HBS2 Headless Bone Screw is used for the treatment of intraarticular and extraarticular fractures and pseudoarthroses of smaller bones and bone fragments as well as for arthrodeses on small joints, especially in the scope of: HBS2 MIDI: Scaphoid fractures, Scaphoid pseudoarthroses, Proximal pole fractures of the scaphoid, DIP arthrodeses, Metacarpal fractures, Metatarsal fractures, Fractures of the processus styloideus ulnae, Proximal radial head fractures HBS2 MINI: Scaphoid fractures, Scaphoid pseudoarthroses, Proximal pole fractures of the scaphoid, DIP arthrodeses, Metacarpal fractures, Fractures of the processus styloideus ulnae, Proximal radial head fractures, Fractures of the processus styloideus radii
Device Story
HBS2 Headless Bone Screws are cannulated, self-drilling, self-cutting metallic fasteners used in hand, accident, and reconstructive surgery. Designed for internal fixation of small bone fractures and arthrodesis. Operated by surgeons in clinical settings. Screws feature a reversing thread to provide compression across fracture sites or joints. Delivered non-sterile; require reprocessing (cleaning/sterilization) by the facility before use. Output is mechanical stabilization of bone fragments, facilitating healing and joint fusion.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by non-clinical bench testing, including biological safety (cytotoxicity, bioburden, endotoxin, particle testing) and mechanical testing (torsional properties, driving torque, axial pullout strength, and static/dynamic 3-point bending) per ASTM standards.
Technological Characteristics
Material: Ti6Al4V (3.7165). Design: Cannulated, self-drilling, self-cutting headless bone screws with reversing thread. Diameters: 1.7 mm (MINI) and 2.0 mm (MIDI). Mechanical testing per ASTM F543-07, F116-00, F1839-08, F1264-03. Biological safety per EN ISO 10993 series. Delivered non-sterile; requires facility-validated reprocessing.
Indications for Use
Indicated for treatment of intraarticular and extraarticular fractures, pseudoarthroses of small bones/fragments, and arthrodeses of small joints in patients requiring bone fixation.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
HBS HEADLESS BONE SCREW, SCHOENING & ASSOC., INC. (K030302)
Reference Devices
SYNTHES 2.4 MM CANNULATED COMPRESSION SCREW, SYNTHES (USA) (K021556)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of the symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
January 12, 2018
Stuckenbrock Medizintechnik GmbH Fabian Stuckenbrock President Lessingstrasse 50 78532 Tuttlingen Germany
Re: K171628
Trade/Device Name: HBS2 Headless Bone Screw Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: December 26, 2017 Received: December 18, 2017
Dear Fabian Stuckenbrock:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Katherine D. Kavlock -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K171628
Device Name HBS2 Headless Bone Screw
#### Indications for Use (Describe)
HBS2 Headless Bone Screw is used for the treatment of intraarticular fractures and pseudoarthroses of smaller bones and bone fragments as well as for arthrodeses on small joints, especially in the scope of:
HBS2 MIDI: Scaphoid fractures, Scaphoid pseudoarthroses, Proximal pole fractures of the scaphoid, DIP arthrodeses, Metacarpal fractures, Metatarsal fractures of the processus styloideus ulnae, Proximal radial head fractures
HBS2 MINI: Scaphoid fractures, Scaphoid pseudoarthroses, Proximal pole fractures of the scaphoid, DIP arthrodeses, Metacarpal fractures, Fractures of the processus styloideus ulnae, Proximal radial head fractures of the processus styloideus radii
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Stuckenbrock Medizintechnik ein Unternehmen der KLS MOITEIN Group
K171628
# 510(k) Summary
Issue Date:
#### 18-12-2017
510(k) Owner´s Name: Stuckenbrock Medizintechnik GmbH Lessingstrasse 50 78532 Tuttlingen Germany Tel: + 49 (0)7461/165880 Fax: +49 (0)7461/4194 Internet: www.stuckenbrock.de
Contact Person:
Fabian Stuckenbrock President of Stuckenbrock Medizintechnik GmbH Tel .: +49 7461 165880 E-Mail: f.Stuckenbrock@stuckenbrock.de
{4}------------------------------------------------
## 1. General Info
## 1.1. Device Name
Trade Name: HBS2 Headless Bone Screw
Common Name: Bone fixation screw
Classification Name: Screw, fixation, bone
## 1.2. Classification Product Code
Stuckenbrock HBS2 Headless Bone Screw can be classified according to following device name and product code:
| Device | Regulation<br>Description | Regulation<br>Medical<br>Specialty | Review<br>Panel | Product<br>Code | Regulation<br>Number | Device<br>Classification |
|--------------------------|----------------------------------------------------------|------------------------------------|-----------------|-----------------|----------------------|--------------------------|
| Screw, fixation,<br>bone | Smooth or threaded<br>metallic bone<br>fixation fastener | Orthopedic | Orthopedic | HWC | 888.3040 | 2 |
{5}------------------------------------------------
## 2. Indication for Use
HBS2 Headless Bone Screw is used for the treatment of intraarticular and extraarticular fractures and pseudoarthroses of smaller bones and bone fragments as well as for arthrodeses on small joints, especially in the scope of:
#### □ HBS2 MIDI:
- י Scaphoid fractures
- Scaphoid pseudoarthroses .
- I Proximal pole fractures of the scaphoid
- DIP arthrodeses I
- . Metacarpal fractures
- I Metatarsal fractures
- Fractures of the processus styloideus ulnae .
- I Proximal radial head fractures
#### 0 HBS2 MINI:
- Scaphoid fractures ■
- Scaphoid pseudoarthroses .
- I Proximal pole fractures of the scaphoid
- DIP arthrodeses '
- . Metacarpal fractures
- Fractures of the processus styloideus ulnae .
- Proximal radial head fractures
- Fractures of the processus styloideus radii ■
{6}------------------------------------------------
## 3. Device Description
The HBS2 Headless Bone Screws are used in the field of hand surgery, accident surgery and reconstructive surgery, and in orthopedics for the treatment of intraarticular and extra articular fractures and pseudo arthroses of small bone fragments, and for arthrodesis on small joints.
The system consists of various screws and instruments for different treatment options.
HBS2 screws are cannulated, self-drilling, self-cutting, have a reversing thread and are available in two diameters with different thread lengths.
HBS2 MIDI screws (magenta) are available with a diameter of 2.0 mm (shaft) and in different lengths. HBS2 MINI screws (green) are available with a diameter of 1.7 mm (shaft) and in different lengths too.
| Image: Screw with arrows | Ø 3.9 mm<br>Thread pitch 1.0 mm | Image: Screw with arrows |
|--------------------------|----------------------------------|--------------------------|
| | Ø 2.0 mm | |
| Image: Screw with arrows | Ø 3.0 mm<br>Thread pitch 1.25 mm | Image: Screw with arrows |
| | Ø 3.2 mm<br>Thread pitch 0.75 mm | |
| Image: Screw with arrows | Ø 1.7 mm | |
| | Ø 2.5 mm<br>Thread pitch 1.0 mm | |
These devices are delivered in non-sterile condition and must be reprocessed before use. All implants are made out of Ti6Al4V 3.7165.
{7}------------------------------------------------
## 4. Predicate Devices
### 4.1. Primary Predicate Device
- □ HBS HEADLESS BONE SCREW, SCHOENING & ASSOC., INC., K030302
### 4.2. Secondary Predicate or Reference Devices
- □ SYNTHES 2.4 MM CANNULATED COMPRESSION SCREW, SYNTHES (USA), K021556
- SYNTHES 3.0MM HEADLESS COMPRESSION SCREWS, SYNTHES (USA), K050636
### 4.2.1. Substantial Equivalence Claimed to predicate Devices
The Stuckenbrock HBS2 headless bone screws are substantially equivalent to the predicate devices in terms of intended use, design, materials used, mechanical safety and performance.
{8}------------------------------------------------
## 5. Clinical and Non-Clinical Test Summary
### 5.1. Clinical Tests
No clinical studies were performed.
#### Non-Clinical Test Summary 5.2.
The follow tests have been conducted:
| Topic | Standard | Test |
|----------------------|--------------------|---------------------------------------------------|
| Biological<br>Safety | EN ISO<br>10993-5 | Cytotoxicity |
| | EN ISO<br>10993-12 | Chemical Analysis |
| | EN ISO<br>10993-18 | Characterization of Organic Extractable |
| | EN ISO<br>11737-1 | Bioburden |
| | USP 85 | LAL Endotoxin |
| | USP 788 | Particle |
| Mechanical<br>Safety | ASTM<br>F543-07 | Torsional properties |
| | ASTM<br>F116-00 | Driving torque |
| | ASTM<br>F1839-08 | Axial pullout strength |
| | ASTM<br>F1264-03 | Static 3-point bending<br>Dynamic 3-point bending |
Sensible manufacturing processes have been validated (e.g. Final Cleaning). Furthermore, reprocessing process has been validated in terms of:
- 0 Pre-Treatment
- O Manual Cleaning
- □ Automated Cleaning/Disinfection (thermal disinfection)
- □ Drying
- □ Sterilization
### 5.2.1. Conclusions Non-Clinical Tests
Stuckenbrock considers the HBS2 headless bone screws to be equivalent to the predicate devices listed above. This conclusion is based upon the device's similarities in principles of operation, technology, materials, and indications for use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.