K171655 · Roche Diagnostics Hematology, Inc. · GKZ · Mar 2, 2018 · Hematology
Device Facts
Record ID
K171655
Device Name
cobas m 511 integrated hematology analyzer
Applicant
Roche Diagnostics Hematology, Inc.
Product Code
GKZ · Hematology
Decision Date
Mar 2, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5220
Device Class
Class 2
Attributes
Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K171655 · Mar 2, 2018
cobas m 511 integrated hematology analyzer
Roche Diagnostics Hematology, Inc.
Residual whole blood samples from clinical laboratory sites
Residual clinical samples were used to perform method comparison studies against predicate devices to establish analytical performance and substantial equivalence.
The cobas m 511 integrated hematology analyzer is a quantitative, automated analyzer with cell locating capability. It is intended for in vitro diagnostic use by a skilled operator in the clinical laboratory. The system prepares a stained microscope slide from EDTA-anticoagulated whole blood. It utilizes computer imaging to count the formed elements of blood and provide an image-based assessment of cell morphology, which may be reviewed by the operator, and also allows for manual classification of unclassified cells. The instrument reports the following parameters: RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, %NRBC, #NRBC, WBC, %NEUT, %LYMPH, #LYMPH, %MONO, #MONO, %EO, #EO, %BASO, #BASO, PLT, MPV, %RET, #RET, HGB-RET.
Device Story
The cobas m 511 is a stand-alone, automated hematology analyzer for clinical laboratories. It processes EDTA-anticoagulated whole blood samples via closed-tube (rack) or open-tube modes. The system performs automated sample mixing, slide printing, and Romanowsky-type staining using DigiMAC3 reagents. It utilizes digital imaging with 10x (low magnification) and 50x (high magnification) objectives to locate, count, and classify blood cells (WBC, RBC, platelets, reticulocytes). The system generates a digital monolayer slide and provides numerical results for CBC, 5-part differential, and reticulocyte parameters. Results and cell images are transmitted to a viewing station for operator review. A skilled medical technologist uses the viewing station to monitor processing, review flagged samples, and perform manual morphological assessments. This automated imaging workflow reduces the need for manual slide preparation and microscopy, potentially improving laboratory throughput and standardizing morphological review.
Clinical Evidence
Clinical performance evaluated across five sites (3 US, 2 EU) using 1,859 residual K2EDTA venous whole blood samples. Compared against Sysmex XN-10. Accuracy assessed via Deming/Passing-Bablok regression; correlation coefficients (r) ranged from 0.559 to 0.999. Sensitivity/specificity for distributional/morphological abnormalities evaluated against 400-cell manual differential (N=426). Sensitivity for 'Any Morphological or Distributional Abnormality' was 95.9% (95% CI: 93.3-97.5). Precision/reproducibility studies (N=120 per control level) met acceptance criteria. Bench testing included linearity, carryover, interference, and stability.
Indicated for use by skilled clinical laboratory operators to perform complete blood counts, 5-part WBC differentials, and reticulocyte enumeration on EDTA-anticoagulated venous or capillary whole blood from pediatric and adult patients, including those with hematologic conditions or distributional abnormalities.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION MEMORANDUM
A. 510(k) Number:
K171655
B. Purpose for Submission:
Clearance of a new device
C. Manufacturer and Instrument Name:
Roche Diagnostics GmbH, cobas m 511 integrated hematology analyzer
D. Type of Test or Tests Performed:
Complete blood count and 5-part differential: RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, %NRBC, #NRBC, WBC, %NEUT, #NEUT, %LYMPH, #LYMPH, %MONO, #MONO, %EO, #EO, %BASO, #BASO, PLT, MPV, %RET, #RET, and HGB-RET parameters.
White blood cell (WBC), red blood cell (RBC), and platelet classification and characterization
E. System Descriptions:
1. Device Description:
The cobas m 511 system is an automated hematology analyzer with integrated slide making capability and digital cell imaging technology. It uses digital morphology to provide a complete blood count, 5-part differential, and reticulocyte enumeration from capillary or venous whole blood samples collected in K₂EDTA or K₃EDTA. It also uses digital imaging to classify red blood cells, white blood cells and platelets based on morphological characteristics.
The cobas m 511 system consists of the following major components:
- The analyzer
- Viewing station
- Associated consumables and components
- Cobas m 511 system software for the analyzer, viewing station and image analysis
- Printer
cobas m 511 analyzer
The cobas m 511 analyzer is a stand-alone hematology analyzer with integrated slide
{1}
making capability and digital cell imaging that contains a rack transport system for sample tubes. It has five processing stations: sample mixing, slide printing, slide staining, slide imaging (low and high magnification) and slide output. It also includes a separate compartment, which houses the necessary consumables and waste containers.
## cobas m 511 viewing station
The viewing station hardware is a computer with software that provides the graphical user interface for the system. It is designed to interface with a standard Laboratory Information System (LIS) for result reporting. It can be configured as the control station or as a review station
## Consumables and components
The DigiMAC³ stain pack is intended to fix and stain cells from a whole blood sample or control/calibrator material that have been applied to a slide by the cobas m 511 analyzer. The DigiMAC³ stain pack is comprised of the following four separate solutions, which are individually applied to each processed slide: DigiMAC³ fix, DigiMAC³ eosin, DigiMAC³ methylene blue, and DigiMAC³ rinse. The application of these stain solutions results in a Romanowsky type stain.
DigiMAC³ reticulocyte is intended to stain a whole blood sample or control material before it is applied to a slide by the cobas m 511 analyzer, in order to make the reticulocytes suitable for automated imaging or manual microscopy.
DigiMAC³ wash is used to clean all specimen-contacting surfaces, including blood fluid pathways, following processing of each blood sample.
DigiMAC³ clean is used to remove protein build-up from the surfaces of the cobas m 511 analyzer components that come in contact with blood samples.
The DigiMAC³ slide is a glass substrate on which a whole blood sample or control/calibrator material is applied and stained by the cobas m 511 analyzer, in order to enable automated imaging or manual microscopy.
## Software Used in cobas m 511 system
The cobas m 511 system includes three major software components: viewing station software, analyzer software, and the image analysis application.
## Viewing Station Software
The viewing station software provides the user interface to the cobas m 511 analyzer, including: managing system configuration and consumables; reporting analyzer and sample processing status; troubleshooting and error recover; quality control processing and results; reviewing sample results; classifying/reclassifying cells; and evaluating cell morphology.
## Analyzer Software
The analyzer software runs the analyzer computer and manages all hardware elements, sample processing, data management, and viewing station communications.
## Image Analysis Application
The image analysis application runs on the analyzer computer and communicates with the
2
{2}
cameras and motion-control hardware to acquire images at both the low- and high-magnification imaging stations. The program identifies and measures cells in the images, and it reports the cell counts and measurements to the analyzer software. It also saves images for display on the viewing station.
2. Principles of Operation:
The cobas m 511 system uses digital morphology from an automatically produced Romanowsky stained monolayer slide, to provide the complete blood count (CBC), automated white blood cell (WBC) differential, and the enumeration of reticulocytes. This is done through the use of proprietary imaging algorithms that locate, count and evaluate red and white blood cells, platelets and nucleated red blood cells.
The cobas m 511 system prints a consistent sample volume (1 uL) onto a glass microscope slide using a precision application method. In the stainer module, four reagents are applied to the cells on the slide: a fixative, an eosin stain, a methylene blue stain, and a rinse. Following staining, the slide is imaged at low magnification using a 10x lens to locate, image, and count white and red blood cells, nucleated red blood cells, and platelets. The system selects a specific number of white blood cell locations and provides their coordinates to the high magnification module for additional analysis. In the high magnification module, a 50x lens is used to classify white blood cell types, evaluate red blood cells, and platelets and to evaluate cellular morphology. The white blood cells are categorized into five types (NEUT, LYM, MONO, EO, BASO), or as unclassified cells by the instrument. This allows a skilled technologist to review images of the cells on the viewing station or to perform a full microscopic review of the slide. This 50x magnification is also used to identify and count the reticulocytes when a reticulocyte test is requested. For reticulocyte processing, an additional slide is produced from blood in the sample cup that is first incubated with a supravital stain. The cells are then printed onto the slide, stained using DigiMAC³ stain and imaged to determine the number and percentage of reticulocytes and reticulocyte hemoglobin.
Following processing, results and images can be viewed on the viewing station. The viewing station allows the operator to monitor system processes and status, and to view sample processing and results. The viewing station also allows a skilled medical technologist to perform a full white blood cell differential and a morphological review of white blood cells, red blood cells and platelets for samples requiring further review.
3. Modes of Operation:
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ☐ or No ☑
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ☐ or No ☑
4. Specimen Identification:
Specimen identification is manual entry by keyboard, hand-held barcode reader or automated identification by barcode reader.
{3}
4
5. Specimen Sampling and Handling:
Samples can be introduced onto the cobas m 511 system using either the closed tube (automated rack mode) or open tube modes. In the closed tube mode, the operator loads the sample tubes into a custom rack. The rack is then mechanically transported into the cobas m 511 system for processing. In this mode the cobas m 511 system automatically mixes, aspirates, and analyzes samples while the sample caps remain intact. In the open tube mode, the operator processes individual samples by manually mixing the sample collection tube by hand and then introducing the sample to the open port aspiration probe presented by the cobas m 511 system. The operator will position the tube such that the aspiration probe is immersed into the approximate center of the blood volume.
6. Calibration:
The DigiMAC³ calibrator is used for calibration of the cobas m 511 system. The calibrator is a stable suspension of cells of human or animal origin. It is used for calibrating the following hematology parameters of the cobas m 511 system: WBC, RBC, MCH, MCV, PLT, and MPV.
7. Quality Control:
DigiMAC³ controls are hematology quality control (QC) materials which consist of stable suspensions of red blood cells, white blood cells and platelets of human or animal origin. There are three levels of controls (L1, L2 and L3), which are used in conjunction with each other for monitoring the performance of the cobas m 511 analyzer.
8. Software:
FDA has reviewed applicant’s Hazard Analysis and Software Development processes for this line of product types:
Yes ☐ X ☐ or No ☐
F. Regulatory Information:
1. Regulation section:
21 CFR 864.5220, Differential Cell Counter
21 CFR 864.5260, Automated Cell Locating Device
2. Classification:
Class II
3. Product code:
GKZ – Counter, Differential Cell
JOY – Device, Automated Cell Locating
{4}
4. Panel:
Hematology 81
# G. Intended Use:
1. Indication(s) for Use:
The cobas m 511 integrated hematology analyzer is a quantitative, automated analyzer with cell locating capability. It is intended for in vitro diagnostic use by a skilled operator in the clinical laboratory. The system prepares a stained microscope slide from EDTA-anticoagulated whole blood. It utilizes computer imaging to count the formed elements of blood and provide an image-based assessment of cell morphology, which may be reviewed by the operator, and also allows for manual classification of unclassified cells. The instrument reports the following parameters: RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, %NRBC, #NRBC, WBC, %NEUT, #NEUT, %LYMPH, #LYMPH, %MONO, #MONO, %EO, #EO, %BASO, #BASO, PLT, MPV, %RET, #RET, HGB-RET.
2. Special Conditions for Use Statement(s):
For prescription use only.
# H. Substantial Equivalence Information:
1. Predicate Device Name(s) and 510(k) numbers:
Sysmex® XN-Series (XN-10, XN-20) Automated Hematology Analyzer (Sysmex Analyzer) - K112605
CellaVision® DM1200 Automated Hematology Analyzer (CellaVision Analyzer) - K092868
2. Comparison with Predicate Device:
| Similarities | | |
| --- | --- | --- |
| Item | Device cobas m 511 system | Predicate Sysmex XN Series |
| Indications for use | The cobas m 511 integrated hematology analyzer is a quantitative, automated analyzer with cell locating capability. It is intended for in vitro diagnostic use by a skilled operator in the clinical laboratory. The system prepares a stained microscope slide from EDTA-anticoagulated whole blood. It utilizes computer imaging to count the formed elements of blood and provide an image-based assessment of cell morphology, which may be reviewed by the operator, and also allows for manual classification of unclassified cells. The instrument reports the following parameters: RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, %NRBC, #NRBC, WBC, %NEUT, #NEUT, %LYMPH, #LYMPH, %MONO, #MONO, %EO, #EO, %BASO, #BASO, PLT, MPV, %RET, #RET, HGB-RET. | The XN-Series modules (XN-10, XN-20) are quantitative multi-parameter automated hematology analyzers intended for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XN-Series modules classify and enumerate the |
{5}
| Similarities | | |
| --- | --- | --- |
| Item | Device
cobas m 511 system | Predicate
Sysmex XN Series |
| | slide from EDTA-anticoagulated whole blood. It utilizes computer imaging to count the formed elements of blood and provide an image-based assessment of cell morphology, which may be reviewed by the operator, and also allows for manual classification of unclassified cells. The instrument reports the following parameters: RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, %NRBC, #NRBC, WBC, %NEUT, #NEUT, %LYMPH, #LYMPH, %MONO, #MONO, %EO, #EO, %BASO, #BASO, PLT, MPV, %RET, #RET, HGB-RET. | following parameters in whole blood: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT, NEUT%/#, LYMPH%/#, MONO%/#, EO%/#, BASO%/#, IG%/#, RDW-CV, RDW-SD, MPV, NRBC%/#, RET%/#, IPF, IRF, RET-He and has a Body Fluid mode for body fluids. The Body Fluid mode enumerates the WBC-BF, RBC-BF, MN%/#, PMN%/#, and TC-BF# parameters in cerebrospinal fluid (CSF),serous fluids (peritoneal, pleural) and synovial fluids. Whole blood should be collected in K2 or K3 EDTA anticoagulant and, Serous and Synovial fluids in K2 EDTA anticoagulant to prevent clotting of fluid. The use of anticoagulants with CSF specimens is neither required nor recommended. |
| Sample identification | Automated barcode reading of sample tube identifier. Manual keyboard entry. Bi-directional instrument to LIS interface: patient demographics, orders, results. | Same |
| Anticoagulant | K2 and K3 ethylenediaminetetraacetic acid (EDTA) | Same |
| Service diagnostics | On-board system diagnostics. Manufacturer can perform web-based diagnostics. | Same |
| Information transfer to and from instrument | Through LIS or manual entry on a computer screen. | Same |
{6}
| Differences | | |
| --- | --- | --- |
| Item | Device: cobas m 511 system | Predicate: Sysmex Analyzer |
| Parameters | Not available | Whole blood parameters - IG%/#, RDW-CV, IPF, IRF Body fluid parameters parameters in cerebrospinal fluid (CSF),serous fluids (peritoneal, pleural) and synovial fluids - WBC-BF, RBC-BF, MN%/#, PMN%/#, and TC-BF# |
| Principle of operation | Characterizes and identifies cells using digital imaging of stained cells on a microscope slide. Low magnification location and imaging of white and red blood cells and platelets using a 10x objective, combined with illumination optics and a camera. High magnification imaging of white blood cell types and cell morphology using a 50x objective, combined with illumination optics and a camera. Hemoglobin measurement using LED light absorption. | Characterizes and identifies cells based on detection of direct-current resistance light scatter, fluorescence, and adaptive cluster analysis. Flow cytometry method using a semiconductor laser to analyze physiological and chemical characteristics of cells and other biological particles Hydro Dynamic Focusing (DC Detection) with aperture to count the RBC and PLT and calculate the HCT via the RBC pulse height detection method. SLS-Hemoglobin Method using hemoglobin absorption after chemically lysing cells and using LED light. |
| Sample preparation procedure | Printing of blood onto a microscope slide followed by staining of the slide. | Flow analysis after dilution and mixing with reagents. |
| Modes of operation | Closed tube (automated rack) mode Open tube (manual) mode | Sampler Analysis Mode(Closed Cap) Manual Analysis Mode (Closed and Open Cap) Pre-dilute Analysis Mode Low WBC Analysis Mode Body Fluid Mode |
| Calibrator | DigiMAC3 calibrator | XN CAL XN CAL PF |
{7}
| Differences | | |
| --- | --- | --- |
| Item | Device: cobas m 511 system | Predicate: Sysmex Analyzer |
| Quality control | DigiMAC3-control L1, L2, L3 | XN Check (3 levels) |
| Sample types | Whole blood | Whole blood and body fluids |
| Sample aspiration volume | 30μL (closed-tube and open-tube mode) | 88μL (Sampler Mode-Whole Blood) 88μL (Manual Mode- Whole Blood) 70μL (Manual Mode- Diluted Blood) |
| Throughput | Approximately 60 samples/hour maximum | Approximately 100 samples/hour maximum |
| Reagents | DigiMAC3 stain pack, including: DigiMAC3 fix (fixative) DigiMAC3 eosin (stain) DigiMAC3 methylene blue (stain) DigiMAC3 rinse (rinse) DigiMAC3 reticulocyte (stain) | CELLPACK DCL (Diluent) CELLPACK DFL (Diluent) LYSERCELL WNR (Lyse) LYSERCELL WDF (Lyse) LYSERCELL WPC (Lyse) FLUOROCELL WNR (Stain) FLUOROCELL WDF (Stain) FLUOROCELL RET (Stain) FLUOROCELL PLT (Stain) FLUOROCELL WPC (Stain) SULFOLYSER (Lyse) |
| Cleaning solutions | DigiMAC3 wash DigiMAC3 clean | CELLCLEAN AUTO |
| Operational conditions | 18 to 27°C ambient temperature 20–60% relative humidity | 15 to 30°C ambient temperature 20–85% relative humidity |
<table><tr><td colspan="3">Similarities</td></tr><tr><td>Item</td><td>Device cobas m 511 system</td><td>Predicate CellaVision DM1200</td></tr><tr><td>Indications for use</td><td>The cobas m 511 integrated hematology analyzer is a quantitative, automated analyzer with cell locating capability. It is intended for in vitro diagnostic use by a skilled operator in the clinical laboratory. The system prepares a stained microscope slide from EDTA-anticoagulated whole blood. It utilizes computer imaging to</td><td>DM1200 is an automated cell-locating device. DM1200 automatically locates and presents images of blood cells on peripheral blood smears. The operator identifies and verifies the suggested classification of each cell according to type. DM1200 is intended to be used by skilled operators, trained in the use of the device and in recognition</td></tr><tr><td></td><td>the blood. It is used to identify and identify the blood and blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to identify and identify blood vessels. It is used to
{8}
| Similarities | | |
| --- | --- | --- |
| Item | Device
cobas m 511 system | Predicate
CellaVision DM1200 |
| | count the formed elements of blood and provide an image-based assessment of cell morphology, which may be reviewed by the operator, and also allows for manual classification of unclassified cells. The instrument reports the following parameters: RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, %NRBC, #NRBC, WBC, %NEUT, #NEUT, %LYMPH, #LYMPH, %MONO, #MONO, %EO, #EO, %BASO, #BASO, PLT, MPV, %RET, #RET, HGB-RET. | of blood cells. |
| Methodology | Performs automated analyses of whole blood cells using a combination of low and high magnification imaging of stained cells on a glass microscope slide, with subsequent operator review of computerized images or glass microscope slide for flagged cases. | Same |
| Sample identification | Automated barcode reading of sample identifier. Manual keyboard entry. | Same |
| Stains | Romanowsky type stain | Same |
| Analysis technique | Locates, identifies, and counts the various types of white blood cells under the automated microscope. Red blood cell and platelet morphology can be assessed by the examiner. | Same |
| Presentation of samples | Cells can be observed on a computer display. Cells can be observed through a microscope. | Same |
9
{9}
| Similarities | | |
| --- | --- | --- |
| Item | Device cobas m 511 system | Predicate CellaVision DM1200 |
| Analyte targets | Stained components of cells such as DNA, RNA, and proteins. | Same |
| Intrinsic color compensation | Algorithms and classifiers tolerate stain variations. Display color cannot be adjusted. | Same |
| Information transfer to/from instrument | Through LIS or manual entry on a computer screen. | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device cobas m 511 system | Predicate CellaVision DM1200 |
| Principle of operation | Low magnification location and imaging of white and red blood cells and platelets. High magnification imaging of white blood cell types and cell morphology. Displays analysis results, graphics, and images on a computer screen. Results can be printed on available printer or transmitted to a host computer. | The analysis process consists of an overview image processing and a cell- location step. The overview image is used to find cells of interest and to obtain an overall impression of the sample. The overview image can have one 10x zoom level or both 10x and 50x zoom levels. The cell-location step uses the optical unit and a camera to obtain images of the identified images and stores the images in a database. |
| Specimen types | Whole Blood | Whole Blood Body Fluids |
| Sample preparation | Automated slide preparation and staining. | Automated or manual slide preparation and staining. |
| Light sources and detector | LED light sources with specific blue, green, yellow and red wavelengths. Black and White cameras. | White light source. Color camera with RGB filters on individual pixels. |
# I. Special Control/Guidance Document Referenced (if applicable):
Class II Special Controls Guidance Document: Premarket Notifications for Automated
{10}
Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA
CLSI EP05-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline – Third Edition
CLSI EP06-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline
CLSI EP07-A2, Interference Testing In Clinical Chemistry; Approved Guideline – Second Edition
CLSI EP09-A3, Measurement Procedure Comparison and Bias Estimation Using Patient Samples; Approved Guideline – Third Edition
CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition
CLSI EP28-A3c, Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline – Third Edition
CLSI H20-A2, Reference Leukocyte (WBC) Differential Count (Proportional) and Evaluation of Instrumental Methods; Approved Standard – Second Edition
CLSI H26-A2, Validation, Verification, and Quality Assurance of Automated Hematology Analyzers; Approved Standard – Second Edition
IEC 61010-1:2010, Safety requirements for electrical equipment for measurement, control, and laboratory use – Part 1: General requirements
IEC 62304:2006 Medical device software – Software life cycle processes
ISO 14971:2007 Medical devices – Application of risk management to medical devices
## J. Performance Characteristics:
1. Analytical Performance:
a. Accuracy:
**Accuracy and Comparison at Medical Decision Levels**
This study was conducted to evaluate accuracy of the cobas m 511 system compared to reference methods and the Symex XN-10 for samples at medical decision levels for hemoglobin (HGB), hematocrit (HCT), mean cell volume (MCV), platelet (PLT) and reticulocytes (RET). The results of 50 samples from the cobas m 511 system were compared to the following methods: cyanmethemoglobin for the HGB parameter, packed cell volume (PCV) for the HCT parameter, and a ratio using the PCV and RBC count to determine the MCV parameter and Sysmex analyzer for PLT and RET paremeters. Bias to the applicable method and its 95% CI were determined both separately and combined using data obtained from each cobas m 511 system and each medical decision level studied. All results were within the predefined acceptance criteria.
11
{11}
12
# Accuracy for Leukopenia Samples
This study was conducted to evaluate accuracy of the cobas m 511 system compared to the 200-cell manual differential of leukopenia samples performed by two medical technologists using the reference method. Ten granulocytopenia samples with granulocytes ≤ 1.5 x 10³/μL and 20 leukopenia samples with white blood cell counts ≤ 4.0 x 10³/μL were included in the analysis. For both the test and reference methods, the five WBC differential parameters, nucleated red blood cells and immature/abnormal cell types (e.g. blast and immature granulocytes) were differentiated and enumerated. Sensitivity and specificity were calculated for the detection of distributional and morphological abnormalities for the cobas m 511 image, automated differential, system messages and slides versus the reference method. The test method results were compared with the results of the reference method to assess bias. The results met the overall acceptance criteria.
## b. Method Comparison:
### Method Comparison of Subjects with Medical Conditions
This study was conducted to demonstrate that results produced by the cobas m 511 system are comparable to those produced by the Sysmex XN-10 analyzer when samples from patients with specific medical conditions are evaluated. One-hundred thirty residual whole blood samples were obtained for 23 targeted medical conditions: leukopenia, leukocytosis, neutropenia, neutrophilia, lymphopenia, lymphocytosis, eosinophilia, basophilia, thrombocytopenia, thrombocytosis, polycythemia, chronic myeloid leukemia, chronic lymphocytic leukemia, acute myeloid leukemia, acute lymphocytic leukemia, myelodysplastic syndrome, lymphoma, sickle cell anemia, thalassemia, hemoglobin SC disease, hemolytic anemia, multiple myeloma and septicemia. Each sample was processed in singlicate on the cobas m 511 system and Sysmex XN-10. Deming regression (slope, intercept, 95% confidence intervals and correlation coefficient) was performed according to CLSI EP09-A3. The results met the predefined acceptance criteria.
### Method Comparison
Method comparison studies were performed at five clinical sites (three sites in U.S. and two sites in the European Union) to evaluate the performance of the cobas m 511 system compared to the Sysmex XN-10 analyzer using a total of 1,859 residual K₂EDTA venous whole blood samples. This study included both pediatric (neonate to ≤ 21 years) and adult subjects, as well as samples from patients with hematologic conditions and distributional abnormalities (anemia, thrombocytopenia, leukopenia, leukocytosis, elevated hematocrit, elevated nucleated red blood cells).
Bias was determined based on the results of either a Passing-Bablok or Deming regression model. For all parameters, except those with an expected large number of valid results of zero (i.e., %EO, #EO, %BASO and #BASO), the Passing-Bablok regression model was used to determine the intercept and slope of the regression line.
{12}
For those parameters with a large number of valid results of zero, the Deming (orthogonal) regression model was used. For either regression model used, the sample size (N), Pearson's correlation coefficient (r), slope and intercept along with their $95\%$ CIs, and the range of comparative measurement procedure values are reported. Bias was determined based on the results of either a Passing-Bablok or Deming regression model. The calculated bias along with its $95\%$ CI are also reported for the critical bias points at the low end, high end, and crossover point. All results were within the pre-defined acceptance criteria.
Correlation of Cobas m 511 vs. Sysmex XN-10 - Combined Sites
| Parameter [Units] | Sample Range | Pearson's (r) | Intercept (95% CI) | Slope (95% CI) |
| --- | --- | --- | --- | --- |
| WBC [103/μL] | 0.04–247.04 | 0.999 | 0.02 (-0.01, 0.05) | 1.012 (1.007, 1.017) |
| RBC [106/μL] | 1.15–7.21 | 0.974 | 0.02 (-0.01, 0.04) | 0.991 (0.985, 1.000) |
| HGB [g/dL] | 4.20–21.20 | 0.970 | -0.33 (-0.41, -0.24) | 1.046 (1.039, 1.053) |
| HCT [%] | 13.50–66.00 | 0.953 | -0.72 (-1.06, -0.35) | 1.043 (1.033, 1.053) |
| MCV [fL] | 58.20–119.20 | 0.887 | -3.03 (-5.06, -0.87) | 1.060 (1.035, 1.083) |
| MCH [pg] | 17.58–40.75 | 0.956 | 1.37 (0.73, 1.80) | 0.974 (0.960, 0.996) |
| MCHC [g/dL] | 26.59–36.80 | 0.559 | 15.71 (14.73, 16.45) | 0.522 (0.500, 0.552) |
| RDW [%] | 10.70–29.40 | 0.913 | 2.46 (2.15, 2.74) | 0.870 (0.848, 0.892) |
| RDW-SD [fL] | 31.70–97.10 | 0.912 | 5.48 (4.48, 6.46) | 0.940 (0.917, 0.963) |
| PLT [103/μL] | 1.00–1061.00 | 0.973 | -11.03 (-13.21, -8.94) | 1.020 (1.008, 1.031) |
| MPV [fL] | 8.00–13.00 | 0.772 | -0.91 (-1.40, -0.20) | 1.063 (1.000, 1.111) |
| %NRBC [/100 WBC] | 0.00–186.10 | 0.981 | N/A | N/A |
| #NRBC [103/μL] | 0.00–9.59 | 0.995 | N/A | N/A |
| %NEUT [%] | 10.10–94.00 | 0.989 | 1.62 (1.12, 2.15) | 1.012 (1.004, 1.019) |
| %LYMPH [%] | 0.70–83.00 | 0.989 | -0.23 (-0.37, -0.08) | 0.977 (0.971, 0.983) |
| %MONO [%] | 0.60–24.50 | 0.913 | -0.60 (-0.82, -0.50) | 1.000 (1.000, 1.026) |
| %EO [%] | 0.00–32.90 | 0.973 | -0.08 (-0.13, -0.03) | 1.042 (1.030, 1.054) |
{13}
Estimated Bias of Cobas m 511 vs. Sysmex XN-10 - Combined Sites
| Parameter [Units] | Sample Range | Evaluation Range | | | Bias (95% CI) | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | Low End | Crossover Point | High End | At Low Limit | Crossover Point | | At High Limit |
| WBC [103/μL] | 0.04–247.04 | 0.50 | 4.00 | 30.00 | 0.03 (0.00, 0.05) | 0.07 (0.05, 0.08) [103/μL] | 1.7 (1.34, 2.09) [%] | 1.26 (0.87, 1.69) |
| RBC [106/μL] | 1.15–7.21 | 4.00 | N/A | 5.50 | -0.41 (-0.59, -0.25) | N/A | | -0.56 (-0.82, -0.18) |
| HGB [g/dL] | 4.20–21.20 | 4.50 | 10.00 | 21.20 | -0.12 (-0.17, -0.07) | 0.14 (0.11, 0.15) [g/dL] | 1.35 (1.15, 1.53) [%] | 3.08 (2.70, 3.42) |
| HCT [%] | 13.50–66.00 | 14.00 | N/A | 66.00 | -0.87 (-2.40, 0.71) | N/A | | 3.20 (2.68, 3.68) |
| MCV [fL] | 58.20–119.20 | 80.00 | N/A | 100.00 | 2.25 (1.97, 2.58) | N/A | | 3.01 (2.65, 3.36) |
| MCH [pg] | 17.58–40.75 | 28.00 | N/A | 34.00 | 2.33 (2.14, 2.52) | N/A | | 1.47 (1.19, 1.76) |
| MCHC [g/dL] | 26.59–36.80 | 32.00 | N/A | 36.00 | 1.26 (1.09, 1.43) | N/A | | -4.19 (-4.44, -3.92) |
| RDW [%] | 10.70–29.40 | 12.00 | N/A | 14.60 | 7.43 (6.98, 7.86) | N/A | | 3.78 (3.39, 4.11) |
| RDW-SD [fL] | 31.70–97.10 | 40.00 | N/A | 60.00 | 7.70 (7.31, 8.13) | N/A | | 3.13 (2.42, 3.90) |
| PLT [103/μL] | 1.00–1061.00 | 10.00 | 75.00 | 1000.00 | -10.83 (-12.94, -8.85) | -9.54 (-11.04, -8.10) [103/μL] | -12.73 (-14.72, -10.80) [%] | 0.88 (-0.08, 1.84) |
| MPV [fL] | 8.00–13.00 | 8.00 | N/A | 10.20 | -5.08 (-6.62, -2.50) | N/A | | -2.63 (-3.03, -1.96) |
| %NRBC | 0.00–186.10 | 0.00 | N/A | 1.50 | -0.03 | N/A | | ‡mean |
{14}
| Parameter [Units] | Sample Range | Evaluation Range | | | Bias (95% CI) | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | Low End | Crossover Point | High End | At Low Limit | Crossover Point | | At High Limit |
| [/100 WBC] | | | | | (-0.06, 0.01) | | | difference |
| #NRBC [103/μL] | 0.00–9.59 | 0.00 | N/A | 0.10 | 0.00 (0.00, 0.00) | N/A | | \( \overline{\mathbf{x}}_{\text{mean}} \) difference |
| %NEUT [%] | 10.10–94.00 | 40.00 | N/A | 85.00 | 5.21 (4.60, 5.84) | N/A | | 3.06 (2.82, 3.30) |
| %LYMPH [%] | 0.70–83.00 | 25.00 | 40.00 | 65.00 | -0.81 (-0.91, -0.70) | -1.15 (-1.32, -0.98) [%LYMPH] | -2.89 (-3.30, -2.44) [%] | -2.66 (-3.13, -2.15) |
| %MONO [%] | 0.60–24.50 | 2.00 | N/A | 10.00 | -0.60 (-0.78, -0.50) | N/A | | -0.60 (-0.70, -0.50) |
| %EO [%] | 0.00–32.90 | 0.00 | N/A | 5.00 | -0.08 (-0.13, -0.03) | N/A | | 0.13 (0.09, 0.17) |
| %BASO [%] | 0.00–3.10 | 0.00 | N/A | 1.00 | -0.30 (-0.35, -0.25) | N/A | | 0.35 (0.31, 0.39) |
| #NEUT [103/μL] | 0.37–37.66 | 1.00 | 1.50 | 10.00 | 0.15 (0.12, 0.17) | 0.16 (0.14, 0.19) [103/μL] | 10.98 (9.18, 12.37) [%] | 3.92 (3.43, 4.36) |
| #LYMPH [103/μL] | 0.02–12.66 | 0.50 | 1.50 | 3.00 | -0.04 (-0.05, -0.03) | -0.01 (-0.02, 0.00) [103/μL] | -0.71 (-1.30, -0.16) [%] | 1.26 (0.58, 1.87) |
| #MONO [103/μL] | 0.01–6.14 | 0.10 | N/A | 1.50 | -0.03 (-0.04, -0.02) | N/A | | -0.03 (-0.06, -0.03) |
| #EO [103/μL] | 0.00–7.17 | 0.00 | N/A | 0.50 | -0.01 (-0.01, 0.00) | N/A | | 0.03 (0.03, 0.04) |
| #BASO [103/μL] | 0.00–0.46 | 0.00 | N/A | 0.10 | -0.02 (-0.03, -0.02) | N/A | | 0.04 (0.04, 0.05) |
| %RET [%] | 0.05–12.93 | 0.50 | 1.67 | 2.50 | -0.31 (-0.34, -0.28) | -0.20 (-0.21, -0.18) [%RET] | -11.91 (-12.75, -10.83) [%] | -4.84 (-5.80, -3.81) |
| #RET [106/μL] | 0.00–0.42 | 0.02 | N/A | 0.15 | -0.01 (-0.01, -0.01) | N/A | | 0.00 (0.00, 0.00) |
| HGB-RET [pg] | 16.23–45.00 | 23.00 | N/A | 40.00 | -0.23 (-2.11, 1.61) | N/A | | 5.84 (4.94, 6.83) |
bias was calculated using the mean difference between the cobas m 511 system result and the Sysmex XN-10
# Sensitivity and Specificity
For the sensitivity and specificity analysis, the cobas m 511 automated differential, cobas m 511 system messages and cobas m 511 images were compared to a 400-cell differential derived from two independent 200-cell microscopic reviews of a blood smear (reference method). For this analysis, separate $2 \times 2$ tables were constructed in order to determine sensitivity for both morphological and distributional abnormalities. Results for sample size (N) and numbers of true positives (TP), false positives (FP), true negatives (TN), and false negatives (FN) are reported, as well as sensitivity, specificity, and efficiency.
{15}
Results Summary for cobas m 511 Images versus Reference Method for All Sites Combined
| Distributional Abnormalities | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | N | TP | FP | FN | TN | Sensitivity (95%CI) | Specificity (95%CI) | Efficiency (95%CI) |
| %NEUT | 426 | 246 | 18 | 23 | 139 | 91.40% (87.5, 94.2) | 88.50% (82.6, 92.6) | 90.40% (87.2, 92.8) |
| %LYMPH | 426 | 273 | 12 | 35 | 106 | 88.60% (84.6, 91.7) | 89.80% (83.1, 94.1) | 89.00% (85.6, 91.6) |
| %MONO | 426 | 93 | 36 | 48 | 249 | 66.00% (57.8, 73.3) | 87.40% (83.0, 90.7) | 80.30% (76.2, 83.8) |
| %EO | 426 | 85 | 75 | 20 | 246 | 81.00% (72.4, 87.3) | 76.60% (71.7, 80.9) | 77.70% (73.5, 81.4) |
| %BASO | 426 | 8 | 67 | 6 | 345 | 57.10% (32.6, 78.6) | 83.70% (79.9, 87.0) | 82.90% (79.0, 86.1) |
| Any Distributional Abnormality or Combination of Distributional Abnormalities | 426 | 341 | 19 | 16 | 50 | 95.50% (92.8, 97.2) | 72.50% (61.0, 81.6) | 91.80% (88.8, 94.0) |
| Morphological Abnormalities | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | N | TP | FP | FN | TN | Sensitivity (95%CI) | Specificity (95%CI) | Efficiency (95%CI) |
| Blasts | 426 | 59 | 8 | 9 | 350 | 86.80% (76.7, 92.9) | 97.80% (95.7, 98.9) | 96.00% (93.7, 97.5) |
| Immature Granulocytes | 426 | 45 | 8 | 14 | 359 | 76.30% (64.0, 85.3) | 97.80% (95.8, 98.9) | 94.80% (92.3, 96.6) |
| NRBCs | 426 | 14 | 1 | 7 | 404 | 66.70% (45.4, 82.8) | 99.80% (98.6, 100.0) | 98.10% (96.3, 99.0) |
| Any Morphological Abnormality or Combination of Morphological Abnormalities | 426 | 99 | 6 | 14 | 307 | 87.60% (80.3, 92.5) | 98.10% (95.9, 99.1) | 95.30% (92.9, 96.9) |
| Overall – Acceptance Criteria Only Applied to These Results | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | N | TP | FP | FN | TN | Sensitivity (95%CI) | Specificity (95%CI) | Efficiency (95%CI) |
| Any Morphological or Distributional Abnormality, or combination | 426 | 351 | 16 | 15 | 44 | 95.90% (93.3, 97.5) | 73.30% (61.0, 82.9) | 92.70% (89.9, 94.8) |
Summary of Results for cobas m 511 Automated Differential versus Reference Method for All Sites Combined
| Distributional Abnormalities | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | N | TP | FP | FN | TN | Sensitivity (95% CI) | Specificity (95% CI) | Efficiency (95% CI) |
| %NEUT | 355 | 186 | 18 | 21 | 130 | 89.9% (85.0, 93.3) | 87.8% (81.6, 92.2) | 89.0% (85.3, 91.9) |
| %LYMPH | 355 | 228 | 8 | 23 | 96 | 90.8% (86.6, 93.8) | 92.3% (85.6, 96.1) | 91.3% (87.9, 93.8) |
| %MONO | 355 | 57 | 31 | 38 | 229 | 60.0% (49.9, 69.3) | 88.1% (83.6, 91.5) | 80.6% (76.1, 84.3) |
| %EO | 355 | 64 | 44 | 15 | 232 | 81.0% (71.0, 88.1) | 84.1% (79.3, 87.9) | 83.4% (79.2, 86.9) |
{16}
| Distributional Abnormalities | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | N | TP | FP | FN | TN | Sensitivity (95% CI) | Specificity (95% CI) | Efficiency (95% CI) |
| %BASO | 355 | 6 | 31 | 5 | 313 | 54.5% (28.0, 78.7) | 91.0% (87.5, 93.6) | 89.9% (86.3, 92.6) |
| Overall – Acceptance Criteria Only Applied to These Results | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | N | TP | FP | FN | TN | Sensitivity (95% CI) | Specificity (95% CI) | Efficiency (95% CI) |
| Any Distributional Abnormality or Combination of Distributional Abnormalities | 355 | 272 | 17 | 16 | 50 | 94.40% (91.2, 96.6) | 74.60% (63.1, 83.5) | 90.70% (87.2, 93.3) |
Summary of Results for cobas m 511 System Messages versus Reference Method for All Sites Combined
| Message | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | N | TP | FP | FN | TN | Sensitivity (95% CI) | Specificity (95% CI) | Efficiency (95% CI) |
| "Suspect Blasts?" | 439 | 76 | 5 | 0 | 358 | 100.00% (95.2, 100.0) | 98.60% (96.8, 99.4) | 98.90% (97.4, 99.5) |
| "Suspect Immature Granulocytes?" | 439 | 53 | 10 | 8 | 368 | 86.90% (76.2, 93.2) | 97.40% (95.2, 98.6) | 95.90% (93.6, 97.4) |
| "Suspect Atypical Lymphocytes?" | 439 | 13 | 8 | 1 | 417 | 92.90% (68.5, 98.7) | 98.10% (96.3, 99.0) | 97.90% (96.2, 98.9) |
| Overall – Acceptance Criteria Only Applied to These Results | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | N | TP | FP | FN | TN | Sensitivity (95% CI) | Specificity (95% CI) | Efficiency (95% CI) |
| Any Message or Combination of Messages | 439 | 118 | 10 | 9 | 302 | 92.90% (87.1, 96.2) | 96.80% (94.2, 98.2) | 95.70% (93.3, 97.2) |
## Morphological Grading
To determine the performance of the cobas m 511 system (images and slides) with respect to reporting of morphology grading, data was analyzed using percent agreement analysis. The morphology grading of 163 sample results from cobas m 511 images and cobas m 511 slides was compared to morphology grading obtained by a trained independent reviewer (IR) from a single blood smear. WBC morphology, platelet morphology and RBC size, color and shape characteristics that were graded using a quantitative grading system (e.g., 0 to 4+ grading scale), the sample size (N), the number of results that agree (A), the number of results that disagree (D), and the percent agreement were reported. RBC inclusions (parasites, basophilic stippling, Howell-jolly bodies, and Pappenheimer bodies) were graded using a present/not present grading system, the sample size (N), the number of results that agree (A and D), the number of results that disagree (B and C), the percent positive and negative agreements (PPA and NPA, respectively), and total agreement are reported. The results met the predefined acceptance criteria.
{17}
Morphology Assessment Results for cobas m 511 Images (Graded Criteria) for All Sites Combined
| Morphological Characteristic | Sample Size (N) | A | D | Percent Agreement |
| --- | --- | --- | --- | --- |
| WBC Morphology | | | | |
| Hypogranular/ Agranular Granulocytes | 163 | 162 | 1 | 99.4 |
| Hyposegmentation | 163 | 163 | 0 | 100.0 |
| Toxic Vacuolation | 163 | 158 | 5 | 96.9 |
| Döhle Bodies | 163 | 157 | 6 | 96.3 |
| Toxic Granulation | 163 | 148 | 15 | 90.8 |
| Hypersegmentation | 163 | 161 | 2 | 98.8 |
| Smudge Cells | 163 | 162 | 1 | 99.4 |
| Platelet Satellitosis | 163 | 163 | 0 | 100.0 |
| RBC Morphology | | | | |
| Anisocytosis | 163 | 143 | 20 | 87.7 |
| Microcytosis | 163 | 141 | 22 | 86.5 |
| Macrocytosis | 163 | 148 | 15 | 90.8 |
| Hypochromia | 163 | 145 | 18 | 89.0 |
| Polychromasia | 163 | 155 | 8 | 95.1 |
| Poikilocytosis | | | | |
| Tear Drops | 163 | 161 | 2 | 98.8 |
| Blister Cells | 163 | 163 | 0 | 100.0 |
| Target Cells | 163 | 140 | 23 | 85.9 |
| Schistocytes | 163 | 162 | 1 | 99.4 |
| Echinocytes | 163 | 158 | 5 | 96.9 |
| Elliptocytes/Ovalocytes | 163 | 151 | 12 | 92.6 |
| Stomatocytes | 163 | 162 | 1 | 99.4 |
| Sickle Cells | 163 | 162 | 1 | 99.4 |
| Spherocytes | 163 | 163 | 0 | 100.0 |
| Acanthocytes | 163 | 161 | 2 | 98.8 |
| PLT Morphology | | | | |
| Large Platelets | 163 | 143 | 20 | 87.7 |
| Giant Platelets | 163 | 155 | 8 | 95.1 |
| Agranular/Hypogranular Platelets | 163 | 163 | 0 | 100.0 |
| Platelet Clumps | 163 | 151 | 12 | 92.6 |
| Megakaryocytic Cells | 163 | 163 | 0 | 100.0 |
18
{18}
Morphology Assessment Results for cobas m 511 Images (Present/Not Present Criteria) for All Sites Combined
| Morphological Characteristic | Sample Size (N) | A | B | C | D | PPA | NPA | Percent Agreement |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Parasites | 203 | 12 | 9 | 1 | 181 | 57.1 | 99.5 | 95.1 |
| Basophilic Stippling | 163 | 15 | 28 | 1 | 119 | 34.9 | 99.2 | 82.2 |
| Howell-Jolly Bodies | 163 | 9 | 2 | 3 | 149 | 81.8 | 98.0 | 96.9 |
| Pappenheimer Bodies | 163 | 4 | 2 | 3 | 154 | 66.7 | 98.1 | 96.9 |
| Auer Rods | 211 | 5 | 6 | 2 | 198 | 45.5 | 99.0 | 96.2 |
c. Precision/Reproducibility:
Two studies were conducted to assess repeatability on the cobas m 511 system. The first study evaluated inter-instrument precision of whole blood samples and the second study evaluated within-run precision.
Inter-instrument precision was evaluated using whole blood samples from five healthy volunteer donors and 21 residual abnormal samples. Eleven replicates were processed from each sample on three different cobas m 511 systems (33 replicates total per sample) at one site. Abnormal samples included low and high reticulocyte counts (%RET ≤ 0.5% and ≥ 5.0%, respectively) and medical decision level samples including anemia (HGB 6–10 g/dL), thrombocytopenia (PLT ≤ 10 × 10³/μL), and severe leukopenia (WBC 0–2 × 10³/μL). The mean, standard deviation (SD) and percent coefficient of variation (% CV) with corresponding 95% confidence intervals were calculated for each parameter using a variance component analysis. Within-run precision (repeatability), between-instrument variability, and total variability (inter-instrument precision) were analyzed. All results met the predefined acceptance criteria.
To evaluate repeatability (i.e., within-run precision) whole blood samples were processed 31 consecutive times on the cobas m 511 system. Residual whole blood samples were selected for the WBC, RBC, HGB and PLT parameters at targeted low, middle and high ranges. In addition, samples at medical decision levels of anemia, thrombocytopenia, severe leukopenia, and nucleated red blood cells were evaluated. At each of the four sites, 48 samples were targeted for collection (36 residual whole blood samples and 12 medical decision level samples). Repeatability results were calculated for all sites combined and all results met their predefined acceptance criteria.
Reproducibility
Two studies were conducted to assess reproducibility on the cobas m 511 system. The first study evaluated total precision of a single cobas m 511 instrument at one site and
{19}
the second study evaluated total precision of the cobas system across multiple sites.
To evaluate single-site reproducibility, testing was performed using a single cobas m 511 system with three lots of DigiMAC3 stain packs, wash solution, and reticulocyte stain and three different lots of DigiMAC3 controls. Three different operators conducted the study over the course of 20 operating days with wo runs per day and two replicates per run. The following components of precision were evaluated: repeatability (within-run), between-run, between-day, between-lot, between-operator, and within-laboratory (total precision). All results met the predefined acceptance criteria.
To evaluate precision at all sites, one lot of control material (L1, L2, L3) was used across four sites for five days with two runs per day and three replicates per run. The data generated from this assessment was used to calculate repeatability (within-run precision), between-day and between-run precision, between-laboratory precision (includes variability of different systems and operators), and reproducibility (total precision). For each reported parameter and for each level of control tested, the sample size (N), mean, SD and $\% \mathrm{CV}$ of the various components of precision were calculated along with the $95\%$ CI of the SD and $\% \mathrm{CV}$ for repeatability and reproducibility (total precision). All results met the predefined acceptance criteria.
Reproducibility for Combined Sites
| Parameter [Units] | Control Level | N | Mean | Repeatability | | Between-Run | | Between-Day | | Between-Laboratory | | Reproducibility (Total Precision) | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| WBC [103/μL] | L1 | 120 | 16.93 | 0.259 | 1.53 | 0.000 | 0.00 | 0.151 | 0.89 | 0.272 | 1.61 | 0.404 | 2.39 |
| | L2 | 120 | 8.00 | 0.209 | 2.62 | 0.022 | 0.28 | 0.065 | 0.82 | 0.158 | 1.97 | 0.271 | 3.39 |
| | L3 | 120 | 2.64 | 0.150 | 5.66 | 0.000 | 0.00 | 0.077 | 2.93 | 0.096 | 3.64 | 0.194 | 7.34 |
| RBC [106/μL] | L1 | 120 | 2.56 | 0.018 | 0.72 | 0.006 | 0.25 | 0.019 | 0.75 | 0.064 | 2.50 | 0.070 | 2.72 |
| | L2 | 120 | 4.29 | 0.053 | 1.23 | 0.000 | 0.00 | 0.039 | 0.92 | 0.102 | 2.38 | 0.121 | 2.83 |
| | L3 | 120 | 5.58 | 0.056 | 1.00 | 0.000 | 0.00 | 0.052 | 0.94 | 0.116 | 2.08 | 0.139 | 2.50 |
| HGB [g/dL] | L1 | 120 | 6.26 | 0.058 | 0.92 | 0.014 | 0.22 | 0.050 | 0.79 | 0.216 | 3.45 | 0.230 | 3.67 |
| | L2 | 120 | 12.31 | 0.143 | 1.16 | 0.019 | 0.15 | 0.112 | 0.91 | 0.374 | 3.04 | 0.417 | 3.38 |
| | L3 | 120 | 17.45 | 0.216 | 1.24 | 0.000 | 0.00 | 0.145 | 0.83 | 0.461 | 2.64 | 0.530 | 3.04 |
| HCT [%] | L1 | 120 | 18.11 | 0.154 | 0.85 | 0.065 | 0.36 | 0.144 | 0.79 | 0.409 | 2.26 | 0.465 | 2.57 |
| | L2 | 120 | 34.95 | 0.406 | 1.16 | 0.118 | 0.34 | 0.283 | 0.81 | 0.639 | 1.83 | 0.817 | 2.34 |
| | L3 | 120 | 49.24 | 0.596 | 1.21 | 0.000 | 0.00 | 0.331 | 0.67 | 0.735 | 1.49 | 1.003 | 2.04 |
| MCV [fL] | L1 | 120 | 70.77 | 0.305 | 0.43 | 0.265 | 0.37 | 0.226 | 0.32 | 0.268 | 0.38 | 0.535 | 0.76 |
| | L2 | 120 | 81.47 | 0.342 | 0.42 | 0.356 | 0.44 | 0.239 | 0.29 | 0.474 | 0.58 | 0.725 | 0.89 |
| | L3 | 120 | 88.27 | 0.413 | 0.47 | 0.285 | 0.32 | 0.464 | 0.53 | 0.499 | 0.56 | 0.846 | 0.96 |
| MCH [pg] | L1 | 120 | 24.48 | 0.129 | 0.53 | 0.087 | 0.36 | 0.115 | 0.47 | 0.288 | 1.18 | 0.347 | 1.42 |
| | L2 | 120 | 28.68 | 0.110 | 0.38 | 0.114 | 0.40 | 0.110 | 0.38 | 0.232 | 0.81 | 0.301 | 1.05 |
| | L3 | 120 | 31.28 | 0.134 | 0.43 | 0.092 | 0.29 | 0.102 | 0.33 | 0.216 | 0.69 | 0.289 | 0.92 |
| MCHC | L1 | 120 | 34.58 | 0.122 | 0.35 | 0.110 | 0.32 | 0.213 | 0.62 | 0.448 | 1.29 | 0.522 | 1.51 |
{20}
| Parameter [Units] | Control Level | N | Mean | Repeatability | | Between-Run | | Between-Day | | Between-Laboratory | | Reproducibility (Total Precision) | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| [g/dL] | L2 | 120 | 35.20 | 0.107 | 0.30 | 0.116 | 0.33 | 0.138 | 0.39 | 0.441 | 1.25 | 0.489 | 1.39 |
| | L3 | 120 | 35.43 | 0.097 | 0.28 | 0.084 | 0.24 | 0.171 | 0.48 | 0.415 | 1.17 | 0.467 | 1.32 |
| RDW [%] | L1 | 120 | 15.62 | 0.277 | 1.77 | 0.090 | 0.57 | 0.060 | 0.38 | 0.236 | 1.51 | 0.379 | 2.43 |
| | L2 | 120 | 13.18 | 0.248 | 1.88 | 0.203 | 1.54 | 0.000 | 0.00 | 0.107 | 0.81 | 0.338 | 2.56 |
| | L3 | 120 | 13.12 | 0.277 | 2.11 | 0.000 | 0.00 | 0.028 | 0.21 | 0.112 | 0.86 | 0.300 | 2.29 |
| RDW-SD [fL] | L1 | 120 | 39.79 | 0.587 | 1.48 | 0.307 | 0.77 | 0.263 | 0.66 | 0.744 | 1.87 | 1.030 | 2.59 |
| | L2 | 120 | 38.65 | 0.628 | 1.63 | 0.487 | 1.26 | 0.000 | 0.00 | 0.480 | 1.24 | 0.929 | 2.40 |
| | L3 | 120 | 41.71 | 0.759 | 1.82 | 0.000 | 0.00 | 0.211 | 0.51 | 0.449 | 1.08 | 0.907 | 2.18 |
| PLT [103/μL] | L1 | 120 | 470.53 | 4.611 | 0.98 | 1.885 | 0.40 | 3.705 | 0.79 | 5.187 | 1.10 | 8.090 | 1.72 |
| | L2 | 120 | 215.97 | 3.213 | 1.49 | 0.771 | 0.36 | 2.524 | 1.17 | 2.262 | 1.05 | 4.734 | 2.19 |
| | L3 | 120 | 76.69 | 1.281 | 1.67 | 0.916 | 1.19 | 0.697 | 0.91 | 0.935 | 1.22 | 1.960 | 2.56 |
| MPV [fL] | L1 | 120 | 7.65 | 0.089 | 1.17 | 0.000 | 0.00 | 0.020 | 0.27 | 0.042 | 0.55 | 0.101 | 1.32 |
| | L2 | 120 | 7.49 | 0.103 | 1.38 | 0.039 | 0.52 | 0.045 | 0.60 | 0.058 | 0.77 | 0.133 | 1.77 |
| | L3 | 120 | 7.36 | 0.173 | 2.35 | 0.028 | 0.39 | 0.000 | 0.00 | 0.043 | 0.58 | 0.180 | 2.45 |
| %NRBC [/100 WBC] | L2 | 120 | 18.30 | 0.931 | 5.09 | 0.250 | 1.37 | 0.000 | 0.00 | 1.185 | 6.48 | 1.528 | 8.35 |
| | L3 | 120 | 8.84 | 1.011 | 11.44 | 0.266 | 3.01 | 0.000 | 0.00 | 0.427 | 4.83 | 1.129 | 12.78 |
| #NRBC [103/μL] | L2 | 120 | 1.46 | 0.067 | 4.59 | 0.023 | 1.59 | 0.000 | 0.00 | 0.110 | 7.50 | 0.131 | 8.94 |
| | L3 | 120 | 0.23 | 0.027 | 11.62 | 0.008 | 3.42 | 0.000 | 0.00 | 0.016 | 6.65 | 0.032 | 13.82 |
| %NEUT [%] | L1 | 120 | 58.35 | 1.734 | 2.97 | 0.161 | 0.28 | 0.432 | 0.74 | 1.247 | 2.14 | 2.185 | 3.74 |
| | L2 | 120 | 59.79 | 1.867 | 3.12 | 0.000 | 0.00 | 1.010 | 1.69 | 1.256 | 2.10 | 2.466 | 4.13 |
| | L3 | 120 | 58.65 | 1.957 | 3.34 | 0.485 | 0.83 | 0.856 | 1.46 | 1.133 | 1.93 | 2.466 | 4.20 |
| %LYMPH [%] | L1 | 120 | 39.37 | 1.667 | 4.23 | 0.000 | 0.00 | 0.266 | 0.68 | 1.280 | 3.25 | 2.119 | 5.38 |
| | L2 | 120 | 37.32 | 1.798 | 4.82 | 0.000 | 0.00 | 1.078 | 2.89 | 1.252 | 3.35 | 2.442 | 6.54 |
| | L3 | 120 | 37.47 | 1.997 | 5.33 | 0.000 | 0.00 | 0.721 | 1.93 | 1.088 | 2.90 | 2.386 | 6.37 |
| %MONO [%] | L1 | 120 | 2.02 | 0.584 | 28.89 | 0.000 | 0.00 | 0.000 | 0.00 | 0.227 | 11.21 | 0.626 | 30.99 |
| | L2 | 120 | 2.45 | 0.663 | 27.02 | 0.167 | 6.80 | 0.096 | 3.90 | 0.279 | 11.39 | 0.744 | 30.35 |
| | L3 | 120 | 3.12 | 0.761 | 24.43 | 0.093 | 2.99 | 0.181 | 5.80 | 0.244 | 7.85 | 0.825 | 26.47 |
| %EO* [%] | L1 | 120 | 0.17 | 0.141 | N/A | 0.040 | N/A | 0.061 | N/A | 0.085 | N/A | 0.180 | N/A |
| | L2 | 120 | 0.31 | 0.191 | N/A | 0.076 | N/A | 0.062 | NA | 0.110 | N/A | 0.241 | N/A |
| | L3 | 120 | 0.68 | 0.351 | N/A | 0.000 | N/A | 0.078 | N/A | 0.247 | N/A | 0.436 | N/A |
| %BASO* [%] | L1 | 120 | 0.09 | 0.127 | N/A | 0.000 | N/A | 0.000 | N/A | 0.059 | N/A | 0.140 | N/A |
| | L2 | 120 | 0.13 | 0.129 | N/A | 0.039 | N/A | 0.037 | N/A | 0.083 | N/A | 0.163 | N/A |
| | L3 | 120 | 0.08 | 0.114 | NA | 0.000 | N/A | 0.041 | N/A | 0.021 | N/A | 0.123 | N/A |
| #NEUT [103/μL] | L1 | 120 | 9.87 | 0.307 | 3.11 | 0.052 | 0.52 | 0.103 | 1.04 | 0.090 | 0.92 | 0.340 | 3.44 |
| | L2 | 120 | 4.78 | 0.191 | 4.00 | 0.040 | 0.84 | 0.091 | 1.90 | 0.073 | 1.54 | 0.228 | 4.76 |
| | L3 | 120 | 1.55 | 0.098 | 6.31 | 0.000 | 0.00 | 0.058 | 3.76 | 0.035 | 2.28 | 0.119 | 7.70 |
| #LYMPH [103/μL] | L1 | 120 | 6.67 | 0.314 | 4.71 | 0.000 | 0.00 | 0.076 | 1.14 | 0.321 | 4.81 | 0.455 | 6.83 |
| | L2 | 120 | 2.99 | 0.167 | 5.58 | 0.000 | 0.00 | 0.089 | 2.99 | 0.151 | 5.05 | 0.242 | 8.10 |
| | L3 | 120 | 0.99 | 0.078 | 7.85 | 0.000 | 0.00 | 0.027 | 2.72 | 0.062 | 6.24 | 0.103 | 10.39 |
{21}
| Parameter [Units] | Control Level | N | Mean | Repeatability | | Between-Run | | Between-Day | | Between-Laboratory | | Reproducibility (Total Precision) | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| #MONO [103/μL] | L1 | 120 | 0.34 | 0.098 | 28.72 | 0.000 | 0.00 | 0.000 | 0.00 | 0.032 | 9.46 | 0.104 | 30.24 |
| | L2 | 120 | 0.20 | 0.054 | 27.46 | 0.009 | 4.64 | 0.007 | 3.55 | 0.017 | 8.89 | 0.058 | 29.45 |
| | L3 | 120 | 0.08 | 0.022 | 26.57 | 0.000 | 0.00 | 0.006 | 7.35 | 0.005 | 6.43 | 0.023 | 28.30 |
| #EO* [103/μL] | L1 | 120 | 0.03 | 0.023 | N/A | 0.007 | N/A | 0.011 | N/A | 0.013 | N/A | 0.029 | N/A |
| | L2 | 120 | 0.02 | 0.015 | N/A | 0.007 | N/A | 0.005 | N/A | 0.008 | N/A | 0.019 | N/A |
| | L3 | 120 | 0.02 | 0.010 | N/A | 0.000 | NA | 0.002 | N/A | 0.006 | N/A | 0.012 | N/A |
| #BASO* [103/μL] | L1 | 120 | 0.02 | 0.021 | N/A | 0.005 | N/A | 0.000 | N/A | 0.010 | N/A | 0.023 | N/A |
| | L2 | 120 | 0.01 | 0.010 | N/A | 0.004 | N/A | 0.003 | N/A | 0.006 | N/A | 0.013 | N/A |
| | L3 | 120 | 0.00 | 0.004 | N/A | 0.000 | N/A | 0.001 | N/A | 0.001 | N/A | 0.004 | N/A |
| %RET [%] | L1 | 120 | 7.46 | 0.525 | 7.03 | 0.310 | 4.16 | 0.155 | 2.07 | 0.397 | 5.32 | 0.744 | 9.97 |
| | L2 | 120 | 3.28 | 0.261 | 7.97 | 0.180 | 5.50 | 0.000 | 0.00 | 0.199 | 6.05 | 0.375 | 11.42 |
| #RET [106/μL] | L1 | 120 | 0.19 | 0.014 | 7.40 | 0.008 | 4.06 | 0.004 | 2.09 | 0.010 | 4.99 | 0.019 | 10.03 |
| | L2 | 120 | 0.14 | 0.012 | 8.66 | 0.006 | 4.41 | 0.005 | 3.23 | 0.010 | 7.01 | 0.017 | 12.42 |
| HGB-RET [pg] | L1 | 120 | 25.22 | 0.320 | 1.27 | 0.213 | 0.84 | 0.000 | 0.00 | 0.327 | 1.30 | 0.504 | 2.00 |
| | L2 | 120 | 26.11 | 0.369 | 1.41 | 0.146 | 0.56 | 0.149 | 0.57 | 0.362 | 1.39 | 0.557 | 2.14 |
*Parameters were evaluated based on SD.
# c. Linearity:
Linearity for RBC, HGB, HCT and WBC was evaluated using fresh whole blood samples. PLT linearity was evaluated using fresh platelet apheresis samples and corresponding whole blood samples. Reticulocyte (RET) linearity was evaluated using residual whole blood samples. Each sample was concentrated and diluted as necessary to create three separate linearity series, each consisting of a high pool, a low pool, and seven intermediate concentrations. The low pool concentrations for WBCs ranged from $0.04 - 0.07 \times 10^{3} / \mu \mathrm{L}$ and the high pool concentrations ranged from $404.8 - 561.1 \times 10^{3} / \mu \mathrm{L}$ . In the linearity series for PLT the low pool concentration ranged from $0 - 9 \times 10^{3} / \mu \mathrm{L}$ and the high pool concentration ranged from $5000 - 5130 \times 10^{3} / \mu \mathrm{L}$ . The low pool concentrations in each of the three separate linearity series for RET ranged from $0.01 - 0.03 \times 10^{6} / \mu \mathrm{L}$ and the high pool concentrations ranged from $0.63 - 0.75 \times 10^{6} / \mu \mathrm{L}$ . Each of the nine concentrations, in each of the three linearity series, was processed to obtain six replicates in each of the closed tube and open tube modes. Each linearity series was processed in both open and closed tube modes on three cobas m 511 systems.
Regression analysis was used to assess linearity for first order model (e.g. linear), and weighted polynomial regression was used to assess linearity for second and third order models (e.g., quadratic and cubic). RBC achieved linearity with a first order model while for all other parameters linearity was achieved through polynomial regression.
{22}
| Parameter | Linear Range |
| --- | --- |
| WBC (x103/μL) | 0.07–404.8 |
| RBC (x106/μL) | 0.37–8.26 |
| HGB (g/dL) | 1.1–24.2.0 |
| HCT (%) | 3.2–72.2 |
| PLT (x103/μL) | 0–5130 |
| RET (x106/μL) | 0.01–0.63 |
# d. Carryover:
Carryover study was conducted at targeted thresholds for WBC, RBC, PLT and samples with a high percentage of blasts. For each of the four sample types, three independent carryover experiments were conducted at each site using residual high-target value (HTV) whole blood samples for a total of 48 samples [12 samples for each of the four categories]. Filtered serum samples were used as low-target value (LTV) samples. For each experiment at each site, the HTV samples were run three consecutive times immediately followed by three LTV serum samples. Slides from the LTV serum samples were reviewed by an external hematopathologist to determine WBC, RBC, PLT and blast percent carryover from the HTV samples. Carryover for the cobas m 511 system met all predefined acceptance criteria.
# e. Interfering Substances:
The purpose of this study was to evaluate potential interference effects of hemolysis, lipemia, icterus (unconjugated or conjugated bilirubin), high levels of white blood cells $(\mathrm{WBC} > 100 \times 10^{9} / \mathrm{L})$ and high levels of platelets $(\mathrm{PLT} > 1000 \times 10^{9} / \mathrm{L})$ . Each interfering substance was tested in a series of six incremental concentrations.
Results showed there were no significant interference effects of unconjugated bilirubin or conjugated bilirubin up to the maximum tested concentration of $40\mathrm{mg / dL}$ for the evaluated parameters.
There were no significant hemolysis interference effects up to the maximum tested concentration of $1000\mathrm{mg / dL}$ for evaluated parameters.
Clinically significant hemolysis interference effects were observed at $\geq 672\mathrm{mg / dL}$ for HGB and $\geq 792\mathrm{mg / dL}$ for HCT.
There were no significant lipemia interference effects up to the maximum tested concentration of $3000\mathrm{mg / dL}$ for evaluated parameters.
Clinically significant lipemia interference effects were observed at $\geq 1646\mathrm{mg / dL}$ for WBC and $\geq 2459\mathrm{mg / dL}$ for LYMPH #.
High concentrations of white blood cells (up to $100.2 \times 10^{3} / \mu \mathrm{L}$ ) and high concentrations of platelets (up to $1166 \times 10^{3} / \mu \mathrm{L}$ ) did not cause any clinically significant bias in any of the evaluated parameters.
# 2. Other Supportive Instrument Performance Data Not Covered Above:
{23}
a. Reference Intervals
The normal adult reference interval for all parameters reported by the cobas m 511 system was established by analyzing whole blood samples collected in K_{2}EDTA from a total population of 489 healthy individuals (N=240 males, N=249 females).
A literature reference interval verification study was performed for the pediatric population. A total of 504 samples were collected from apparently healthy pediatric patients that ranged from < 1 day old to 18 years (258 samples) and 18 to ≤ 22 years (246 samples) to support the Intended Use. All samples were collected in K_{2}EDTA anticoagulant and analyzed in singlet in the automated whole blood sampling mode.
b. Mode to Mode Comparison
To demonstrate equivalency between results of samples processed in the closed tube (i.e. automated) mode of operation and the open tube (i.e. manual) mode of operation on the cobas m 511 system, a comparison study was conducted using 140 residual samples collected in K_{2}EDTA. Correlation and bias between the closed tube and open tube processing mode results were determined based on the results of either a Passing-Bablok or Deming regression model. For the %NRBC and #NRBC parameters the mean difference was used to calculate bias. All reportable parameters that were evaluated met their predefined bias acceptance criteria.
c. Open Mode studies
Precision (repeatability)
To evaluate repeatability in open tube mode performance, 12 residual K_{2}EDTA whole blood samples selected at random from the routine laboratory population as well as two residual whole blood samples with reticulocytes ≥ 5% were used for this study. Eleven replicates were processed from each random sample on three different cobas m 511 systems (33 replicates per sample) in the open tube mode. The results from all three systems were pooled for analysis. The mean, standard deviation (SD), and percent coefficient of variation (% CV) with corresponding 95% confidence intervals were calculated for each parameter using a variance component analysis in accordance with the CLSI EP05-A3 guideline. Within-run precision (repeatability), between-instrument variability, and total variability (inter-instrument precision) were analyzed. All results were within the predefined acceptance criteria.
Morphology
To assess the WBC differential parameters and WBC, RBC and PLT morphology in open tube mode, two studies were conducted using 24 targeted residual K_{2}EDTA whole blood samples. The samples were processed in both open tube and closed tube modes on each of three cobas m 511 systems. Two separate analyses were conducted. In Part IIA the cobas m 511 system automated WBC differential results for the 24 samples were compared between the open tube and closed tube modes. Correlation and bias were determined in accordance with the CLSI EP09-A3 guideline. In Part IIB, each of three medical technologists (MTs) reviewed on the viewing station all of the cobas m 511 images generated from the 24 samples from all three systems, which were run in both open tube and closed tube modes. This analysis was conducted to determine whether WBC, RBC and PLT morphology were consistent between the
{24}
two modes and between systems. Two analysis methods were used to evaluate results from this portion of the study. Agreement analysis was performed for each morphological characteristic using either a qualitative grading system (e.g., Present / Not Present) or a quantitative grading system (e.g., 0 to $4+$ grading scale). All results were within the predefined acceptance criteria.
# Carryover
Carryover in the open tube mode was evaluated using four residual $\mathrm{K}_2\mathrm{EDTA}$ whole blood samples, each selected as a high target value (HTV) sample type. Three separate experiments were run for each of the WBC, RBC, PLT and blasts samples. In each experiment, the HTV sample was processed three consecutive times immediately followed by three tubes of low target value (LTV) samples. The HTV and LTV samples were run in open tube mode on a single cobas m 511 system. The LTV samples, were comprised of filtered serum equilibrated to room temperature. By design, when serum samples are run on the cobas m 511 system, the instrument will make a slide but will ultimately reject the sample when it is presented at the low-magnification imaging station because it does not contain sufficient cells for the imaging microscopes to focus. The LTV slides, however, were retrieved from the instrument and manually reviewed under a microscope to assess if any carryover was evident from the three prior runs of HTV samples. Carryover in the open tube mode for the cobas m 511 system met all predefined acceptance criteria.
# d. Limits of Detection, Blank, and Quantitation (LoD, LoB, and LoQ)
This study was conducted to evaluate LoB, LoD and LoQ WBC and PLT parameters on the cobas m 511 system. The study was conducted using three cobas m 511 systems. To determine LoB, testing was performed on three individual test days, using five preserved Streck RBC samples which contained no white blood cells or platelets. These samples were processed two times per test day on each of two cobas m 511 systems, yielding a total of 60 measurements. To determine LoD, residual $\mathrm{K}_2\mathrm{EDTA}$ whole blood samples with targeted low level values that were greater than the LoB were used. The LoD samples were processed four consecutive times on each of three cobas m 511 systems, for a total of 72 measurements for both WBC and PLT samples. To determine LoQ, residual whole blood samples with targeted low level values that were greater than the LoD were used. The LoB, LoD and LoQ were successfully determined and met all pre-specified acceptance criteria.
| Parameter | LoB | LoD | LoQ |
| --- | --- | --- | --- |
| WBC (x103/μL) | 0.05 | 0.08 | 0.24 |
| PLT (x103/μL) | 1 | 3 | 6 |
# e. $\mathrm{K}_2\mathrm{EDTA}$ and $\mathrm{K}_3\mathrm{EDTA}$ Performance Comparison
This study was conducted to demonstrate comparability between whole blood samples collected into $\mathrm{K}_2$ and $\mathrm{K}_3$ EDTA. This study was performed using 44 whole blood samples from healthy volunteer donors. For each sample, the $\mathrm{K}_2\mathrm{EDTA}$ results
{25}
were compared to $\mathrm{K}_3\mathrm{EDTA}$ results for all parameters. All acceptance criteria were met, demonstrating equivalency of results between samples collected into $\mathrm{K}_2\mathrm{EDTA}$ and $\mathrm{K}_3\mathrm{EDTA}$.
To evaluate whole blood samples collected with $\mathrm{K}_3\mathrm{EDTA}$ a method comparison study was conducted with 40 abnormal samples on the cobas m 511 and Sysmex XN-10. Bias was determined based on the results of either a Passing-Bablok or Deming regression model, except for the $\%$NRBC and #NRBC parameters. For these two parameters, no regression analysis was used; rather, the bias was calculated as the mean difference between the cobas m 511 system and Sysmex Analyzer. Results were within predefined critical bias limits.
f. Venous and Capillary Blood Performance Comparisons
This study was conducted to demonstrate comparability between capillary and venous blood processed on the cobas m 511 system from the same donor. This study was performed using 40 paired whole blood samples from healthy volunteer donors. Blood was drawn twice from each donor, once via venipuncture (venous) and once via finger-stick (capillary). For all samples, results from the cobas m 511 system were comparable for capillary and venous blood. Overall, the data demonstrate that both capillary and venous blood can be processed on the cobas m 511 system and produce similar results.
To evaluate abnormal capillary whole blood samples on the cobas m 511 system a method comparison study was conducted with 40 abnormal capillary samples on the cobas m 511 and Sysmex XN-10. Bias was determined based on the results of either a Passing-Bablok or Deming regression model, except for the $\%$NRBC and #NRBC parameters. For these two parameters, no regression analysis was used; rather, the bias was calculated as the mean difference between the cobas m 511 system and Sysmex Analyzer. Results were within predefined bias limits.
g. Sample Stability
Sample stability was determined for normal and abnormal samples at two clinical sites. Normal samples were processed in duplicate at each of the following time points: baseline (time zero (0)) and after 9, 20, 26, 38, and 50 hours in both ambient $(15^{\circ}\mathrm{C} - 25^{\circ}\mathrm{C})$ and refrigerated $(2^{\circ}\mathrm{C} - 8^{\circ}\mathrm{C})$ storage conditions. Abnormal samples were processed in duplicate at each of the following time points: baseline (time zero (0)) and after 9, 26, and 50 hours in both ambient $(15^{\circ}\mathrm{C} - 25^{\circ}\mathrm{C})$ and refrigerated $(2^{\circ}\mathrm{C} - 8^{\circ}\mathrm{C})$ storage conditions. For each time point, results were compared to the respective baseline (zero (0) hour) results. The combined results from both protocols demonstrated stability for normal and abnormal samples for 48 hours.
h. Stability of cobas m 511 Slides
This study was conducted to demonstrate that cobas m 511 slides stored for 30 days produce similar results compared to original (baseline) results, with respect to manual microscopic reviews for the WBC differential and morphology. The results from all evaluations performed during this study were successful and met predefined acceptance criteria. This demonstrates that slides generated by the cobas m 511 system are stable for manual microscopic review for a minimum of 30 days.
26
{26}
K. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
L. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
27
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.