1854 (tensile bond strength, shear strength, shear fatigue, solubility, microstructural analysis). No clinical data provided.
Technological Characteristics
Implants manufactured from PEEK Optima LT1 (ASTM F2026). Endplates coated with Commercially Pure Titanium (ASTM F1580) and Hydroxyapatite (ASTM F1185-03). Supplied sterile. Mechanical stabilization via interbody cage geometry.
Indications for Use
Indicated for intervertebral body fusion in skeletally mature patients with degenerative disc disease (DDD) and up to Grade I spondylolisthesis at one or two contiguous levels (L2-S1). Requires supplemental fixation and autograft. Patients must have failed at least six months of non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Precision Spine, Inc. Mr. Michael C. Dawson Sr. Director of Regulatory Affairs / Deputy General Counsel 2050 Executive Drive Pearl, Mississippi 39208
August 11, 2017
Re: K171657
Trade/Device Name: ShurFit 2C Lumbar Interbody Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX Dated: May 15, 2017 Received: June 5, 2017
Dear Mr. Dawson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K171657
Device Name ShurFit 2C Lumbar Interbody Fusion System
#### Indications for Use (Describe)
The ShurFit 2C Interbody Fusion Devices (LLF and TLIF) are indicated for intervertebral body fusion of the spine in skeletally mature patients. The device is designed to be used with autograft to facilitate fusion. One device is used per intervertebral space for the LLIF and TLIF.
The ShurFit 2C Interbody Fusion Devices (LLIF and TLIF) are intended for use at either one level or two contiguous levels in the lumbar spine, for L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The lumbar device is to be used with supplemental fixation and autograft. Patients should have at least six months of non-operative treatment with a lumbar intervertebral fusion device.
Type of Use (Select one or both, as applicable)
| <div style="display:flex; align-items:center;"><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> |
|-------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY
| Submitter's Name: | Precision Spine, Inc. |
|----------------------------|-------------------------------------------|
| Submitter's Address: | 2050 Executive Drive<br>Pearl, MS 39208 |
| Submitter's Telephone: | 973-455-7150 ext. 128 |
| Contact Person: | Michael C. Dawson |
| Date Summary was Prepared: | August 9, 2017 |
| Trade or Proprietary Name: | ShurFit 2C Lumbar Interbody Fusion System |
| Common or Usual Name: | Intervertebral Body Fusion Device |
| Classification: | Class II per 21 CFR §888.3080 |
| Product Code: | MAX |
| Classification Panel: | 87 Orthopedic Panel |
## DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
The ShurFit 2C Interbody Fusion System (LLIF and TLIF) consists of implants with various widths, heights, and lengths to accommodate individual patient anatomy and graft material size. It is to be packed with autogenous bone graft to facilitate fusion. The device is intended to provide mechanical support to the implanted level until biologic fusion is achieved. All components are manufactured from medical grade polyetheretherketone (PEEK Optima, LT1), and coated with Commercially Pure Titanium in compliance with ASTM F1580, and Hydroxyapatite in compliance with ASTM F1185-03. The products are supplied clean and "STERILE". The cages are included without a coating (already cleared K092193) or with a Cp Ti-HA (Commercially Pure Titanium - Hydroxyapatite) coating (this submission).
## CHANGE FROM PREDICATE:
The purpose of this submission is to add CpTi-HA (Commercially Pure Titanium -Hydroxyapatite) coated cages to the ShurFit Interbody Fusion System cleared in K092193. This submission seeks to add a line of ShurFit Lumbar devices (LLIF and TLIF) that have a coating on their endplates. The coating consists of a commercially pure Titanium (CpTi) based layer with a Hydroxyapatite (HA) layer on top of the CpTi. Precision Spine relies on the Master File of its contract manufacturer, APS Biomaterials Inc. located at 4011 Riverside Drive, Dayton, Ohio, for the composition and other technical characteristics of the HA and the CpTi coating. A copy of the
{4}------------------------------------------------
MAF authorization letter from APS is attached in Appendix A. The coated products will be supplied sterile in validated sterile packaging.
## INDICATIONS FOR USE
The ShurFit 2C Interbody Fusion Devices (LLIF and TLIF) are indicated for intervertebral body fusion of the spine in skeletally mature patients. The device is designed to be used with autograft to facilitate fusion. One device is used per intervertebral space for the LLIF and TLIF.
The ShurFit 2C Interbody Fusion Devices (LLIF and TLIF) are intended for use at either one level or two contiguous levels in the lumbar spine, for L2 to S1, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The lumbar device is to be used with supplemental fixation and autogenous autograft. Patients should have at least six months of non-operative treatment prior to treatment with a lumbar intervertebral fusion device.
## TECHNOLOGICAL CHARACTERISTICS
The intended use and technological features of the modifications/additions to the components of the ShurFit Lumbar Interbody Fusion System do not substantially differ from the legally marketed predicate devices.
All ShurFit 2C Interbody Fusion System components are manufactured from medical grade polyetheretherketone as described by ASTM F2026, specifically PEEK Optima LT1. The coated implants are provided sterile. The subject and predicate devices have identical technological characteristics and differ only in the fact that a CpTi-HA coating has been added and the coated devices will be provided supplied sterile. Additional predicates that contain CpTi-HA coating as well as CpTi or HA are listed below.
| 510k Number | Trade or Proprietary or Model Name | Manufacturer |
|-----------------------------|------------------------------------------------------------------|------------------------------|
| K092193 | Spinal USA Interbody Fusion Device | Precision Spine (Spinal USA) |
| K161809 (Primary Predicate) | ShurFit CpTi-HA ACIF Interbody Fusion System (Coating Predicate) | Precision Spine |
| K150061 | Lucent Ti Bond Systems | Spinal Elements |
| K133967 | Aurora Spine Interbody Fusion System | Aurora Spine |
| K150321 | EVOS HA Lumbar Interbody System | Cutting Edge Spine |
Table 1 Predicate Devices
{5}------------------------------------------------
The following is a brief summary of non-clinical tests provided to the agency for review:
Mechanical performance of the ShurFit 2C Interbody Fusion System (LLIF and TLIF) was evaluated under the standards of ASTM F2077-11, F2267-04, and the draft standard F-04.25.02.02. The following properties were tested:
- Static Axial Compression ●
- Dynamic Axial Compression
- Static Compression Shear
- Dynamic Compression Shear
- Static Expulsion
- o Static Subsidence
Coating properties were evaluated using the standards ASTM 1147, ASTM F1044, ASTM F1160, and ASTM F1854. The following properties were tested:
- Static Tensile Bond Strength ●
- Static Shear Strength
- Shear Fatigue ●
- Solubility and Dissolution
- Microstructural and Image Analysis of coated surface ●
Additional information on both the CpTi and the HA coatings can be found in the Master Access File of the coatings vendor, APS Materials, attached in Appendix A as well as Coating Characterization reports in Appendix B.
## CONCLUSION
The overall technology characteristics and mechanical engineering analysis lead to the conclusion that the ShurFit 2C Interbody Fusion System (LLIF and TLIF) is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.